User Manual
GDS-1000A-U Series User Manual
142
Common specifications
Vertical Sensitivity 2mV/div~10V/Div (1-2-5 increments)
Accuracy ± (3% x |Readout|+0.1div + 1mV)
Bandwidth See model-specific specifications
Rise Time See model-specific specifications
Input Coupling AC, DC, Ground
Input Impedance
1MΩ±2%, ~15pF
Polarity Normal, Invert
Maximum Input 300V (DC+AC peak), CAT II
Math Operation +, –, ×, FFT, FFT rms
Offset Range 2mV/div~50mV/div: ±0.4V
100mV/div~500mV/div: ±4V
1V/div~5V/div: ±40V
10V/div : ±300V
Trigger Sources CH1, CH2, Line, EXT
Modes Auto, Normal, Single, TV, Edge, Pulse
Coupling AC, DC, LF rej, HF rej, Noise rej
Sensitivity See model-specific specifications
Holdoff 40ns ~ 2.5s
External trigger Range DC: ±15V, AC: ±2V
Sensitivity See model-specific specifications
Input Impedance
1MΩ±2%, ~15pF
Maximum Input 300V (DC+AC peak), CATII
Horizontal Range 1ns/div~50s/div, 1-2.5-5 increment
Roll: 50ms/div – 50s/div
Modes Main, Window, Window Zoom, Roll, X-Y
Accuracy ±0.01%
Pre-Trigger 10 div maximum
Post-Trigger 1000 div
X-Y Mode X-Axis Input Channel 1
Y-Axis Input Channel 2
Phase Shift ±3° at 100kHz
Signal Acquisition Real-Time 1G Sa/s maximum
Equivalent 25G Sa/s maximum
Vertical
Resolution
8 bits
Record Length Maximum; 2M points (1 channel), 1M
points (2 channels)
Acquisition Normal, Peak Detect, Average
Peak Detection 10ns (500ns/div ~ 50s/div)
Average 2, 4, 8, 16, 32, 64, 128, 256