Manual
GDS-1000-U Programming Manual
22
Parameter
<X>
Channel
<NR3>
Vertical scale
1/2
CH1/2
2e–3 ~ 5e+0
2mV ~ 5V
(Probe x1)
2e–2 ~ 5e+1
20mV ~ 50V
(Probe x10)
2e–1 ~ 5e+2
200mV ~ 500V
(Probe x100)
Example
:channel1:probe 0
:channel1:scale 2.00e–3
Sets the Channel 1 probe
attenuation factor to x1
Sets the Channel 1
vertical scale to 2mV/div