760000 Alternating Pressure / Low Air Loss Mattress System User Manual Important: Do not operate the Mattress System without first reading and understanding this manual! Save this manual for future use.
Contents Introduction...................................................................................... 3 Intended use of this device......................................................... 3 Important safety precautions................................................... 3 Danger.................................................................................................. 4 Warning................................................................................................ 4 Caution............
Introduction Use this manual for set-up and operation of the Alternating Pressure / Low Air Loss Mattress System. Read all instructions before using the mattress system. Save this manual for future reference. Intended use of this device The intended use of the Alternating Pressure / Low Air Loss Mattress System is: • The treatment of pressure ulcers while optimizing user comfort. • Pain management as prescribed by a physician. • As described above, in either a homecare or long-term care setting.
Danger To reduce the risk of electrocution: DANGER: Always unplug this product immediately after using. DANGER: Do not use this product while bathing. DANGER: Do not place or store this product where it can fall or be pulled into a tub or sink. DANGER: Do not place this product in or drop into water or other liquid. DANGER: Do not reach for a product that has fallen into water. Unplug it immediately.
WARNING: Never block the air openings of this product or place it on a soft surface, such as a bed or couch, where the openings may be blocked. Keep the air opening free of lint, hair, and other similar particles. WARNING: Never drop or insert any object into any opening or hose. WARNING: Do not use this product in the presence of flammable anesthesia, or where oxygen or nitrous oxide is being administered. WARNING: Connect this product only to a properly grounded power outlet.
For U.S. and Canada only Medical Equipment: Air Pump with respect to electrical shock, fire and mechanical hazards only in accordance with UL60601-1 and CAN/CSA C22.2 No. 601.
Pump: rear fuse main power switch power cord receptacle hanger filter Pump: front panel decrease pressure button; pressure level indicators at right of button cycle time adjust button; cycle time indicators at right of button increase pressure button therapy adjust button; therapy mode indicators at right of button panel lock button; panel lock indicator at right of button power ON/OFF button; ON/OFF indicator at right of button service indicator low pressure indicator (upper) power failure indicator
increase pressure under the buttocks). Press the therapy adjust button at the left end of the row to cycle between the settings. The selected therapy mode setting's indicator will illuminate. Front panel fourth row Panel lock button: The panel lock protects the panel settings from accidental change. Press to lock; press for three seconds to unlock. When the panel remains untouched for three minutes, it locks automatically. When the panel is locked, the panel lock indicator will illuminate.
Service indicator: Mechanical failure, service required; contact your Graham-field equipment provider. Installation Unpacking 1. Before unpacking, check for obvious damage to the carton or its contents. If damage is evident, please notify the carrier and your Graham-Field equipment provider. 2. Remove all loose packing from the carton. 3. Carefully remove all components from the carton. 4. Inspect all components. If damage is evident, please notify the carrier and your Graham-Field equipment provider.
s Caution: Place the device in a position where the caregiver and/or doctor can access it easily. 5. Connect the mattress-pump air hose quick connector; ensure that the connector has clicked securely into place before continuing. s Caution: Ensure that the air hoses are not kinked or tucked under the mattress. WARNING: Locate all cords so that they will not be stepped on, tripped over, or otherwise subjected to damage or stress.
Note: The mattress system will automatically enter Autofirm mode for the quickest inflation at set-up; initial inflation takes approximately twenty minutes. 3. The pump will begin delivering air into the mattress. 4. When the selected pressure level is reached, the pump will enter alternating mode. The mattress can then be adjusted to the desired pressure. 5. Replace all sheets, blankets, and pillows on the bed.
quick connector Transport cap In case of power failure or transport: Disconnect the quick connector and cap the mattress with the transport cap to slow deflation. CPR function If CPR must be performed on the patient while the mattress is in use, to immediately deflate the mattress, either: 1. Pull the CPR valves from the mattress in the location of the arrow shown at right, or 2. Disconnect the quick connector from the pump in the location of the arrow shown at right.
GENERAL Maintenance 1. Check main power cord for abrasion or excessive wear. 2. Check mattress cover for signs of wear or damage. Ensure mattress cover and tubes are connected correctly. 3. Check airflow from the air hose connector. The airflow should alternate between each connector every halfcycle time if it’s in alternating mode. 4. Check the air hoses for any kink or break. For replacement, please contact your GF equipment provider. Fuse replacement tool needed: small screwdriver 1.
Storage 1. Lay the mattress out flat and upside down. 2. Roll from the head end toward the foot end. 3. Stretch the foot-end strap around the rolled mattress to prevent unrolling. 4. Store mattress and pump in a dust-free environment with no exposure to direct sunlight. s Caution: Do not fold, crease or stack mattress.
Mattress Specification Model 8" mattress Dimensions (L x W x H) 78.7" x 35.4" x 8" (200 x 90 x 20.3 cm) Weight 16.5 lb (7.5 kg) Warranty GF Health Products, Inc. warrants the Lumex AltaDyne Plus 760000 Alternating Pressure / Low Air Loss Mattress System for a period of twelve months for defects in workmanship and materials. During the warranty period, defective items will be repaired or replaced at GF Health Products' option at no charge.
USA, Corporate Headquarters: Graham-Field Health Products 2935 Northeast Parkway Atlanta, Georgia 30360 telephone: 800-347-5678, 770-447-1609 fax: 800-726-0601, 678-291-3232 www.grahamfield.com 760000-INS-LAB-RevA08 © 2008 GF Health Products, Inc.