1. Technical Data
1.1 General
Compatibility: Signet 2720 Twist-Lock Preampli er
Pipe Size Range: ½ in. and larger
Use Signet installation ttings to 4 in.
Use pipe adapter in pipes over 4 in.
Submersion with 2720 pre-ampli er requires ¾ in. NPT or
ISO 7-R ¾ in. male threaded extensions
Ef ciency: >97% @ 25°C (77° F)
Wetted Materials: CPVC Body
Glass
FPM O-rings
Porous UHMW Polyethylene reference
junctions
ORP: Platinum sensing surface
Primary Functions: 2714, 2715: Flat surface resists fouling
2716, 2717: Bulb surface for general
use
2714-HF: Extended use in applications
with trace hydro uoric acid (<2%)
2716-DI: Extended use in pure waters
(<100 μS)
Secondary Junction: Nylon Filament
Element: Ag/AgCl
Shipping Weight: 0.2 kg (0.4 lb)
Chinese RoHS (Go to www.gfsignet.com for details)
1.2 2714 and 2716 pH Electrodes
Operating Range: 0 to 14 pH (2714-HF: 0 to 12 pH)
Reference: Electrolyte: Solidi ed Acrylamide Gel
3.5M KCI (2714, 2714-HF, 2716)
0.1 M KCl (2716-DI)
Temperature Sensor: 3KΩ Balco (3000Ω = 25°C)
ResponseTime: <5 s for 95% of signal change
Response Time, τ: 140 s (2714), 196 s (2716)
1.3 2715 and 2717 ORP Electrodes
Operating Range: -999 to +999 mV
Reference: Electrolyte: Solidi ed Acrylamide Gel
3.5M KCL
Response Time, τ: Application dependent
1.5 Dimensions
Flat electrode Bulb electrode
1.4 Temperature and pressure speci cations
ORP electrode operating Temperature/Pressure
6.89 bar @ 0°C to 65ºC (100 psi @ 32° to 149°F)
4.00 bar @ 66ºC to 85°C (58 psi @150° to 185°F)
2716 and 2716-DI pH electrode operating Temperature/Pressure
6.89 bar @ 0°C to 65ºC (100 psi @ 32° to 149°F)
4.00 bar @ 66ºC to 85°C (58 psi @150° to 185°F)
2714 pH electrode operating Temperature/Pressure
6.89 bar @ 10°C to 65ºC (100 psi @ 50° to 149°F)
4.00 bar @ 66ºC to 85°C (58 psi @150° to 185°F)
2714-HF pH electrode operating Temperature/Pressure
6.89 bar @ 0°C to 50ºC (100 psi @ 32° to 122°F)
Storage Temperature: > -10°C (14°F)
1.0 in.
(27 mm)
0.9 in.
(23 mm)
4.1 in.
(104 mm)
3.4 in.
(86 mm)
1.2 in.
(30.5 mm)
2.6 in.
(65 mm)
1.0 in.
(27 mm)
1.2 in.
(30.5 mm)
2.6 in.
(65 mm)
0.9 in.
(23 mm)
3.4 in.
(86 mm)
3.8 in.
(97 mm)
50
3
PSI BAR
50
122
75
167
25
77
0
10
32
50
25
2
75
5
100
7
°F
°C
2714
2714-HF
SAFETY INSTRUCTIONS
1. Depressurize and vent system prior to installation or removal.
2. Con rm chemical compatibility before use.
3. Do not exceed maximum temperature/pressure speci cations.
4. Wear safety goggles or faceshield during installation/service.
5. Do not alter product construction.
6. When using chemicals or solvents care should be taken and appropriate eye, face, hand, body,
and/or respiratory protection should be used.
*3-2714.090-1*
3-2714.090-1 Rev. J 3/08 English