3.6 ALARM TEST
What is the alarm test?
EXO’s alarm test is an audio/visual assessment used to ensure the siren,
speaker, LED lights and microphones are operating correctly on the area
monitor. The top lights will ash and you will hear pitches that gradually
get higher. The volume of the alarm test sounds cannot be adjusted.
When is a alarm test run?
A alarm test is run as part of the startup sequence, before all bump
tests, before each calibration and after every 24 hour period of
continuous run time.
What does a failed alarm test mean?
If EXO detects a problem while running the alarm test, the full test
will immediately be attempted two more times. If EXO still detects
a problem after the third attempt, a failed alarm test results will be
recorded and sent to Blackline Live.
NOTE: Muing the sounds during a alarm test will result in a failure. Too
loud of an environment will also result in a failed alarm test.
3.5 MAINTENANCE CODE
EXO features an optional maintenance code function to prevent
unauthorized individuals from changing the settings when the device
is unmanned. EXO’s entire menu, device power-down and volume
change actions are locked when a maintenance code is enabled.
Entering the code will allow you to access locked features.
Enabling the maintenance code and setting the four-digit number
passcode is done using the EXO configuration page available on the
Blackline Live Portal.
When the device is locked, a lock icon will be shown in the
gas status screen banner. Pressing any button while EXO is
locked will display the maintenance code entry screen.
To unlock EXO,
Press any button to display the maintenance code entry screen
You will be prompted to enter a four-digit number. Use the up and
down buttons to select a number for the current digit
Press OK to move on to the next digit
After selecting the nal digit, press OK to enter the entire code
If successful, the entered code will become green and you will be
taken to the unlocked gas detection screen.
If unsuccessful, the entered code will become red. You can either
select Retry to enter another code or Back to return to the locked
gas status screen.
Enter maintenance code