6. To start the bump test, select by pressing the right
button.
NOTE:
To cancel the bump test and exit the workflow at
any time, select by pressing the left button.
G6 performs a hardware self test of your device’s audio,
visual, and vibration capabilities.
7. Verify the gas and your device’s connection to the
calibration cap and turn on the gas.
Your device automatically detects the gas and starts the
bump test.
8. Once the bump test is successfully completed, turn off the
gas and disconnect the calibration cap from your G6.
9. Select by pressing the right button.