User Manual

6
5
Pressing the Auto key will automatically adjust the
ECG amplitude to 10mm.
The automatic adjustment will function only when the
key is pressed.
The automatic adjustment will not function when the
monitor is learning arrhythmia.
4. Select the lead.
The leads can be selected from 3 leads, 6 leads, 7 leads depending on the connected ECG relay
cable.
ECG Relay Cable Lead
3-electrode
4-electrode
5-electrode
5. Set the baseline position.
If the waveform is difficult to see due to ECG
amplitude, set the 0mV baseline position.
The baseline position for the waveform display and
recording will be adjusted.
CAUTION
?
The threshold level for arrhythmia detection changes with ECG
waveform size. Set a proper waveform size for monitoring.
When the ECG waveform size is
?
1/4,
?
1/2, or
?
1, the detection
threshold is 250 µV.
When the ECG waveform size is
?
2, or
?
4, the detection threshold is
150 µV.
?
Automatic size/position of the ECG is effective only at the time the
AUTO
key is pressed. This does not continually adjust size and
position.
6
ECG