Datasheet

Output
@ 25 °C (77 °F) =1.00 mV (reference junction @ 0 °C)
Seebeck Coefficient
@ 25 °C (77 °F) = 40.50 μV / °C
Measurement Time
(Time Constant): 3.0 seconds in 100 °C still water at sea level
pressure. Complete step change equals 5 time constants (3.75
seconds).
Maximum Voltage
24 V ac rms or 60 V dc
Maximum Temperature of Tip
1090 °C (1994 °F)
Sheath
Material: Alloy 600
Dimensions:
Diameter: 3.175 mm (1/8 in)
Length: 212.725 mm (8.375 in)
Grounding
Junction welded to sheath.
Cable
Length: 40 inches (1 meter)
Insulation
Material: PVC
Maximum Temperature: 105 °C (220 °F)
Jacket Color: Gray
Conductors
Type: K
Size: AWG #24 stranded (7 strands of #32)
Handle
Material: Hytrel
Maximum Temperature: 125 °C (257 °F)
Connector
Type: Mini-thermocouple connector with .792 mm (0.312 in)
pin spacing
Material: Yellow Hytrel
Maximum Temperature: 125 °C (257 °F)
Overall Length
31.75 cm (12.5 in) from probe tip to end of cable strain relief.
Protection
Class 3. Relates solely to insulation and grounding properties
defined in IEC 348.