User's Manual
Table Of Contents
- Chapter 1
- Beginning Your Dexcom G5® Mobile Continuous Glucose Monitoring (CGM) System Journey
- Chapter 2
- Indications for Use and Safety Statement
- Chapter 3
- Risks and Benefits
- Chapter 4
- What’s the G5 Mobile?
- Chapter 5
- Set Up Your Display Devices
- Chapter 6
- Starting a Sensor Session: Inserting Sensor, AttachingTransmitter, and Starting Your Session
- Chapter 7
- Calibration
- Chapter 8
- Ending Your Sensor Session and Transmitter Session
- Chapter 9
- Home Screen, Rate of Change Arrows, and Errors
- Chapter 10
- Daily Events Affect Your Glucose Trends and Patterns
- Chapter 11
- Sensor Glucose Alarm and Alerts
- Chapter 12
- On the Go With G5: Customizing Your Alarm and Alerts
- EVERYTHING ELSE G5
- Chapter 13
- Warranty: The Fine Print
- Chapter 14
- How to Take Care of Your G5 Mobile
- Chapter 15
- On the Go With G5 Mobile: Getting Through Security
- Chapter 16
- Need Help? You’re Not Alone!
- Chapter 17
- Technical Information
- Chapter 18
- Troubleshooting
- Chapter 19
- Symbols on Package Labels
- SHARING IS CARING
Dexcom G5 Mobile System User Guide
Technical Information
275
Table 3-C. Concurrence of CGM Readings and YSI Values (Original Pediatric
Study)
CGM
(mg/dL)
YSI (mg/dL)
Row Percentage of Matched Pairs in Each CGM Glucose Range
< 40
40-
60
61-
80
81-
120
121-
160
161-
200
201-
250
251-
300
301-
350
351-
400
> 400
Number
of Paired
CGM-YSI
< 40 0% 0% 0% 54% 31% 15% 0% 0% 0% 0% 0% 13
40-60 0% 21% 58% 16% 5% 0% 0% 0% 0% 0% 0% 19
61-80 0% 21% 45% 30% 4% 0% 0% 0% 0% 0% 0% 76
81-120 0% 1% 20% 66% 12% 1% 0% 0% 0% 0% 0% 338
121-160 0% 0% 1% 36% 54% 7% 1% 0% 0% 0% 0% 511
161-200 0% 0% 0% 4% 40% 48% 6% 1% 0% 0% 0% 596
201-250 0% 0% 0% 1% 9% 44% 41% 5% 0% 0% 0% 658
251-300 0% 0% 0% 0% 2% 7% 50% 36% 3% 0% 2% 432
301-350 0% 0% 0% 0% 0% 2% 18% 59% 21% 0% 0% 206
351-400 0% 0% 0% 0% 0% 0% 3% 28% 50% 16% 2% 86
> 400
0% 0% 0% 0% 0% 0% 1% 14% 41% 36% 7% 70