User's Instructions

solutions, may influence the wear of a device or equipment.
1. Soak a clean cloth in the disinfectant solution. Squeeze the cloth thoroughly to remove all excess
solution.
2. Thoroughly wipe down all surfaces (including crevices) of the remote keypad ensuring that a slightly
wet surface is maintained for five minutes.
3. Sterile gauze or a clean, soft cloth moistened with purified water should be used to remove the
disinfectant from the device. Repeat four additional times to thoroughly remove residue.
4. Dry the remote keypad with a clean, soft cloth and ensure the cloth is inserted into the crevices of the
device.
Disinfecting the Remote Keypad: Option 2
Using Process Chemical Super Sani-Cloth (Manufacturer: PDI):
1. Use the specified wipe to thoroughly wipe down the device, including crevices, ensuring that a wet
surface is maintained for 2 minutes.
Note: Pay special attention to crevices.
Note: Additional wipes may be used to maintain the 2 minute contact time.
2. Allow the keypad to air dry.
NOTE: In the applied cleaning and disinfection process, ensure that detergent and disinfectant residuals
are sufficiently removed. Purified or deionized water should be used during final rinsing
processes; where applicable. (Multiple rinses may be required.) Refer to the manufacturer’s
recommendations for the removal of disinfectant residuals.