User Manual

Motion Features
Product Reference Guide
297
Motion Features
Motionless Timeout
This setting specifies the amount of time that the reader takes to assume
that it is in a motionless condition. The range for this setting is from 500
msec to 25.5 seconds, in 100 millisecond increments.
Follow these instructions to set this feature.
1. Determine the desired setting.
2. Pad the result with leading zeroes to yi
eld three digits. For example: 0.5
= 0005 = 00, 5 = 050, 20 = 200, etc.
3. Go to page 262 and scan the ENTER/EXIT PROGRAMMING MODE bar
code to enter Programming Mode.
4. Scan the bar code: SELECT MOT
IONLESS TIMEOUT SETTING.
5. Scan the appropriate three digits from the keypad in Appendix D, Key-
pad representing the duration which was determined in the steps
above. You will hear a two-beep indic
ation after the last character.
If you make a mistake before the last character, scan the CANCEL bar
code to abort and not save the entry string. You can then start again
at the beginning.
6. Scan the ENTER/EXIT PROGRAMMING MODE bar code to exit Program-
ming Mode.
This completes the procedure. See Table 24 for examples of how to set this
feature.
Table 24. Motionless Timeout Setting Examples
STEP ACTION
EXAMPLES
1 Desired Setting
500ms 1 sec. 10 sec 9,900ms (9.9 sec.)
2
Divide by 100 (and pad with
le
ading zeroes to yield two
digits)
005 010 100 250
3 Scan ENTER/EXIT PROGRAMMING MODE
4 Scan SELECT MOTIONLESS TIMEOUT SETTING
5
Scan Two Characters From
Appendix D, Keypad
‘0’, ‘0’ and ‘5’ ‘0’, ‘1’ and ‘0’ ‘1’, ‘0’, and ‘0’ ‘2’, ‘5’, and ‘0
6 Scan ENTER/EXIT PROGRAMMING MODE