Cover TM For use with List Numbers 12348 and 12618 Technical Service Manual Hospira, Inc.
This page intentionally left blank.
Change History Part Number Description of Change 430-95424-008 (A, 2014-08) Eighth Issue Update front and back covers, table of contents, and lists of figures and tables. Make various clerical corrections. Section 5 - Add instruction to Section 5.2.4, Door Roller Inspection and Test - Update Section 5.3.5, Unrestricted Flow Test - Update Section 5.3.9, Keypad Lockout Switch Test - Relocate Figure 5-52 and step in Section 5.3.13, Distal Occlusion Test Section 7 - Add new procedure: “Section 7.2.14.8.
CHANGE HISTORY This page intentionally left blank.
CONTENTS Contents SECTION 1 INTRODUCTION . . . . . . . . . . . 1.1 SCOPE . . . . . . . . . . . . 1.2 CONVENTIONS . . . . . . . . . 1.3 COMPONENT DESIGNATORS . . . 1.4 ACRONYMS AND ABBREVIATIONS . 1.5 USER QUALIFICATION . . . . . . 1.6 ARTIFACTS . . . . . . . . . . 1.7 INFUSION SYSTEM INSTALLATION . 1.7.1 UNPACKING . . . . . . . . 1.7.2 INSPECTION . . . . . . . . 1.7.3 SELF TEST . . . . . . . . 1.8 BIOMED SETTINGS . . . . . . . 1.8.1 IV PARAMETERS . . . . . . 1.8.2 ALARMS LOG . . . . . . . 1.8.
CONTENTS 4.2.2 POWER SUPPLY SUBSYSTEM . . . . . . . . . . . . . . . . 4.2.2.1 MAIN SWITCHING REGULATOR. . . . . . . . . . . . 4.2.2.2 MAIN REGULATOR FAULT DETECTION . . . . . . . . . 4.2.2.3 SYSTEM POWER . . . . . . . . . . . . . . . . . . 4.2.2.4 AUXILIARY SUPPLIES . . . . . . . . . . . . . . . 4.2.2.5 POWER CONTROL . . . . . . . . . . . . . . . . . 4.2.2.6 BATTERY VOLTAGE MEASUREMENT . . . . . . . . . 4.2.2.7 BATTERY CHARGE/DISCHARGE CURRENT MEASUREMENT . 4.2.2.8 BATTERY CHARGER . . . . . . . . . . . . .
CONTENTS 5.2.12 DISPLAY AND INDICATORS INSPECTION . . . . . 5.2.13 KEYPAD LOCKOUT SWITCH INSPECTION . . . . 5.2.14 PREVENTIVE MAINTENANCE CHECKLIST . . . . 5.3 PERFORMANCE VERIFICATION TEST . . . . . . . . 5.3.1 PVT EQUIPMENT LIST . . . . . . . . . . . . 5.3.2 TEST SETUPS . . . . . . . . . . . . . . . 5.3.2.1 BASIC TEST SETUP . . . . . . . . . . 5.3.2.2 PROXIMAL AIR-IN-LINE TEST SETUP . . . 5.3.2.3 DISTAL AIR-IN-LINE TEST SETUP . . . . 5.3.2.4 PRIMING A RUN-IN CASSETTE ASSEMBLY . 5.3.2.
CONTENTS 7.2.10 FRONT/REAR ENCLOSURE GASKET REPLACEMENT . . . . 7.2.11 LOWER FRONT ENCLOSURE GASKET REPLACEMENT . . . 7.2.11.1 EMI GASKET REPLACEMENT. . . . . . . . . . 7.2.11.2 KEYPAD GASKET REPLACEMENT . . . . . . . . 7.2.11.3 TOP SEAL GASKET REPLACEMENT . . . . . . . 7.2.12 REAR ENCLOSURE ASSEMBLY COMPONENT REPLACEMENT . 7.2.12.1 POLE CLAMP ASSEMBLY AND BACKING PLATE REPLACEMENT . . . . . . . . . . . . . . . 7.2.12.2 INTERNAL AC POWER CORD REPLACEMENT . . . 7.2.12.3 AC CONNECTOR REPLACEMENT . . . . . .
CONTENTS Figures Figure 1-1 Figure 1-2 Figure 1-3 Figure 1-4 Figure 1-5 Figure 1-6 Figure 1-7 Figure 4-1 Figure 4-2 Figure 4-3 Figure 4-4 Figure 4-5 Figure 4-6 Figure 4-7 Figure 4-8 Figure 4-9 Figure 5-1 Figure 5-2 Figure 5-3 Figure 5-4 Figure 5-5 Figure 5-6 Figure 5-7 Figure 5-8 Figure 5-9 Figure 5-10 Figure 5-11 Figure 5-12 Figure 5-13 Figure 5-14 Figure 5-15 Figure 5-16 Figure 5-17 Figure 5-18 Figure 5-19 Figure 5-20 Figure 5-21 Figure 5-22 Figure 5-23 Figure 5-24 Figure 5-25 Figure 5-26 Figure 5-27 Fig
CONTENTS Figure 5-40 Figure 5-41 Figure 5-42 Figure 5-43 Figure 5-44 Figure 5-45 Figure 5-46 Figure 5-47 Figure 5-48 Figure 5-49 Figure 5-50 Figure 5-51 Figure 5-52 Figure 5-53 Figure 5-54 Figure 5-55 Figure 6-1 Figure 7-1 Figure 7-2 Figure 7-3 Figure 7-4 Figure 7-5 Figure 7-6 Figure 7-7 Figure 7-8 Figure 7-9 Figure 7-10 Figure 7-11 Figure 7-11 Figure 7-12 Figure 7-13 Figure 7-13 Figure 7-14 Figure 7-15 Figure 7-16 Figure 7-17 Figure 7-18 Figure 7-19 Figure 9-1 Figure 9-2 Figure 9-3 Figure 9-4 Figure 9-5 F
CONTENTS Tables Table 1-1 Table 1-2 Table 4-1 Table 4-2 Table 4-3 Table 4-4 Table 4-5 Table 4-6 Table 4-7 Table 4-8 Table 4-9 Table 4-10 Table 5-1 Table 5-2 Table 6-1 Table 6-2 Table 6-3 Table 6-4 Table 9-1 Table 9-2 Table A-1 Table A-2 Table A-3 Table A-4 Conventions . . . . . . . . . . . . . . . . . . . . . . System Configuration Data . . . . . . . . . . . . . . . . Analog Inputs . . . . . . . . . . . . . . . . . . . . . Keypad Map . . . . . . . . . . . . . . . . . . . . . CPU-Power Supply Interface . .
CONTENTS This page intentionally left blank.
Section 1 INTRODUCTION The Plum A+™ 3 infusion system is designed to meet the growing demand for hospital wide device standardization, and serves a wide range of general floor and critical care needs. The infusion system consists of three component infusers, designated line 1, line 2, and line 3. By incorporating three lines into one unit, the Plum A+3 provides three primary lines, three secondary lines, and piggyback fluid delivery capabilities.
SECTION 1 INTRODUCTION 1.2 CONVENTIONS The conventions listed in Table 1-1 are used throughout this manual. Table 1-1. Convention Conventions Application Example Blue italic Reference to a section, figure, table, or website (see Section 6.1) [ALL CAPS] in brackets Touchswitches, keys, buttons [START] ALL CAPS Bold Screens and displayed messages CASSETTE TEST IN PROGRESS Red Bold Warnings and Cautions CAUTION: Use proper ESD grounding techniques when handling components.
1.4 ACRONYMS AND ABBREVIATIONS 1.
SECTION 1 INTRODUCTION LED Light emitting diode L/S Line select MB Megabyte MHz Megahertz min Minute mL Milliliter mL/hr Milliliter per hour MMIO Memory-mapped input/output MOSFET Metal-oxide semiconductor field-effect transistor ms Millisecond nF nanofarad Op-amp Operational amplifier pF picofarad PROM Programmable read-only memory PVT Performance verification test PWA Printed wiring assembly PWM Pulse width modulator RAM Random-access memory rms Root-mean-square RTC Real-time clock SCC Serial communicati
1.5 USER QUALIFICATION 1.5 USER QUALIFICATION The Plum A+3 must be used at the direction of or under the supervision of licensed physicians or certified healthcare professionals who are trained in the use of the infusion system and the administration of parenteral and enteral fluids and drugs, and whole blood or red blood cell components. Training should emphasize preventing related IV complications, including appropriate precautions to prevent accidental infusion of air.
SECTION 1 INTRODUCTION 1.7.1 UNPACKING Inspect the shipping container as detailed in Section 1.7.2. Use care when unpacking the infusion system. Retain the packing slip and save all packing materials in the event it is necessary to return the infusion system to the factory. Verify the shipping container contains a copy of the System Operating Manual. 1.7.2 INSPECTION Inspect the shipping container for damage. Should any damage be found, contact the delivering carrier immediately.
1.7 INFUSION SYSTEM INSTALLATION 3. The LCD screen briefly displays the SELF TEST screen (see Figure 1-1). If the SELF TEST screen does not appear, contact Hospira. 4. After the self test is complete, the message INSERT PLUM SET CLOSE LEVER appears. 5. Verify the time and date. To set the time and date, see Section 1.8.3. Open the cassette door and insert a primed cassette. Close the cassette door. The cassette test is complete when the CASSETTE TEST IN PROGRESS message disappears.
SECTION 1 INTRODUCTION 1.8 BIOMED SETTINGS The BIOMED SETTINGS screens contain the following options that can be changed or reviewed by qualified personnel: - IV screen parameters - Alarms log - Set time and date All infusers (new or refurbished) are shipped with factory settings (see Table 1-2). Note: Biomed screens do not time out for the Infuser Idle alarm or No Action alarm. Note: The battery will not be detected in the Biomed service mode. To access the Biomed settings, proceed as follows: 1.
1.8 Table 1-2. BIOMED SETTINGS System Configuration Data Data Options Range Factory Setting Maximum macro IV mode delivery rate 0.1 - 99.
SECTION 1 INTRODUCTION 1.8.1 IV PARAMETERS The IV Parameters screen contains Common IV Parameters and Macro IV Parameters (see Figure 1-3). To change the IV parameters see Figure 1-4 and Figure 1-5, then proceed as follows: 1. Access the BIOMED SETTINGS screen as described in Section 1.8. 2. Select IV Screen Parameters, and press [CHOOSE]. 3. Select the parameters to be changed, and press [CHOOSE]. 4. Using the [CHANGE VALUE] softkey, select the desired value, and press [ENTER]. 5.
1.8 BIOMED SETTINGS BIOMED SETTINGS Common IV Parameters Continue Rate KVO Deliver Together Concurrent Enable Delay/Standby Yes Callback Default No Select using Change Value Change Value Figure 1-4. Enter Cancel/ Back Common IV Parameters BIOMED SETTINGS Macro IV Parameters 6.0 psi Default Distal Press Max Rate 999 mL/hr Enter Value using keypad Enter Figure 1-5.
SECTION 1 INTRODUCTION 1.8.2 ALARMS LOG The Alarms Log will retain the latest 40 alarm and malfunction codes, listed in order from the most current to the oldest. To view the Alarms Log see Figure 1-6, then proceed as follows: 1. Access the BIOMED SETTINGS screen as described in Section 1.8. 2. Select Alarms Log, and press [CHOOSE]. Use the [PAGE UP] and [PAGE DOWN] softkeys to view the Alarms Log. 3. Press [BACK] to exit the Alarms Log and return to the main BIOMED SETTINGS screen.
1.8 BIOMED SETTINGS 1.8.3 SETTING THE TIME AND DATE Note: The infusion system will automatically display February 29 on leap year. Note: Daylight savings and time zone changes must be made manually. To set the time and date, see Figure 1-7, then proceed as follows: 1. Access the BIOMED SETTINGS screen as described in Section 1.8. 2. Select Set Time and Date, and press [CHOOSE]. 3. Select the parameter to be changed, then enter the desired value. 4.
SECTION 1 INTRODUCTION This page intentionally left blank.
Section 2 WARRANTY Subject to the terms and conditions herein, Hospira, Inc., hereinafter referred to as Hospira, warrants that (a) the product shall conform to Hospira's standard specifications and be free from defects in material and workmanship under normal use and service for a period of one year after purchase, and (b) the replaceable battery shall be free from defects in material and workmanship under normal use and service for a period of 90 days after purchase.
SECTION 2 WARRANTY This page intentionally left blank.
Section 3 SYSTEM OPERATING MANUAL A copy of the System Operating Manual is included with every Plum A+3 infusion system. If a copy is not available, contact Hospira (see Section 6.1).
SECTION 3 SYSTEM OPERATING MANUAL This page intentionally left blank.
Section 4 THEORY OF OPERATION This section describes the Plum A+3 theory of operation. The theory of operation details the general description, electronic subsystem overview, printed wiring assemblies, remote mounted peripherals, and mechanical overview of the infusion system. 4.
SECTION 4 THEORY OF OPERATION 4.2 ELECTRONIC SUBSYSTEM OVERVIEW This section describes the function and electronic circuitry of the following three main subsystems in the infusion system: CPU subsystem, power supply subsystem, and mechanism subsystem (see Figure 4-1). Note: An asterisk (*) denotes an active low or negative true logic signal.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW 4.2.1 CPU SUBSYSTEM The CPU subsystem contains the main microcontroller that is responsible for controlling the display/keyboard interface, external communications interfaces, barcode reader (BCR) interface, and system management.
SECTION 4 THEORY OF OPERATION 4.2.1.3 PROGRAMMABLE READ-ONLY MEMORY The CPU subsystem has two 512 K x 8 bit programmable read-only memory (PROM) memory devices that provide a total of 1024 KB. The PROM space is expandable up to 2 MB. The PROM memory devices operate off the 3.3 VDC supply. The CPU chip-select 0 pin (CS0*), is connected to the PROM chip-enable (CE*) pin (signal CSROM*). This special chip-select signal can support bootstrap operation after reset.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW 4.2.1.7 LCD BACKLIGHT CONTROL The LCD panel is backlit by a cold cathode fluorescent tube (CCFT) lamp. The CCFT lamp requires 300 Vrms to operate; a current controlled DC-to-AC voltage inverter circuit is used to deliver a current regulated sine wave to the lamp. A switching regulator regulates the CCFT current by monitoring feedback pin 3, and varies its output duty cycle to drive a DC/AC inverter.
SECTION 4 THEORY OF OPERATION 4.2.1.10 VOLTAGE MONITOR WATCHDOG TIMER It is important to protect the system during power transitions, and the CPU is reset after the VCC power supply is applied. The microprocessor supervisory circuit generates an automatic reset output during power up, power down, or brownout conditions. When the VCC falls below the reset threshold voltage of 2.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW U4 CPU I/O CS* A0 A1 I/O CLOCK SPCLK ADDRESS SPTXD DATA OUT SPRXD A2 SCP A3 A4 ANALOG INPUTS A5 EOC I/O (OR INTERRUPT) A6 A7 A8 REF+ 2.5V A9 REFA10 GND GANA GDIG Figure 4-2. Serial Interface to ADC Table 4-1.
SECTION 4 THEORY OF OPERATION 4.2.1.12 DIGITAL-TO-ANALOG CONVERTER The dual 8 bit digital-to-analog converter (DAC) generates two analog signals to control the phase A and phase B motor coil currents. The interface between the DAC device and the CPU is the 8 bit data bus that is connected to DATA15:8. All the control signals for this DAC are generated by FPGA logic devices. Buffer amplifier/ground compensation circuits condition the DAC outputs. 4.2.1.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW 4.2.1.17 NURSE CALL INTERFACE A nurse call relay switch on the peripheral interface PWA indicates alarm conditions to a remote operator. A memory-mapped output signal (NURSE) activates the relay during alarm conditions. The relay has both normally-open and normally-closed contacts. A jumper on the peripheral interface board selects the contact type. The factory setting is normally-open. 4.2.1.18 AUDIBLE INDICATORS There are two audible indicators on the CPU subsystem.
SECTION 4 THEORY OF OPERATION The DataPort system conforms to the EIA-232-D standard, with the following exceptions: - DataPort uses non-standard DB-15 and 6 pin modular connectors in addition to the standard DB-25 and DB-9 connectors - With DataPort, more than one infuser is allowed on the line - The minimum line impedance is 2 K (EIA-232-D standard: 3 K min.) - The maximum line impedance is 30 K (EIA-232-D standard: 7 K max.
4.2 Table 4-3. ELECTRONIC SUBSYSTEM OVERVIEW CPU-Power Supply Interface Signal Name Type IBATT A, I Voltage proportional to integration of battery charge/ discharge current VBATT A, I Divided battery terminal voltage CHG* D, O Battery charger enable VFLOAT* D, O Set the main regulator voltage to battery float charge level ITGRST D, O Reset the charge current integrator Legend: Description P = Power A = Analog D = Digital I = Input O = Output 4.2.1.
SECTION 4 THEORY OF OPERATION Table 4-4.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW 4.2.2 POWER SUPPLY SUBSYSTEM The power supply subsystem provides DC power to system circuits and interface software controlled power and battery management.
SECTION 4 THEORY OF OPERATION 4.2.2.1.3 Current Limit Loop The current limit loop is activated when the primary current, sensed by R71, exceeds 3 A. Resistor R70 and capacitor C46 filter the voltage across R71 and feed it back to the current sense input (1.5 VDC threshold) of U12. The duty cycle of U12 is reduced until the excessive load is removed. 4.2.2.2 MAIN REGULATOR FAULT DETECTION If the switching regulator’s main loop fails, the secondary voltage limit loop takes over.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW The infuser is activated when the [ON/OFF] key is pressed or the AC line is plugged in. The [ON/OFF] key activates the STRTUP signal, triggering a three second one-shot circuit that will temporarily turn the system power on. This three second one-shot period allows the CPU enough time to power up, initialize, and turn on the PWRHLD signal. The CPU monitors the STRTUP signal, via interrupt, to signal a user request for turning off the infuser.
SECTION 4 THEORY OF OPERATION 4.2.2.8 BATTERY CHARGER The software battery management algorithm controls the battery charger. The charging scheme is a current limit/two stage voltage limit charger. The charge current is limited to 1.3 A and the voltage is limited to either 6.9 VDC or 7.5 VDC. The source of the charge current is power MOSFET transistor Q7 operating in the linear mode.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW 4.2.3 MECHANISM SUBSYSTEM The mechanism subsystem includes the electronics and electromechanical components that interface with the pumping mechanism.
SECTION 4 THEORY OF OPERATION The unipolar motor windings have a center tap connected on each of the two coils as shown in Figure 4-4. Unidirectional current enters the center tap and is steered to one end of the coil or the other end by the driver electronics, creating positive or negative flux lines in the motor coil. With two coils each with a choice of flux polarity, four electrical combinations or phases are possible. A ACOM A B Figure 4-4. BCOM B Stepper Motor Coils 4.2.3.1.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW Each optical switch consists of an infrared LED that shines through a rectangular aperture, across a slot, to illuminate a photo-transistor. The photo-transistor is activated as long as the beam is on and not blocked (by a flag in the slot). The optical switches are distributed throughout the mechanism, near their associated flags. The motor rotational optical switches are mounted on the driver PWA along with the control circuitry.
SECTION 4 THEORY OF OPERATION 4.2.3.4 AIR SENSORS The mechanism subsystem includes two air sensors, used to detect air passage into (proximal) or out of (distal) the cassette. Both sensors are piezoelectric crystal transmitter receiver pairs. Liquid between the transmitter and receiver will conduct the ultrasonic signal, while air will not (see Figure 4-5). XTL TX VSO VCO/PLL CPU Figure 4-5. G_RX G_TX OUTPUT AMP A/D XTL RX PEAK DETECTOR Air Sensor Block Diagram 4.2.3.4.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW 4.2.3.4.2 Receiver Circuitry When the cassette’s test port is filled with fluid, the transmitted signal will be coupled to an identical piezoelectric crystal, where it is amplified and detected by the receiver circuitry. The receiver circuitry consists of an amplifier, a peak detector, and an adjustable gain buffer stage. There is a separate, symmetrical receiver circuit for each channel (proximal and distal).
SECTION 4 THEORY OF OPERATION PRESSURE SENSOR ENABLE (FROM CPU) BRIDGE EXCITATION 3.75 V WHEATSTONE BRIDGE REFERENCE 2.5 V DIFFERENTIAL AMPLIFIER AND OFFSET ADJUST OUTPUT AMPLIFIER AND FILTER Figure 4-6. PRESSURE SIGNAL OUTPUT (TO CPU) Pressure Sensor Excitation and Amplifier Block Diagram 4.2.3.5.1 Bridge Excitation Supply The bridge excitation voltage is 3.75 VDC, and is derived from the 2.5 VDC reference signal (V2_5), gained 1.5 times by an amplifier.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW 4.2.3.6 PRESSURE SENSOR CALIBRATION Pressure sensors are calibrated for offset and gain during mechanism calibration. A trimming potentiometer is used to adjust the initial, zero pressure offset. The proximal and distal pressure sensors have independent offset adjustments. The final system gain (cassette pressure to corrected amplifier output) is adjusted in software.
SECTION 4 THEORY OF OPERATION 4.3 PRINTED WIRING ASSEMBLIES Infusion system electronics are packaged into six printed wiring assemblies (PWA) and several remote mounted peripherals (see Section 4.4). The following sections provide a brief description of the functional interfaces of each PWA. 4.3.
4.3 PRINTED WIRING ASSEMBLIES 4.3.2 PERIPHERAL PWA The peripheral PWA contains part of the CPU subsystem circuitry, including system program and data memories (PROM and SRAM), and external communication interface circuits. The peripheral PWA is designed to be field replaceable, to facilitate software upgrades or additional external interfaces. The peripheral PWA is a four layer board that includes the following: one ground plane, one power plane, and two signal layers.
SECTION 4 THEORY OF OPERATION 4.3.4 CPU PWA The CPU PWA contains most of the CPU subsystem functions, with the exception of main memory and communications ports that are located on the peripheral PWA. The CPU PWA also accommodates system interconnect. The CPU PWA is an eight layer board with the following: one ground plane, one power plane, and six signal layers. The CPU PWA primarily contains SMT components.
4.3 PRINTED WIRING ASSEMBLIES See Table 4-9 for driver PWA interface connections. Table 4-9. Connector Driver PWA Interface Connections Type Interface J7 6 pin header Plunger motor J8 6 pin header Input/output motor J9 6 pin header Line select motor J10 20 pin SMT Flat flex cable to APP PWA J11 50 pin header Ribbon cable to CPU PWA J12 6 pin SMT FSR flex circuit J13 4 pin header Motor power, from power supply PWA J14 8 pin SMT Flat flex cable to switch PWA 4.3.
SECTION 4 THEORY OF OPERATION 4.4 REMOTE MOUNTED PERIPHERALS Remote mounted peripherals include the LCD and barcode reader (BCR) wand. 4.4.1 LCD The infuser uses a graphic LCD module with a CCFT. The CCFT provides a backlight source for the LCD. The LCD requires a nominal -16 VDC supply for contrast control that is controlled by the CPU. The infuser’s graphic display data is shifted out to the LCD by the CPU LCD controller that interfaces directly with the CPU (see Section 4.2.1.6).
4.5 MECHANICAL OVERVIEW 4.5.1 CASSETTE The cassette (see Figure 4-7 and Figure 4-8) operates on a fluid displacement principle to volumetrically deliver fluid. See the system operating manual for a description of the major cassette functions. The pumping cycle begins when the outlet valve is opened and the inlet valve is closed. The plunger extends to deflect the cassette diaphragm and expel fluid.
SECTION 4 THEORY OF OPERATION SECONDARY PORT (Y-RESEAL OR LOCKING CAP) FROM PRIMARY CONTAINER PRECISION GRAVITY FLOW REGULATOR (CONTROL NOT SHOWN) FINGER GRIP B VALVE A VALVE AIR-IN-LINE SENSOR (PROXIMAL) AIR-IN-LINE SENSOR (DISTAL) PRESSURE SENSOR (PROXIMAL) PRESSURE SENSOR (DISTAL) PUMPING CHAMBER INLET VALVE RIGHT VIEW LEFT VIEW REAR VIEW Figure 4-7.
4.5 MECHANICAL OVERVIEW 4.5.2 MECHANISM ASSEMBLY The mechanism assembly is a fully self-contained unit consisting of the motor and valve assemblies, A/B valve subsystem, inlet/outlet valve subsystem, plunger drive subsystem, air bubble (ultrasonic) sensor assemblies, cassette door, and pressure sensor assemblies. The motor and valve assemblies, A/B valve subsystem, inlet/outlet valve subsystem, and plunger drive subsystem are detailed in the following sections.
SECTION 4 THEORY OF OPERATION B VALVE A VALVE REGULATOR ACTUATOR AIR-IN-LINE SENSOR (DISTAL) AIR-IN-LINE SENSOR (PROXIMAL) PRESSURE SENSOR (DISTAL) PRESSURE SENSOR (PROXIMAL) CASSETTE LOCATOR FORCE SENSING RESISTOR OUTLET VALVE PLUNGER Figure 4-9. INLET VALVE Mechanism Valve Pins and Sensor Locations 4.5.2.3 INLET/OUTLET VALVE SUBSYSTEM The inlet/outlet valve subsystem is similar in function and build to the A/B valve subsystem (see Section 4.5.2.2). 4.5.2.
Section 5 MAINTENANCE AND SERVICE TESTS A complete maintenance program promotes infusion system longevity and trouble-free operation. Such a program should include cleaning, Preventive Maintenance, and the Performance Verification Test (PVT). 5.1 CLEANING AND SANITIZING As a minimum requirement, inspect and clean the infuser after each use. In addition, establish a regular cleaning schedule for the device. Practice the cleaning and sanitizing guidelines in this section.
SECTION 5 MAINTENANCE AND SERVICE TESTS Table 5-1. Cleaning Solution Cleaning Solutions Manufacturer Preparation DispatchTM Hospital Cleaner Disinfectant with Bleach The Clorox Company Per manufacturer's recommendation PreciseTM Hospital Foam Cleaner Disinfectant The Clorox Company Per manufacturer's recommendation Sani-ClothTM HB Wipe Professional Disposables, Inc. Per manufacturer's recommendation Sani-ClothTM Bleach Wipe Professional Disposables, Inc.
5.2 PREVENTIVE MAINTENANCE 5.2 PREVENTIVE MAINTENANCE The Plum A+3 is three infusers in a single housing. This Preventive Maintenance section describes the inspections and tests that must be performed at least once every 12 months. The inspections and tests are described for one infuser system. The preventive maintenance process must be performed three times - once for each infuser system.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.2.1 LABELS INSPECTION Visually inspect the infuser labels at least once every 12 months. To inspect the labels, see Figure 5-1, and proceed as follows: 1. Place the infuser on a flat, stable surface. 2.
5.2 TOP NOTE LABEL PREVENTIVE MAINTENANCE CAUTION LABEL OPERATING INSTRUCTIONS LABEL CLAMP WARNING LABEL POLE CLAMP LABEL SWITCHPORT LABEL REAR CAUTION LABEL PRODUCT I.D. LABEL SWITCHPORT DESCRIPTION LABEL SERVICE REVISION LEVEL LABEL (2) MAC ADDRESS LABELS (WIRELESS) BATTERY LABEL (3x) CLOSE LEVER LABEL (3x) (LEVER DOOR OPEN) LOGO LABEL (3x) CLOSE LEVER w/ARROW LABEL (3x) Figure 5-1.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.2.2 AC POWER CORD, RETAINER, AND VELCRO STRAP INSPECTION Inspect the power cord, retainer, and Velcro strap at least once every 12 months. The recommended tool for this inspection is a #2 Phillips screwdriver. To inspect the power cord and retainer, see Figure 5-2, and proceed as follows: 1. Turn off the infuser and disconnect the device from AC power. 2. Using the Phillips screwdriver, remove the screw that holds the power cord retainer in place.
5.2 PREVENTIVE MAINTENANCE 5.2.3 FRONT ENCLOSURE AND REAR ENCLOSURE INSPECTION Visually inspect the front enclosures and rear enclosure at least once every 12 months. To inspect the front and rear enclosures, see Figure 5-3, and proceed as follows: 1. Place the infuser on a flat, stable surface. 2. Inspect the front and rear enclosures for cracks, chips, and gouges. If an enclosure is cracked, or has a significant chip or gouge, replace the damaged part (see Section 7.2.6 and Section 7.2.12). 3.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.2.4 DOOR LEVER ASSEMBLY, DOOR ROLLER, DOOR INSPECTION, AND TEST Inspect and test each door lever, door roller, and door at least once every 12 months. To inspect and test each door lever and door, proceed as follows: 1. Inspect the door lever for cracks and damage. Replace the door lever if it is damaged (see Section 7.2.14.11). 2. Move the door lever to the OPEN position. Confirm that the door opens smoothly.
5.2 PREVENTIVE MAINTENANCE 5.2.5 FLUID SHIELD INSPECTION Visually inspect the Plum A+3 fluid shields at least once every 12 months. Equipment required for the fluid shield inspection is a 0.025 inch (0.65 mm) feeler gauge (plastic or metal), To inspect each fluid shield, see Figure 5-5, Figure 5-6, and Figure 5-7, and proceed as follows: 1. Place the infuser on a flat, stable surface. 2. Press the door release tab to the right and open the cassette door all the way so that it lays flat (see Figure 5-5).
SECTION 5 MAINTENANCE AND SERVICE TESTS GAP GAP FEELER GAUGE Figure 5-6. Inspecting the Fluid Shield with Feeler Gauge A VALVE PIN B VALVE PIN CASSETTE PRESENCE DETECTOR OUTLET VALVE INLET VALVE Figure 5-7.
5.2 PREVENTIVE MAINTENANCE 5.2.6 DISTAL PRESSURE PIN INSPECTION Visually inspect the distal pressure pins at least once every 12 months, and each time the infuser is serviced. Note: The distal pressure pin is the black pin (see Figure 5-8). To inspect the distal pressure pin, proceed as follows: 1. Place the infuser on a flat, stable surface. 2. Release the door so that it lays flat. Press the door release tab to the right and open the cassette door all the way (see Figure 5-5). 3.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.2.7 PROXIMAL PRESSURE PIN INSPECTION Visually inspect the proximal pressure pins at least once every 12 months and each time the infuser is serviced. Note: The proximal pressure pin is the white pin (see Figure 5-9). To inspect the proximal pressure pin, proceed as follows: 1. Place the infuser on a flat, stable surface. 2. Release the door so that it lays flat. Press the door release tab to the right and open the cassette door all the way (see Figure 5-5).
5.2 PREVENTIVE MAINTENANCE 5.2.8 RUBBER FOOT PAD INSPECTION Perform a visual inspection of the rubber foot pads at least once every 12 months. The recommended tool for this inspection is a #2 Phillips screwdriver. To inspect the rubber foot pads, see Figure 5-10, and proceed as follows: 1. Rotate the infuser to access the bottom as shown in Figure 5-10. 2. Inspect for missing, loose, or damaged rubber foot pads. Replace foot pads as described in Section 7.2.3.
SECTION 5 MAINTENANCE AND SERVICE TESTS Note: The A+3 infuser utilizes a heavy duty clamp that has a slide which is driven by the lead screw to clamp the pole. 5. Mount the infuser on a Hospira-approved IV pole and fully tighten the clamp. Ensure that the infuser is held firmly and does not slide on the IV pole. If the clamp does not hold the infuser securely, replace the pole clamp assembly (see Section 7.2.12.1). RUBBER PAD Clamp slides to grip the pole Figure 5-11. Pole Clamp Assembly 5.2.
5.2 PREVENTIVE MAINTENANCE 3. Carefully pull the battery and wire harness assembly out of the enclosure and disconnect it from the infuser’s internal wiring at the inline connector. 4. Inspect the battery compartment for any debris. If debris is present, wipe or brush the debris out of the compartment. 5. Inspect the battery door and replace, if damaged or cracked. 6. Inspect the battery door pad on the battery door to ensure the pad is attached and is not damaged.
SECTION 5 MAINTENANCE AND SERVICE TESTS BATTERY DOOR (3) Figure 5-12. Battery Doors BATTERY WITH WIRE HARNESS ASSEMBLY (3) DOOR PAD (3) RETAINING RING (3) BATTERY DOOR (3) Figure 5-13.
5.2 PREVENTIVE MAINTENANCE 5.2.11 KEYPAD INSPECTION Visually inspect the keypads at least once every 12 months. To inspect each keypad, see Figure 5-14, and proceed as follows: 1. Place the infuser on a flat, stable surface. 2. Inspect each keypad for tears, cracks, or edges lifting away from the infuser. If damage is observed, replace the keypad (see Section 7.2.14.2). 3. Inspect each keypad for worn or illegible numbers or letters. If letters or numbers are not readable, replace the keypad. 4.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.2.12 DISPLAY AND INDICATORS INSPECTION Visually inspect the displays and indicators at least once every 12 months. To inspect each display and its indicators, see Figure 5-15, and proceed as follows: 1. Place the infuser on a flat, stable surface. 2. Connect the power cord to the mains supply, and confirm that the CHARGE LED is lit. 3. Rotate the infuser so that the rear of the device is facing to the front. 4.
5.2 PREVENTIVE MAINTENANCE LINE FLOW LEDs Back Prime A B Options/ Vol Inf CHARGE LED Figure 5-15.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.2.13 KEYPAD LOCKOUT SWITCH INSPECTION Inspect the keypad lockout switch at least once every 12 months. The keypad lockout switch is located on the CE module on the rear of the infuser. To inspect the keypad lockout switch, see Figure 5-16, and proceed as follows: 1. Inspect for the presence of the keypad lockout switch and ensure the switch is not broken. 2. Inspect for a loose or dislodged switch.
Plum A+3 Infusion System Technical Service Manual PASS / FAIL PASS / FAIL PASS / FAIL PASS / FAIL PASS / FAIL PASS / FAIL PASS / FAIL PASS / FAIL PASS / FAIL PASS / FAIL PASS / FAIL PASS / FAIL PASS / FAIL Distal Pressure Pin Inspection Proximal Pressure Pin Inspection Rubber Foot Pad Inspection Pole Clamp Inspection and Test Battery Inspection and Replacement (indicate replacement below) Keypad Inspection Display and Indicators Inspection Keypad Lockout Switch Inspection 5 - 21 PASS / FAIL PAS
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.3 PERFORMANCE VERIFICATION TEST The Performance Verification Test (PVT) consists of the tests described in this Section. Use the PVT for the following: • Preventive maintenance – Perform the PVT at least once every 12 months as part of Preventive Maintenance, which also includes the visual inspections in Section 5.2. This ensures that the infusion system is operating properly.
5.3 PERFORMANCE VERIFICATION TEST 5.3.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.3.2.1 BASIC TEST SETUP The Basic test setup consists of primed primary and secondary lines attached to fluid bags. The cassette is inserted into the infusion unit being tested and the distal (patient) end of the tubing is placed in a collection container. The Basic test setup is shown in Figure 5-17, with the cassette installed to test infusion unit number 1. SCORE MARK DRIP CHAMBER 12” - 24” Figure 5-17.
5.3 PERFORMANCE VERIFICATION TEST The following tests can be performed using one Basic test setup. Not all tests will use every part of the setup. • Section 5.3.3, Self Test • Section 5.3.5, Unrestricted Flow Test • Section 5.3.6, Display Test • Section 5.3.7, Keypad Verification/Functional Test • Section 5.3.8, Alarm Loudness Test • Section 5.3.9, Keypad Lockout Switch Test • Section 5.3.10, Proximal Occlusion Test • Section 5.3.13, Distal Occlusion Test • Section 5.3.14, Delivery Accuracy Test 5.3.2.1.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.3.2.1.2 Preparing the Primary Line To prepare the primary line, proceed as follows to fill the cassette and tubing on the primary PlumSet with liquid (that is, prime it), eliminating all air, and then load the cassette into the infuser. 1. Place the infuser on a bench or attach it to an IV pole. 2. Press the cassette flow regulator in to ensure it is closed (see Figure 5-18). Figure 5-18. Closing the Flow Regulator 3.
5.3 PERFORMANCE VERIFICATION TEST 5. Suspend the container on an IV pole. 6. Check for leaks. If any part of the container is leaking, replace it. 7. Squeeze the drip chamber to fill it about 1/2 full or to the score mark (see Figure 5-21). Do not completely fill the drip chamber. Figure 5-21. Squeezing the Drip Chamber 8. Invert the cassette so that the secondary port is pointing down (see Figure 5-22). Figure 5-22. Secondary Port 9.
SECTION 5 MAINTENANCE AND SERVICE TESTS 10. Tap and clear air from the cassette, Y-site, and tubing to remove all air from the remainder of the administration set (see Figure 5-24). Figure 5-24. Removing Air from the Administration Set 11. Push in the flow regulator to close it (see Figure 5-25). Check the distal end of the tubing to confirm that there is no flow. Figure 5-25. Closing the Flow Regulator Note: If there is flow or leaks, close all clamps and replace the administration set. 12.
5.3 PERFORMANCE VERIFICATION TEST 2. Grasp the cassette by the finger grip (see Figure 5-27). FINGER GRIP Figure 5-27. Cassette Finger Grip 3. Slide the cassette into the door guide (see Figure 5-28). DOOR GUIDES Figure 5-28. Cassette and Door Guide 4. Press the lever down to close the cassette door. 5. Open all clamps. 6. Check the distal end of the tubing to confirm that there is no flow and that no kinks appear in the tubing.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.3.2.1.4 Preparing the Secondary Line To prepare the secondary line, proceed as follows to prime the line and attach it to the cassette: 1. Insert the piercing pin into the secondary container outlet using a twisting motion (see Figure 5-29). pin. Note: Do not position the container above the infuser while inserting the piercing Figure 5-29. Inserting the Piercing Pin 2. Suspend the container on an IV pole. 3. Check the secondary container for leaks.
5.3 PERFORMANCE VERIFICATION TEST 5. Slowly open the roller clamp to allow fluid to flow into the secondary tubing (see Figure 5-31). TO OPEN CAIR CLAMP OPEN POSITION ROLLER ROLLER Figure 5-31. Opening the Roller Clamp 6. After all air is removed, close the roller clamp (see Figure 5-32). TO CLOSE CAIR CLAMP CLOSED POSITION ROLLER ROLLER Figure 5-32. Closing the Roller Clamp 7.
SECTION 5 MAINTENANCE AND SERVICE TESTS • If the cassette has a capped secondary port: Confirm that the cassette door is closed, and then loosen and remove the white cap. Discard the cap. Insert the end of the secondary line into the port and twist clockwise to secure the line to the port (see Figure 5-34). Figure 5-34.
5.3 PERFORMANCE VERIFICATION TEST The test setup is complete, as shown in Figure 5-35. SCORE MARK DRIP CHAMBER 12” - 24” Figure 5-35.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.3.2.2 PROXIMAL AIR-IN-LINE TEST SETUP This section describes the steps for the proximal air-in-line test setup, including modifying a run-in cassette, priming it, and loading the cassette into the infuser. This setup is used for the proximal air-in-line test in Section 5.3.11. 5.3.2.2.1 Equipment Required for the Proximal Air-in-Line Test Setup • Run-in cassette • Sterile or tap water • X-Acto knife or equivalent • Permanent marker 5.3.2.2.
5.3 PERFORMANCE VERIFICATION TEST 5.3.2.3 DISTAL AIR-IN-LINE TEST SETUP This section describes the steps for the Distal Air-in-Line test setup, including modifying a run-in cassette, priming it, and loading the cassette into the infuser. This setup is used for the Distal Air-in-Line test in Section 5.3.12. 5.3.2.3.1 Equipment Required for the Distal Air-in-Line Test Setup • Run-in Cassette • Sterile or tap water • X-Acto knife or equivalent • Permanent marker 5.3.2.3.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.3.2.4 PRIMING A RUN-IN CASSETTE ASSEMBLY Primed run-in cassettes are required for Proximal Air-in-Line and Distal Air-in-Line tests. The run-in cassette has tubing that is arranged so that fluid is pumped in a continuous loop, as shown in Figure 5-38. CAP TOP PROXIMAL TUBING DRIP CHAMBER FLOW REGULATOR FINGER GRIP PUMPING CHAMBER DISTAL TUBING LOOP BACK OF PUMPING CHAMBER Figure 5-38.
5.3 PERFORMANCE VERIFICATION TEST 5. Press and hold [BACKPRIME] to pump water from the drip chamber into the proximal lines and cassette. 6. When bubbles are no longer being pushed into the drip chamber, release the [BACKPRIME] key. The cassette test will proceed. 7. If the cassette test fails again, repeat Steps 5 and 6. 8. When the cassette test completes with no alarms, replace the white cap. 5.3.2.4.
SECTION 5 MAINTENANCE AND SERVICE TESTS 6. Press in firmly on the pumping chamber to pump air out of the chamber (see Figure 5-40). Figure 5-40. Pressing the Pumping Chamber 7. Continue to press on the pumping chamber as you use your other hand to push the flow regulator closed (see Figure 5-41). This prevents the air from returning to the pumping chamber. Figure 5-41. Preventing Air from Returning to the Pumping Chamber 8. Release the pumping chamber and flow regulator. 9.
5.3 PERFORMANCE VERIFICATION TEST 5.3.2.5 DISTAL OCCLUSION TEST SETUP This section describes the distal occlusion test setup, including adding a stopcock and Digital Pressure Meter (DPM) to the distal tubing on the Basic test setup. The distal occlusion test setup is shown in Figure 5-42. UNUSED PORT ON STOPCOCK CONTROL ARM DISTAL TUBING STOPCOCK ATTACHED TO DPM PORT UNIVERSAL PRESSUREMETER 0.00 cmOFHO 2 INCHESOFHO 2 mmHg -13.5TO15 PSI -13.5TO75 OFF PRESSUREINPUT Figure 5-42.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.3.2.5.2 Setup Procedure The three-way stopcock has two ports that are opposite each other. in addition to a third, perpendicular port for attachment to the DPM, as shown in Figure 5-43. MALE PORT (WITH COVER ATTACHED) FEMALE PORT FEMALE PORT (WITH COVER ATTACHED) CONTROL ARM Figure 5-43. Three-Way Stopcock Ports 1. Remove the protective caps from the stopcock ports. 2.
5.3 PERFORMANCE VERIFICATION TEST 4. Place the DPM connector at a height of 0 12 inches from the midline of the pumping chamber on the cassette (see Figure 5-46). UNIVERSAL PRESSUREMETER 0.00 cmOFHO 2 INCHESOFHO 2 mmHg -13.5TO15 PSI -13.5TO75 OFF +12” PRESSUREINPUT 0” UNIVERSAL PRESSUREMETER 0.00 cmOFHO 2 INCHESOFHO 2 mmHg -13.5TO15 PSI -13.5TO75 OFF -12” Figure 5-46.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.3.3 SELF TEST The self test procedure uses the Basic test setup in Section 5.3.2.1. CAUTION: Do not place the infuser in service if the self test fails. To perform the self test, proceed as follows: 1. Plug the power cord into a grounded AC outlet. Verify that the charge/line indicator is lit and an alarm sounds. 2. Without a cassette installed, press [ON/OFF] to turn on the infuser. The LCD screen briefly displays the SELF TEST screen.
5.3 PERFORMANCE VERIFICATION TEST 5.3.5 UNRESTRICTED FLOW TEST To perform the unrestricted flow test, use the Basic test setup in Section 5.3.2.1, and proceed as follows: 1. Insert the primed cassette into the infuser and close the cassette door. 2. If an Area Selection or CCA Selection screen appears, choose a care area and press [ENTER]. Otherwise, skip this step. 3. With the cassette door closed, check the distal end of the tubing for fluid flow.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.3.7 KEYPAD VERIFICATION/FUNCTIONAL TEST To perform the keypad verification/functional test, use the Basic test setup in Section 5.3.2.1, and proceed as follows: 1. While the infuser displays the DELIVERY screen, press [A] to select Line A. 2. Verify that the PROGRAM screen is displayed. Enter a rate of 123 mL/hr and VTBI of 4567.
5.3 PERFORMANCE VERIFICATION TEST 5.3.8 ALARM LOUDNESS TEST To perform the alarm loudness test, use the Basic test setup in Section 5.3.2.1, and proceed as follows: 1. Press [A] to select Line A. 2. If the message CLEAR LINE A SETTINGS? appears, press [YES]. 3. Enter a rate of 400 mL/hr and VTBI of 1 mL. 4. Press [START]. If a CONFIRM PROGRAM? message appears, confirm that the rate and VTBI are correct, and then press [YES]. 5.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.3.9 KEYPAD LOCKOUT SWITCH TEST To perform the keypad lockout switch test, use the Basic test setup in Section 5.3.2.1, and proceed as follows: 1. Press [A] to select Line A. If the message CLEAR LINE A SETTINGS? appears, press [YES]. 2. Enter a rate of 400 mL/hr and VTBI of 50 mL. 3. Press [START]. If a CONFIRM PROGRAM? message appears, confirm that the rate and VTBI are correct, and then press [YES]. 4.
5.3 PERFORMANCE VERIFICATION TEST 5.3.10 PROXIMAL OCCLUSION TEST To perform the Proximal Occlusion test, use the Basic test setup in Section 5.3.2.1 and the programming from the Keypad Lockout Switch test in Section 5.3.9, and proceed as follows: Note: If performing this section as a standalone test, select Line A and enter a rate of 400 mL/hr and a VTBI of 50 mL. Go to Step 1. 1. Press [START]. If a CONFIRM PROGRAM? message appears, confirm that the rate and VTBI are correct, and then press [YES].
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.3.12 DISTAL AIR-IN-LINE TEST The Distal Air-in-Line test uses the Distal Air-in-Line test setup in Section 5.3.2.3 and the programming from the Proximal Air-in-Line test in Section 5.3.11. prepared Note: If performing this section as a standalone test, insert the test cassette in Section 5.3.2.3, select Line A, and enter a rate of 400 mL/hr and a VTBI of 50 mL. Go to Step 4. 1. Insert the distal test cassette into the infuser and close the cassette door. 2.
5.3 PERFORMANCE VERIFICATION TEST 11. Position the control arm of the three-way stopcock over the DPM connector (see Figure 5-49). CONTROL ARM DISTAL TUB- DPM CONNECTOR Figure 5-49. Positioning the Control Arm Over the DPM Connector 12. Press [START]. If a CONFIRM PROGRAM? message appears, confirm that the rate and VTBI are correct, and then press [YES]. 13.
SECTION 5 MAINTENANCE AND SERVICE TESTS 16. Open the three-way stopcock to air by positioning the control arm over the distal tubing (see Figure 5-51). This releases the pressure that was built up during the test. CONTROL ARM DISTAL TUBDPM CONNECTOR Figure 5-51. Opening the Three-Way Stopcock to Air 17. Position the control arm of the three way stopcock over the DPM connector (see Figure 5-52). CONTROL ARM DISTAL TUB- DPM CONNECTOR Figure 5-52.
5.3 PERFORMANCE VERIFICATION TEST 24. Set the three-way stopcock to measure pressure by positioning the control arm over the open port (see Figure 5-53). OPEN PORT ON STOPCOCK CONTROL ARM DISTAL TUBDPM CONNECTOR Figure 5-53. Positioning the Control Arm to Measure Pressure 25. Verify that the distal occlusion audible alarm occurs at 10 psi ± 3 psi. Confirm that the DISTAL OCCLUSION message is flashing on the screen and that pumping is stopped. 26. Turn off the infuser. 27.
SECTION 5 MAINTENANCE AND SERVICE TESTS SCORE MARK DRIP CHAMBER 12” - 24” CANNULA 25 mL GRADUATED CYLINDER Figure 5-54. Delivery Accuracy Test Setup 2. Turn on the infuser. 3. If an Area Selection or CCA Selection screen appears, choose a care area and press [ENTER]. 4. Press [YES] at the CLEAR SETTINGS? or NEW PATIENT? prompt. 5. Press [A] to select Line A. 6. Enter a rate of 200 mL/hr and VTBI of 10 mL. 7. Press [START].
5.3 PERFORMANCE VERIFICATION TEST 11. Enter a rate of 200 mL/hr and a VTBI of 10 mL. 12. Press [START]. If a CONFIRM PROGRAM? message appears, confirm that the rate and VTBI are correct, and then press [YES]. 13. Verify that fluid is pumping; the message PUMPING is displayed in the Line B status bar, and the LED over Line B flashes. pumping Note: Line A will be stopped (DELAYED) while Line B is pumping, and will resume when Line B delivery is complete. 14.
SECTION 5 MAINTENANCE AND SERVICE TESTS 3. Turn on the DMM and set it to measure resistance. 4. On the infuser, press [A] to select Line A. If a CLEAR SETTINGS? or NEW PATIENT? prompt appears, press [YES]. 5. Set the delivery rate to 400 mL/hr, and the VTBI to 1 mL. 6. Press [START] and verify that the infuser is pumping. 7. After KVO flashes and the LINE A VTBI COMPLETE message appears, check for a short circuit on the DMM (approximately 1 on a scale of 0 to 100 ).
5.3 PERFORMANCE VERIFICATION TEST 5.3.17 END OF THE PVT If any tests fail, see Section 6, or contact Hospira. If the infuser passed all performance verification tests, follow these instructions to clear all programming and prepare the device to be put back into service: 1. Press ON/OFF on all three keypads to turn all infusion units on. 2. If an Area Selection or CCA Selection screen appears, choose a care area and press [ENTER] on each device. 3.
SECTION 5 MAINTENANCE AND SERVICE TESTS This page intentionally left blank.
Section 6 TROUBLESHOOTING This section contains information on technical assistance, warning messages, alarm messages and error codes, and troubleshooting procedures. 6.1 TECHNICAL ASSISTANCE For technical assistance, product return authorization, and to order parts, accessories, or manuals within the United States, contact Hospira. 1-800-241-4002 For additional technical assistance, technical training, and product information, visit the website at www.hospira.com.
SECTION 6 TROUBLESHOOTING Table 6-1. Message Warning Messages Possible Cause Corrective Action Stop delivery, then turn off Attempting to turn off the infuser while a delivery is in progress Stop all lines, then turn off the infuser Warning: Low Battery Battery is discharged so that only approximately 30 minutes of battery life remain Plug into AC power Warning: Replace Battery Battery service needed Replace the battery (see Section 7.2.
6.3 Table 6-2.
SECTION 6 TROUBLESHOOTING Table 6-2. Alarm Code Operational Alarm Messages and Corrective Actions Alarm Description Possible Cause Corrective Action N182 or E182 Prox. Occl B, Air or Prox. Occl B Negative proximal occlusion B, non-delivery Proximal occlusion detected on line B during non-delivery Backprime the cassette and restart line B or Stop all lines, backprime the cassette, and restart all lines N183 or E183 Prox. Occl B, Air or Prox.
6.3 Table 6-2. Alarm Code ALARM MESSAGES AND ERROR CODES Operational Alarm Messages and Corrective Actions Alarm Description Possible Cause Corrective Action N189 or E189 Prox. Occl B, Air Peak proximal occlusion B, delivery Proximal occlusion detected during delivery on line B Fix occlusion and restart line B or Stop all lines, fix occlusion and restart the infuser N190 or E190 Prox.
SECTION 6 TROUBLESHOOTING Table 6-2. Alarm Code Operational Alarm Messages and Corrective Actions Alarm Description Possible Cause Corrective Action N233 or E233 Distal air cumulative Distal air cumulative 500 mL of air detected in the last 5.
6.3 Table 6-2.
SECTION 6 TROUBLESHOOTING 6.3.2 ERROR CODES REQUIRING TECHNICAL SERVICE Table 6-3 lists infusion system error codes that require technical service. Also listed in Table 6-3 are malfunction descriptions, possible causes, and corrective actions. Perform corrective actions in the order listed in the Corrective Action column. Table 6-3.
6.3 Table 6-3. ALARM MESSAGES AND ERROR CODES Error Codes Requiring Technical Service Error Code Malfunction E321 Failure of charging current to drop below end of charge current threshold within eight hours of charging, or below one ampere within six hours of charging Possible Cause Defective or worn out battery Corrective Action Turn power off, then on to clear the error on the device and remove the infuser from service.
SECTION 6 TROUBLESHOOTING Table 6-3.
6.3 Table 6-3. Error Code E378 ALARM MESSAGES AND ERROR CODES Error Codes Requiring Technical Service Malfunction I/O valve phase loss Possible Cause Generic I/O valve failure Corrective Action Turn power off, then on, to reset the infuser Replace CPU PWA (see Section 7.2.14.6) Replace mechanism assembly (see Section 7.2.14.8) Reset time and date, if required (see Section 1.8.
SECTION 6 TROUBLESHOOTING Table 6-3. Error Code E436 Error Codes Requiring Technical Service Malfunction ROM failure Possible Cause Corrective Action ROM checksum failure Turn power off, then on, to reset the infuser Replace peripheral assembly (see Section 7.2.7) Reset time and date, if required (see Section 1.8.
6.3 Table 6-3. Error Code ALARM MESSAGES AND ERROR CODES Error Codes Requiring Technical Service Malfunction Possible Cause Corrective Action Replace mechanism assembly (see Section 7.2.14.8) E448 SEEP write failure SEEP data write failed E449 SEEP calibration data corrupted Calibration data block corrupted Replace CPU PWA (see Section 7.2.14.6) Replace CPU/driver cable (see Section 7.2.14.4) Reset time and date, if required (see Section 1.8.
SECTION 6 TROUBLESHOOTING The following error codes are not generated in the Biomed service mode: E320 E323 E326 E346 E373 E376 E379 E431 E441 E321 E324 E343 E371 E374 E377 E380 E432 E447 E322 E325 E345 E372 E375 E378 E430 E433 Note: Some error codes include sub-ID codes. These sub-ID codes are intended for Hospira internal use only, and should be included when contacting Hospira.
6.4 TROUBLESHOOTING PROCEDURES 6.4 TROUBLESHOOTING PROCEDURES This section details recommended procedures for problems not associated with malfunction alarms. Note: See Section 6.4.1 for unrestricted flow. Before performing any troubleshooting procedure, turn the infuser off, then on. Allow the self test to complete and proceed as follows: 1. If a malfunction exists, carefully inspect the infuser for damage as described in Section 5.2.2. 2.
SECTION 6 TROUBLESHOOTING Table 6-4. Test Failure Proximal Occlusion Test Section 5.3.10 Proximal Air-in-Line Test Section 5.3.11 Distal Air-in-Line Test Section 5.3.12 Distal Occlusion Test Section 5.3.13 Delivery Accuracy Test Section 5.3.14 Electrical Safety Test Section 5.3.
6.4 TROUBLESHOOTING PROCEDURES 6.4.1 UNRESTRICTED FLOW WARNING: UNRESTRICTED FLOW MAY BE LIFE-THREATENING. CAUTION: Prevent unrestricted flow by ensuring that the regulator closer is properly seated. If unable to ensure this, discontinue use and contact Hospira. Unrestricted flow may occur if the infuser’s regulator closer is not seated correctly and the cassette is removed from the device without ensuring the roller clamp or slide clamp on the administration set is in the closed position.
SECTION 6 TROUBLESHOOTING This page intentionally left blank.
Section 7 REPLACEABLE PARTS AND REPAIRS This section itemizes all parts and subassemblies of the infusion system that are repairable within the scope of this manual. In addition, this section details replacement procedures for all listed parts. 7.1 REPLACEABLE PARTS Replaceable parts for the infusion system are itemized in the Illustrated Parts Breakdown (IPB) and are identified in Figure 9-1. Table 9-2 identifies each part by an index number that correlates to Figure 9-1.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.2 REQUIRED TOOLS AND MATERIALS The following tools and materials, or equivalents, are required for the replacement procedures in this section. In addition, the beginning of each procedure lists tools and materials required for that specific procedure.
7.2 REPLACEMENT PROCEDURES RUBBER FOOT PAD (4) BATTERY DOOR (3) 6-32 x 1/2 PAN HEAD SCREW (4) Figure 7-1. Bottom View 7.2.
SECTION 7 REPLACEABLE PARTS AND REPAIRS To replace a battery, wire harness, door, or door pad, see Figure 7-2, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Place the infuser face down on a soft flat surface. 3. Using the flat blade screwdriver, remove the screw and retaining ring that attach the battery door to the infuser, and remove the door. 4. Inspect the battery door and replace, if required. 5.
7.2 REPLACEMENT PROCEDURES AC POWER CORD BATTERY WITH WIRE HARNESS ASSEMBLY (3) VELCRO STRAP DOOR PAD (3) 4-40 x 3/8 PAN HEAD SCREW EQUIPOTENTIAL TERMINAL RETAINING RING (3) POWER CORD RETAINER BATTERY DOOR (3) 6-32 x 1/2 HEX HEAD SCREW (3) Figure 7-2.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.5 AC POWER CORD, RETAINER, AND VELCRO STRAP REPLACEMENT The recommended tool for this procedure is a #2 Phillips screwdriver. Replacement parts for this procedure are: Cordset, AC Power, Hospital Grade, Detachable Retainer, AC Power Cord Strap, Velcro, AC Power Cord Screw, 4-40 x 3/8, Pan Head, Phillips To replace the AC power cord, retainer, or Velcro strap, see Figure 7-2, then proceed as follows: 1.
7.2 REPLACEMENT PROCEDURES 7.2.6 SEPARATING THE FRONT ENCLOSURE, REAR ENCLOSURE, AND MAIN CHASSIS ASSEMBLY The front enclosure consists of an upper assembly and a lower assembly. The main chassis assembly consists of an upper chassis and a lower chassis. The recommended tool for this procedure is a #2 Phillips screwdriver.
SECTION 7 REPLACEABLE PARTS AND REPAIRS UPPER FRONT ENCLOSURE FRONT/REAR ENCLOSURE GASKET (3) PERIPHERAL INTERFACE ASSEMBLY 6-32 x 1 1/4 PAN HEAD SCREW (5) 6-32 x 2 3/4 PAN HEAD SCREW (2) #6 FLAT WASHER (8) 6-32 x 1/2 PAN HEAD SCREW LOWER FRONT ENCLOSURE MAIN CHASSIS ASSEMBLY 8-32 x 3 1/2 PAN HEAD SCREW (3) REAR ENCLOSURE #8 FLAT WASHER (3) Figure 7-3.
7.2 REPLACEMENT PROCEDURES 8 1 4 6 3 2 9 7 Figure 7-4. 5 Screw Placement Sequence 7.2.7 PERIPHERAL INTERFACE ASSEMBLY REPLACEMENT CAUTION: Peripheral interface assembly replacement should only be performed after receiving approval from Hospira. Note: Replacing the peripheral interface assembly does not change the existing Biomed settings.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 6. Depress the retention clip and carefully pull the peripheral interface assembly away from the infuser. Note: When removing the peripheral interface assembly, note the placement guides where the peripheral interface PWA rests. 7. Install the replacement peripheral interface assembly in the exact reverse order of removal. Note: Verify the peripheral interface PWA is placed properly between the guides and fits correctly into the CPU PWA. 8.
7.2 REPLACEMENT PROCEDURES PERIPHERAL PWA #2 PERIPHERAL PWA #1 PERIPHERAL CABLE #2 PERIPHERAL CABLE #1 RETENTION CLIPS PERIPHERAL INTERFACE ASSEMBLY Figure 7-5. Peripheral Interface Assembly and Peripheral PWAs 7.2.9 PERIPHERAL COMPONENT REPLACEMENT Peripheral component replacement includes the replacement of the volume control knob and the peripheral cover. To replace peripheral components, see Figure 7-6, then proceed as detailed in the following sections. 7.2.9.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 4. Using the X-acto knife, lift the volume control knob end cap away from the knob, exposing a flat head screw. 5. Using the flat blade screwdriver, remove the screw that secures the knob. 6. Using long needle nose pliers, remove the knob cap, knob cover, and volume control knob. 7. Install the replacement volume control knob in the exact reverse order of removal. 8. Install the peripheral interface assembly in the exact reverse order of removal. 9.
7.2 REPLACEMENT PROCEDURES 9. Using a 3/16 nutdriver, remove the screws from the DataPort connector. 10. Inspect the ESD spring and replace, if required. 11. Install the replacement peripheral cover in the exact reverse order of removal. 12. Install the volume control knob and nurse call jack nut in the exact reverse order of removal. 13. Install the peripheral interface assembly as described in Section 7.2.7. 14. Reinstall the batteries, and connect the device to AC power. 15.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.10 FRONT/REAR ENCLOSURE GASKET REPLACEMENT The recommended tool for this procedure is needle nose pliers. The replacement part for this procedure is: Gasket, Front/Rear Enclosure Note: Clean and remove any foreign matter on the replacement gasket or in the spaces where the replacement gasket is to be installed. To replace a front/rear enclosure gasket, see Figure 7-3, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power.
7.2 REPLACEMENT PROCEDURES LOWER FRONT ENCLOSURE TOP SEAL GASKET (3) KEYPAD GASKET (3) EMI D-SHAPE GASKET (4) Figure 7-7. Lower Front Enclosure Gaskets 7.2.11.1 EMI GASKET REPLACEMENT The recommended tool for this procedure is needle nose pliers. The replacement part for this procedure is: Gasket, EMI, D-Shape, 6.62 in. Gasket, EMI, D-Shape, 13 in. To replace an EMI gasket, see Figure 7-7, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.11.2 KEYPAD GASKET REPLACEMENT The recommended tool for this procedure is needle nose pliers. The replacement part for this procedure is: Gasket, Keypad To replace a keypad gasket, see Figure 7-7, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3. Separate the front enclosure, rear enclosure, and main chassis assembly as described in Section 7.2.6. 4.
7.2 REPLACEMENT PROCEDURES 7.2.12 REAR ENCLOSURE ASSEMBLY COMPONENT REPLACEMENT Rear enclosure assembly component replacement includes replacement of the following: - Rear enclosure gaskets - Pole clamp assembly, backing plate, and insulator - Equipotential terminal - Internal AC power cord - AC connector - Fuses To replace the rear enclosure assembly components, see Figure 7-8 and Figure 7-9, then proceed as detailed in the following sections.
SECTION 7 REPLACEABLE PARTS AND REPAIRS Figure 7-9.
7.2 REPLACEMENT PROCEDURES 7.2.12.1 POLE CLAMP ASSEMBLY AND BACKING PLATE REPLACEMENT Recommended tools for this procedure are a set of nutdrivers and a medium size flat blade screwdriver.
SECTION 7 REPLACEABLE PARTS AND REPAIRS Replacement parts for this procedure are: Cord, Internal, AC Power Clamp, Internal Power Cord Wire, Ground, AC Power Screw, 4-40 x 3/8, Hex Head, Slotted Washer, Flat, 1/4 Washer, Lock, 1/4 To replace the internal AC power cord, see Figure 7-9, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3. Separate the front and rear enclosures as described in Section 7.2.6. 4.
7.2 REPLACEMENT PROCEDURES 8. Join the rear enclosure and main chassis assembly in the exact reverse order of separation. 9. Reinstall the batteries and connect the infuser to AC power. To verify successful AC connector replacement, perform the PVT in Section 5.3. 7.2.12.4 FUSE REPLACEMENT Recommended tools for this procedure are a #2 Phillips screwdriver and a small flat blade screwdriver. Replacement parts for this procedure are: Fuse, 1.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.12.5 REAR ENCLOSURE GASKET REPLACEMENT The recommended tool for this procedure is needle nose pliers. Replacement parts for this procedure are: Gasket, Rear Enclosure, 20 13/16 Gasket, Rear Enclosure, 45 7/8 To replace a rear enclosure gasket, see Figure 7-9, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3. Separate the rear in Section 7.2.6.
7.2 REPLACEMENT PROCEDURES CLUTCH HOUSING CLUTCH SPRING BAG HANGER COTTER RING Figure 7-10. Minipole Assembly 7.2.14 MAIN CHASSIS ASSEMBLY COMPONENT REPLACEMENT Main chassis assembly component replacement includes replacement of the following: - Power supply PWA - Mechanism assembly - Keypad - Cassette door - Display - Fluid shield - CPU PWA - Opener handle - Piezo alarm To replace the main chassis assembly components, see Figure 7-11, then proceed as detailed in the following sections.
SECTION 7 REPLACEABLE PARTS AND REPAIRS MECHANISM/CHASSIS BUMPER (6) MECHANISM ASSEMBLY (3) 6-32 x 1/2 HEX HEAD SCREW (9) INSULATION TAPE (3) UPPER CHASSIS KEYPAD (3) CPU/DRIVER CABLE (3) 4-24 x 1/4 PAN HEAD SCREW (3) 4-40 x 3/16 HEX HEAD SCREW (12) LOWER CHASSIS DISPLAY ASSEMBLY (3) Figure 7-11.
7.2 REPLACEMENT PROCEDURES 4-40 x 3/8 HEX HEAD SCREW (6) PIEZO ALARM (3) CPU/DRIVER CABLE (3) INSULATION TAPE (3) POWER SUPPLY/ BATTERY CABLE (3) MOTOR POWER CABLE (3) PERIPHERAL PWA (2) POWER SUPPLY (3) PERIPHERAL CABLE #2 4-40 x 1/4 HEX HEAD SCREW (3) CPU PWA (3) PERIPHERAL INTERFACE PWA PERIPHERAL CABLE #1 Figure 7-11.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.14.1 POWER SUPPLY PWA REPLACEMENT The recommended tool for this procedure is a medium size flat blade screwdriver. Replacement parts for this procedure are: PWA, Power Supply Assembly, Cable, Power Supply/Battery To replace a power supply PWA, see Figure 7-11 (2 of 2), then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3.
7.2 REPLACEMENT PROCEDURES 6. Using the Phillips screwdriver, remove the screw that secures the keypad and display to the lower main chassis. 7. Carefully disconnect the flex ribbon cable from the display by pushing the connector locking tabs down. 8. Using the flat blade screwdriver, separate the keypad and display by removing the screws that secure the keypad to the display. 9. Install the replacement keypad in the exact reverse order of removal. 10.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.14.4 CPU/DRIVER CABLE REPLACEMENT The recommended tool for this procedure is an X-acto knife. Replacement parts for this procedure are: Assembly, Cable, CPU/Driver Tape, Ferrite Tape, Insulation To replace a CPU/driver cable, see Figure 7-11 (1 of 2), Figure 7-12, and Figure 7-13, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3.
7.2 REPLACEMENT PROCEDURES TOP TAB CABLE CENTER TAB FERRITE LOWER TAB Figure 7-12.
SECTION 7 REPLACEABLE PARTS AND REPAIRS FERRITE INSULATION TAPE Figure 7-13. Ferrite Tape Positioning (1 of 2) FERRITE CENTER TAB TAPE Figure 7-13.
7.2 REPLACEMENT PROCEDURES 7.2.14.5 MOTOR POWER CABLE REPLACEMENT The recommended tool for this procedure is a medium size flat blade screwdriver. The replacement part for this procedure is: Assembly, Cable, Motor Power To replace a motor power cable, see Figure 7-11 (2 of 2), then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 10. Disconnect the piezo alarm cable from the CPU PWA. 11. Using the Phillips screwdriver, remove the screw from the bottom of the lower main chassis assembly. 12. Slide the CPU PWA out of the main chassis until J5 on the CPU PWA is accessible. 13. Using the flat blade screwdriver, release the locking tabs that secure the flex cable to the CPU PWA. 14. Install the replacement CPU PWA in the exact reverse order of removal. 15.
7.2 REPLACEMENT PROCEDURES 7.2.14.8 MECHANISM ASSEMBLY REPLACEMENT Note: Replacing the mechanism changes the biomed settings to those stored in the replacement mechanism assembly. Recommended tools for this procedure are a medium size flat blade screwdriver and #2 Phillips screwdriver.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.14.8.1 ENTER THE SERIAL NUMBER After the mechanism assembly has been replaced, the device serial number must be entered. To enter the serial number: 1. Press the [ON/OFF] key to turn on the device and view the setup screen and prompt for the device serial number. 2. In the SETUP screen, enter the serial number of the infuser. The serial number is found on the Product Identification Label (see Figure 5-1). Increment the serial number entry for each infuser.
7.2 REPLACEMENT PROCEDURES To verify successful mechanism chassis gasket tape installation, perform the PVT in Section 5.3. ALIGNMENT FIXTURE GASKET ALIGNMENT GASKET TAPE LOCATION Figure 7-14. Installing the Mechanism Chassis Gasket Tape 7.2.14.10 CASSETTE DOOR AND FLUID SHIELD REPLACEMENT WARNING: UNRESTRICTED FLOW MAY BE LIFE-THREATENING. CAUTION: To ensure that the fluid shield tab does not obstruct or unseat the regulator closer, the fluid shield tab must be visually inspected.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 6. Disengage the cassette door from the opener handle assembly and remove the door. 7. Disengage the clips on the back side of the fluid shield that retain the upper portion of the fluid shield to the mechanism assembly. 8. Lift the locking pins to release the fluid shield/driver flex connector, and disconnect the flex connector from the driver PWA. 9. Pull the shield away from the top of the mechanism assembly at an approximate 15 degree angle.
7.2 REPLACEMENT PROCEDURES 14. Align the mechanism assembly pins, then install the replacement fluid shield in the exact reverse order of removal. CAUTION: If the fluid shield assembly is misaligned, snapping the fluid shield assembly in place may unseat the regulator closer, which may result in unrestricted flow. CAUTION: Use extreme caution when aligning the mechanism assembly pins and installing the replacement fluid shield. 15.
SECTION 7 REPLACEABLE PARTS AND REPAIRS Normal Tab Figure 7-17. Bent Tab Close-up Views of Normal (left) and Bent (right) Fluid Shield Tab 18. Install the replacement cassette door in the exact reverse order of removal. 19. Replace the mechanism assembly in the exact reverse order of removal. 20. Join the front enclosure, rear enclosure, and main chassis assembly in the exact reverse order of separation. 21. Reinstall the batteries and connect the infuser to AC power.
7.2 ROX APP REPLACEMENT PROCEDURES 15 MECHANISM ASSEMBLY GASKET, .72 IN. (2) FLUID SHIELD GASKET, 1.09 IN. CASSETTE DOOR Figure 7-18.
SECTION 7 REPLACEABLE PARTS AND REPAIRS FLEX CONNECTOR TORSION SPRING RETAINING RING MECHANISM ASSEMBLY CASSETTE DOOR OPENER HANDLE 4-40 x 3/8 HEX HEAD SCREW WITH WASHER 3/32 PUSH-ON RETAINING RING DOOR PIVOT CAP Figure 7-19.
7.2 REPLACEMENT PROCEDURES 7.2.14.11 OPENER HANDLE ASSEMBLY REPLACEMENT The recommended tool for this procedure is a medium size flat blade screwdriver. Replacement parts for this procedure are: Assembly, Opener Handle Link, Door Ring, Retaining, Push-On To replace the opener handle assembly, see Figure 7-19, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3.
SECTION 7 REPLACEABLE PARTS AND REPAIRS This page intentionally left blank.
Section 8 SPECIFICATIONS The following specifications apply to the Plum A+3 infusion system. PHYSICAL Dimensions: Approximately 19 H x 15 W x 14 D inches (excluding pole clamp and power cord storage) Weight: Approximately 28 lbs (with batteries) Casing: High-impact plastic Expected Service Life*: 10 years ELECTRICAL Power Requirements: 95-132 VAC; 47-62 Hz; 90 W Power Cord: Hospital-grade AC cord; 10 feet; with transparent plug and retainer plate Fuses: 0.
SECTION 8 SPECIFICATIONS ENVIRONMENT Operating: 41 to 104 F (5 to 40 C); 10 % to 90 % relative humidity Transporting and Storage: -4 to 140 F (-20 to 60 C); 10 % to 90 % relative humidity Atmospheric Pressure: 0 - 10,000 feet (0 - 3000 meters) or equivalent atmospheric pressure Relative Humidity: 10 - 90 % (104 F max) DELIVERY RATE RANGE Lines A and B: 0.1 to 99.9 mL/hr (in 0.1 mL/hr increments) 100 to 999 mL/hr (in 1 mL/hr increments) Concurrent Delivery: 0.
Section 9 DRAWINGS Figure 9-1 through Figure 9-10 show the Illustrated Parts Breakdown (IPB) and assembly drawings. Table 9-1 lists drawings by figure number and title. Table 9-2 identifies parts by index numbers which correlate to Figure 9-1. Drawings in Section 9 are provided as information only, and may not exactly reflect current product configuration. PWA drawings are available from Hospira upon request. Table 9-1.
SECTION 9 DRAWINGS Table 9-2. Index Number IPB for the Infuser Nomenclature Replacement Procedure 7 PWA, Power Supply Section 7.2.14.1 8 Assembly, Mechanism Section 7.2.14.8 9 Assembly, Fluid Shield Section 7.2.14.10 10 Assembly, Opener Handle Section 7.2.14.11 11 Assembly, Cassette Door Section 7.2.14.10 12 PWA, Peripheral Section 7.2.8 13 Assembly, Peripheral Interface Section 7.2.7 14 Cover, Peripheral Section 7.2.9.2 15 Shield, Spring, ESD Section 7.2.9.
SECTION 9 DRAWINGS Table 9-2. Index Number IPB for the Infuser Nomenclature Replacement Procedure 37 Assembly, Minipole A: Hanger, Bag B: Housing, Clutch C: Ring, Cotter D: Spring, Clutch Section 7.2.13 38 Assembly, Cable, Peripheral #2 Section 7.2.7 39 Assembly, Cable, Peripheral #1 Section 7.2.7 40 Assembly, Cable, CPU/Driver Section 7.2.14.4 41 Assembly, Piezo Alarm Section 7.2.14.7 42 Battery Section 7.2.4 43 Assembly, Wire Harness, Battery Section 7.2.
SECTION 9 DRAWINGS Table 9-2. Index Number IPB for the Infuser Nomenclature Replacement Procedure 65 Screw, 4-24 x 1/4, Pan Head, Phillips, SS As applicable 66 Screw, 4-40 x 1/4, Hex Head, Slotted, with Washer As applicable 67 Screw, 4-40 x 3/8, Pan Head, Phillips, SS As applicable 68 Screw, 4-40 x 3/8, Hex Head, Nylon Section 7.2.9.
1 8 3 2 9 23 10 11 24 25 26 4 12 14 13 27 28 15 29 6 5 16 17 30 31 18 32 21 20 19 33 7 34 35 22 36 37 37A 37B HOSPIRA, INC. 37D 37C Figure 9-1. Illustrated Parts Breakdown DRAWING NO. NOT APPLICABLE Technical Service Manual 9-5 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
39 38 56 57 40 41 58 59 42 43 61 60 44 63 62 66 65 64 47 46 45 48 68 67 49 50 70 69 51 71 52 53 72 54 73 55 74 87 75 76 77 78 79 80 81 82 83 84 85 86 HOSPIRA, INC. Figure 9-1. Illustrated Parts Breakdown DRAWING NO. NOT APPLICABLE Technical Service Manual 9-7 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
50 3 78 75 1 2 71 4 79 76 78 HOSPIRA, INC. Figure 9-2. Front Enclosures, Rear Enclosure, and Main Chassis Assembly 72 DRAWING NO. NOT APPLICABLE Technical Service Manual 9-9 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
3 50 54 2 20 48 HOSPIRA, INC. 78 Figure 9-3. Front Enclosure Assemblies 53 49 71 Technical Service Manual 9 - 11 DRAWING NO. NOT APPLICABLE Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
26 74 28 85 52 82 30 68 77 33 82 84 1 82 84 29 HOSPIRA, INC. 51 Figure 9-4. Rear Enclosure Assembly DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 13 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
1 27 59 67 57 20 78 HOSPIRA, INC. Figure 9-4. Rear Enclosure Assembly 58 56 71 DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 15 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
14 15 12 68 60 34 80 86 61 67 HOSPIRA, INC. Figure 9-5. Peripheral Interface Assembly DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 17 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
69 41 40 47 46 32 31 12 38 39 7 18 66 HOSPIRA, INC. Figure 9-6. Main Chassis Assembly 13 DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 19 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
73 19 6 8 46 16 65 5 70 HOSPIRA, INC. Figure 9-6. Main Chassis Assembly 17 DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 21 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
18 70 17 16 40 HOSPIRA, INC. Figure 9-7. CPU PWA, Display, and Keypad DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 23 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
4 18 77 66 HOSPIRA, INC. Figure 9-8. CPU PWA and Main Chassis DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 25 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
37 25 23 42 67 24 45 64 43 44 73 HOSPIRA, INC. Figure 9-9. AC Power Cord, Retainer, Batteries, and Minipole DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 27 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
9 11 8 10 64 55 HOSPIRA, INC. Figure 9-10. Mechanism Assembly 69 DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 29 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
APPENDIX USE OF THE INFUSION SYSTEM IN ELECTROMAGNETIC ENVIRONMENTS The Plum A+3 Infusion System is intended for use in the electromagnetic environment specified in Table A-1, Table A-2, Table A-3, and Table A-4. The user of the infusion system should assure that it is used only in the appropriate environment. ELECTROMAGNETIC EMISSIONS Table A-1 details electromagnetic emissions compliance and guidance. Table A-1.
APPENDIX ELECTROMAGNETIC IMMUNITY Table A-2 details guidance for the electromagnetic environment. Table A-2.
APPENDIX ELECTROMAGNETIC IMMUNITY FOR LIFE-SUPPORTING EQUIPMENT AND SYSTEMS Table A-3 provides guidance for use of the infusion system near communications equipment. Table A-3.
APPENDIX a The industrial, scientific and medical (ISM) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.660 MHz to 40.700 MHz. b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas.
APPENDIX RECOMMENDED SEPARATION DISTANCES FOR COMMUNICATIONS EQUIPMENT The infusion system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The recommendations provided in Table A-4 help the user of the infusion system to prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the infuser, according to the maximum output power of the communications equipment.
APPENDIX This page intentionally left blank.
Back page For technical assistance, product return authorization, and to order parts, accessories, or manuals within the United States, contact Hospira. 1-800-241-4002 For additional services and technical training courses, visit the website: www.hospira.com For technical assistance and services outside the United States, contact the local Hospira sales office. WARNING: EXPLOSION HAZARD EXISTS IF THE INFUSION SYSTEM IS USED IN THE PRESENCE OF FLAMMABLE SUBSTANCES.