R INFUSION SYSTEM For use with List Numbers 12348 and 12618 Technical Service Manual 430-95424-003 (Rev.
©Hospira, Inc. This document and the subject matter disclosed herein are proprietary information. Hospira retains all the exclusive rights of dissemination, reproduction, manufacture, and sale.
Contents Section 1 INTRODUCTION . . . . . . . . . . . . . . . 1.1 SCOPE . . . . . . . . . . . . . . . . 1.2 CONVENTIONS . . . . . . . . . . . . . 1.3 COMPONENT DESIGNATORS . . . . . . . 1.4 ACRONYMS AND ABBREVIATIONS . . . . . 1.5 USER QUALIFICATION . . . . . . . . . . 1.6 ARTIFACTS . . . . . . . . . . . . . . 1.7 INSTRUMENT INSTALLATION PROCEDURE . 1.7.1 UNPACKING . . . . . . . . . . . 1.7.2 INSPECTION . . . . . . . . . . . 1.7.3 SELF TEST . . . . . . . . . . . . 1.8 BIOMED SETTINGS . . . . . . . . . . .
CONTENTS 4.2.2 POWER SUPPLY SUBSYSTEM . . . . . . . . . 4.2.2.1 MAIN SWITCHING REGULATOR . . . . 4.2.2.2 MAIN REGULATOR FAULT DETECTION . 4.2.2.3 SYSTEM POWER . . . . . . . . . . . 4.2.2.4 AUXILIARY SUPPLIES . . . . . . . . . 4.2.2.5 POWER CONTROL . . . . . . . . . . 4.2.2.6 BATTERY VOLTAGE MEASUREMENT . . 4.2.2.7 BATTERY CHARGE/DISCHARGE CURRENT MEASUREMENT . . . . . . . . . . 4.2.2.8 BATTERY CHARGER . . . . . . . . . 4.2.3 MECHANISM SUBSYSTEM . . . . . . . . . . 4.2.3.1 MOTORS/MOTOR DRIVE . . . . . . .
CONTENTS 5.2.11 PROXIMAL OCCLUSION TEST . 5.2.12 PROXIMAL AIR-IN-LINE TEST . 5.2.13 DISTAL AIR-IN-LINE TEST. . . 5.2.14 DISTAL OCCLUSION TEST . . 5.2.15 DELIVERY ACCURACY TEST . . 5.2.16 NURSE CALL TEST . . . . . . 5.2.17 ELECTRICAL SAFETY TEST . . 5.2.18 END OF THE PVT . . . . . . 5.5 PERIODIC MAINTENANCE INSPECTION 5.6 BATTERY OPERATION OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CONTENTS 7.2.14 MAIN CHASSIS ASSEMBLY COMPONENT REPLACEMENT . . 7.2.14.1 POWER SUPPLY PWA REPLACEMENT. . . . . . . 7.2.14.2 KEYPAD REPLACEMENT. . . . . . . . . . . . 7.2.14.3 DISPLAY ASSEMBLY REPLACEMENT . . . . . . . 7.2.14.4 CPU/DRIVER CABLE REPLACEMENT . . . . . . . 7.2.14.5 MOTOR POWER CABLE REPLACEMENT. . . . . . 7.2.14.6 CPU PWA REPLACEMENT . . . . . . . . . . . 7.2.14.7 PIEZO ALARM ASSEMBLY REPLACEMENT. . . . . 7.2.14.8 MECHANISM ASSEMBLY REPLACEMENT . . . . . 7.2.14.
CONTENTS Figure 7-11. Figure 7-11. Figure 7-12. Figure 7-13. Figure 7-13. Figure 7-14. Figure 7-15. Figure 9-1. Figure 9-1. Figure 9-2. Figure 9-3. Figure 9-4. Figure 9-4. Figure 9-5. Figure 9-6. Figure 9-6. Figure 9-7. Figure 9-8. Figure 9-9. Figure 9-10. Main Chassis Components (1 of 2) . . . . . . . . . . Main Chassis Components (2 of 2) . . . . . . . . . . CPU/Driver Cable Routing . . . . . . . . . . . . . Ferrite Tape Positioning (1 of 2) . . . . . . . . . . . Ferrite Tape Positioning (2 of 2) . . .
CONTENTS This page intentionally left blank.
Section 1 INTRODUCTION The Plum A+®3 infusion system is designed to meet the growing demand for hospital wide device standardization, and serves a wide range of general floor and critical care needs. The infusion system consists of three component infusers, designated line 1, line 2, and line 3. By incorporating three lines into one unit, the Plum A+3 provides three primary lines, three secondary lines, and piggyback fluid delivery capabilities.
SECTION 1 INTRODUCTION 1.2 CONVENTIONS The conventions listed in Table 1-1 are used throughout this manual. Table 1-1. Convention Conventions Application Example Italic Reference to a section, figure, table, or publication (see Section 6.1) [ALL CAPS] In-text references to keys, touchswitches, and display messages [START] Emphasis CAUTION: Use proper ESD grounding techniques when handling components. Screen displays Select Set Time and Date.
1.4 ACRONYMS AND ABBREVIATIONS 1.
SECTION 1 INTRODUCTION LED Light emitting diode L/S Line select MB Megabyte MHz Megahertz min Minute mL Milliliter mL/hr Milliliter per hour MMIO Memory-mapped input/output MOSFET Metal-oxide semiconductor field-effect transistor ms Millisecond nF nanofarad Op-amp Operational amplifier pF picofarad PROM Programmable read-only memory PVT Performance verification test PWA Printed wiring assembly PWM Pulse width modulator RAM Random-access memory rms Root-mean-square RTC Real-time clock SCC Serial communicati
1.5 USER QUALIFICATION 1.5 USER QUALIFICATION The Plum A+3 must be used at the direction of or under the supervision of licensed physicians or certified healthcare professionals who are trained in the use of the infusion system and the administration of parenteral and enteral fluids and drugs, and whole blood or red blood cell components. Training should emphasize preventing related IV complications, including appropriate precautions to prevent accidental infusion of air.
SECTION 1 INTRODUCTION 1.7.1 UNPACKING Inspect the shipping container as detailed in Section 1.7.2. Use care when unpacking the infusion system. Retain the packing slip and save all packing material in the event it is necessary to return the infuser to the factory. Verify the shipping container contains a copy of the system operating manual. 1.7.2 INSPECTION Inspect the shipping container for damage. Should any damage be found, contact the delivering carrier immediately.
1.7 INSTRUMENT INSTALLATION PROCEDURE Note: The message MECHANISM INITIALIZATION IN PROGRESS may briefly appear prior to the CASSETTE TEST IN PROGRESS message. 7. The CLEAR SETTINGS? message may appear. Press the [YES] softkey. 8. Press [ON/OFF] to turn off the infuser. LINE FLOW INDICATORS A B HOSPIRA Plum A+ Version X.
SECTION 1 INTRODUCTION 1.8 BIOMED SETTINGS The BIOMED SETTINGS screens contain the following options that can be changed or reviewed by qualified personnel: - IV screen parameters - Alarms log - Set time and date All infusers (new or refurbished) are shipped with factory settings (see Table 1-2). Note: Biomed screens do not time out for the Infuser Idle alarm or No Action alarm. Note: The battery will not be detected in the Biomed service mode. To access the Biomed settings, proceed as follows: 1.
1.8 Table 1-2. BIOMED SETTINGS System Configuration Data Data Options Range Factory Setting Maximum macro IV mode delivery rate 0.1 - 99.
SECTION 1 INTRODUCTION 1.8.1 IV PARAMETERS The IV Parameters screen contains Common IV Parameters and Macro IV Parameters (see Figure 1-3). To change the IV parameters see Figure 1-4 and Figure 1-5, then proceed as follows: 1. Access the BIOMED SETTINGS screen as described in Section 1.8. 2. Select IV Screen Parameters, and press [CHOOSE]. 3. Select the parameters to be changed, and press [CHOOSE]. 4. Using the [CHANGE VALUE] softkey, select the desired value, and press [ENTER]. 5.
1.8 BIOMED SETTINGS BIOMED SETTINGS Common IV Parameters Continue Rate KVO Deliver Together Concurrent Enable Delay/Standby Yes Callback Default No Select using Change Value Change Value Figure 1-4. Enter Cancel/ Back Common IV Parameters BIOMED SETTINGS Macro IV Parameters 6.0 psi Default Distal Press Max Rate 999 mL/hr Enter Value using keypad Enter Figure 1-5.
SECTION 1 INTRODUCTION 1.8.2 ALARMS LOG The Alarms Log will retain the latest 40 alarm and malfunction codes, listed in order from the most current to the oldest. To view the Alarms Log see Figure 1-6, then proceed as follows: 1. Access the BIOMED SETTINGS screen as described in Section 1.8. 2. Select Alarms Log, and press [CHOOSE}. Use the [PAGE UP] and [PAGE DOWN] softkeys to view the Alarms Log. 3. Press [BACK] to exit the Alarms Log and return to the main BIOMED SETTINGS screen.
1.8 BIOMED SETTINGS 1.8.3 SETTING THE TIME AND DATE Note: The infusion system will automatically display February 29 on leap year. Note: Daylight savings and time zone changes must be made manually. To set the time and date, see Figure 1-7, then proceed as follows: 1. Access the BIOMED SETTINGS screen as described in Section 1.8. 2. Select Set Time and Date, and press [CHOOSE]. 3. Select the parameter to be changed, then enter the desired value. 4.
SECTION 1 INTRODUCTION This page intentionally left blank.
Section 2 WARRANTY Subject to the terms and conditions herein, Hospira, Inc., hereinafter referred to as Hospira, warrants that (a) the product shall conform to Hospira's standard specifications and be free from defects in material and workmanship under normal use and service for a period of one year after purchase, and (b) the replaceable battery shall be free from defects in material and workmanship under normal use and service for a period of 90 days after purchase.
SECTION 2 WARRANTY This page intentionally left blank.
Section 3 SYSTEM OPERATING MANUAL A copy of the System Operating Manual is included with every Plum A+3 infusion system. If a copy is not available, contact Hospira (see Section 6.1).
SECTION 3 SYSTEM OPERATING MANUAL This page intentionally left blank.
Section 4 THEORY OF OPERATION This section describes the Plum A+3 theory of operation. The theory of operation details the general description, electronic subsystem overview, printed wiring assemblies, remote mounted peripherals, and mechanical overview of the infusion system. 4.
SECTION 4 THEORY OF OPERATION 4.2 ELECTRONIC SUBSYSTEM OVERVIEW This section describes the function and electronic circuitry of the following three main subsystems in the infusion system: CPU subsystem, power supply subsystem, and mechanism subsystem (see Figure 4-1). Note: An asterisk (*) denotes an active low or negative true logic signal.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW 4.2.1 CPU SUBSYSTEM The CPU subsystem contains the main microcontroller that is responsible for controlling the display/keyboard interface, external communications interfaces, barcode reader (BCR) interface, and system management.
SECTION 4 THEORY OF OPERATION 4.2.1.3 PROGRAMMABLE READ-ONLY MEMORY The CPU subsystem has two 512 K x 8 bit programmable read-only memory (PROM) memory devices that provide a total of 1024 KB. The PROM space is expandable up to 2 MB. The PROM memory devices operate off the 3.3 VDC supply. The CPU chip-select 0 pin (CS0*), is connected to the PROM chip-enable (CE*) pin (signal CSROM*). This special chip-select signal can support bootstrap operation after reset.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW 4.2.1.7 LCD BACKLIGHT CONTROL The LCD panel is backlit by a cold cathode fluorescent tube (CCFT) lamp. The CCFT lamp requires 300 Vrms to operate; a current controlled DC-to-AC voltage inverter circuit is used to deliver a current regulated sine wave to the lamp. A switching regulator regulates the CCFT current by monitoring feedback pin 3, and varies its output duty cycle to drive a DC/AC inverter.
SECTION 4 THEORY OF OPERATION 4.2.1.10 VOLTAGE MONITOR WATCHDOG TIMER It is important to protect the system during power transitions, and the CPU is reset after the VCC power supply is applied. The microprocessor supervisory circuit generates an automatic reset output during power up, power down, or brownout conditions. When the VCC falls below the reset threshold voltage of 2.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW U4 CPU A1 A2 A3 A4 ANALOG INPUTS I/O CS* A0 I/O CLOCK SPCLK ADDRESS SPTXD DATA OUT SPRXD A5 EOC SCP I/O (OR INTERRUPT) A6 A7 A8 REF+ A9 2.5V REF- A10 GND GANA GDIG Figure 4-2. Serial Interface to ADC Table 4-1.
SECTION 4 THEORY OF OPERATION 4.2.1.12 DIGITAL-TO-ANALOG CONVERTER The dual 8 bit digital-to-analog converter (DAC) generates two analog signals to control the phase A and phase B motor coil currents. The interface between the DAC device and the CPU is the 8 bit data bus that is connected to DATA15:8. All the control signals for this DAC are generated by FPGA logic devices. Buffer amplifier/ground compensation circuits condition the DAC outputs. 4.2.1.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW 4.2.1.17 NURSE CALL INTERFACE A nurse call relay switch on the peripheral interface PWA indicates alarm conditions to a remote operator. A memory-mapped output signal (NURSE) activates the relay during alarm conditions. The relay has both normally-open and normally-closed contacts. A jumper on the peripheral interface board selects the contact type. The factory setting is normally-open. 4.2.1.18 AUDIBLE INDICATORS There are two audible indicators on the CPU subsystem.
SECTION 4 THEORY OF OPERATION 4.2.1.20 DATAPORT INTERFACE The CPU communicates with an external computer by way of a DataPort interface. The DataPort interface provides for remote monitoring of up to four infusers using a host computer with a modified RS-232-D serial interface. Infusers are either connected directly to the host or in a daisy chain configuration using junction boxes that provide a 5 bit hard ID via DIP switches on the junction box.
4.2 Table 4-3.
SECTION 4 THEORY OF OPERATION Table 4-4.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW 4.2.2 POWER SUPPLY SUBSYSTEM The power supply subsystem provides DC power to system circuits and interface software controlled power and battery management.
SECTION 4 THEORY OF OPERATION 4.2.2.1.3 Current Limit Loop The current limit loop is activated when the primary current, sensed by R71, exceeds 3 A. Resistor R70 and capacitor C46 filter the voltage across R71 and feed it back to the current sense input (1.5 VDC threshold) of U12. The duty cycle of U12 is reduced until the excessive load is removed. 4.2.2.2 MAIN REGULATOR FAULT DETECTION If the switching regulator’s main loop fails, the secondary voltage limit loop takes over.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW The infuser is activated when the [ON/OFF] key is pressed or the AC line is plugged in. The [ON/OFF] key activates the STRTUP signal, triggering a three second one-shot circuit that will temporarily turn the system power on. This three second one-shot period allows the CPU enough time to power up, initialize, and turn on the PWRHLD signal. The CPU monitors the STRTUP signal, via interrupt, to signal a user request for turning off the infuser.
SECTION 4 THEORY OF OPERATION 4.2.2.8 BATTERY CHARGER The software battery management algorithm controls the battery charger. The charging scheme is a current limit/two stage voltage limit charger. The charge current is limited to 1.3 A and the voltage is limited to either 6.9 VDC or 7.5 VDC. The source of the charge current is power MOSFET transistor Q7 operating in the linear mode.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW 4.2.3.1.1 Stepper Motors Each motor is named by its function, as follows: - Plunger motor for driving the plunger screw - I/O valve motor for moving the input-output valve pins - L/S valve motor for moving the line select valve pins A and B All three motors are four phase stepper types. One electrical revolution is accomplished after four motor steps (phases) are completed. The step-angle (the number of steps per shaft revolution) resolutions are 3.
SECTION 4 THEORY OF OPERATION 4.2.3.2 MOTOR POSITION SENSORS Motor position is estimated by counting the motor steps, relative to a position reference. Optical switches and flags serve as position references that are used to find the motor home positions and to verify proper motion. Flag positions are anticipated by software.
4.2 ELECTRONIC SUBSYSTEM OVERVIEW 4.2.3.4 AIR SENSORS The mechanism subsystem includes two air sensors, used to detect air passage into (proximal) or out of (distal) the cassette. Both sensors are piezoelectric crystal transmitter receiver pairs. Liquid between the transmitter and receiver will conduct the ultrasonic signal, while air will not (see Figure 4-5). XTL TX VSO VCO/PLL CPU Figure 4-5. G_RX G_TX OUTPUT AMP A/D XTL RX PEAK DETECTOR Air Sensor Block Diagram 4.2.3.4.
SECTION 4 THEORY OF OPERATION 4.2.3.4.2 Receiver Circuitry When the cassette’s test port is filled with fluid, the transmitted signal will be coupled to an identical piezoelectric crystal, where it is amplified and detected by the receiver circuitry. The receiver circuitry consists of an amplifier, a peak detector, and an adjustable gain buffer stage. There is a separate, symmetrical receiver circuit for each channel (proximal and distal).
4.2 PRESSURE SENSOR ENABLE (FROM CPU) ELECTRONIC SUBSYSTEM OVERVIEW BRIDGE EXCITATION 3.75 V WHEATSTONE BRIDGE REFERENCE 2.5 V DIFFERENTIAL AMPLIFIER AND OFFSET ADJUST OUTPUT AMPLIFIER AND FILTER Figure 4-6. PRESSURE SIGNAL OUTPUT (TO CPU) Pressure Sensor Excitation and Amplifier Block Diagram 4.2.3.5.1 Bridge Excitation Supply The bridge excitation voltage is 3.75 VDC, and is derived from the 2.5 VDC reference signal (V2_5), gained 1.5 times by an amplifier.
SECTION 4 THEORY OF OPERATION 4.2.3.6 PRESSURE SENSOR CALIBRATION Pressure sensors are calibrated for offset and gain during mechanism calibration. A trimming potentiometer is used to adjust the initial, zero pressure offset. The proximal and distal pressure sensors have independent offset adjustments. The final system gain (cassette pressure to corrected amplifier output) is adjusted in software.
4.3 PRINTED WIRING ASSEMBLIES 4.3 PRINTED WIRING ASSEMBLIES Infusion system electronics are packaged into six printed wiring assemblies (PWA) and several remote mounted peripherals (see Section 4.4). The following sections provide a brief description of the functional interfaces of each PWA. 4.3.
SECTION 4 THEORY OF OPERATION 4.3.2 PERIPHERAL PWA The peripheral PWA contains part of the CPU subsystem circuitry, including system program and data memories (PROM and SRAM), and external communication interface circuits. The peripheral PWA is designed to be field replaceable, to facilitate software upgrades or additional external interfaces. The peripheral PWA is a four layer board that includes the following: one ground plane, one power plane, and two signal layers.
4.3 PRINTED WIRING ASSEMBLIES 4.3.4 CPU PWA The CPU PWA contains most of the CPU subsystem functions, with the exception of main memory and communications ports that are located on the peripheral PWA. The CPU PWA also accommodates system interconnect. The CPU PWA is an eight layer board with the following: one ground plane, one power plane, and six signal layers. The CPU PWA primarily contains SMT components.
SECTION 4 THEORY OF OPERATION See Table 4-9 for driver PWA interface connections. Table 4-9. Connector Driver PWA Interface Connections Type Interface J7 6 pin header Plunger motor J8 6 pin header Input/output motor J9 6 pin header Line select motor J10 20 pin SMT Flat flex cable to APP PWA J11 50 pin header Ribbon cable to CPU PWA J12 6 pin SMT FSR flex circuit J13 4 pin header Motor power, from power supply PWA J14 8 pin SMT Flat flex cable to switch PWA 4.3.
4.4 REMOTE MOUNTED PERIPHERALS 4.4 REMOTE MOUNTED PERIPHERALS Remote mounted peripherals include the LCD, sealed lead acid battery, and barcode reader (BCR) wand. 4.4.1 LCD The infuser uses a graphic LCD module with a CCFT. The CCFT provides a backlight source for the LCD. The LCD requires a nominal -16 VDC supply for contrast control that is controlled by the CPU.
SECTION 4 THEORY OF OPERATION 4.5.1 CASSETTE The cassette (see Figure 4-7 and Figure 4-8) operates on a fluid displacement principle to volumetrically deliver fluid. See the system operating manual for a description of the major cassette functions. The pumping cycle begins when the outlet valve is opened and the inlet valve is closed. The plunger extends to deflect the cassette diaphragm and expel fluid.
4.5 MECHANICAL OVERVIEW LINE A A VALVE LINE B B VALVE AIR TRAP CHAMBER AIR-IN-LINE SENSOR (PROXIMAL) INLET VALVE PRESSURE SENSOR (PROXIMAL) OUTLET VALVE PUMPING CHAMBER PRESSURE SENSOR (DISTAL) AIR-IN-LINE SENSOR (DISTAL) PRECISION GRAVITY FLOW REGULATOR (AND SHUT OFF) Figure 4-7.
SECTION 4 THEORY OF OPERATION 4.5.2 MECHANISM ASSEMBLY The mechanism assembly is a fully self-contained unit consisting of the motor and valve assemblies, A/B valve subsystem, inlet/outlet valve subsystem, plunger drive subsystem, air bubble (ultrasonic) sensor assemblies, cassette door, and pressure sensor assemblies. The motor and valve assemblies, A/B valve subsystem, inlet/outlet valve subsystem, and plunger drive subsystem are detailed in the following sections.
4.5 MECHANICAL OVERVIEW B VALVE A VALVE REGULATOR ACTUATOR AIR-IN-LINE SENSOR (DISTAL) AIR-IN-LINE SENSOR (PROXIMAL) PRESSURE SENSOR (DISTAL) PRESSURE SENSOR (PROXIMAL) CASSETTE LOCATOR FORCE SENSING RESISTOR OUTLET VALVE PLUNGER Figure 4-9. INLET VALVE Mechanism Valve Pins and Sensor Locations 4.5.2.3 INLET/OUTLET VALVE SUBSYSTEM The inlet/outlet valve subsystem is similar in function and build to the A/B valve subsystem (see Section 4.5.2.2). 4.5.2.
SECTION 4 THEORY OF OPERATION This page intentionally left blank.
Section 5 MAINTENANCE AND SERVICE TESTS A complete maintenance program promotes infusion system longevity and trouble-free operation. Such a program should include routine maintenance, periodic maintenance inspection, and following any repair procedure, performance verification testing. 5.1 ROUTINE MAINTENANCE Routine maintenance consists of basic inspection and cleaning procedures. As a minimum requirement, inspect and clean the infuser after each use.
SECTION 5 MAINTENANCE AND SERVICE TESTS Note: Disinfecting properties of cleaning solutions vary, and not all cleaning solutions are sanitizers. Check product labeling or consult the manufacturer for specific information. Table 5-1.
5.2 PERFORMANCE VERIFICATION TEST 5.2.1 EQUIPMENT REQUIRED The PVT requires the following equipment, or equivalents: - Graduated cylinder, 25 mL, with 0.2 mL graduations (Type A) - Sterile water or tap water in an IV bag/container - Digital pressure meter (DPM), 0 to 50 psi (Fluke® Biomedical DPM3) - Three-way stopcock, latex-free (List No. 3233-01) - IV Set (List No. 1642) - Secondary Piggyback Set (List No. 1832) - 21-gauge butterfly needle, latex-free (List No.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.2.3 TEST SETUP WARNING: A PATIENT SHOULD NEVER BE CONNECTED TO THE INFUSER DURING DEVICE TESTING. To set up the infuser for the PVT, proceed as follows: 1. Confirm the infuser and appropriate accessories are assembled. 2. Hang two sterile water containers at a height of 18 ± 6 inches above the pumping chamber of the infuser. 3. Connect the infuser to AC power, and press [ON/OFF] to turn on the device. 4. Verify the infuser is in the unlocked mode.
5.2 PERFORMANCE VERIFICATION TEST LINE FLOW INDICATORS A B HOSPIRA Plum A+ Version X.XX - MM/DD/YY Copyright HOSPIRA 2009 STATUS REGION System Self Test In Progress WORKING REGION MESSAGE REGION SOFTKEY LABEL REGION START STOP CHARGE LINE INDICATOR ON / OFF 1 4 7 CLEAR Figure 5-1. Technical Service Manual 2 5 8 0 3 6 9 .
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.2.5 CASSETTE ALARM TEST To perform the cassette alarm test, proceed as follows: 1. Verify the infuser is on. Insert an empty cassette and close the door. 2. Verify the CASSETTE TEST FAIL message is flashing on the display and the alarm sounds after the cassette test is complete. 3. Open the door and remove the cassette. 4. Turn off the infuser. 5.2.6 FREE FLOW TEST To perform the free flow test, proceed as follows: 1.
5.2 A STOPPED PERFORMANCE VERIFICATION TEST STOPPED B Dopamine 0.0 Rate mL/hr 0.0 0.0 Vol/Inf mL 0.0 Back Prime A Figure 5-2. B Options/ Vol Inf Primary Screen 5.2.8 KEYPAD VERIFICATION/FUNCTIONAL TEST To perform the keypad verification/functional test, see Figure 5-1, then proceed as follows: 1. With the infuser in the PRIMARY screen, press the [A] softkey to select line A. 2. Verify the PROGRAM screen is displayed. 3. Enter a rate of 123 mL/hr and VTBI of 4567 mL. 4.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.2.9 ALARM LOUDNESS TEST To perform the alarm loudness test, proceed as follows: 1. Press the [A] softkey to select line A. 2. Enter a rate of 400 mL/hr and VTBI of 1 mL. 3. Press [START] and verify fluid is pumping, the message PUMPING is displayed in the line A status bar, and the line A LED flashes. 4. Verify the alarm sounds when the dose has been delivered. 5. Turn the volume control knob between HIGH and LOW.
5.2 PERFORMANCE VERIFICATION TEST 5.2.10 LOCKOUT SWITCH TEST To perform the lockout switch test, proceed as follows: 1. Press the [A] softkey to select line A. 2. Enter a rate of 400 mL/hr and VTBI of 50 mL. 3. Press [START] and verify fluid is pumping, the message PUMPING is displayed in the line A status bar, and the line A LED flashes. 4. Toggle the lockout alarm switch up (ON) to engage the alarm (see Figure 5-3). 5.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.2.12 PROXIMAL AIR-IN-LINE TEST To perform the proximal air-in-line alarm test, see Figure 5-4, then proceed as follows: 1. Install the special cassette marked proximal, and close the cassette door. Note: Confirm the special cassette proximal bubble sensor tips are removed. 2. Press [YES] to clear settings. 3. Press the [A] softkey to select line A. 4. Enter a rate of 400 mL/hr and VTBI of 50 mL. 5.
5.2 PERFORMANCE VERIFICATION TEST 5.2.13 DISTAL AIR-IN-LINE TEST To perform the distal air-in-line alarm test, see Figure 5-4, then proceed as follows: 1. Install the special cassette marked distal, and close the cassette door. Note: Confirm the special cassette distal bubble sensor tips are removed. 2. Press [YES] to clear settings. 3. Press the [A] softkey to select line A. 4. Enter a rate of 400 mL/hr and VTBI of 50 mL. 5.
SECTION 5 MAINTENANCE AND SERVICE TESTS 14. Open and close the door. Press [NO] at the CLEAR SETTINGS? prompt. 15. Press [OPTIONS/VOL INF] to select the OPTIONS screen. 16. Select Pressure/Post Infusion Rate, and press [CHOOSE]. 17. Select Distal Pressure Limit. Enter 10 psi, and press [ENTER]. 18. Set the three-way stopcock to measure pressure, then press [START]. 19. Verify the distal occlusion audible alarm occurs at 10 ± 3 psi. Verify the DISTAL OCCLUSION message is flashing on the screen. 20.
5.2 PERFORMANCE VERIFICATION TEST 5.2.15 DELIVERY ACCURACY TEST Note: Accuracy testing is for informational purposes only, and is not to be used as a re-release test. If there is any concern as to infuser accuracy, contact Hospira. CAUTION: Do not remove the protective cover from the 21-gauge needle. To perform the delivery accuracy test, proceed as follows: 1. Open the cassette door and insert a primed cassette. Close the cassette door. 2. Press [YES] to clear settings. 3.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.2.16 NURSE CALL TEST Note: The nurse call test may be bypassed if the nurse call function is not used. To perform the nurse call test, attach the nurse call test cable and proceed as follows: 1. Set the primary delivery rate to 400 mL/hr, and the primary dose limit to 1 mL. 2. Connect the DMM to the nurse call test cable. 3. Press [START] and verify pumping action. 4.
5.3 PERIODIC MAINTENANCE INSPECTION 5.2.18 END OF THE PVT If all performance verification tests have been successful, proceed as follows: 1. Press [OPTIONS/VOL INF]. Select Volume Infused, and press [CHOOSE]. 2. Press [CLEAR] to clear the volume infused, then press [ENTER]. 3. Press the [A] softkey. Press the [CLEAR PROGRAM] softkey. Note: [CLEAR PROGRAM] will appear only if there is a program in the A or B channel. 4. Press [YES] at the CLEAR LINE A SETTINGS? prompt. 5.
SECTION 5 MAINTENANCE AND SERVICE TESTS 5.4 BATTERY OPERATION OVERVIEW The infusion system is intended to operate on battery power on an exception basis only, such as emergency backup or temporary portable operation. Examples of emergency backup include AC power failure or inadvertent disconnection of the AC power cord. An instance of temporary portable operation includes patient transfer from one location to another.
Section 6 TROUBLESHOOTING This section contains information on technical assistance, warning messages, alarm messages and error codes, and troubleshooting procedures. 6.1 TECHNICAL ASSISTANCE For technical assistance, product return authorization, and to order parts, accessories, or manuals within the United States, contact Hospira. 1-800-241-4002 For additional technical assistance, technical training, and product information, visit the website at www.hospira.com.
SECTION 6 TROUBLESHOOTING Table 6-1. Message Warning Messages Possible Cause Corrective Action Stop delivery, then turn off Attempting to turn off the infuser while a delivery is in progress Stop all lines, then turn off the infuser Warning: Low Battery Battery is discharged so that only approximately 30 minutes of battery life remain Plug into AC power Warning: Replace Battery Battery service needed Replace the battery (see Section 7.2.
6.3 Table 6-2.
SECTION 6 TROUBLESHOOTING Table 6-2. Alarm Code Operational Alarm Messages and Corrective Actions Alarm Description Possible Cause Corrective Action N183 or E183 (OP2) Prox. Occl B, Air or Prox. Occl B Peak proximal occlusion B, non-delivery Proximal occlusion detected on line B during non-delivery Backprime the cassette and restart line B or Stop all lines, backprime the cassette, and restart all lines N184 or E184 (OP1) Prox. Occl A, Air or Prox.
6.3 Table 6-2. Alarm Code ALARM MESSAGES AND ERROR CODES Operational Alarm Messages and Corrective Actions Alarm Description Possible Cause Corrective Action N191 or E191 (OP1) Prox. Occl A, Air Peak proximal occlusion A, delivery Proximal occlusion detected during delivery on line A Fix occlusion and restart line A or Stop all lines, fix occlusion and restart the infuser N230 or E230 (APT) Prox.
SECTION 6 TROUBLESHOOTING Table 6-2. Alarm Code Operational Alarm Messages and Corrective Actions Alarm Description Possible Cause Corrective Action N252 or E252 (BDP) Depleted battery Low battery Battery terminal voltage is less than 5.
6.3 ALARM MESSAGES AND ERROR CODES 6.3.2 ERROR CODES REQUIRING TECHNICAL SERVICE Table 6-3 lists infusion system error codes that require technical service. Also listed in Table 6-3 are malfunction descriptions, possible causes, and corrective actions. Table 6-3. Error Code Error Codes Requiring Technical Service Malfunction Possible Cause Corrective Action E300 ADC failure Analog to digital converter failure Replace CPU PWA (see Section 6) Reset time and date, if required (see Section 1.8.
SECTION 6 TROUBLESHOOTING Table 6-3. Error Code Error Codes Requiring Technical Service Malfunction Possible Cause Corrective Action E326 Battery disconnected Battery disconnected while the infuser is on Check for loose battery connections Replace battery (see Section 7.2.4) Reset time and date, if required (see Section 1.8.3) E327 Brownout condition Brownout condition detected Replace power supply PWA (see Section 7.2.14.1) Reset time and date, if required (see Section 1.8.
6.3 Table 6-3.
SECTION 6 TROUBLESHOOTING Table 6-3. Error Code Error Codes Requiring Technical Service Malfunction Possible Cause Corrective Action E443 LCD failure LCD bias is out of range Replace display assembly (see Section 7.2.14.3) Reset time and date, if required (see Section 1.8.3) E444 CPU timebase inaccurate CPU timer 2 and RTC measured times disagree Turn power off, then on, to reset the infuser Replace CPU PWA (see Section 7.2.14.6) Reset time and date, if required (see Section 1.8.
6.4 Table 6-3. Error Code TROUBLESHOOTING PROCEDURES Error Codes Requiring Technical Service Malfunction Possible Cause Corrective Action E455 Invalid device configuration Incorrect flash memory on peripheral PWA Turn power off, then on, to reset the infuser Replace peripheral PWA (see Section 7.2.
SECTION 6 TROUBLESHOOTING Table 6-4. Test Failure Troubleshooting with the PVT Probable Cause Corrective Action Self Test Section 5.2.4 Cassette not properly installed Reseat cassette Defective CPU PWA Replace CPU PWA (see Section 7.2.14.6) Cassette Alarm Test Section 5.2.5 Cassette not properly seated Reseat cassette Defective cassette Replace cassette Free Flow Test Section 5.2.
6.4 Table 6-4. Test Failure Distal Occlusion Test Section 5.2.14 Delivery Accuracy Test Section 5.2.15 Electrical Safety Test Section 5.2.16 Technical Service Manual TROUBLESHOOTING PROCEDURES Troubleshooting with the PVT Probable Cause Corrective Action Cassette not properly primed Re-prime cassette Defective cassette Replace cassette Dirty sensor pin Clean sensor pin Defective APP PWA Replace mechanism assembly (see Section 7.2.14.
SECTION 6 TROUBLESHOOTING This page intentionally left blank.
Section 7 REPLACEABLE PARTS AND REPAIRS This section itemizes all parts and subassemblies of the infusion system that are repairable within the scope of this manual. In addition, this section details replacement procedures for all listed parts. 7.1 REPLACEABLE PARTS Replaceable parts for the infusion system are itemized in the Illustrated Parts Breakdown (IPB) and are identified in Figure 9-1. Table 9-2 identifies each part by an index number that correlates to Figure 9-1.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.2 REQUIRED TOOLS AND MATERIALS The following tools and materials, or equivalents, are required for the replacement procedures in this section. In addition, the beginning of each procedure lists tools and materials required for that specific procedure.
7.2 REPLACEMENT PROCEDURES RUBBER FOOT PAD (4) BATTERY DOOR (3) 6-32 x 1/2 PAN HEAD SCREW (4) Figure 7-1. Bottom View 7.2.
SECTION 7 REPLACEABLE PARTS AND REPAIRS To replace a battery, wire harness, door, or door pad see Figure 7-2, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Place the infuser face down on a soft flat surface. 3. Using the flat blade screwdriver, remove the screw and retaining ring that attach the battery door to the infuser, and remove the door. 4. Inspect the battery door and replace, if required. 5.
7.2 REPLACEMENT PROCEDURES AC POWER CORD BATTERY WITH WIRE HARNESS ASSEMBLY (3) VELCRO STRAP DOOR PAD (3) 4-40 x 3/8 PAN HEAD SCREW EQUIPOTENTIAL TERMINAL RETAINING RING (3) POWER CORD RETAINER BATTERY DOOR (3) 6-32 x 1/2 HEX HEAD SCREW (3) Figure 7-2.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.5 AC POWER CORD, RETAINER, AND VELCRO STRAP REPLACEMENT The recommended tool for this procedure is a #2 Phillips screwdriver. Replacement parts for this procedure are: Cordset, AC Power, Hospital Grade, Detachable Retainer, AC Power Cord Strap, Velcro, AC Power Cord Screw, 4-40 x 3/8, Pan Head, Phillips To replace the AC power cord, retainer, or Velcro strap see Figure 7-2, then proceed as follows: 1.
7.2 REPLACEMENT PROCEDURES 7.2.6 SEPARATING THE FRONT ENCLOSURE, REAR ENCLOSURE, AND MAIN CHASSIS ASSEMBLY The front enclosure consists of an upper assembly and a lower assembly. The main chassis assembly consists of an upper chassis and a lower chassis. The recommended tool for this procedure is a #2 Phillips screwdriver.
SECTION 7 REPLACEABLE PARTS AND REPAIRS UPPER FRONT ENCLOSURE FRONT/REAR ENCLOSURE GASKET (3) PERIPHERAL INTERFACE ASSEMBLY 6-32 x 1 1/4 PAN HEAD SCREW (5) 6-32 x 2 3/4 PAN HEAD SCREW (2) #6 FLAT WASHER (8) 6-32 x 1/2 PAN HEAD SCREW LOWER FRONT ENCLOSURE MAIN CHASSIS ASSEMBLY 8-32 x 3 1/2 PAN HEAD SCREW (3) REAR ENCLOSURE #8 FLAT WASHER (3) Figure 7-3.
7.2 REPLACEMENT PROCEDURES 8 1 4 6 3 2 9 7 Figure 7-4. 5 Screw Placement Sequence 7.2.7 PERIPHERAL INTERFACE ASSEMBLY REPLACEMENT CAUTION: Peripheral interface assembly replacement should only be performed after receiving approval from Hospira. Replacement parts for this procedure are: Assembly, Peripheral Interface Assembly, Cable, Peripheral #1 Assembly, Cable, Peripheral #2 To replace the peripheral interface assembly proceed as follows: 1.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7. Install the replacement peripheral interface assembly in the exact reverse order of removal. Note: Verify the peripheral interface PWA is placed properly between the guides and fits correctly into the CPU PWA. 8. Reinstall the batteries and connect the infuser to AC power. 9. Turn on the infuser, and verify completion of the self test. To verify successful peripheral interface assembly replacement, perform the PVT in Section 5.2. 7.2.
7.2 REPLACEMENT PROCEDURES PERIPHERAL PWA #2 PERIPHERAL PWA #1 PERIPHERAL CABLE #2 PERIPHERAL CABLE #1 RETENTION CLIPS PERIPHERAL INTERFACE ASSEMBLY Figure 7-5. Peripheral Interface Assembly and Peripheral PWAs 7.2.9 PERIPHERAL COMPONENT REPLACEMENT Peripheral component replacement includes the replacement of the volume control knob and the peripheral cover. To replace peripheral components see Figure 7-6, then proceed as detailed in the following sections.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.9.1 VOLUME CONTROL KNOB REPLACEMENT Recommended tools for this procedure are a medium size flat blade screwdriver, an X-acto knife, and long needle nose pliers. Replacement parts for this procedure are: Assembly, Volume Control Knob Cap, Knob Cover, Knob To replace the volume control knob see Figure 7-6, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3.
7.2 REPLACEMENT PROCEDURES 3. Remove the peripheral interface assembly as described in Section 7.2.7. 4. Remove the volume control knob as described in Section 7.2.9.1. 5. Using a 5/16 nutdriver, remove the nut that secures the potentiometer to the peripheral cover. Using the needle nose pliers, remove the lock washer. 6. Using the custom nutdriver, remove the hex nut that secures the nurse call jack to the peripheral cover. 7. Inspect the rectangle seal and round seals and replace, if required. 8.
SECTION 7 REPLACEABLE PARTS AND REPAIRS PERIPHERAL COVER 4-40 HEX NUT (4) FLAT WASHER (2) ESD SPRING 4-40 x 3/8 HEX SCREW (2) POTENTIOMETER ROUND SEAL (2) KNOB COVER KNOB CAP VOLUME CONTROL KNOB RECTANGLE SEAL NURSE CALL JACK NUT PERIPHERAL INTERFACE PWA 4-40 x 3/8 PAN HEAD SCREW (2) Figure 7-6.
7.2 REPLACEMENT PROCEDURES 7.2.10 FRONT/REAR ENCLOSURE GASKET REPLACEMENT The recommended tool for this procedure is needle nose pliers. The replacement part for this procedure is: Gasket, Front/Rear Enclosure Note: Clean and remove any foreign matter on the replacement gasket or in the spaces where the replacement gasket is to be installed. To replace a front/rear enclosure gasket see Figure 7-3, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2.
SECTION 7 REPLACEABLE PARTS AND REPAIRS LOWER FRONT ENCLOSURE TOP SEAL GASKET (3) KEYPAD GASKET (3) EMI D-SHAPE GASKET (4) Figure 7-7. Lower Front Enclosure Gaskets 7.2.11.1 EMI GASKET REPLACEMENT The recommended tool for this procedure is needle nose pliers. The replacement part for this procedure is: Gasket, EMI, D-Shape, 6.62 in. Gasket, EMI, D-Shape, 13 in. To replace an EMI gasket see Figure 7-7, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2.
7.2 REPLACEMENT PROCEDURES 7.2.11.2 KEYPAD GASKET REPLACEMENT The recommended tool for this procedure is needle nose pliers. The replacement part for this procedure is: Gasket, Keypad To replace a keypad gasket see Figure 7-7, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3. Separate the front enclosure, rear enclosure, and main chassis assembly as described in Section 7.2.6. 4.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.12 REAR ENCLOSURE ASSEMBLY COMPONENT REPLACEMENT Rear enclosure assembly component replacement includes replacement of the following: - Rear enclosure gaskets - Pole clamp assembly, backing plate, and insulator - Equipotential terminal - Internal AC power cord - AC connector - Fuses To replace the rear enclosure assembly components see Figure 7-8 and Figure 7-9, then proceed as detailed in the following sections.
7.2 Figure 7-9.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.12.1 POLE CLAMP ASSEMBLY AND BACKING PLATE REPLACEMENT Recommended tools for this procedure are a set of nutdrivers and a medium size flat blade screwdriver.
7.2 REPLACEMENT PROCEDURES 7.2.12.2 INTERNAL AC POWER CORD REPLACEMENT Recommended tools for this procedure are a medium size flat blade screwdriver, needle nose pliers, and a set of nutdrivers. Replacement parts for this procedure are: Cord, Internal, AC Power Clamp, Internal Power Cord Wire, Ground, AC Power Screw, 4-40 x 3/8, Hex Head, Slotted Washer, Flat, 1/4 Washer, Lock, 1/4 To replace the internal AC power cord see Figure 7-9, then proceed as follows: 1.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 5. Using the needle nose pliers, remove the internal power cord wires and the AC ground wire from the AC connector. 6. Using the Phillips screwdriver, remove the screws that secure the AC connector to the rear enclosure. 7. Install the replacement AC connector in the exact reverse order of removal. 8. Join the rear enclosure and main chassis assembly in the exact reverse order of separation. 9. Reinstall the batteries and connect the infuser to AC power.
7.2 REPLACEMENT PROCEDURES 7.2.12.5 REAR ENCLOSURE GASKET REPLACEMENT The recommended tool for this procedure is needle nose pliers. Replacement parts for this procedure are: Gasket, Rear Enclosure, 20 13/16 Gasket, Rear Enclosure, 45 7/8 To replace a rear enclosure gasket see Figure 7-9, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3. Separate the rear in Section 7.2.6.
SECTION 7 REPLACEABLE PARTS AND REPAIRS CLUTCH HOUSING CLUTCH SPRING BAG HANGER COTTER RING Figure 7-10. Minipole Assembly 7.2.
7.2 REPLACEMENT PROCEDURES MECHANISM/CHASSIS BUMPER (6) MECHANISM ASSEMBLY (3) 6-32 x 1/2 HEX HEAD SCREW (9) INSULATION TAPE (3) UPPER CHASSIS KEYPAD (3) CPU/DRIVER CABLE (3) 4-24 x 1/4 PAN HEAD SCREW (3) 4-40 x 3/16 HEX HEAD SCREW (12) LOWER CHASSIS DISPLAY ASSEMBLY (3) Figure 7-11.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 4-40 x 3/8 HEX HEAD SCREW (6) PIEZO ALARM (3) CPU/DRIVER CABLE (3) INSULATION TAPE (3) POWER SUPPLY/ BATTERY CABLE (3) MOTOR POWER CABLE (3) PERIPHERAL PWA (2) POWER SUPPLY (3) PERIPHERAL CABLE #2 4-40 x 1/4 HEX HEAD SCREW (3) CPU PWA (3) PERIPHERAL INTERFACE PWA PERIPHERAL CABLE #1 Figure 7-11.
7.2 REPLACEMENT PROCEDURES 7.2.14.1 POWER SUPPLY PWA REPLACEMENT The recommended tool for this procedure is a medium size flat blade screwdriver. Replacement parts for this procedure are: PWA, Power Supply Assembly, Cable, Power Supply/Battery To replace a power supply PWA see Figure 7-11 (2 of 2), then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 6. Using the Phillips screwdriver, remove the screw that secures the keypad and display to the lower main chassis. 7. Carefully disconnect the flex ribbon cable from the display by pushing the connector locking tabs down. 8. Using the flat blade screwdriver, separate the keypad and display by removing the screws that secure the keypad to the display. 9. Install the replacement keypad in the exact reverse order of removal. 10.
7.2 REPLACEMENT PROCEDURES 7.2.14.4 CPU/DRIVER CABLE REPLACEMENT The recommended tool for this procedure is an X-acto knife. Replacement parts for this procedure are: Assembly, Cable, CPU/Driver Tape, Ferrite Tape, Insulation To replace a CPU/driver cable see Figure 7-11 (1 of 2), Figure 7-12, and Figure 7-13, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3.
SECTION 7 REPLACEABLE PARTS AND REPAIRS TOP TAB CABLE CENTER TAB FERRITE LOWER TAB Figure 7-12.
7.2 REPLACEMENT PROCEDURES FERRITE INSULATION TAPE Figure 7-13. Ferrite Tape Positioning (1 of 2) FERRITE CENTER TAB TAPE Figure 7-13.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.14.5 MOTOR POWER CABLE REPLACEMENT The recommended tool for this procedure is a medium size flat blade screwdriver. The replacement part for this procedure is: Assembly, Cable, Motor Power To replace a motor power cable see Figure 7-11 (2 of 2), then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3.
7.2 REPLACEMENT PROCEDURES 10. Disconnect the piezo alarm cable from the CPU PWA. 11. Using the Phillips screwdriver, remove the screw from the bottom of the lower main chassis assembly. 12. Slide the CPU PWA out of the main chassis until J5 on the CPU PWA is accessible. 13. Using the flat blade screwdriver, release the locking tabs that secure the flex cable to the CPU PWA. 14. Install the replacement CPU PWA in the exact reverse order of removal. 15.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.14.8 MECHANISM ASSEMBLY REPLACEMENT Note: Replacing the mechanism changes the biomed settings to those stored in the replacement mechanism assembly. Recommended tools for this procedure are a medium size flat blade screwdriver and #2 Phillips screwdriver.
7.2 REPLACEMENT PROCEDURES 7.2.14.9 CASSETTE DOOR AND FLUID SHIELD REPLACEMENT The recommended tool for this procedure is a medium size flat blade screwdriver. Replacement parts for this procedure are: Assembly, Cassette Door Assembly, Fluid Shield Cap, Door Pivot Spring, Extension Screw, 4-40 x 3/8, Hex Head, Slotted, with Washer To replace the cassette door and fluid shield see Figure 7-14 and Figure 7-15, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2.
SECTION 7 REPLACEABLE PARTS AND REPAIRS APPROX 15 DEGREE FLUID SHIELD MECHANISM ASSEMBLY CASSETTE DOOR Figure 7-14.
7.2 REPLACEMENT PROCEDURES FLEX CONNECTOR TORSION SPRING RETAINING RING MECHANISM ASSEMBLY CASSETTE DOOR OPENER HANDLE 4-40 x 3/8 HEX HEAD SCREW WITH WASHER 3/32 PUSH-ON RETAINING RING DOOR PIVOT CAP Figure 7-15.
SECTION 7 REPLACEABLE PARTS AND REPAIRS 7.2.14.10 OPENER HANDLE ASSEMBLY REPLACEMENT The recommended tool for this procedure is a medium size flat blade screwdriver. Replacement parts for this procedure are: Assembly, Opener Handle Link, Door Ring, Retaining, Push-On To replace the opener handle assembly see Figure 7-15, then proceed as follows: 1. Turn off the infuser, and disconnect the device from AC power. 2. Remove the batteries as described in Section 7.2.4. 3.
Section 8 SPECIFICATIONS The following specifications apply to the Plum A+3 infusion system. PHYSICAL Dimensions: Approximately 19 H x 15 W x 14 D inches (excluding pole clamp and power cord storage) Weight: Approximately 28 lbs (with batteries) Casing: High-impact plastic ELECTRICAL Power Requirements: 95-132 VAC; 47-62 Hz; 90 W Power Cord: Hospital-grade AC cord; 10 feet; with transparent plug and retainer plate Fuses: 0.
SECTION 8 SPECIFICATIONS ENVIRONMENT Operating: 41° to 104° F (5° to 40° C); 10 % to 90 % relative humidity Transporting and Storage: -4° to 140° F (-20° to 60° C); 10 % to 90 % relative humidity Atmospheric Pressure: 0 - 10,000 feet (0 - 3000 meters) or equivalent atmospheric pressure Relative Humidity: 10 - 90 % (104° F max) DELIVERY RATE RANGE Lines A and B: 0.1 to 99.9 mL/hr (in 0.1 mL/hr increments) 100 to 999 mL/hr (in 1 mL/hr increments) Concurrent Delivery: 0.
SECTION 9 DRAWINGS Section 9 DRAWINGS Figure 9-1 through Figure 9-10 show the Illustrated Parts Breakdown (IPB) and assembly drawings. Table 9-1 lists drawings by figure number and title. Table 9-2 identifies parts by index numbers which correlate to Figure 9-1. Drawings in Section 9 are provided as information only, and may not exactly reflect current product configuration. Table 9-1.
SECTION 9 DRAWINGS Table 9-2. Index No. IPB for the Infuser Nomenclature Replacement Procedure 9 Assembly, Fluid Shield Section 7.2.14.9 10 Assembly, Opener Handle Section 7.2.14.10 11 Assembly, Cassette Door Section 7.2.14.9 12 PWA, Peripheral Section 7.2.8 13 Assembly, Peripheral Interface Section 7.2.7 14 Cover, Peripheral Section 7.2.9.2 15 Shield, Spring, ESD Section 7.2.9.2 16 Assembly, Keypad Section 7.2.14.2 17 Assembly, Display Section 7.2.14.
SECTION 9 DRAWINGS Table 9-2. Index No. IPB for the Infuser Nomenclature Replacement Procedure 37 Assembly, Minipole A: Hanger, Bag B: Housing, Clutch C: Ring, Cotter D: Spring, Clutch Section 7.2.13 38 Assembly, Cable, Peripheral #2 Section 7.2.7 39 Assembly, Cable, Peripheral #1 Section 7.2.7 40 Assembly, Cable, CPU/Driver Section 7.2.14.4 41 Assembly, Piezo Alarm Section 7.2.14.7 42 Battery Section 7.2.4 43 Assembly, Wire Harness, Battery Section 7.2.
SECTION 9 DRAWINGS Table 9-2. Index No. IPB for the Infuser Nomenclature Replacement Procedure 65 Screw, 4-24 x 1/4, Pan Head, Phillips, SS As applicable 66 Screw, 4-40 x 1/4, Hex Head, Slotted, with Washer As applicable 67 Screw, 4-40 x 3/8, Pan Head, Phillips, SS As applicable 68 Screw, 4-40 x 3/8, Hex Head, Nylon Section 7.2.9.
1 8 3 2 9 23 10 11 24 25 26 4 12 14 13 27 28 15 29 6 5 16 17 30 31 18 32 21 20 19 33 7 35 34 22 36 37 37A 37B HOSPIRA, INC. 37D Figure 9-1. Illustrated Parts Breakdown 37C DRAWING NO. NOT APPLICABLE Technical Service Manual 9-5 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
39 38 56 57 40 41 58 59 42 43 61 60 75 44 63 62 76 77 66 65 64 78 47 46 45 79 68 67 80 48 81 49 70 69 82 50 83 51 71 84 85 52 53 54 73 72 55 74 86 HOSPIRA, INC. Figure 9-1. Illustrated Parts Breakdown DRAWING NO. NOT APPLICABLE Technical Service Manual 9-7 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
50 3 78 75 1 2 71 4 79 76 78 HOSPIRA, INC. Figure 9-2. Front Enclosures, Rear Enclosure, and Main Chassis Assembly 72 DRAWING NO. NOT APPLICABLE Technical Service Manual 9-9 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
3 50 54 2 20 48 HOSPIRA, INC. 78 53 71 Figure 9-3. Front Enclosure Assemblies 49 DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 11 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
26 74 28 85 52 82 30 68 77 33 82 84 1 82 84 29 HOSPIRA, INC. 51 Figure 9-4. Rear Enclosure Assembly DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 13 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
1 27 59 67 57 20 HOSPIRA, INC. 78 58 Technical Service Manual 56 Figure 9-4. Rear Enclosure Assembly 71 DRAWING NO. NOT APPLICABLE 9 - 15 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
14 15 12 68 60 34 80 86 61 67 HOSPIRA, INC. Figure 9-5. Peripheral Interface Assembly DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 17 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
69 41 40 47 46 32 31 12 38 39 7 18 66 HOSPIRA, INC. Figure 9-6. Main Chassis Assembly 13 Technical Service Manual DRAWING NO. NOT APPLICABLE 9 - 19 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
73 19 6 8 46 16 65 5 HOSPIRA, INC. 70 Figure 9-6. Main Chassis Assembly DRAWING NO. NOT APPLICABLE 17 Technical Service Manual 9 - 21 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
18 70 17 16 40 HOSPIRA, INC. Figure 9-7. CPU PWA, Display, and Keypad DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 23 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
4 18 77 66 HOSPIRA, INC. Figure 9-8. CPU PWA and Main Chassis DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 25 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
37 25 23 42 67 24 45 64 44 43 73 HOSPIRA, INC. Figure 9-9. AC Power Cord, Retainer, Batteries, and Minipole DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 27 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
9 11 8 10 64 55 HOSPIRA, INC. Figure 9-10. Mechanism Assembly 69 DRAWING NO. NOT APPLICABLE Technical Service Manual 9 - 29 Rev.
SECTION 9 DRAWINGS This page intentionally left blank.
APPENDIX USE OF THE INFUSION SYSTEM IN ELECTROMAGNETIC ENVIRONMENTS EN-2 The Plum A+3 Infusion System is intended for use in the electromagnetic environment specified in Table A-1, Table A-2, Table A-3, and Table A-4. The user of the infusion system should assure that it is used only in the appropriate environment. ELECTROMAGNETIC EMISSIONS Table A-1 details electromagnetic emissions compliance and guidance. Table A-1.
APPENDIX ELECTROMAGNETIC IMMUNITY Table A-2 details guidance for the electromagnetic environment. Table A-2.
APPENDIX ELECTROMAGNETIC IMMUNITY FOR LIFE-SUPPORTING EQUIPMENT AND SYSTEMS Table A-3 provides guidance for use of the infusion system near communications equipment. Table A-3.
APPENDIX a The industrial, scientific and medical (ISM) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.660 MHz to 40.700 MHz. b The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas.
APPENDIX RECOMMENDED SEPARATION DISTANCES FOR COMMUNICATIONS EQUIPMENT The infusion system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The recommendations provided in Table A-4 help the user of the infusion system to prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the infuser, according to the maximum output power of the communications equipment.
APPENDIX This page intentionally left blank.
INDEX A C A/B valve subsystem, 4-30 AC connector replacement, 7-21 AC power cord Replacement, 7-6 Retainer, 7-6 Velcro strap, 7-6 Acronyms and abbreviations, 1-3 Air sensors, 4-19 Receiver circuitry, 4-20 Transmitter circuitry, 4-19 Alarm loudness test, 5-8 Alarm messages and error codes, 6-2 Error codes requiring technical service, 6-7 Operational alarm messages, 6-2 Alarms log, 1-12 Analog-to-digital converter, 4-6 APP PWA, 4-26 Artifacts, 1-5 Audible indicators, 4-9 Auxiliary supplies, 4-14 Cassette,
INDEX D I DataPort interface, 4-10 Delivery accuracy test, 5-13 Digital-to-analog converter, 4-8 Display assembly replacement, 7-28 Display test, 5-6 Distal air-in-line test, 5-11 Distal occlusion test, 5-11 Drawings, 9-1 Driver PWA, 4-25 Illustrated parts breakdown, 9-1 Inlet/outlet valve subsystem, 4-31 Inspection, 1-6, 5-3 Instrument installation procedure, 1-5 Inspection, 1-6 Self test, 1-6 Unpacking, 1-6 Internal AC power cord replacement, 7-21 Introduction, 1-1 Acronyms and abbreviations, 1-3 Arti
INDEX P Maintenance and service tests, 5-1 Battery operation overview, 5-16 Performance verification test, 5-2 Periodic maintenance inspection, 5-15 Routine maintenance, 5-1 Mechanical overview, 4-27 Cassette, 4-28 Mechanism assembly, 4-30 Mechanism assembly, 4-30 A/B valve subsystem, 4-30 Inlet/outlet valve subsystem, 4-31 Motor and valve assemblies, 4-30 Plunger drive subsystem, 4-31 Replacement, 7-34 Mechanism interface, 4-11 Mechanism subsystem, 4-16 Air sensors, 4-19 Cassette type/presence selection,
INDEX Peripheral components, 7-11 Peripheral cover, 7-12 Peripheral interface assembly, 7-9 Peripheral PWA, 7-10 Piezo alarm assembly, 7-33 Pole clamp assembly and backing plate, 7-20 Power supply PWA, 7-27 Rear enclosure assembly components, 7-18 Rear enclosure gasket, 7-23 Required tools and materials, 7-2 Rubber foot pad, 7-2 Safety and equipment precautions, 7-1 Separating the front enclosure, rear enclosure, and main chassis assembly, 7-7 Top seal gasket, 7-17 Volume control knob, 7-12 Required tools
INDEX Top seal gasket replacement, 7-17 Troubleshooting, 6-1 Alarm messages and error codes, 6-2 Procedures, 6-11 Technical assistance, 6-1 Warning messages, 6-1 U Unpacking, 1-6 User qualification, 1-5 V V2_5 reference voltage, 4-18 Velcro strap replacement, 7-6 Voltage monitor watchdog timer, 4-6 Volume control knob replacement, 7-12 W Warning messages, 6-1 Warranty, 2-1 Technical Service Manual I-5 430-95424-003
INDEX This page intentionally left blank.
For technical assistance, product return authorization, and to order parts, accessories, or manuals within the United States, contact Hospira. 1-800-241-4002 For additional services and technical training courses, visit the website at www.hospira.com. For technical assistance and services outside the United States, contact the local Hospira sales office. CAUTION: Federal (USA) law restricts this infuser to sale by or on the order of a physician or other licensed practitioner.