The following content is the common problems and questions that our customers encountered before. Please read carefully before use. If any problem occurs during use, please CONTACT US FIRST. We promise we will help you solve them and give you a satisfied solution. Notice: All of our scales has already been calibrated and are ready for use! If the scale get inaccurate after some time of use, please calibrate it following the instructions.
Ⅰ. Common Questions (1) Q: What to do if the scale get inaccurate after some time of use? A: Please calibrate it according to the instruction. (see in (4)) (2) Q: Why the reading is always fluctuating and it can’t set zero or get a sable weighing results? A: Please check whether your under-plate is installed reversed. The incorrect installation will cause fluctuating reading. See in part 1, how to install the under-plate correctly.
A: It means that the scale is out of power now. If you have powered it by battery, please change battery. If you power the scale by power cord, then the cord may be damaged. Please check it by trying to power it by battery first. If the scale can work properly with the battery, then it’s the power cord problem. Please contact us and we will help you free replace the power cord within one year. You can also purchase power cord from us if it’s out of one year warranty.
Part 1 Under-plate Pan Installation CGOLDENWALL® HZ Electronic balance Brand Store: https://www.amazon.
Ⅱ.Overview and Specification 1. Overview This analytical electronic balance has the functions of peeling weight, self correcting, memory, counting, fault display and the like. And it features automatic fault detection, over-load alarm, a large easy-to-read digital display, and most importantly fast accurate readings.
3. Specifications and Performance Indicators JM series electronic balance is divided into following specifications according to different ranges. Model Maximum range Resolution Verification index Peeling range Calibration weight Scale size Size Power Operating temperature 2002 3002 5002 10002 20002 30002 50002 5001 10001 12001 20001 200g 300g 500g 1000g 2000g 3000g 5000g 500g 1000g 1200g 2000g 0.01g 0.01g 0.01g 0.01g 0.01g 0.01g 0.01g 0.1g 0.1g 10d 10d 10d 10d 10d 10d 10d 10d 10d 0.1g 0.
message when "--CAL--" is displayed. Wait until the standard weight value is displayed. ", place the weight of the "standard weight value" that shines on the plate, display "---" waiting state, wait a moment, display "standard weight value", take the weight, the display becomes "---" wait Status, wait, "0 or 0.0 or 0.00" is displayed and the calibration is completed. If the weighing after calibration is still inaccurate, repeat the calibration several times as described above.
counted are placed on the weighing pan, and then the "PCS count" key is pressed. The display changes to "---" wait state, and the display changes later. The averaging operation is done for “10”, and the balance can perform the counting operation.(In this operation, the zero point must be: “0 or 0.0 or 0.00”, the weight of a single count must be greater than the balance minimum readable, ie resolution) 6. Balance Unit Conversion Operation Press “UNIT” unit conversion key.
Pan Size (mm) Windbre ak Weight Amazon ASIN 200g/0.1g Ф140 × × B074PMMNVF 500g/0.1g Ф140 × × B074PNTTC9 1000g/0.1g Ф140 × × B074PNWXLX 2000g/0.1g 180*140 × × B074PNH7ZJ 3000g/0.1g 180*140 × × B074PNF8C3 5kg/0.1g 180*140 × × B074PQL6F9 10kg/0.1g 180*140 × × B074PR3FTY 15kg/0.1g 180*140 × × B074PQNWZR 20kg/0.1g 180*140 × × B074Q693FJ 30kg/0.1g 180*140 × × B074PR2QCH 200g/0.01g Ф130 √ 100g*1 B074QGCR7P 300g/0.01g Ф130 √ 100g*2 B074Q9V5HQ 500g/0.