Specifications

6
CANNON
®
CT-2000 Constant Temperature Bath
Revision 1.0g—February, 2012; CANNON
®
Instrument Company
2139 High Tech Road • State College, PA 16803 • USA
Assembly procedure
1. Unpack the bath unit housing
and move it to its permanent
location on a stable labora-
tory bench or table.
2. Remove all eight screws from
the stainless steel top covers
(see Figure 2).
3. Disconnect all external
cabling (probes, heaters, and
float switch) from the upper
rear panel of the CT-2000.
4. Remove the front top cover and rear top cover (see Figures 3 and 4).
Use caution when removing the rear top cover because the tempera-
ture control probes and heating elements are attached to it.
Glass jar installation
1. Remove the glass jar from its box.
WARNING
The glass jar is heavy. Use caution when lifting it.
2. Make sure the rubber
support ring is seated
properly around the
bottom jar opening in the
bath unit (see Figure 5).
3. Place the rubber gasket
around the top rim of the
jar (see Figure 6). The
Figure 5: Seating the support ring
Figure 2: Detaching top cover
Figure 3: Removing
front top cover
Figure 4: Removing
back top cover