4
5
[1] Deb (-): Red [2] Deb (+): Blue [3] RA: White [4] LV: Green [5] RV: White [6] Deb(+)[7]Deb (-) [8] RA
(-) [9] RV (-) [10] LV (-) [11] Suture Hole
Figure 6. Lead connections an d setscrew locations, RA : IS-1, RV: IS-1/DF-1, LV: IS-1
NOTE: The pulse generator case is used as a debrillating electrode unless the pulse generator has been
programmed to the Distal Coil to Proximal Coil (or “Cold Can”) Shock Vector.
IMPLANTING THE PULSE GENERATOR
Implant the pulse generator by performing the following steps in the sequence provided. Some patients
may require pacing therapies immediately upon connecting the leads to the pulse generator. In such cases,
consider programming the pulse generator before or in parallel with implanting the lead system and forming
the implantation pocket.