User Manual

11
12
Example of the print report
Program: wrapped
Temperature: 134
Pressure: 39 PSI
Vacuum Num.: 3
Dry Time: 15min
Ster Time: 12min
Cycle No: 00009
MAX. Temp: 135.1
MIN. Temp: 134.1
AVG. Temp: 134.6
MAX. Pres.: 41PSI
MIN. Pres.: 38PSI
AVG. Pres.: 39 PSI
Start Time: 22:38
End Time: 23:17
Date: 2008-07-19
¨¨¨¨¨¨¨¨¨¨¨¨
¨¨¨¨¨¨¨¨¨¨¨¨¨
¡
¨¨¨¨¨¨¨¨¨¨¨¨¨
¨¨¨¨¨¨¨¨¨¨¨¨
7. Maintenance
10 Maintenance
Instructions manual
Instructions manual
9. Printer (Optional)
9.1 If you have purchase the BioClave printer device, you can connect the
printer cable to the back of the sterilizer.
9.2 Connect the printer power cord to the socket with the proper electrical
specifications.
9.3 Power on the sterilizer and press the “sel” button.
9.4 After the sterilization cycle has completed, the printer will begin printing
the details of the sterilization cycle (see below).
Frequency
Operation
Daily
Clean the door seal
Wipe dry the inner chamber
Weekly
Clean the clean water reservoir
Clean the inner-chamber
Replace the door seal
Annually
8.1 Cleaning the door seal
Clean the door seal daily with a softy cloth saturated in
distilled water
8.2 Cleaning the clean water tank
Clean the clean water tank every
week with medical disinfectant .