User's Manual

TD 93021US Rev A
10 FEB 2014 / Ver A 57
Installation Guide
teleCARE IP
Figure 70. External LED inputs to the NICL corridor lamp
External Corridor Lamp Outputs
When the NICL corridor lamp is used as a passive slave to the room controller, a cable with 6-pole
connectors is used to connect the NICL with the NIRC External Corridor Lamp Outputs bus. The +V
pole from the bus applies continuous 24Vdc from the NIRC to the four LEDs on the NICL. Each LED is
individually switched to 0 volts by the NIRC via the L1, L2, L3, L4 poles. In this configuration, the NICL
LEDs operate in parallel to the LEDs on the NIRC.
5.8 System Manager (NISM2)
The teleCARE IP System Manager (NISM2) is an Ascom ELISE3 application. The ELISE3 module serves
as the interface to a 100base-T Ethernet LAN and it contains a Linux based web server.
The NISM2 is the management tool for centrally managing the teleCARE IP system. It is used to setup
and maintain the teleCARE IP system and it also includes a staff GUI for creating and maintaining staff
assignments.
The NISM2 is connected to the IP network and stores the configurations for each of the teleCARE IP
Room Controllers that are connected to the IP network.
In teleCARE IP installations which include NIRD Room Displays but do not include Ascom Messaging,
the NISM2 is used as the Interactive Messaging server for the room displays.
A Java based graphical user interface (GUI) is supplied with the NISM2. The setup of the teleCARE IP
system is done using a web browser. Any Internet browser that can interpret Java ™ script 1.2 (CSS-2)
is acceptable. Microsoft Internet Explorer 7.0 or higher with Sun Java ™ Runtime Environment 1.6 or
later is recommended.