5) For example, V=2.55ml after calculation. Input 2.55 into the space
of Real Volume. The pump may calculate and update the accuracy
factor from 4840 to 4745 automatically as figure below.
Fig 7.3.5.1-4 Accuracy Factor Updated
Caution
Only trained personnel can perform pump calibration.
7.3.5.2 RFID Switch
Touch the RFID switch and enter the password to switch on/off the RFID
identification of the dedicated feeding set.
Caution
To avoid using incorrect tubing for the pump, there is a separate password for
RFID switch function. Only authorized person can switch on/off the RFID
identification.
7.3.5.3 Sensor
Enter the “Sensor” page to check the sensor real time return values, which
may help technicians diagnose the issues of pump.
Note: The threshold of each sensor cannot be adjusted.
7.3.5.4 Self-test
The self-test has been performed during pump start up process. If the pump
self-test fails, the feeding tasks cannot be performed.
The results of the starting self-test can be reviewed on the "System Self-Test"
page under "System Maintenance". After reviewing, the pump maintenance
engineer can touch the "Retest" button on the page to double check whether
the pump is normal. If it still fails, please contact the customer service.