System Description

File Name
IFU of AMSure
®
Enteral Feeding
Pump and Feeding Set
REF#
L10196_US
Version
B
Page 30 of 81
AMSINO Confidential
Touch the “Confirm” button to start prime. After prime, the page returns to the
setting page (Figure 7.2.1-1). Repeat the prime process until the distal
connector leaking the liquid and then user can terminate prime by touching
“Stop” button.
Connect distal connector of the feeding set to the patients gastric tube as
shown below:
Fig 7.2.2-2 Connect to Patient
Caution
Before the connection to the patient, it must be ensured that there is no air in
the feeding set.
When the set has been primed and connected to the patient’s gastric tube,
user can start feeding.
Note:
1) Prime rate is 1200mL/h and cannot be adjusted.
2) When prime is in processing, the function of infrared sensor, ultrasonic
sensor will be bypassed, however the pressure sensor will keep monitor
the tube pressure.
3) The prime warning page will prompt (Fig 7.2.2-1 A) in any conditions
once Prime button is pressed, which is to prevent the injury to patient
caused by mis-operation.
7.2.3 Set Feeding Parameters
Touch the numbers in orange to enter keyboard and type the feeding
parameters as below (Fig 7.2.3-1). Touch the “Enter” button to confirm
setting.