gastric tube → start feeding → complete feeding → remove the feeding set →
switch off.
7.1 Preparation before Use
7.1.1 Pump Installation
Before fix the pump to the I.V. pole, hang the formula bag and feeding set to
I.V. pole and adjust the pump head height to the fluid level in the formula bag.
For better accuracy, the initial fluid level shall be at 45±5cm (17.7±1.96
inches) higher than the pump and the pump shall be placed higher than the
feeding site to the patient.
Note: To reach the required head height, fold the upstream tube in half, the
pump position can be adjusted on the I.V. pole with the door's upper edge
flushing to the tube middle point.
Fig 7.1.1-1 Initial Fluid Level (Head Height)
As shown in Figure 7.1.1-2, loosen the locking bolts of the pole clamp, mount
the pole clamp to the I.V. pole, then tight the clamp to fix the pump.
45±5 cm
(17.7±1.96 inches)