System Description

File Name
IFU of AMSure
®
Enteral Feeding
Pump and Feeding Set
REF#
L10196_US
Version
B
Page 11 of 81
AMSINO Confidential
5. Technical Specifications
5.1 Basic Parameters
Dimensions
133mm×110mm×170mm
Weight
1.08 Kg
Power adapter
No.
Parameter
Specification
1
Power adapter
specification
US market
LXCP26-150160A
EU market
LXCP26-150160B
2
Frequency
50/60Hz
3
Input voltage
AC 100240V
4
Input current
0.8A
5
Output voltage
15V
6
Output current
2.0A
Built-in battery
No.
Parameter
Specification
1
Quantity
1pcs
2
Nominal battery
voltage
DC 11.1V
3
Battery capacity
2200mAh
4
Working time
Continuously working for 7 hours with a feeding rate of
120ml/h after fully charged.
5
Battery power
24.42Wh
Requirement of enteral
feeding set
Refer to “9. Dedicated Enteral Feeding Sets (Disposable)”
5.2 Main Performance
Range of feeding rate
1~400mL/hstep value = 1mL/h
Range of feeding volume
1~9999mLstep value = 1mL
Feeding accuracy
±10%
Interval waiting
(Intermittent Mode)
00h01min ~ 23h59min
Feed count
(Intermittent Mode)
2~8
Feeding mode
Continuous mode or Intermittent mode
Recoverable alarm sound
and automatic recovery time
1min50s 2mins
No-operation timeout alarm
time
1min50s 2mins
Occlusion pressure
50±25kPa
Classification
Type II BF equipment
IPX4 level; not AP or APG type.