ADC Satin Standard, Fiber Optic and Disposable Laryngoscope Blades and Handles ® Directions for Use TM
Thank you for choosing an ADC® Satin™ Laryngoscope. Device Description and Intended Use A Laryngoscope is a two-part, hand held device consisting of a handle that contains batteries and a detachable blade. Laryngoscopes are designed to provide illumination within the larynx during the process of performing intubations. This instruction book covers both standard, fiberoptic type and disposable laryngoscopes.
2 standards. They are compatible with virtually all major brands of standard illumination hook on blades and handles. Fiber optic ADC® Satin™ Laryngoscopes and Disposable Laryngoscope blades conform to current ASTM F1195, ISO 7376-1, and ISO 7376-3 standards. They are compatible with virtually all major brands of fiber optic illumination hook on blades and handles. For easy visual identification, ALL Fiber Optic models are clearly delineated by a green indicator on both handle and blade.
sion/autoclave, we recommend using only ADC #4500-1 and #4501-1 replacement lamps for standard blades. 1) To remove lamp on standard blades, grip lamp collar and rotate lamp counterclockwise until free. 2) Replace the lamp with a new ADC lamp. 3) Verify that lamp is sufficiently tightened before use. CAUTION: Be sure lamp’s glass surface is clean and free from fingerprints after assembly. If necessary, the glass may be cleaned with a soft cloth or cotton ball moistened in alcohol.
kept on hand for emergency situations. Basic Cleaning Procedure for Standard and Fiberoptic Blades Prepare a Klenzyme®, Manu-klenz™, or equivalent cleaning solution as per manufacturer’s instructions and soak for ten minutes. Rinse devices with lukewarm running tap water for one minute to remove any residual detergent after cleaning. Dry with a clean, lint-free cloth. Note: Always allow blades to completely dry before disinfection or use.
Steam Sterilization for Standard and Fiberoptic Blades and Handles Sterilizer Type: Prevacuum Sterilizer Type: Gravity Preconditioning Pulses: 3 Minimum Temperature: 249.8°F Minimum Temperature: 269.6°F (132°C) (121°C) Full Cycle Time: 4 minutes Full Cycle Time: 30 minutes Minimum Dry Time: 20 minutes Minimum Dry Time: 15 minutes LIMITED WARRANTY American Diagnostic Corporation (ADC®) warrants its products against defects in materials and workmanship under normal use and service as follows: 1.
REPLACEMENT PARTS CHART DESCRIPTIONS STANDARD MODELS FIBEROPTIC MODELS Fiberoptic Fiberoptic Standard Laryngoscope Standard Blade Replacement Lamp Blade Replacement Lamp Item N/A 4070F 4500 4070 Mac 0 N/A 4071F 4500 4071 Mac 1 N/A 4072F 4500 4072 Mac 2 N/A 4073F 4501 4073 Mac 3 N/A 4074F 4501 4074 Mac 4 N/A N/A 4500 4080-0 Miller 00 N/A 4080F 4500 4080 Miller 0 N/A 4081F 4500 4081 Miller 1 N/A 4082F 4501 4082 Miller 2 N/A 4083F 4501 4083 Miller 3 N/A 4084F 4501 4084 Miller 4 N/A N/A 4500 4090 Wisconsin 0