Non-isolated inputs and outputs to set voltage and current limit
Control input and output
scaling:
Rear panel control inputs (CV and CC) permit external 0V to 5V or 0 to10V
signals to set 0 to 100% of rated output voltage and current.
Set values of 0 to 100% of rated output voltage and current generate
0 to 5V signals at the rear panel V
out
and I
out
outputs.
These signals are referenced to the positive output.
Control input accuracy:
Voltage: 0.3% ±10mV. Input impedance 100kΩ; protected to 60V.
Current: 0.5% ±5mA. Input impedance 64k; protected to 60V.
Control output accuracy:
Voltage: 0.3% ±10mV
Current: 0.5% ±5mA.
Output impedance: 125Ω; output short-circuit protected.
Remote Off:
Rear panel socket allows a switch closure or logic low to turn output off.
DIGITAL INTERFACES (PL-P models only)
Full digital remote control facilities are available through the RS232, USB, LAN and GPIB
interfaces.
Voltage Setting:
Resolution: 1mV; Accuracy: ± (0·05% + 5mV).
Voltage Readback:
Resolution: 1mV; Accuracy: ± (0·1% + 5mV).
Current Setting and
Current Readback:
Resolution: 0.1mA; 0.01mA on Low I range, except 6V O/P: 1mA & 0.1mA.
Accuracy: ± (0·3% + 3mA) to 3A, ± (0·5% + 3mA) to 6A, ± (0·3% + 0·3mA)
Low I range, except 6V O/P: ± (0·3% + 3mA) to 8A and
± (0·3% + 0·3mA) on Low I range.
RS232:
Standard 9-pin D-connector. Baud rate 9600.
GPIB (optional):
Conforming with IEEE488.1 and IEEE488.2
USB:
Standard USB 2.0 hardware connection. Operates as a virtual COM port.
LAN:
Ethernet 100/10base-T hardware connection. 1.4 LXI Core 2011.
Remote Command
Processing Time:
Typically <25ms between receiving the command terminator for a step
voltage change at the instrument and the output voltage beginning to
change.
GENERAL
AC Input:
230V AC or 115V AC ± 10%, 50/60Hz. Installation Category II
Power Consumption:
Single: 280VA max. Dual: 560VA max. Triple: 840VA max.
Operating Range:
+5ºC to +40ºC, 20% to 80% RH
Storage Range:
40ºC to + 70ºC
Environmental:
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Cooling: Intelligent variable-speed low noise fan assists convection. Over-
temperature trip shuts down output if internal temperatures exceed
predetermined thresholds.
Safety:
Complies with EN610101
EMC:
Complies with EN613261
Size:
(excluding feet, knobs
and terminals)
Single: 107mm x 131mm (¼ rack 3U) x 288mm L (PL-P: 343mm L),
Dual: 214mm x 131mm (½ rack 3U) x 288mm L
Triple: 321mm x 131mm (¾ rack 3U) x 288mm L
Weight: Single: 4·5kg (PL-P: 4.9kg); Dual: 8.5kg; Triple: 12.6kg