User`s guide

Table Of Contents
284 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:CALC{1-4}:PAR:COUN
Syntax :CALCulate{[1]|2|3|4}:PARameter:COUNt <numeric>
:CALCulate{[1]|2|3|4}:PARameter:COUNt?
Description Sets the number of traces of channel 1 (:CALC1) to channel 4 (:CALC4).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:PAR:COUN 4"
20 OUTPUT 717;":CALC1:PAR:COUN?"
30 ENTER 717;A
Equivalent key [Display] - Num of Traces
<numeric>
Description Number of traces
Range 1 to 4
Preset value 1
Resolution 1