FILE No.
Content Designation .................................................................................................................... 3 Introduction to Features of Product ............................................................................... 4 Product appearance ........................................................................................................ 5 Dimensions ....................................................................................................................
Designation Regulations for type naming: HYC XXX X Design No. is expressed with pinyin character sequence Rated volume value is shown with L Pharmacy Refrigerator Note: rated volume can be the gross volume or effective volume; the manufacturer can decide it by him according to the actual situation. The effective volume value must be marked on the nameplate whether effective volume or gross volume is marked in the product name. HYC-68A is the replacement of the sight glass doors on the basis of HYC-68.
Introduction to Features of Product 1. Medical pharmaceutical refrigerator: Temperature can be between 2℃ and 8℃; 2. Micro-computer control and digital display for refrigerator temperature; 3. Automatic defrosting; 4. Visible acoustic-optical alarm for over-temperature alarm, door-opening alarm and sensor trouble alarm; 5.
Produ uct appea arance Con ntrol panel Light Fan n Sheelf Door handle Door lock Foo oting 5
Dimensions HYC-68/68A 7
product nameplate For example: HYC‐68 For example: HYC‐68A 8
For example: HYC‐118 For example: HYC‐118A 9
Technical Data Model HYC-68/68A HYC-118/118A Weather type N N Protection type against electric shock I I 68 118 Rated voltage 220V/50Hz 220V/50Hz Fan power (W) 1.3 1.3 Total input power(W) 84 160 0.57/0.93 1.
Circuit Diagram HYC‐68/68A/118 HYC‐118A 11
Compressor Parameters (Under Standard Operating Conditions) Product model HYC-68/68A HYC-118/118A Compressor model Zanussi HVY57AA TL5G Refrigeration capacity (W) 95 Input power(W) 68.4 Rated voltage(V) 220-240 220-240 Rated frequency(Hz) 50 50 Current(A) 0.
Fan motor Parameters (Under Standard Operating Conditions) Fan model: AD0912LB‐A71GL Voltage(V): DC 12 Rotate speed(rpm): 2300 线路描述: 红色线为正极,黑色线为负极 Current(A): 0.
Computer Board 1. The computer board consists of display board and main control board for close cooperation. 2. Basic parameters (for power):AC current 2.1. Rated voltage: 220~240V 2.2. Rated frequency: 50Hz as universal frequency 2.3. Rated power: 350W 2.4. Automatic voltage identification: With 2.5. Automatic frequency identification: With 2.6. The maximum operating load of controller: 500W 3. Safety certification requirements: CE certification, 3C certification 4.
Operation panel 1. Appearance: (Display panel and display board are shown in following figure.) 2. Key 2.1 Temperature up/down key: To press up/down key for temperature setting. 2.2 Setting key: To set functions and enquire the maximum and minimum refrigerator temperature. 2.3 Lighting switch key: To press lighting key so that lighting lamp is on. To press it again so that it is off. 2.
Peripheral Signal of Computer Board 1. System input quantity: A、 Temperature sensor: With ordinary refrigeration sensor in the refrigerator; B、 Key signal: Please refer to key definition for details. C、 Door switch signal: With door switch signal; Fan is connected when door body is closed and it is disconnected when door body is opened. Door switch is fitted with normally closed switch. That is, 2 terminals of switch are closed when opening the door body.
Function Control Display and Key Control 1. Display: 1.1 Refrigerator shall display actual temperature during initial power supply and after temperature reduces to the minimum value and before increasing to setting temperature which corresponds to such temperature gear. 1.2 Current actual temperature is ±2°C of setting temperature for gradual variation to setting temperature.
bootstrap point. There is no alarm if refrigerator temperature is within scope of alarm limit. There is instant temperature alarm and buzzer will ring if refrigerator temperature is out of control limit scope. Alarm lamp will flicker and alarm sound can be cancelled by pressing alarm key. However, alarm indicator lamp will keep flickering and it is off only when refrigerator temperature is within scope of alarm limit. 4. Temperature control 4.
and display refrigerator temperature again (without flickering) if there is no operation within 10 seconds. 4.5 Setting of Internal Parameters Several internal parameters can be adjusted by following method to guarantee normal operation of refrigerator: To press “regulate” key for more than 10 seconds to display original design value (rd) with flickering; To press “+” or “-” to adjust setting value. Adjustment scope is 1℃ to 4℃ because setting value can vary by 1℃ by pressing the key every time.
Form 1:Internal parameter form under control contents Internal parameter Parameter instructions Scope Default value Instructions /unit rd Adjustment gain is 1 for setting temperature difference 1~6 2 ℃ r1 Adjustment gain is 1 for the maximum setting temperature LE~20 8 ℃ 3 r2 Adjustment gain is 1 for the minimum setting temperature -40~HE 0 ℃ 4 c1 Press delay startup period during controller operation 0~10 0 min 5 c2 The minimum compressor shutdown period 0~10 5 min c3 Long
4.6 Inquiry of the Maximum and Minimum Refrigerator Temperature The maximum and minimum refrigerator temperature can be enquired by following way: The maximum or minimum temperature is actual temperature detected by sensor, not the maximum or minimum displayed temperature. Enquiry of the maximum value: In the meanwhile, users can press “SET” key and “+” to display the maximum actual value without flickering in the window. Please loosen any key to return to display refrigerator temperature.
key at the same time to show higher alarm limit in the window without flickering. It can return to refrigerator temperature display when loosening the key. Enquiry of lower alarm limit: To press alarm key and “-” key at the same time to show lower alarm limit in the window without flickering. It can return to refrigerator temperature display when loosening the key. 5.
9. Delay Protection Function Shutdown and bootstrap interval of compressor shall be not less than c2 minutes.
Trouble Display and Treatment Display temperature is “EE” for display board and it will flicker by frequency of 1Hz during trouble (such as short circuit or broken circuit) of temperature sensor. In the meanwhile, it will buzz and enter into protection mode for compressor operation control according to internal parameters dO and dF.
Details of Main Replacement Parts HYC‐68/68A Product model HYC-68/68A No. Parts details Specification or model Special No.
HYC‐118/118A Product model HYC-68/68A No. Parts details Specification or model Special No.
Common Trouble Analysis and Repair Measures Problem 1. No startup of press Reason analysis Repair measures 1. Broken fuse To replace fuse 2. Damaged connectors for cabin wiring To replace connectors 3. Poor wiring of electric control box To install and position after inspection 4. Trouble of control board To replace control board 5. Broken protector 2. Uneven refrigerator temperature and large temperature difference 3. Display board can display “EE” 4. Large noise of refrigerator 5.
Reply of Common Problems Problem Answer Treatment measures 1. Higher or lower refrigerator temperature It relates to ambient temperature. Higher or lower refrigerator temperature can be solved by adjusting control temperature. Please refer to Appendix 1 for detailed operation method. 2. Water accumulation in the refrigerator after a certain operation period To clean it by towel periodically if there is a little of water. 3.
Structural Principle Sketch of Refrigeration System 蒸发器 冷凝器 干燥过滤器 毛细管 回气管 排气管 压缩机 压机油 evaporator condenser dry filter capillary air return tube compressor press oil air exhaust tube 蒸发器 冷凝器 干燥过滤器 毛细管 回气管 压缩机 压机油 排气管 29
Im mportant compon nents to install an nd remov ve 1. Demolition D n, installattion consid derations Be confirm c that unplug the power wh hen Demolittion and insstallation; When installed d in accordaance with th he demolitio on of the asssembly seq quence conttrary, oduct must be installed in the saame place tthe assembly in wheen any partt of the pro placce. D bly and insttallation guide g 2. Disassemb 2.
※Step2:open the top cover, and then disconnect the display panel wire-line; ※Step3:remove 3 screws on the display panel,so that can separate the top cover.
※Step4:open the cover so that can replace the display board.
2.
※Step2:Disconnect the fan wire‐line, so that the fan can be replaced; 34
※Step3:remove these 4 screws, so that the evaporimeter can be moved Note: Be careful about the ecaporimeter and remove it slowly. 2.3 installation and removal of the compressor Step1.
Step2.remove these 2 screws。 Step3. Removed the four hexagonal bolts connected to the main body, so that the compressor can be complete removal.
Testing Date 37
Inspired living Haier Medical & Laboratory Products Co., Ltd. Room 403D, Brand Building, Haier Industrial Park, No.1 Haier Road Qingdao China Website: www.haiermedical.