11 8 Monitor Configuration Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111 About Factory Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112 About Default Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112 About Temporary Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
112 Monitor Configuration Welch Allyn Propaq LT Vital Signs Monitor About Factory Configuration When the monitor arrives from the factory, it is fully configured and ready for operation. The factory configuration defines every setting and behavior for the monitor. Each setting and behavior defined in the factory configuration remains in effect until it is changed by a custom configuration or by a temporary adjustment.
Directions for Use Monitor Configuration 113 About Temporary Configuration While monitoring a patient, you can use the monitor buttons to temporarily adjust many of the monitor settings to accommodate the needs of a specific patient. These custom settings persist through a power cycle if you select to save patient data on shut-down. The persist until they are changed manually or by Acuity, or until the monitor is powered down and then powered up to monitor another patient.
114 Monitor Configuration Welch Allyn Propaq LT Vital Signs Monitor Table 11.
Directions for Use Monitor Configuration 115 Table 11. Parameter Configuration Matrix (continued) Parameter Factory Setting Possible Values Where Set Configuration Name Factory 1.
116 Monitor Configuration Welch Allyn Propaq LT Vital Signs Monitor Table 11. Parameter Configuration Matrix (continued) Parameter Factory Setting Possible Values Where Set Display All ECG Vectors Yes Yes, No Configuration Utility ECG I Wave Size 1 mV/cm 8 mV/cm, 4 mV/cm, 2 mV/cm, 1 mV/cm, Monitor 0.5 mV/cm, 0.2 mV/cm ECG II Wave Size 1 mV/cm 8 mV/cm, 4 mV/cm, 2 mV/cm, 1 mV/cm, Monitor 0.5 mV/cm, 0.2 mV/cm ECG III Wave Size 1 mV/cm 8 mV/cm, 4 mV/cm, 2 mV/cm, 1 mV/cm, Monitor 0.5 mV/cm, 0.
Directions for Use Monitor Configuration 117 Table 11.
118 Monitor Configuration Welch Allyn Propaq LT Vital Signs Monitor Table 11.
Directions for Use Monitor Configuration Table 11. Parameter Configuration Matrix (continued) Parameter Factory Setting Possible Values Where Set Systolic Upper Alarm Limit 145 mmHg 19.3 kPa 32 - 160 mmHg 4.3 - 21.3 kPa Monitor, Acuity Systolic Lower Alarm Limit 75 mmHg 10.0 kPa 30 - 158 mmHg 4.0 - 21.1 kPa Monitor, Acuity Diastolic Upper Alarm Limit 100 mmHg 13.3 kPa 17 - 130 mmHg 2.3 - 17.3 kPa Monitor, Acuity Diastolic Lower Alarm Limit 35 mmHg 4.7 kPa 15 - 128 mmHg 2.0 - 17.
120 Monitor Configuration Welch Allyn Propaq LT Vital Signs Monitor Table 11.
121 9 PC Utility Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .121 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .122 Configuring the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .126 About AutoPrint . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 PC Utility Welch Allyn Propaq LT Vital Signs Monitor Propaq LT Monitor AutoPrint Utility With a PC connected to a printer and with a cradle configured with the optional data link capability, you can configure the PC to print, manually or automatically, all patient data stored in the monitor each time the monitor is placed in the cradle. (See “Printing Patient Data” on page 103.
Directions for Use PC Utility 123 Figure 101. Windows Logo Message Usually, the installation completes quickly. However, if the utility also has to install .NET Framework on your computer, this can add several minutes. 3. Click through the Welcome screen and the license agreement. You must accept the terms of the license agreement to proceed with the installation. Figure 102. Accepting the Terms of the Software License 4.
124 PC Utility Welch Allyn Propaq LT Vital Signs Monitor Figure 103. Complete or Custom Installation • To install both utilities, click Complete. • To install only one of the utilities, click Custom and, in the next screen (Figure 104), select the utility you want to install. Figure 104. Selecting to Install Either the AutoPrint Utility or the Configuration Utility If you click the next to your choice, you can also select one of the installation options available for that choice (Figure 105).
Directions for Use PC Utility 125 Figure 105. Menu of Installation Options 5. The installation prompts you occasionally for information. At each prompt, provide the requested information and then click Next. The installation continues until it is finished.
126 PC Utility Welch Allyn Propaq LT Vital Signs Monitor Configuring the Monitor Making a Configuration File 1. On a PC, start the Propaq LT Monitor Configuration Utility (Figure 106): Start > Programs > Welch Allyn > Propaq LT Monitor > Configuration Figure 106. Starting the Configuration Utility - Start Menu You can also start the utility by double-clicking the Configuration icon on the computer desktop. The configuration utility Welcome screen (Figure 107) appears. Figure 107.
Directions for Use PC Utility 127 Figure 108. Selecting a Configuration File for Display 3. Select DefaultPropaqLTConfiguration.mnt and click Open. The file is opened for viewing in the configuration display (Figure 109). (See “About the Configuration Display” on page 130.) Figure 109. Configuration Display 4. The program does not permit you to modify DefaultPropaqLTConfiguration.mnt. To make a new configuration file, save DefaultPropaqLTConfiguration.
128 PC Utility Welch Allyn Propaq LT Vital Signs Monitor Figure 110. Creating a New Configuration File (Save As) The utility displays the new file (Figure 111). 5. To open the new file for editing, click Edit. You can now enter text and select parameter options. Figure 111.
Directions for Use PC Utility 129 6. Define the monitor settings for the new configuration. Examine each field in the configuration utility. Change any value that is not appropriate for your clinical situation. 7. When the form is complete and the value in each field is correct, click Save. Alternatively, to save the file and close it, click Save/Close. Printing a Configuration File To print this configuration file, click Print.
130 PC Utility Welch Allyn Propaq LT Vital Signs Monitor Modifying a Configuration File To change a configuration file that has already been completed, follow these steps: 1. Start the utility. 2. Select the configuration file that you want to modify. 3. Click Edit. 4. Using the section tabs, navigate to the setting you want to change. 5. Enter the new value. 6. Repeat steps 4 and 5 until you have updated all of the settings you want to change. 7. Click Save or Save/Close. Configuring a Monitor 1.
Directions for Use PC Utility The configuration display contains the following elements: • • • • Windows command bar Configuration command bar Page tabs Configuration settings and parameters. The Windows command bar includes File, Configuration, and Help. File Save, Close, Save and Close, Exit Configuration Select, Save As, Edit, Load, Print, Print Preview Help About Propaq LT Monitor Configuration Utility The configuration commands include the following: Select Open a file-selection dialog.
132 PC Utility Welch Allyn Propaq LT Vital Signs Monitor The page tabs link you to the sections of the configuration file: Identification Name the configuration, define monitor ownership and support responsibility, and define parameters related to your locale. Display Select and define the displays. Neonatal Mode Define default settings for neonatal patients. Pediatric Mode Define default settings for pediatric patients. Adult Mode Define default settings for adult patients.
Directions for Use PC Utility 133 Figure 115. Opening a Drop-down Menu Click a choice (Figure 116) to select it... Figure 116. Selecting From a Drop-Down Menu The new selection appears in the field (Figure 117). Figure 117. New Selection Displayed To set a value for a Yes-No item, click the box to insert a mark for Yes (Figure 118), or click the box to remove the mark for No (Figure 119). Figure 118.
134 PC Utility Welch Allyn Propaq LT Vital Signs Monitor Figure 119. Selecting No (No Check Mark) To view the settings for a topic, click the tab for that topic. For example, to see the Default Resp Lead setting for adults, click Adult Mode (Figure 120). Figure 120. Settings Under Adult Mode Error Messages • Communication with the monitor failed. Verify that the monitor is powered on and properly seated in the cradle. Verify that the cable is connected.
Directions for Use PC Utility 135 About the Configuration Worksheet WARNING You must read and understand the warning statements in “Enable Wireless Communications (with Acuity)” on page 147 before any configuration file is created based on a worksheet. The first step in customizing a monitor configuration is to complete a configuration worksheet. This defines the settings to be stored in the monitor. Note A worksheet is not required.
136 PC Utility Welch Allyn Propaq LT Vital Signs Monitor After you reconfigure the monitor, the settings defined in the new configuration become the ‘system defaults’, and the factory defaults are no longer in effect. The new system defaults remain in effect until another configuration is downloaded to the monitor. If you reconfigure the monitor and then want to restore the factory defaults, download the configuration file “DefaultPropaqLTConfiguration.mnt.
Configuration Worksheet 137 Configuration Worksheet 1 Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138 Name the configuration, define monitor ownership and support responsibility, and define parameters related to your locale. 2 Display Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141 Select and define the displays. 3 Neonatal Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
138 Welch Allyn Propaq LT Vital Signs Monitor Note All configuration parameters are set at the factory. You do not have to change any settings for the monitor to work properly. 1 Identification 1.1 Configuration Identity For each configuration, enter a unique name that identifies the following: • the hospital and the care unit or department that owns the monitor • the date on which this configuration is created The name can include any standard keyboard characters other than the comma (,).
Configuration Worksheet 139 1.2 Facility This information further clarifies ownership of the monitor, and it identifies a person who should be notified if the monitor needs attention. What facility, department, and individual is responsible for this monitor? Facility Name (Up to 40 characters.) Department or Unit (Up to 40 characters.) Contact Name (Up to 40 characters.) Contact Telephone (Up to 40 characters.) 1.3 Country Settings Note All configuration parameters are set at the factory.
140 Welch Allyn Propaq LT Vital Signs Monitor 1.3.3 Date Format How do you want the monitor to display the date? __ mm/dd/yy __ dd.mm.yy __ yy/mm/dd 1.3.4 Decimal Format You can configure the monitor to use a dot (for example, 195.18) or a comma (195,18) to display the decimal point. How do you want the monitor to display the decimal point? __ Dot __ Comma 1.3.5 Power Source Filter Set the power-source interference filter to the appropriate frequency for your location.
Configuration Worksheet 141 2 Display Configuration The monitor interface includes a Display button: . Pressing quickly through the three views you specify below. cycles the display In the configuration, you can choose any three of the five display choices shown below for quick display with .
142 Welch Allyn Propaq LT Vital Signs Monitor Display Format Which display format do you want to use for ...
Configuration Worksheet 143 Display Format (continued) If Dual Waveform display is enabled, which signal is to be displayed as the default bottom waveform? Display 1 Display 2 Display 3 __ Lead I __ Lead I __ Lead I __ Lead II __ Lead II __ Lead II __ Lead III __ Lead III __ Lead III __ Lead V __ Lead V __ Lead V __ aVR __ aVR __ aVR __ aVL __ aVL __ aVL __ aVF __ aVF __ aVF __ SpO2 __ SpO2 __ SpO2 __ Resp __ Resp __ Resp If Tabular Trends display is enabled, what is the defa
144 Welch Allyn Propaq LT Vital Signs Monitor 3 Neonatal Mode For each parameter, the factory default value is shown in bold. The default value is used by the monitor unless you specify another value in this worksheet. 3.1 Use Smartcuf as Default for NIBP Note Smartcuf will be available in 2006.
Configuration Worksheet 145 4 Pediatric Mode For each parameter, the factory default value is shown in bold. The default value is used by the monitor unless you specify another value in this worksheet. 4.1 Use Smartcuf as Default for NIBP Note Smartcuf will be available in 2006.
146 Welch Allyn Propaq LT Vital Signs Monitor 5 Adult Mode For each parameter, the factory default value is shown in bold. The default value is used by the monitor unless you specify another value in this worksheet. 5.1 Use Smartcuf as Default for NIBP Note Smartcuf will be available in 2006.
Configuration Worksheet 147 6 Feature Enable These settings specify whether or not certain features are available to the clinician. If you enable a feature in the configuration, then it is available to the clinician. If you disable a feature in the configuration, then it is not available to the clinician. For each parameter, the factory value is shown in bold. Factory values are used until other values are uploaded to the monitor. 6.
148 Welch Allyn Propaq LT Vital Signs Monitor 6.2 Enable Lockouts If lockouts are enabled, simultaneously pressing the Left Arrow, Right Arrow, and Down Arrow buttons locks out one or more of the following: Buttons Block access to the monitor buttons Back Light Extend the battery charge life Display Prevent unauthorized viewing of confidential patient information For the following questions, select which lockouts, if any, to enable for this monitor configuration. 6.2.
Configuration Worksheet 6.3 149 Display Time Out Display Time Out greatly increases monitor run time on battery power. Display Time Out completely shuts off the display if the monitor detects no button press and no alarm for a number of minutes defined below. When the display is off, no monitor information is visible. Any alarm condition, alert condition, or button press immediately restores the display.
150 Welch Allyn Propaq LT Vital Signs Monitor 6.5 SpO2 Spot Check Enable If SpO2 Spot Check is enabled, a clinician can perform a quick SpO2 measurement when SpO2 monitoring is turned off. For detailed information, see Propaq LT Vital Signs Monitor Directions for Use. Do you want to enable SpO2 Spot Check? (Yes) (No) 6.6 NIBP Turbo Mode Enable When the monitor takes automatic NIBP measurements in Turbo Mode, it takes as many measurements as possible within the 5-minute Turbo Mode measurement period.
Configuration Worksheet 151 7 Monitor Behavior For each parameter, the factory value is shown in bold. The factory value is used until another value is uploaded to the monitor. 7.1 Patient Mode Default In what default patient mode should the monitor be set for new patients? __ Adult __ Pediatric __ Neonatal 7.2 Display All ECG Vectors If a 5-lead ECG cable is connected to the monitor, should the monitor be able to display the optional aVR, aVL, and aVF waveforms? (Yes) (No) 7.
152 Welch Allyn Propaq LT Vital Signs Monitor 7.4 Pacer Indicator On Default If a patient being monitored has a pacemaker, the monitor detects the occurrence of pacemaker signals and optionally indicates them with vertical dashed lines in the waveform. Pacemaker signal indicators If the pacemaker signal is strong enough, the monitor displays it as a spike. This spike is displayed whether the pacemaker indicator is on or off. Should the pacemaker indicator be on by default? (Yes) (No) 7.
Configuration Worksheet 153 7.7 Audible Alarms Suspension Time Default While monitoring a patient, the clinician can use the settings menu to suspend all alarm tones for all parameters for a configurable period. Note This parameter does not affect the behavior of of how this parameter is configured, pressing tone for 90 seconds. , (Silence/Reset). Regardless silences a sounding alarm Suspend can be turned on or off by the clinician, but the suspend period cannot be changed by the clinician.
154 Welch Allyn Propaq LT Vital Signs Monitor 8 ParamSet If ParamSet is enabled, the clinician can quickly and easily change the current alarm limits for any vital sign. If you choose to enable ParamSet, you can then define the percentage by which the alarm limit for each vital sign is changed. To Adjust Alarm Limits Quickly with ParamSet: When the ParamSet message appears, press twice. Example: The ParamSet value for the Upper HR/PR alarm limit is configured to be 15%.
Configuration Worksheet 155 For each parameter listed below, the factory value is shown in bold. The factory value is used until another value is uploaded to the monitor.
156 Welch Allyn Propaq LT Vital Signs Monitor 8.1 ParamSet Enable Do you want to enable ParamSet? (Yes) (No) Note 8.2 If you select (No), disregard the rest of the ParamSet settings. ParamSet % for Upper HR/PR ParamSet increases the Upper HR/PR limit by ____%. (5, 10, 15, 20, 25) 8.3 ParamSet % for Lower HR/PR ParamSet decreases the Lower HR/PR limit by ____%. (5, 10, 15, 20, 25) 8.4 ParamSet % for Upper NIBP Systolic ParamSet increases the Upper NIBP Systolic limit by ____%.
Configuration Worksheet 157 Configuration Worksheet Authorization The information on this page relates only to this worksheet, and is not entered into the monitor configuration. Users are encouraged to provide their own procedures for defining, categorizing, storing, and approving configuration files.
158 Welch Allyn Propaq LT Vital Signs Monitor