Product Manual

1
4341 State Street Road
PO Box 220
Skaneateles Falls, NY 13153-0220 USA
Tel: +1 315 685 4560
Fax: +1 315 685 3361
www.welchallyn.com
Part No. 690048-1 Rev D
Fiber Optic Laryngoscope Blades and
Handles
Fiber Optic Laryngoscope Blades
Care and Maintenance
Welch Allyn laryngoscope blades are crafted from enduring stainless steel. To insure
maximum life and performance the following instructions should be strictly adhered to.
Miller, MacIntosh, and E-MacIntosh styles of fiber optic blades all have removable light
pipes. This unique feature eliminates the expense of replacing the entire blade due to
the fiber bundle deterioration that results from repeated exposure to disinfection and
sterilization agents. Now the bundle can simply be replaced at a fraction of the cost of a
new blade.
Cleaning Procedure
Immediately after use, blades should be rinsed in clean tap water to remove any residue.
The fiber optic light pipe should be removed from the blade following the procedure
outlined in the Fiber Optic Light Pipe Replacement section. Both the light pipe and blade
should then be gently scrubbed in soapy water, with a soft brush, to provide a thorough
physical cleaning.
All Welch Allyn blades are compatible with enzymatic cleaners. Refer to manufacturer’s
instructions for recommended exposure times and solution strengths.
After cleaning, thoroughly rinse the light pipe and blade, dry, and reassemble (see Fiber
Optic Light Pipe Replacement section) prior to disinfection or sterilization.
Note
If germicidal solutions are required for cleaning, please contact Welch Allyn for
compatibility.
WARNING Ultrasonic cleaning is not recommended.

Summary of content (8 pages)