User Documentation

Copyright © 2013 Weidmüller Interface GmbH & Co. KG 1 / 1
All rights reserved. Reproduction without permission is prohibited.
Detailed values for temperature converting ACT20M-UI-AO-S 1176030000
RTD temprature sensors
Type
min.
Temperature
max.
Temperature
Basic
accuracy
Temperature
coefficient
PT10
-200°C
+850°C
≤ ±1.4°C
≤ ±0.14°C / °C
PT20
-200°C
+850°C
≤ ±0.8°C
≤ ±0.08°C / °C
PT50
-200°C
+850°C
≤ ±0.6°C
≤ ±0.06°C / °C
PT100
-200°C
+850°C
≤ ±0.2°C
≤ ±0.02°C / °C
PT200
-200°C
+850°C
≤ ±0.2°C
≤ ±0.02°C / °C
PT250
-200°C
+850°C
≤ ±0.6°C
≤ ±0.06°C / °C
PT300
-200°C
+850°C
≤ ±0.6°C
≤ ±0.06°C / °C
PT400
-200°C
+850°C
≤ ±0.6°C
≤ ±0.06°C / °C
PT500
-200°C
+850°C
≤ ±0.3°C
≤ ±0.03°C / °C
PT1000
-200°C
+850°C
≤ ±0.2°C
≤ ±0.02°C / °C
NI50
-60°C
+250°C
≤ ±0.6°C
≤ ±0.06°C / °C
NI100
-60°C
+250°C
≤ ±0.3°C
≤ ±0.03°C / °C
NI120
-60°C
+250°C
≤ ±0.3°C
≤ ±0.03°C / °C
NI1000
-60°C
+250°C
±0.3°C
≤ ±0.03°C / °C
Effect of sensor cable resistance
< 0.002 Ω / Ω
Thermo couple
Type
min.
Temperature
max.
Temperature
Basic
accuracy
Temperature
coefficient
B
0°C
+1820°C
≤ ±4.5°C
(160...400°C)
≤ ±0.45°C / °C
(160...400°C)
≤ ±2°C
(400...1820°C)
≤ ±0.2°C / °C
(400...1820°C)
E
-100°C
+1000°C
≤ ±1°C
≤ ±0.1°C / °C
J
-100°C
+1200°C
≤ ±1°C
≤ ±0.1°C / °C
K
-180°C
+1372°C
≤ ±1°C
≤ ±0.1°C / °C
L
-200°C
+ 900°C
≤ ±1°C
≤ ±0.1°C / °C
N
-180°C
+1300°C
≤ ±1°C
≤ ±0.1°C / °C
R
-50°C
+1760°C
≤ ±2°C
≤ ±0.2°C / °C
S
-50°C
+1760°C
≤ ±2°C
≤ ±0.2°C / °C
T
-200°C
+400°C
≤ ±1°C
≤ ±0.1°C / °C
U
-200°C
+600°C
≤ ±1°C
≤ ±0.1°C / °C
W3
0°C
+2300°C
≤ ±2°C
≤ ±0.2°C / °C
W5
0°C
+2300°C
≤ ±2°C
≤ ±0.2°C / °C
LR
-200°C
+800°C
≤ ±2°C
≤ ±0.2°C / °C

Summary of content (1 pages)