VIVIX-S 1012N User Manual Mini SCU 1 [TBD] If you use the DR Trigger mode, connect the one end of generator interface cable to the EXT_INF port of SCU, and the other to the X-ray generator. If you use the AED mode, a generator interface cable is not needed as the detector operates by detecting X-ray automatically. 2 Connect one end of the LAN cable to one of the LAN ports of SCU, and the other to the LAN Card Connector of workstation assigned for data transfer.
VIVIX-S 1012N User Manual 4.2.2 Booting up SCU 1 Turn on the power switch at the front side of SCU. 2 Check if the power switch of SCU lights up green. 3 Check if the status indicator is blinking in green after the power switch of SCU is permitted. 4 Check if the status indicator of SCU is blinking in green and turns to light up in the same color. 5 Check if the status indicator of SCU is blinking in green and turns to light up in blue.
VIVIX-S 1012N User Manual 4.2.4 Checking Status LED of Detector [TBD] Power LED The power LED indicates the power status information which is permitted to the detector in green. The power LED lights up when the power is permitted normally. If the detector is connected with a tether interface, the power LED lights up when power is permitted to SCU because the detector is supplied power from SCU. The power LED is blinking when the detector is in the sleep mode.
VIVIX-S 1012N User Manual 4.3 Software Installation This section gives information about how to install the software on the workstation (PC) and how to configure the environment for software operation and communication. Check suitability of acquiring, processing and adjusting of images by referring to the recommended workstation specifications before the software installation. 4.3.1 Software Classification [TBD] Vieworks provides clients who purchase our detector system with software as below.
VIVIX-S 1012N User Manual 4.4 Windows Environment Setting This section gives information about configuring Windows to communicate with the detector and SCU after installing the Setup program or Viewer. The contents in this chapter are made on the basis of Windows 7. Configuration environment can be different depending on network adaptor manufacturer or models. 4.4.
VIVIX-S 1012N User Manual 3 Click the chosen network adaptor with the right mouse button and click Properties to display the Properties window. Setting Network Adaptor 1 Click Configure button to open the following dialog box, and then go to the Advanced tab. 2 Set Jumbo Packet to the maximum value. (Recommended value: 9014 Bytes) 3 Choose Performance Options in the list of Settings and click Properties button on the right.
VIVIX-S 1012N User Manual 5 Choose Interrupt Moderation Rate in the list of Settings and select Extreme on the Value list as shown below. 6 Choose Receive Buffers and set it to the maximum value. 7 Click OK button. Deactivate Power-Saving Mode on Network Adaptor 1 Click the Power Management tab and uncheck Allow the computer to turn off this device to save power. 2 Click OK button. Viewer may not work normally if the power save mode is activated.
VIVIX-S 1012N User Manual Protocol Selection and IP Address Setting 1 Choose Internet Protocol Version 4 (TCP/IPv4).and click Properties button. 2 Input the IP address and subnet mask as shown below, and then click OK button. ViVIX Device Driver is the image filter driver for acquiring images from a detector. It is recommended to uncheck the other items on the list except for ViVIX Device Driver and Internet Protocol Version 4(TCP/IPv4), since they are not related with the detector communication.
VIVIX-S 1012N User Manual 4.4.2 Disabling Sleep Mode of Monitor If you use the sleep mode, viewer may not work normally. 1 Click Start Control Panel Power Options and then move to the Choose when to turn off the display tab. Rev.1.0 2 Set Put the computer to sleep to Never to disable the sleep mode. 3 Click Save changes button.
VIVIX-S 1012N User Manual 5. Setting [TBD] This section gives information about the product setting with using the Setup program. Start Setting VIVIX SCU Setting Detector Setting Rev.1.
VIVIX-S 1012N User Manual 5.1 Start Setting 5.1.1 Setup Program The Setup program provides functions for setting and managing the internal data to make VIVIX-S 1012N wireless detector and SCU work normally. In addition, The Setup program has diagnostic function for checking performance and abnormality of the devices as well as the image calibration function for improving the image quality. 5.1.2 Checking Devices 5.1.3 Getting into the Devices Rev.1.
VIVIX-S 1012N User Manual 5.2 Rev.1.
VIVIX-S 1012N User Manual 5.3 Detector Setting 5.3.1 Detector Configuration 5.3.2 Detector Power Save Function Rev.1.
VIVIX-S 1012N User Manual 6. Calibration [TBD] This chapter gives information about the calibration methods after installing a detector. Calibration Dialogue Detector Configuration Calibration Guide Calibrating by Loading the Calibration Data Direct Calibration Rev.1.
VIVIX-S 1012N User Manual 6.1 Rev.1.
VIVIX-S 1012N User Manual 6.2 Rev.1.
VIVIX-S 1012N User Manual 6.3 Calibration Guide The different installation environment of each detector and unique features of the X-ray generator device can affect the acquired images. Therefore, the certified engineer from Vieworks should do the detector calibration after installing it. Otherwise, the image quality can be affected seriously. Vieworks provides two types of calibration for performing calibration. Performing calibration by loading the calibration data CD provided by Vieworks.
VIVIX-S 1012N User Manual 6.4 Calibrating by Loading the Calibration Data Vieworks provides the offset calibration data as a CD with the detector. However, we recommend the service enginner to carry out the calibration by oneself as the detector condition can be different by the operating method or use environment. 6.4.1 Preparing Calibration Data 1 Prepare the calibration data CD provided with the detector. 2 Copy the calibration data files and save them to a safe route such as local HDD.
VIVIX-S 1012N User Manual 6.5 Rev.1.
VIVIX-S 1012N User Manual 7. Diagnosis, Inspection and Maintenance [TBD] This section gives information about diagnosis, inspection & maintenance of the product. Diagnosis Product Inspection Cleaning and Disinfection Product Initialization Replacing the Fuse of VIVIX SCU Rev.1.
VIVIX-S 1012N User Manual 7.1 Diagnosis 7.1.1 Image Diagnosis Check the image quality through Diagnosis tools after installing the detector or before usage. If the problems with regard to products or image occur during diagnosis, try to do a calibration again. If the problems are not solved, consult the sales representative in Vieworks or a service engineer. You can acquire and review an image from the Diagnosis window in Setup program.
VIVIX-S 1012N User Manual Check the Connection Status User can check the connection status of wired or wireless detector through VXvue or VIVIX SDK. Check the Strength of Wireless Communication Signal The signal strength of wireless communication is provided as 5 stages or percentage (%), user can check it through VXvue (Vieworks Viewer) and VIVIX SDK. Since the signal strength of wireless communication at the 1st stage is not good, it is required to check the surrounding wireless communication status.
VIVIX-S 1012N User Manual 7.2 Product Inspection To use products safely, make sure to check the products before use. If problems occur during inspection or the product is impossible to repair, consult the sales representative in Vieworks or a relevant engineer. 7.2.1 Daily Inspection Before or after using the detector and other surrounding devices, check below items daily. Item Description Ensure that there are no loose screws or breaks.
VIVIX-S 1012N User Manual 7.3 Cleaning and Disinfection After using the detector and peripheral equipments for examination, use germicidal disinfecting wipes or cloth with mild diluted disinfectant detergent to clean surfaces of the product.
VIVIX-S 1012N User Manual 7.4 Product Initialization If the connection status of system is not stable or setting value is not correct, user can initialize the products. 7.4.1 SCU Initialization 1 Click Configuration button of SCU after running the VIVIX Setup program. 2 Click Factory Reset button in Configuration dialogue. 3 Wait for SCU to be initialized and rebooted automatically. 4 Check whether SCU initialization is completed.
VIVIX-S 1012N User Manual 7.4.2 Detector Initialization 1 Click Configuration button of Detector after running the VIVIX Setup program. 2 Click Factory Reset button in Configuration dialogue. 3 Wait for Detector to be rebooted automatically. 4 Check whether detector initialization is completed.
VIVIX-S 1012N User Manual 7.4.3 Wireless Initialization of Detector 1 Turn off the detector. 2 Press and hold the power button while installing a battery into the battery compartment of the detector. 3 Keep pressing the power button. Keep pressing the power button until the orange LED flickers three times. (It will take about 15 seconds.) 4 Initialization will be conducted when orange LED is flickering, after that, the detector will be turned off automatically.
VIVIX-S 1012N User Manual 7.5 Replacing the Fuse of SCU (Basic SCU only) There are 2 fuses attached on the standard SCU for the purpose of electrical accident precaution, in case of over current from external power input. Stop using the SCU immediately when the fuse is blown. Fuse Information Item Specifications Model Littelfuse® 218002 (2EA) Type Time Lag Cartridge Fuse Amp Rating 2A Voltage Rating 250V Pull the plug out and turn all the devices off before changing the fuse.
VIVIX-S 1012N User Manual 8. Troubleshooting [TBD] This section gives information about troubleshooting. Troubleshooting Rev.1.
VIVIX-S 1012N User Manual 8.1 Trobleshooting When you encounter problems while using the equipment, search for the table below for the problem or error messages and try the solutions.If the problem persists, turn off the detector and consult your sales representative or a distributor. Please refer to the details of the following symptoms or error messages. Troubleshooting must be performed by service engineer who is authorized by Vieworks.
VIVIX-S 1012N User Manual Solutions 8.1.3 1 Check the connection between AC power cable and basic SCU. 2 Check the connection between DC power cable and mini SCU. 3 Turn off the power switch and turn on again and then check the fan or status of back side. 4 Replace another SCU and check the result of it 5 Replace the fuse of standard SCU. (refer to Change fuse of SCU) 6 Replace corresponding devices.
VIVIX-S 1012N User Manual 8.1.5 4 Check if the surrounding wireless communication is good 5 Check cable connection again when connected with tether interface cabel 6 Replace other devices and check the result of it 7 Replace corresponding devices. Errors in Detector LED Symptom All LED lamps of detector are blinking. 2 LED lamps of detector are blinking and remaining is blinking slowly. Expected Causes Internal hardware errors of the detector. Solutions 8.1.
VIVIX-S 1012N User Manual 2 Rev.1.0 Consult with service engineers of Vieworks.
VIVIX-S 1012N User Manual 9. Regulatory Information This section gives explanation about the regulatory information and standard related to the products Medical Equipment Safety Standards Radio Frequency Compliance Information Labels and Symbols Guidance and Manufacturer Declaration for EMC Rev.1.
VIVIX-S 1012N User Manual 9.1 Medical Equipment Safety Standards 9.1.1 Medical Equipment Classification Item Description Type of protection against electrical shock Class Ⅰ equipment Internally powered Degree of protection against electrical shock Type B applied parts Degree of protection against ingress of water IPX0 Operation mode Continuous operation NOT suitable for use in the presence of a flammable Flammable anesthetics anesthetic mixture with air or with oxygen or nitrous oxide. 9.1.
VIVIX-S 1012N User Manual U.S.A / Canada Item IEC 60601-1(ed.2 am1+ am2+ co1) Medical electrical equipment- Part1: General requirements for safety UL 60601-1(ed.2) CSA-C22.2 No. 601-1-M90 (R2006) IEC 60601-1-2: 2007 (ed.3) IEC 60601-1-4: 2000 (ed.1.
VIVIX-S 1012N User Manual 9.2 Radio Frequency Compliance Information Country Item FCC Part 15.107 Subpart (b) / 15.109(g) Subpart B U.S.A FCC Part 15 Subpart E 15.407 FCC Part 15 Subpart C 15.247 European Union Clause 3, Article 58-2 of Radio Waves Act South Korea Clause 2, Article 58-2 of Radio Waves Act Japan 9.2.
VIVIX-S 1012N User Manual 9.3 Labels and Symbols The ViVIX-S 1012N detector and relevant components have labels attached on them. The contents and locations of each label are indicated below. 9.3.1 Label Detector Battery Rev.1.
VIVIX-S 1012N User Manual VIVIX SCU (VIVIX System Control Unit) Battery Charger Rev.1.
VIVIX-S 1012N User Manual Outer Box 9.3.2 [TBD] Product Serial Number Serial Number Composition The serial numbers for each product or accessory are composed as follows. 1 0 - Revision R - 1 D Item A Composition B Year J 0 Month 0 1 Serial number Revision will be updated in case of follows. Mass production or a large amount of order. Exterior alteration. Item code will be produced based on internal management standard of vieworks. Composition code is like follows.
VIVIX-S 1012N User Manual 9.3.3 Product Symbols Symbol Description Direct current Alternating current Protective earth (Ground) Equipotentiality Power on Power on for part of the equipment Power off Power off for part of the equipment Attention, consult accompanying documents General warning sign Warning sign for electricity This Mark shows compliance with both Canadian and U.S. safety requirements. With Respect to electric shock, fire, and mechanical hazards only.
VIVIX-S 1012N User Manual 9.4 Guidance and Manufacturer’s Declaration for EMC This device has been tested for EMI/EMC compliance, but interference can still occur in an electromagnetically noisy location. Attempt to maintain a suitable distance between electrical devices to prevent malfunction. 9.4.1 Electromagnetic Emissions The Equipment Under Test (EUT) is intended for use in the electromagnetic environment specified below.
VIVIX-S 1012N User Manual 9.4.2 Electromagnetic Immunity The ViVIX-S 1012N system is intended for using in the electromagnetic environment specified below. The user of this system should assure that it is used in the following environment.
VIVIX-S 1012N User Manual Voltage Dips, Short Interruptions and Voltage Variations on Power Supply Input Lines IEC 61000-4-11 Item Description Voltage dips, short interruptions and voltage variations on power Immunity test supply input lines IEC 61000-4-11 <5% Uт (>95% dip in Uт) for 0.5 cycle. IEC 60601 test condition 40% Uт (60% dip in Uт) for 5 cycles. 70% Uт (30% dip in Uт) for 25 cycles. <5% Uт (<95% dip in Uт) for 5 sec. <5% Uт (>95% dip in Uт) for 0.5 cycle.
VIVIX-S 1012N User Manual Conducted RF IEC 61000-4-6 / Radiated RF IEC 61000-4-3 Item Description Immunity test Conducted RF IEC 61000-4-6 Radiated RF IEC 61000-4-3 IEC 60601 test condition 3 Vrms 150 ㎑ to 80 ㎒ 3 V/m 80 ㎒ to 2.5 ㎓ Compliance Level 3 Vrms 150 ㎑ to 80 ㎒ 3 V/m 80 ㎒ to 2.
VIVIX-S 1012N User Manual 10. Information This section gives overview information for service and warranty of the product. Service Information Warranty Revision History Rev.1.
VIVIX-S 1012N User Manual 10.1 Service Information 10.1.1 Product Lifetime The estimated product lifetime may be up to five (5) years under the appropriate regular inspection and maintenance. 10.1.2 Regular Inspection and Maintenance In order to ensure the safety of patients, operating personnel and third parties, and to maintain the performance and reliability of the equipment, be sure to perform regular inspection at least once a year.
VIVIX-S 1012N User Manual 10.2 Warranty Vieworks warrants that this product will be free from defects in materials and workmanship for a period of 24 months from the date of delivery. If any such product proves defective during this warranty period, Vieworks at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
VIVIX-S 1012N User Manual 10.3 Revision History Version Date Descriptions 1.0 2015-01-?? Initial Release Vieworks Co., Ltd. (Gwanyang-dong) 41-3, Burim-ro 170beon-gil, Dongan-gu, Anyang-si, Gyeonggi-do, 431-060 Republic of Korea Telephone: +82-70-7011-6161 Fax: +82-31-386-8631 Homepage: http://www.vieworks.com European representative: DONGBANG ACUPRIME 1 Forrest Units, Hennock Road East, Marsh Barton, Exeter EX2 8RU, UK Telephone: +44(0)-1392-829500 Homepage: http://www.acuprime.com Rev.1.