VIVIX-S 1012N User Manual © Vieworks. 2015 All rights reserved. Under copyright laws, this manual should not be reproduced, in whole or in part, without the written permission of Vieworks.
VIVIX-S 1012N User Manual Contents 1. Instruction ......................................................................................................................................... 8 1.1 Document Guide ..................................................................................................................................... 9 1.1.1 Caution Information .........................................................................................................................................
VIVIX-S 1012N User Manual 2.2.1 Detector............................................................................................................................................................................23 2.2.2 SCU (System Control Unit) ......................................................................................................................................23 2.2.3 Battery & Charger ..................................................................................................
VIVIX-S 1012N User Manual 3.2 Diagram................................................................................................................................................... 49 3.2.1 Block Diagram ...............................................................................................................................................................49 3.2.2 Wiring Diagram ...................................................................................................................
VIVIX-S 1012N User Manual 5.1.2 Checking Devices .........................................................................................................................................................80 5.1.3 Getting into the Devices ..........................................................................................................................................82 5.2 SCU Setting .................................................................................................................
VIVIX-S 1012N User Manual 7.2.1 Daily Inspection ......................................................................................................................................................... 120 7.2.2 Performance Inspection ......................................................................................................................................... 120 7.3 Cleaning and Disinfection ..................................................................................................
VIVIX-S 1012N User Manual 10.1.3 Repair ............................................................................................................................................................................. 149 10.1.4 Replacement Parts Support ................................................................................................................................. 149 10.1.5 Consumables.......................................................................................................
VIVIX-S 1012N User Manual 1. Instruction This section gives basic information of this manual and products with the safety guide. Document Guide Product Use Guide Safety Instruction Product Operation Guide Rev.1.
VIVIX-S 1012N User Manual 1.1 Document Guide This User Manual explains how to use the VIVIX-S 1012N detector made by Vieworks, X-ray interface unit, and other peripheral equipment. With this user manual, you can install, set and manage the VIVIX-S 1012N wired/wireless detectors as well as use its various functions. 1.1.1 Caution Information If the user is not fully aquainted with this manual, the product can be malfunctioned or unsuspected problem can be happened due to carelessness.
VIVIX-S 1012N User Manual 1.1.4 Notations Bold Types We applied bold font style to the words which indicated products terms, or the words and sentences which are needed to transmit clear meaning to the customers. This helps you to easily distinguish the words from other technical ones for explaining functions. 1.1.5 Contact Department For any comments or inquiries regarding this document, contact via email below. Rev.1.
VIVIX-S 1012N User Manual 1.2 Product Use Guide This chapter provides instructions about the use of product and disposal as well as the liability limit of Vieworks. 1.2.1 1.2.2 Product Usage 1 Only a physician or a legally certified operator should use this product. 2 The equipment should be kept in a safe and operable condition by maintenance personnel. 3 Observe the contents written in this user manual when installing and using this product.
VIVIX-S 1012N User Manual 1.2.3 Product Disposal Disposal of this product in an unlawful manner may have a negative impact on human health and the environment. When disposing this product, therefore, be absolutely sure to follow the procedure which is in conformity with the laws and regulations applicable in your area. European Union (and EEA*) only This symbol indicates that this product is not to be disposed with your household waste, according to the WEEE Directive (2002/96/EC) and your national law.
VIVIX-S 1012N User Manual 1.3 Safety Instruction This product is designed and manufactured to ensure maximum safety of operation and to meet all the safety requirements applicable to electronic medical equipment. Follow these safeguards while using the products. If not, severe personal injury or substantial property damage can be happened. It is important for you to read and understand the contents of this manual before attempting to use the product. 1.3.
VIVIX-S 1012N User Manual To make it easy to disconnect the plug at any time, avoid putting any obstacles near the outlet. Otherwise, it may not be possible to disconnect the plug in an emergency. Do not place a heavy object such as medical equipment on cables and cords, and do not pull, bend, bundle, or step on them. Doing so may damage their sheath. If you alter them, it may damage the cords which could result in a fire or electric shock. Securely plug the power cord into the AC outlet.
VIVIX-S 1012N User Manual Do not block the ventilation ports of SCU to prevent overheating. Overheating can cause product’s malfunctions and damages. 1.3.4 Battery Pack and Charger Do not use the battery pack as a power source for equipment other than VIVIX-S 1012N detector. Be sure to use only the dedicated battery pack for the VIVIX-S 1012N detector. The battery charger is designed for the dedicated battery pack. Do not use the battery charger other than the dedicated one.
VIVIX-S 1012N User Manual 1.3.5 Environment of Use Do not install the equipment in any of the locations listed below. Doing so may result in failure or malfuction, equipment falling, fire, or injury. Close to facilities where water is used.
VIVIX-S 1012N User Manual 1.3.6 Problem Management Should any of the following occur, immediately turn OFF the power to each piece of equipment, unplug the power cord from the AC outlet, and contact sales representative or distributor. When there is smoke, an odd smell or abnormal sound no period. When liquid is spilled into the equipment or a metal object is entered through an opening. When the equipment has been dropped and is damaged. 1.3.
VIVIX-S 1012N User Manual 1.4 Product Usage Guide When using the equipment, take the following precautions. Otherwise, problems may occur and the equipment may not function correctly. 1.4.1 Calibration To ensure optimal system performance, it is important to verify that the system is calibrated correctly. Check if the calibration is performed after the equipment is completed to be installed or repaired. Do not try to use the system if the calibration has not been performed.
VIVIX-S 1012N User Manual 1.4.5 1.4.6 Use Limitations Vieworks software has the engineer mode operated by inputting the administrator password only. This mode should be operated by the person who is qualified by Vieworks. Disposal Disposal of this product in an unlawful manner may have negative effects on human health and the environment. Be sure to follow the procedure which is in conformity with the laws and regulations applicable in your area. 1.4.
VIVIX-S 1012N User Manual 1.4.9 During Exposure This equipment is not protected (sealed) against liquids such as blood and medicine in the operating room. If necessary, wrap the equipment in a disposable cover when using it. Do not use the selected frequency channel (2.4 ㎓ and 5 ㎓ dual band) for other wireless devices. Mutual interference may affect the image data transmission rate. Do not use the detector near devices generating a strong magnetic field.
VIVIX-S 1012N User Manual 2. Product This section gives an instruction about the product components and their specifications. Overview Product Components VIVIX-S 1012N Detector SCU Battery Pack Battery Charger Others Rev.1.
VIVIX-S 1012N User Manual 2.1 Overview ViVIX-S 1012N detector is the digital X-ray imaging solution. It acquires images by exposing X-ray which has been penetrated the human body. When X-ray photons pass through scintillator in the detector, the photons convert to visible ray, and the visible ray is converted to electronic signals through TFT (a-Si). Then the detector digitalizes x-ray images and transfers them to the computer (workstation) for radiography diagnostics.
VIVIX-S 1012N User Manual 2.2 Product Components 2.2.1 Detector Component Description VIVIX-S 1012N FXRD-1012NAW (2.2kg) Wireless Detector FXRD-1012NBW (2.2kg) VIVIX-S 1012N FXRD-1012NA (2.2kg) Wired Detector FXRD-1012NB (2.2kg) A deco sheet attached on the detector can be different depending on each client company. 2.2.2 SCU (System Control Unit) Component Description FXRS-02A (2.5kg) SCU Basic (FXRS-02A) AC Power Cable (2m) FXRS-03A (2.
VIVIX-S 1012N User Manual FXRS-04A (1.2kg) SCU mini DC Power Supply (24V) (FXRS-04A) AC Power Cable (2m) You can choose the type of SCU depending on the purpose of use. 2.2.3 Battery & Charger Component Description Battery FXRB-03A (0.12kg, 2EA) FXRC-02A (0.8kg) Charger DC Power Supply (24V) Rev.1.
VIVIX-S 1012N User Manual 2.2.4 Accessaries Component Description Software (Viewer or SDK) Resource Manuals Tether Interface Cable (7m) Ether Con Cable (7m, Option) Cables Generator Interface Cable (7m) UTP LAN Cable (15m, Direct) CAT 5E or CAT6 The use of accessories and cables other than those approved and sold by Vieworks Co., Ltd. may result in increased release of electromagnetic waves or decreased stability of the equipment.
VIVIX-S 1012N User Manual 2.3 VIVIX-S 1012N Detector VIVIX-S 1012N is designed to acquire digital images by collecting x-ray signals and sereval conversion processes. You can use the acquired image (10” x 12” film size) diversely depending on the purpose of use. VIVIX-S 1012N is one of the VIVIX detector models produced by Vieworks. 2.3.
VIVIX-S 1012N User Manual 2.3.2 Drawing Sheet VIVIX-S 1012NAW / 1012NBW (Wireless Detector) Item Description Dimensions (H × W × D) 287.0㎜ × 350.0㎜ × 15.0㎜ Curvature of Edges R6.5 Under the ISO4090 regulation, the allowed tolerance of a thickness of detector is from 2㎜ ~ +1㎜. Rev.1.
VIVIX-S 1012N User Manual VIVIX-S 1012NA / 1012NB (Wired Detector with a tether interface cable) Item Description Dimensions (H × W × D) 287.0㎜ × 350.0㎜ × 15.0㎜ Curvature of Edges R6.5 Under the ISO4090 regulation, the allowed tolerance of a thickness of detector is from 2㎜ ~ +1㎜. Rev.1.
VIVIX-S 1012N User Manual 2.3.3 Functions VIVIX-S 1012NAW / 1012NBW (Wireless) Name Description Used for tighten the tether interface cable. A Tether interface connector B Tether interface holder Fix/release holder of the tether interface cable. C Battery lock lever Lock/unlock lever of the equipped battery pack. D Antenna for wireless LAN Antennas for wireless communication (2EA) Used for wired connection between a detector and SCU.
VIVIX-S 1012N User Manual VIVIX-S 1012NA / 1012NB (Wired) Name A Tether interface connector B Tether interface holder Description Used for tighten the tether interface cable. Used for wired connection between a detector and SCU. Fix/release holder of the tether interface cable. Detector status indicator C Status indicator DATA LED (Blue): Indicates communication and transmission ACTIVE LED (Orange): Indicates ready to work POWER LED (Green): Indicates power On/Off status D Rev.1.
VIVIX-S 1012N User Manual 2.3.4 Deco Sheet VIVIX-S 1012NAW / 1012NBW (Wireless) Indication Info. Description A Wireless communication logo Indicates that this model can be operated wirelessly. B Status indication logo C Image starting point Indicates the starting point of an original image. D Center of the detector Indicates the central position of detector. E Vieworks logo Indicates the logo of Vieworks.
VIVIX-S 1012N User Manual VIVIX-S 1012NA / 1012NB (Wired) Indication Info. Rev.1.0 Description Indicates the operating state of detector. A Status indication logo B Image starting point Indicates the starting point of an original image. C Center of the detector Indicates the central position of detector. D Vieworks logo Indicates the logo of Vieworks. E Certification logo Indicates certification logos relating to a medical device.
VIVIX-S 1012N User Manual 2.3.5 Wireless Communication Item Specifications Wireless standard IEEE802.11n 2.412 ~ 2.472㎓ (13 Channels) Frequency range 5.18 ~ 5.24㎓ (4 Channels) 5.745 ~ 5.805㎓ (4 Channels) Data transmission rate 802.11n: Max. 300Mbps (MIMO 2X2) Modulation BPSK, QPSK, 16-QAM, 64-QAM Transmission power Max. 17dBm Security WPA2-PSK Antenna Dual Band Antennas (2EA, inbuilt) The specification of detector and SCU are same, but the location of antenna is different. 2.3.
VIVIX-S 1012N User Manual 2.4 SCU Basic (FXRS-02A) SCU synchronizes the image and X-ray signal as locating among the detector, workstation and the X-ray generator. You can use the SCU directly when the power supplies to SCU after connecting it under the VIVIX-S 1012N system environment. 2.4.1 Specifications SCU Item Specifications Model FXRS-02A Power Supply Cable Connection port 2.4.2 Rev.1.0 Input: AC100 to 240V, 50/60㎐, Max. 200VA Output: DC +24V 3.
VIVIX-S 1012N User Manual 2.4.3 Functions Front Side No. Name Description 1 Power Switch SCU power swtich (ON or OFF) (Including green LED Lamp) 2 Fan Expels interior air of SCU Rear Side No. Name 1 EXT_INF1 EXT_INF2 Description X-ray generator interface connector (D-SUB 15 pins x 2 EA, Female) EXT_INF1 : 1 ~ 15 EXT_INF2 : 16 ~ 30 Gigabit Ethernet port (1000BASE-T) 2 LAN port Port 1: Communicates between workstation and SCU Basic.
VIVIX-S 1012N User Manual Orange: 100 Mbps 4 PoE (Power over Ethernet) port (1000BASE-T) PoE port Communicates between the detector and SCU Basic. (Port 4, 5) Supplies power to the detector. 5 Detector power supply port 6 P.E Max. DC +24V/15W (2 ports) For FXRD-1717S detector only. Equipotential ground T2AL250V fuse (2 EA) 7 AC input port 100 ~ 240V 50/60㎐ Supplies power to SCU Basic. The P.
VIVIX-S 1012N User Manual 2.5 SCU Basic (FXRS-03A) SCU Basic synchronizes the image and X-ray signal as locating among the detector, workstation and the Xray generator. You can use the SCU Basic directly when the power supplies to SCU Basic after connecting it under the ViVIX-S 1012N system environment 2.5.1 Specifications Item Specifications Model FXRS-03A Power supply Cable connection port 2.5.2 Rev.1.0 Input: AC100 to 240V, 50/60㎐, Max. 200VA Output: DC +24V 3.
VIVIX-S 1012N User Manual 2.5.3 Functions Front Side No. Name Description 1 Power switch Turns on/off the power of SCU Basic. (Including Green LED Lamp) 2 Fan Expels the air inside of SCU Basic. 3 Antenna Assists communications between the detector and SCU Basic. Indicates the operation and connection status of SCU Basic. 4 Status LED Blinking green: Booting Green: Completed to boot up Blue: The software is connected and ready to communicate. Rev.1.
VIVIX-S 1012N User Manual Rear Side No. Name 1 Description X-ray generator interface connector (D-SUB 15 pins x 2 EA, Female) EXT_INF1 EXT_INF1 : 1 ~ 15 EXT_INF2 EXT_INF2 : 16 ~ 30 Gigabit Ethernet port (1000BASE-T) 2 LAN port Port 1: Communicates between workstation and SCU Basic. (Port 1, 2, 3) Port 2, 3: Communicates between FXRD-1717S detector and SCU Basic when configuring multiple detectors.
VIVIX-S 1012N User Manual 2.6 SCU mini (FXRS-04A) SCU mini synchronizes the image and X-ray signal as locating among the detector, workstation and the Xray generator. You can use the SCU mini directly when the power supplies to SCU mini after connecting it under the VIVIX-S 1012N system environment 2.6.1 Specifications Item Specifications Model FXRS-04A Power supply Input: DC +24V 2A Max Cable connection port 2.6.2 Rev.1.
VIVIX-S 1012N User Manual 2.6.3 Functions Front Side No. Name Description 1 Power switch Turns on/off the power of SCU mini. (Including Green LED Lamp) 2 Antenna Assists communications between the detector and SCU mini. Indicates status of SCU mini operation and connection. 3 Status LED Blinking green: Booting Green: Completed to boot up Blue: The detector is connected and ready to communicate. Rev.1.
VIVIX-S 1012N User Manual Rear Side No. Name Description 1 X-ray generator interface connector (D-SUB 15pin, Female) EXT_INF Gigabit Ethernet port (1000BASE-T) 2 LAN port Port 1: Communication between the workstation and SCU mini. (Port 1, 2, 3) Port 2, 3: Communication between FXRD-1717S detector and SCU mini when configuring multiple detectors. Indicats the status of PoE port.
VIVIX-S 1012N User Manual 2.7 Battery Pack 2.7.1 Specifications 2.7.2 Rev.1.0 Item Specifications Model FXRB-03A Type Lithium Ion Rated Power Supply Output: DC +7.6V Capacity 3,100㎃h Number of cell 2S1P (2 Series 1 Parallel) Life Approx. 500 times (Fully charged/Discharged completely, 1cycle) Dimension (H × W × D) 160.0 ㎜ × 61.8 ㎜ × 5.
VIVIX-S 1012N User Manual 2.8 Battery Charger 2.8.1 Specifications 2.8.2 Rev.1.0 Item Specifications Model FXRC-02A Simultaneous Charging 2 battery packs Charging time 3 hours Rated power supply DC +24V, 2A Max. Dimension (H × W × D) 192.0 ㎜ × 101.0 ㎜ × 26.0 ㎜ Weight 0.8 ㎏ Drawing Sheet No. Name Description 1 Top / Bottom LED Indicates the position and status of a battery being charged.
VIVIX-S 1012N User Manual 2.9 Others 2.9.1 X-ray Generator (Recommended Exposure Condition) 2.9.2 Item Recommended condition X-ray energy range 40kVp ~ 150kVp Reliability (Lifetime Dose) 74Gy or higher (35 uGy x 365days x 24hrs x 60min. x 60sec. / 15sec.
VIVIX-S 1012N User Manual 3. System Configuration This section gives information about the various connection / configuration ways among the detector, SCU, workstaton and X-ray generator. You can figure out the most suitable way of system configuration through this instruction. Detector Connection Method Diagram System Configuration Generator Interface Rev.1.
VIVIX-S 1012N User Manual 3.1 Detector Connection Methods You can organize wired or wireless connection between the VIVIX-S 1012N detector and SCU. You can also choose other connection ways suitable for the use environment in case of need. 3.1.1 Wireless Connection The images and data are transmitted between the VIVIX-S 1012N detector and SCU by wireless connection. A battry pack should be installed in the detector to use it under the wireless configuration.
VIVIX-S 1012N User Manual 3.1.2 Wired Connection Connect VIVIX-S 1012N detector and SCU with a tether interface cable to make a wired configuration. As the tether interface cable supplies power to the detector, a battery pack is not needed to be installed in the detector. We recommend you the wired connection method for the following cases. When continuous power supply to the detector is needed. When the detector is fixed in a bucky stand or on a table.
VIVIX-S 1012N User Manual 3.2 Diagram 3.2.1 Block Diagram 3.2.2 Wiring Diagram Rev.1.
VIVIX-S 1012N User Manual 3.3 System Configuration This section gives information about the configuration modes of VIVIX-S 1012N detector, SCU, workstation and the generator. 3.3.1 AP Mode (SCU AP Mode) SCU AP mode is the wireless connection method that connecting the detector with SCU as the center wirelessly. SCU & Detector The VIVIX-S 1012N detector and SCU are connected wirelessly. In this case, SCU and the detector operate as AP (Access Point) and STATION respectively.
VIVIX-S 1012N User Manual 3.3.2 Tether Interface Mode Tether interface mode is the wired connection method that connecting the detector with SCU as the center. SCU & Detector SCU and the VIVIX-S 1012N detector are connected with the tether interface cable. Other connection ways of the tether interface mode are the same as the AP mode. Rev.1.
VIVIX-S 1012N User Manual 3.3.3 External AP Mode External AP mode is the wireless connection method that adding the external AP device instead of SCU. External AP & SCU The external AP and SCU are connected with a LAN cable. External AP & Detector The external AP and VIVIX-S 1012N detector are connected wirelessly. In this case, the external AP and the detector operate as AP (Access Point) and STATION respectively. Other connection ways of the tether interface mode are the same as the AP mode.
VIVIX-S 1012N User Manual 3.3.4 Detector AP Mode Detector AP mode is the wireless connection method that connecting the PC (Workstation) with the detector as the center wirelessly. Detector & PC (Workstation) The VIVIX-S 1012N detector and PC are connected wirelessly. In this case, the detector and the PC operate as AP (Access Point) and STATION respectively. Since SCU is not used for configuring the detector AP mode, you can only use the AED mode without a wired connection.
VIVIX-S 1012N User Manual 3.3.5 Detector Stand-Alone Mode You can use the Stand-Alone mode without making a wired or wireless connection. You should take images with the AED mode only when using the Stand-Alone mode. (Cannot use the DR Trigger mode.) The acquired images are stored in the internal storage of the detector in order and the images can be transmitted to the PC after connecting the detector to the PC. The transmitted images are removed automatically from the detector. Rev.1.
VIVIX-S 1012N User Manual 3.4 Generator Interface VIVIX-S 1012N detector provides DR Trigger interface and AED interface as a generator interface way to acquire images by detecting X-ray. Mode AED Description The detector detects X-ray exposure from the generator automatically and then performs image acquisition without any cable connection. The detector and generator receive and send their signal to each other for image DR Trigger acquisition.
VIVIX-S 1012N User Manual 3.4.2 DR Trigger Interface Connect the detector and X-ray generator with a generator interface cable, and then acquire images by sending and receiving their signal. Signal Processing Steps 1 The detector receives EXP_REQ signal from the X-ray generator. 2 After the detector completes to prepare image acquisition, it sends the EXP_OK signal to the X-ray generator. 3 The X-ray generator confirms EXP_OK signal and generates X-rays.
VIVIX-S 1012N User Manual 3.5 Configuring DR Trigger Interface To configure the DR Trigger interface, SCU and the generator should be connected with a generator interface cable. Connect the one end of generator interface cable to EXT_INF port of SCU, and then connect the other end to the generator. Up to three generators can be connected with SCU Basic. Only one generator can be connected with SCU mini. 3.5.
VIVIX-S 1012N User Manual 3.5.3 Line Trigger Each detector shares a signal signal by being connected with generators separately. Up to three X-ray generators can be connected to SCU. The connection of each detector and generator can be configured from VIVIX Setup. You should comprehend about a detector and X-ray generator to make correct exposure. Otherwise, X-ray can be exposed to the detector in different location.
VIVIX-S 1012N User Manual 3.5.4 EXT_INF Port Pin Map EXT_INF port is D-SUB 15 pin female connectors of SCU, and used for the generator interface. Product EXT_INF port SCU Basic (FXRS-03A) SCU mini (FXRS-04A) EXT_INF1 port pin map definition (1 ~ 15) Rev.1.0 No. Signal name I/O Type Color Ch.
VIVIX-S 1012N User Manual EXT_INF2 port pin map definition (16 ~ 30) No. Signal name I/O Type Color Ch.
VIVIX-S 1012N User Manual 3.5.5 Input / Output Circuit Exposure Request Input Circuit (EXP_REQ) Exposure Respond Output Circuit (EXP_OK) Rev.1.
VIVIX-S 1012N User Manual 4. Installation This section gives information about the installation process and method to use a detector. Battery Pack Product Installation Generator Connection Software Installation Windows Environment Setting Rev.1.
VIVIX-S 1012N User Manual 4.1 Battery Pack Detector and its attached devices should be installed by an authorized service engineer. If you encounter any problems, consult the sales representative in Vieworks or a relevant engineer. 4.1.1 How to insert a battery pack 1 Open the battery cover. 2 Insert a battery into a slot until it clicks and push the lever to the right for fixing the battery. 3 Close the battery cover. Check if the battery is fully charged before using.
VIVIX-S 1012N User Manual 4.1.2 How to take out a battery pack 1 Open the battery cover. 2 Push the lever to the left for taking out the battery. 3 Close the battery cover. Make sure to turn off the detector by pressing and hold the power button for 3 seconds before detaching a battery pack. All status LED lamps are off and then the detector is turned off. Rev.1.
VIVIX-S 1012N User Manual 4.1.3 How to charge a battery pack The detector is battery-operated when it is connected wirelessly. Charge the battery pack fully before inserting it to the detector. Charging type 1 Supply power to the charger with the power adapter before charging a battery. When the power LED is turned to green, it means that the power is permitted. Rev.1.0 1 Open the cover at the lower side of charger. 2 Put the battery on the top of charger in the right direction.
VIVIX-S 1012N User Manual Charging type 2 Supply power to the charger with the power adapter before charging a battery. When the power LED is turned to green, it means that the power is permitted. 1 Open the cover at the lower side of charger. 2 Insert the battery into the slot until it clicks. 3 Close the cover at the lower side of charger. The charging time can be different depending on temperature and charging status of a battery.
VIVIX-S 1012N User Manual 4.2 Product Installation 4.2.1 Connecting Devices Installation of this equipment should be made by licensed and authorized personnel by Vieworks. This equipment must only be connected to the power with protective earth. SCU Basic (FXRS-02A) 1 If you use the DR Trigger mode, connect the one end of generator interface cable to the EXT_INF port of SCU, and the other to the X-ray generator.
VIVIX-S 1012N User Manual SCU Basic (FXRS-03A) 1 If you use the DR Trigger mode, connect the one end of generator interface cable to the EXT_INF port of SCU, and the other to the X-ray generator. If you use the AED mode, a generator interface cable is not needed as the detector operates by detecting X-ray automatically. 2 Connect one end of the LAN cable to one of the LAN ports of SCU, and the other to the LAN Card Connector of workstation assigned for data transfer.
VIVIX-S 1012N User Manual VIVIX SCU mini (FXRS-04A) 1 If you use the DR Trigger mode, connect the one end of generator interface cable to the EXT_INF port of SCU mini, and the other to the X-ray generator. If you use the AED mode, a generator interface cable is not needed as the detector operates by detecting X-ray automatically. 1 Connect one end of the LAN cable to one of the LAN ports of SCU mini, and the other to the LAN Card Connector of workstation assigned for data transfer.
VIVIX-S 1012N User Manual 4.2.2 Booting up SCU 1 Turn on the power switch at the front side of SCU. 2 Check if the power switch of SCU lights up green. 3 Check if the status indicator is blinking in green after the power switch of SCU is permitted. 4 Check if the status indicator of SCU is blinking in green and turns to light up in the same color. When the power switch lights up green, it means that the power is applied normally.
VIVIX-S 1012N User Manual 4.2.4 Checking Status LED of Detector Power LED The power LED indicates the power status information which is permitted to the detector in green. The power LED lights up when the power is permitted normally. If the detector is connected with a tether interface, the power LED lights up when power is permitted to SCU because the detector is supplied power from SCU.
VIVIX-S 1012N User Manual Summary list of the Detector AP LED Information Detector AP LED Detector AP OFF (Communication status: Good) Wireelss communication: 3rd ON level or higher Wired communication: 1000Mbps Detector AP OFF (Communication status: Normal) ON Wireelss communication: Under 2nd level Wired communication: 100Mbps Detector AP ON ON Switching the status of detector AP Blink The wireless setting is being synchronized Blink Power OFF OFF If the LED blinks abnormally, r
VIVIX-S 1012N User Manual 4.3 Software Installation This section gives information about how to install the software on the workstation (PC) and how to configure the environment for software operation and communication. Check suitability of acquiring, processing and adjusting of images by referring to the recommended workstation specifications before the software installation. 4.3.1 Software Classification Vieworks provides clients who purchase our detector system with software as below.
VIVIX-S 1012N User Manual 4.4 Windows Environment Setting This section gives information about configuring Windows to communicate with the detector and SCU after installing the Setup program or Viewer. The contents in this chapter are made on the basis of Windows 7. Configuration environment can be different depending on network adaptor manufacturer or models. 4.4.
VIVIX-S 1012N User Manual Setting Network Adaptor 1 Click Configure button to open the following dialog box, and then go to the Advanced tab. 2 Set Jumbo Packet to the maximum value. (Recommended value: 9014 Bytes) 3 Choose Performance Options in the list of Settings and click Properties button on the right. 4 Choose Flow Control in the list of Settings and select Rx & Tx Enabled on the Value list as shown below. Rev.1.
VIVIX-S 1012N User Manual 5 Choose Interrupt Moderation Rate in the list of Settings and select Extreme on the Value list as shown below. 6 Choose Receive Buffers and set it to the maximum value. 7 Click OK button. Deactivate Power-Saving Mode on Network Adaptor 1 Click the Power Management tab and uncheck Allow the computer to turn off this device to save power. 2 Click OK button. Viewer may not work normally if the power save mode is activated.
VIVIX-S 1012N User Manual Protocol Selection and IP Address Setting 1 Choose Internet Protocol Version 4 (TCP/IPv4).and click Properties button. 2 Input the IP address and subnet mask as shown below, and then click OK button. ViVIX Device Driver is the image filter driver for acquiring images from a detector. It is recommended to uncheck the other items on the list except for ViVIX Device Driver and Internet Protocol Version 4(TCP/IPv4), since they are not related with the detector communication.
VIVIX-S 1012N User Manual 4.4.2 Disabling Sleep Mode of Monitor If you use the sleep mode, viewer may not work normally. 1 Click Start Control Panel Power Options and then move to the Choose when to turn off the display tab. Rev.1.0 2 Set Put the computer to sleep to Never to disable the sleep mode. 3 Click Save changes button.
VIVIX-S 1012N User Manual 5. Setting This section gives information about the product setting with using the Setup program. Start Setting SCU Setting Detector Setting Changing the Wireless Setting Rev.1.
VIVIX-S 1012N User Manual 5.1 Start Setting 5.1.1 Setup Program The VIVIX Setup program provides functions for setting and managing the internal data to make VIVIX-S 1012N wireless detector and SCU work normally. In addition, The VIVIX Setup program has diagnostic function for checking performance and abnormality of the devices as well as the image calibration function for improving the image quality. 5.1.2 Checking Devices Check information of the detector and SCU displayed on the Discovery list.
VIVIX-S 1012N User Manual Common Model No. Model name of SCU or detector Serial No. Serial number of SCU or detector IP Address IP Address of SCU or detector MAC Address MAC Address of SCU or detector The default IP address of SCU is 169.254.2.100. The default IP address of detector is 169.254.1.10. You can change the IP Address of detector or SCU. Refer to Change Configuration on the next page. MAC address is a unique indentifier of the network device. Do not change it randomly.
VIVIX-S 1012N User Manual 5.1.3 Getting into the Devices Click Next button to enter the registered SCU and detector. You can use SCU and the detector normally when the Status of SCU is Connected and the Status of detector is Initialization OK. Button Button 설명 SCU Configuration Checks and sets information of SCU setting. Diagnosis Checks information of wireless AP on SCU and performs the self-diagnosis. Detector Configuration Checks and changes the information of detector settings.
VIVIX-S 1012N User Manual 5.2 SCU Setting 5.2.1 SCU Configuration System Checks and sets system information of SCU. Item Description Model No SCU model name Serial No SCU serial numbers Package Version information of SCU firmware package Network Checks and sets network information of SCU. Rev.1.0 Item Description IP Address IP address of SCU. Net Mask Subnet Mask of SCU. Gateway Gateway of SCU.
VIVIX-S 1012N User Manual AP Checks and sets AP (Access Point) information of SCU. Item Description AP On / Off Selects whether to use SCU AP (Access Point) mode or not. On Turns on SCU AP Off Turns off SCU AP Frequency Frequency channel of wireless network. 2.4 ㎓ 5 ㎓ Uses 2.4 ㎓ Frequency. (13 channels) Uses 5 ㎓ Frequency. (8 channels) Country Country code of using wireless network. (KR, US, EU, JP, CN) Band Wireless network bandwidth. 20 ㎒ Default frequency band.
VIVIX-S 1012N User Manual Trigger Configures trigger information for integration between SCU and X-ray generator. Item Description Method Trigger mode Packet Software Trigger mode Line Hardware Trigger mode Polarity Polarity of trigger signal Auto (Default) Recognizes polarity automatically and handles it. (Default value) Active High Handles polarity of Active High Active Low Handles polarity of Active Low Configuration value of trigger is applied only when Exposure mode is set DR Trigger.
VIVIX-S 1012N User Manual 5.2.2 SCU Diagnosis Self Diagnosis You can perform self diagnosis to check operational status of the parts in SCU. Refer to <7.2 Product Inspection> for the corrective measures on the result of each item. Rev.1.
VIVIX-S 1012N User Manual 5.3 Detector Setting 5.3.1 Detector Configuration System Checks and sets system information of the detector. Item Description Model No Detector model name Serial No Detector serial numbers Package Version information of the detector firmware package Network Checks and sets network information of the detector. Rev.1.
VIVIX-S 1012N User Manual WNetwork Sets AP (Access Point) information of wireless communication from the detector. Item Description SSID Configures wireless network ID AP from the detector. Key Configures wireless network key value of AP from the detector. Wireless Only Configures the wireless communication method of the detector. The detector is operated in a wired way when a tether interface cable is On connected.
VIVIX-S 1012N User Manual Set SSID and Key values not to be duplicated with those of the peripheral system. SSID and Key values should be input as capital / small alphabet letters, numeric characters and only “-“, “_” among special letters up to 20. Wireless network setting should be done by an engineer who understands the wireless communication and its related technique. Unless the network is set properly, a communication error would occur or the image quality would be affected.
VIVIX-S 1012N User Manual If Power Off is set to by SCU, the battery cannot supply power to the detector. In this case, you should connect the tether interface cable to supply power of SCU to the detector. Wireless network setting should be done by an engineer who understands the wireless communication and its related technique. Unless the network is set properly, a communication error would occur or the image quality would be affected. 5.3.
VIVIX-S 1012N User Manual Setting Items of Power Save Function Mode Meaning Normal - Sleep Selects the function On (use) or Off (not use). Sleep after (min.) Sleep mode is activated if the detector is used during the setting time. Shut Down Selects the function On (use) or Off (not use). Shutdown after (min.) The detector is turned off if it is not used during the setting time. This menu is activated while the Shut Down function is used. (30 / 60 / 90 / 120 min.
VIVIX-S 1012N User Manual 5.4 Changing the Wireless Setting 5.4.1 Switching to the Detector AP Mode You can change the detector mode as AP by following the two ways. Choose the AP option as On or Off from the Detector Configuration dialog in VIVIX Setup program. Press the AP button on the detector for 5 seconds to set the Detector AP mode. The LED of detector AP blinks in a blue color while the mode is being switched, and turns on blue after the mode is compeleted to be set. 5.4.
VIVIX-S 1012N User Manual 6. Calibration This chapter gives information about the calibration methods after installing a detector. Calibration Dialogue Detector Configuration Calibration Guide Calibrating by Loading the Calibration Data Direct Calibration Rev.1.
VIVIX-S 1012N User Manual 6.1 Calibration Dialogue Use the following menus to configure the system and process calibrations. Menu Description System Configuration Configures the exposure mode, exposure-related time and Gain type. Offset Calibration Processes Offset calibration. Defect Calibration Processes calibration for defect data of the detector. Gain Calibration Processes calibration for image sensitivity.
VIVIX-S 1012N User Manual Exposure Timing To take images exactly, set the timing information for exposure. The information can be different depening on the characteristics of the X-ray generator. Exposure Timing Diagram Exposure Timing Items Item Default Description Exposure section 500㎳ Time to acquire X-ray by the detector. Pre Exposure section 0㎳ Standby time after the detector sent EXP_OK signal to the generator.
VIVIX-S 1012N User Manual When Exposure section is changed, make sure to create new calibration data by processing Offset calibration and Gain calibration to get optimized images. Pre Exposure section can be used when the time delays until the generator receives EXPOK signal from the detector and generates X-ray. Although Pre Exposure section is set as 0㎳ normally, it is recommended to measure and set the actual delay time of X-ray generation to achieve the best performance of the detector.
VIVIX-S 1012N User Manual 6.1.2 Offset Calibration Dialogue Vieworks provides the Defect Map data stored in the detector. However, we recommend you to carry out the Offset calibration by yourself as the detector condition can be different by the operating method or use environment. Offset Calibration Item Description Current value Shows value when performing the manual Offset calibration. Stage Rev.1.
VIVIX-S 1012N User Manual 6.1.3 Defect Calibration Dialogue Vieworks provides the Defect Map data stored in the detector. However, you should download the data as the calibration cannot be performed in the detector automatically. This Defect Map data has been generated through stringent test in the production stages of detector. However, new defect may be newly formed while using the detector. In this case, make sure to calibrate defects in person and generate a new defect map. Defect Map Rev.1.
VIVIX-S 1012N User Manual 6.1.4 Gain Calibration Dialogue Vieworks provides the Gain data stored in the detector. However, you should download the data as the calibration cannot be performed in the detector automatically. This data is provided in case the Gain calibration cannot be performed. You should do the Gain calibration in person as the detector condition can be different by the operating method or use environment. Gain Calibration Rev.1.
VIVIX-S 1012N User Manual 6.2 Detector Configuration From the Detector Configuration dialog, you can configure the detector and check the images before or after doing calibration. Rev.1.0 Function Description Detector Direction Compensation Sets the displayed direction of image. Effective Area Sets effective area of an image. Offset Refresh Setting Sets conditions of automatic Offset Refresh. Time Interval (min.
VIVIX-S 1012N User Manual You can only change the displayed direction of an image from Detector Direction Compensation. The direction of an original image cannot be changed. Effective Area can be changed within the effective range which has been configured at the time of the first shipment of detector. Once Use offset refresh is used, the offset refresh is processed automatically as follows.
VIVIX-S 1012N User Manual 6.3 Calibration Guide The different installation environment of each detector and unique features of the X-ray generator device can affect the acquired images. Therefore, the certified engineer from Vieworks should do the detector calibration after installing it. Otherwise, the image quality can be affected seriously. Vieworks provides two types of calibration for performing calibration. Performing calibration by loading the calibration data CD provided by Vieworks.
VIVIX-S 1012N User Manual 6.4 Calibrating by Loading the Calibration Data Vieworks provides the calibration data stored in the detector. Although the Offset calibration is performed automatically, you need to download both Defect and Gain calibration data since their calibration cannot be performed in the detector automatically. 6.4.1 Preparing Calibration Data Among the three types of calibration data stored in the detector, download the Defact Map and Gain calibration data to local HDD. 6.4.
VIVIX-S 1012N User Manual Be sure to select the Defect Map data file provided with a detector. If the file has a wrong serial number or you select a wrong file, an error message will be displayed. Vieworks provides the Defect Map data stored in the detector. However, you should download the data as the calibration cannot be performed in the detector automatically. This Defect Map data has been generated through stringent test in the production stages of detector.
VIVIX-S 1012N User Manual 6.5 Direct Calibration 6.5.1 Prepration Precheck the state of X-ray generator and console. Precheck the state of X-ray tube. It is recommended you to check if X-ray dose value of the generator is exact by using the device like a dose meter. Make sure to preheat the detector for 30 minutes before starting calibration. The result of measurement can be incorrect if the detector is not preheated sufficiently. 6.5.2 Rev.1.
VIVIX-S 1012N User Manual 6.5.3 Gain Calibration 1 Put a collimator on center of the detector, and open the collimator completely. 2 Adjust SID as 130㎝ ~ 150㎝ to make X-ray exposure range include the detector. 3 Remove any objects or foreign materials between the tube and the detector. 4 Choose Gain Calibration -Normal. 5 Set the tube voltage of X-ray generator as 70㎸ ~ 80㎸.
VIVIX-S 1012N User Manual 6.5.4 Auto Defect Correction While operating the detector, defect pixels may appear on the image. In this case, use the defect auto correction to calibrate the defect pixels. Be sure to check the followings before acquiring the FLAT image. Preheat the detector for 30 minutes or more. The recommend SID is 150 ㎝ (distance between X-ray tube and detector) Open the collimator of X-ray tube completely. Align the center of the detector with the center of collimator.
VIVIX-S 1012N User Manual Rev.1.0 3 Close the Image dialog box, and open the Calibration dialog box. 4 Choose Defect Calibration as follows and click Manual Defect Correction button. 5 Move to the folder where the raw file is saved and select the file. 6 The Manual Defect Detection screen will be displayed. 7 Click Show and Apply buttons on the top left and then click Auto Defect Finder button.
VIVIX-S 1012N User Manual 8 Input 15% for Threshold, and 100 for Offset at the bottom of the Pixel List screen. 9 Click Search button at the bottom of the screen. The coordinates of defect pixels and Gray level values are displayed on Pixel list. 10 While pressing the Ctrl or Shift key, select the coordinate of defect pixels to be added to Defect Map. 11 Click Add button to add the selected items to Pixel List in Manual Defecct Detection.
VIVIX-S 1012N User Manual 14 Save the newly generated Defect Map data to the path where the Calibration data has been stored. It is recommended to save the newly generated Defect Map data with a different name to preserve the past Defect Map data. Rev.1.
VIVIX-S 1012N User Manual 6.5.5 Manual Defect Correction Execute Manual Defect Detection when the defect pixels are visible to the naked eye even the automatic defect correction is processed. Manual Defect Detection is divided into the pixel type and line type. Manual Defect Correction in Pixel Type 1 Make sure to carry out the automatic defect correction from 1 to 14 steps first. 2 Click Show and Apply buttons on the left top of Manual Defect Detection.
VIVIX-S 1012N User Manual Manual Defect Correction in Line Type 1 Make sure to carry out the automatic defect correction from 1 to 12 steps first. 2 Click Show and Apply buttons on the left top of Manual Defect Detection. 3 Press and hold the right mouse button and move it to left/right/top/bottom in the FLAT image to adjust the brightness until you can verify the image with the naked eyes. 4 Find and move to the start point of defect lines (line type) on the image.
VIVIX-S 1012N User Manual Functions of Manual Defect Detection Rev.1.0 Function Description Show Decides whether to indicate the selected defect on the image Apply Decides whether to apply Defect in the list to the calibrated image. All Applies to all coordinates of defects added to the list Selected Applies to relevant coordinates of the selected defects from the list only. Auto Defect Finder Performs the auto defect correction. Delete Deletes the selected items from the list.
VIVIX-S 1012N User Manual 7. Diagnosis, Inspection and Maintenance This section gives information about diagnosis, inspection & maintenance of the product. Diagnosis Product Inspection Cleaning and Disinfection Product Initialization Replacing the Fuse of SCU Rev.1.
VIVIX-S 1012N User Manual 7.1 Diagnosis 7.1.1 Image Diagnosis Check the image quality through Diagnosis tools after installing the detector or before usage. If the problems with regard to products or image occur during diagnosis, try to do a calibration again. If the problems are not solved, consult the sales representative in Vieworks or a service engineer. You can acquire and review an image from the Diagnosis window in VIVIX Setup program.
VIVIX-S 1012N User Manual Dignostic Functions Item Description Pan Press and drag a mouse button to move the image to the desired position. Press and hold a mouse button to move the image upwards or downwards for Zoom expanding / reducing the image. Press and hold a mouse button to move the image to up/down/right/left for W/L adjusting its window level. This function can be used without clicking W/L button, but with the right mouse button basically.
VIVIX-S 1012N User Manual 7.1.2 Battery Pack Diagnosis Install the battery pack to the detector and check the voltage and remaining amount of the battery pack. Furthermore, always check the remaining amount of the battery pack during use of the detector. If performance of the battery pack has some problems, consult the sales representative in Vieworks or a relevant engineer. The battery pack belongs to consumables which performance will be decreased as time passed.
VIVIX-S 1012N User Manual If the remaining of battery is under 30% or at the 1st level, the system warns low battery and the detector will be turned off automatically if the battery is being consumed for a specific period of time. Therfore, it is recommended to change the battery when a warning message or indicator is displayed. 7.1.
VIVIX-S 1012N User Manual Item Description Connection Type Indicates the connection mode. (Wireless / Tether) Link Speed Connection status in the wired mode. (1000Mbps / 100Mbps) Link Quality Link Quality (Status of the connection quality) value Signal-level Singal strength Bit rate Transmission / reception speed Frequency Frequency of the connected AP (Access Point) The communication is not running smoothly when the strength of wireless communication is under the 2nd stage.
VIVIX-S 1012N User Manual 7.2 Product Inspection To use products safely, make sure to check the products before use. If problems occur during inspection or the product is impossible to repair, consult the sales representative in Vieworks or a relevant engineer. 7.2.1 Daily Inspection Before or after using the detector and other surrounding devices, check below items daily. Item Description Ensure that there are no loose screws or breaks.
VIVIX-S 1012N User Manual Self Diagnosis You can perform self diagnosis of the detector or SCU from the Diagnosis window in the VIVIX Setup program. 1 Choose a desired item to diagnose. 2 Click Execute Test button located at the bottom of the Test Case window. 3 Check the progress and result of diagnosis for each item in the Result window. 4 Click Save to File to save the diagnosis result as a file when the self diagnosis is completed.
VIVIX-S 1012N User Manual Wireless Sensor Voltage Information N/A N/A Remain Information N/A N/A Detection Decision Defective wireless module Contact a service eingineer. Inconsistent environment of the Check obstacles and distance Connection Decision wireless communication between a detector and SCU. Defective wireless module Contact a service eingineer. Impact Sensor Decision Defective shock sensor Contact a service eingineer.
VIVIX-S 1012N User Manual Self-Diagnosis Items of SCU Category Item Form Wireless Detection Decision Defective wireless module Contact a service eingineer. Detection Decision Not available to save logs. Contact a service eingineer. File System Decision Not available to save logs. Contact a service eingineer. Status Decision N/A N/A Memory Switching IC IC Expected problem Decision Not available to connect the detector and PC. Action Contact a service eingineer.
VIVIX-S 1012N User Manual 7.3 Cleaning and Disinfection After using the detector and peripheral equipments for examination, use germicidal disinfecting wipes or cloth with mild diluted disinfectant detergent to clean surfaces of the product.
VIVIX-S 1012N User Manual 7.4 Product Initialization If the connection status of system is not stable or setting value is not correct, user can initialize the products. 7.4.1 SCU Initialization 1 Click Configuration button of SCU after running the VIVIX Setup program. 2 Click Factory Reset button in Configuration dialogue. 3 Wait for SCU to be initialized and rebooted automatically. 4 Check whether SCU initialization is completed.
VIVIX-S 1012N User Manual 7.4.2 Detector Initialization 1 Click Configuration button of Detector after running the VIVIX Setup program. 2 Click Factory Reset button in Configuration dialogue. 3 Wait for Detector to be rebooted automatically. 4 Check whether detector initialization is completed. Default value of detector initialization Item Default Value Network IP Address 169.254.1.10 Subnet Mask 255.255.0.0 Gateway 169.254.2.
VIVIX-S 1012N User Manual 7.4.3 Wireless Initialization of Detector 1 Turn off the detector. 2 Press and hold both the power button and AP button over 3 seconds. 3 Initialization will be conducted when orange LED is flickering, after that, the detector will be turned off automatically. 4 Turn on the detector and check whether detector initialization is completed. Default value of wireless initialization Item Default Value Network IP Address 169.254.1.10 Subnet Mask 255.255.0.0 Gateway 169.
VIVIX-S 1012N User Manual 7.5 Replacing the Fuse of SCU (SCU Basic only) There are 2 fuses attached on the standard SCU for the purpose of electrical accident precaution, in case of over current from external power input. Stop using the SCU immediately when the fuse is blown. Fuse Information Item Specifications Model Littelfuse® 218002 (2EA) Type Time Lag Cartridge Fuse Amp Rating 2A Voltage Rating 250V Pull the plug out and turn all the devices off before changing the fuse.
VIVIX-S 1012N User Manual 8. Troubleshooting This section gives information about troubleshooting. Rev.1.
VIVIX-S 1012N User Manual 8.1 Trobleshooting When you encounter problems while using the equipment, search for the table below for the problem or error messages and try the solutions.If the problem persists, turn off the detector and consult your sales representative or a distributor. Please refer to the details of the following symptoms or error messages. Troubleshooting must be performed by service engineer who is authorized by Vieworks.
VIVIX-S 1012N User Manual 8.1.3 2 Check the connection between DC power cable and SCU mini. 3 Turn off the power switch and turn on again and then check the fan or status of back side. 4 Replace another SCU and check the result of it 5 Replace the fuse of standard SCU. (refer to Change fuse of SCU) 6 Replace corresponding devices. Communication Test is failed Symptom Transmission error is occurred, failure of communication test is occurred.
VIVIX-S 1012N User Manual 7 8.1.5 Replace corresponding devices. Errors in Detector LED Symptom All LED lamps of detector are blinking. 2 LED lamps of detector are blinking and remaining is blinking slowly. Expected Causes Internal hardware errors of the detector. Solutions 8.1.6 1 Boot up the detector again and check the result of it 2 Replace the detector. Rapid Consumption of Battery Symptom Consumption of fully charged battery pack is fast.
VIVIX-S 1012N User Manual 9. Regulatory Information This section gives explanation about the regulatory information and standard related to the products Medical Equipment Safety Standards Radio Frequency Compliance Information Labels and Symbols Guidance and Manufacturer Declaration for EMC Rev.1.
VIVIX-S 1012N User Manual 9.1 Medical Equipment Safety Standards 9.1.1 Medical Equipment Classification Item Description Type of protection against electrical shock Class Ⅰ equipment Internally powered Degree of protection against electrical shock Type B applied parts Degree of protection against ingress of water IPX0 Operation mode Continuous operation NOT suitable for use in the presence of a flammable Flammable anesthetics anesthetic mixture with air or with oxygen or nitrous oxide. 9.1.
VIVIX-S 1012N User Manual U.S.A / Canada Item IEC 60601-1(ed.2 am1+ am2+ co1) Medical electrical equipment- Part1: General requirements for safety UL 60601-1(ed.2) CSA-C22.2 No. 601-1-M90 (R2006) IEC 60601-1-2: 2007 (ed.3) IEC 60601-1-4: 2000 (ed.1.
VIVIX-S 1012N User Manual 9.2 Radio Frequency Compliance Information Country Item FCC Part 15.107 Subpart (b) / 15.109(g) Subpart B U.S.A FCC Part 15 Subpart E 15.407 FCC Part 15 Subpart C 15.247 ETSI EN 301 489-1 V1.8.1 (EMC) ETSI EN 301 489-17 V2.1.1 (EMC) European Union EN 300 328 V.1.7.1; EN 301 893 V1.6.1 (RF) EN 62311:2008 (RF Exposure) ETSI EN 300 328 V1.7.1; EN 301 893, V1.5.
VIVIX-S 1012N User Manual 9.2.2 FCC SAR 9.2.3 OET Bulletin 65, Supplement C (Edition 01-01) CE R&TTE SAR Item EN 62311:2008 Assessment of electronic and electrical equipment related to human exposure restrictions for electromagnetic fields. (0 Hz - 300 GHz) Human exposure to radio frequency fields from hand-held and bodymounted wireless communication devices – Human models, EN 62209-1:2006 instrumentation, and procedures.
VIVIX-S 1012N User Manual 9.3 Labels and Symbols The ViVIX-S 1012N detector and relevant components have labels attached on them. The contents and locations of each label are indicated below. 9.3.1 Label VIVIX-S 1012NA / 1012NB (Wired Detector) VIVIX-S 1012NAW / 1012NBW (Wireless Detector) Rev.1.
VIVIX-S 1012N User Manual Battery Battery Charger Rev.1.
VIVIX-S 1012N User Manual SCU Basic (FXRS-02A) Rev.1.
VIVIX-S 1012N User Manual SCU Basic (FXRS-03A) Rev.1.
VIVIX-S 1012N User Manual SCU mini (FXRS-04A) Rev.1.
VIVIX-S 1012N User Manual Outer Box 9.3.2 [TBD] Product Serial Number Serial Number Composition The serial numbers for each product or accessory are composed as follows. V 1 D Item A Composition B J Year 0 0 Month 1 Serial number Revision will be updated in case of follows. Mass production or a large amount of order. Exterior alteration. Item code will be produced based on internal management standard of vieworks. Composition code is like follows.
VIVIX-S 1012N User Manual 9.3.3 Product Symbols Symbol Description Direct current Alternating current Protective earth (Ground) Equipotentiality Power on Power on for part of the equipment Power off Power off for part of the equipment Attention, consult accompanying documents General warning sign Warning sign for electricity This Mark shows compliance with both Canadian and U.S. safety requirements. With Respect to electric shock, fire, and mechanical hazards only.
VIVIX-S 1012N User Manual 9.4 Guidance and Manufacturer’s Declaration for EMC This device has been tested for EMI/EMC compliance, but interference can still occur in an electromagnetically noisy location. Attempt to maintain a suitable distance between electrical devices to prevent malfunction. 9.4.1 Electromagnetic Emissions The Equipment Under Test (EUT) is intended for use in the electromagnetic environment specified below.
VIVIX-S 1012N User Manual Electrical Fast Transient/Burst IEC 61000-4-4 Item Description Immunity test Electrical fast transient/burst IEC 61000-4-4 IEC 60601 test condition Compliance Level Power supply lines ±2㎸ Input / output lines ±1㎸ Power supply lines ±2㎸ Input / output lines ±1㎸ Electromagnetic Environment - Main power quality should be that of a typical commercial or hospital Guidance environment.
VIVIX-S 1012N User Manual Power Frequency (50/60 ㎐) Magnetic Field IEC 61000-4-8 Item Description Immunity test Power frequency (50/60 ㎐) magnetic field IEC 61000-4-8 IEC 60601 test condition 3 A/m Compliance Level 3 A/m Electromagnetic Environment - Power frequency magnetic fields should be at levels characteristic of a Guidance typical location in a typical commercial or hospital environment.
VIVIX-S 1012N User Manual 10. Information This section gives overview information for service and warranty of the product. Service Information Warranty Revision History Rev.1.
VIVIX-S 1012N User Manual 10.1 Service Information 10.1.1 Product Lifetime The estimated product lifetime may be up to five (5) years under the appropriate regular inspection and maintenance. 10.1.2 Regular Inspection and Maintenance In order to ensure the safety of patients, operating personnel and third parties, and to maintain the performance and reliability of the equipment, be sure to perform regular inspection at least once a year.
VIVIX-S 1012N User Manual 10.2 Warranty Vieworks warrants that this product will be free from defects in materials and workmanship for a period of 24 months from the date of delivery. If any such product proves defective during this warranty period, Vieworks at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
VIVIX-S 1012N User Manual 10.3 Revision History Version Date Descriptions 1.0 2015-02-?? Initial Release Vieworks Co., Ltd. (Gwanyang-dong) 41-3, Burim-ro 170beon-gil, Dongan-gu, Anyang-si, Gyeonggi-do, 431-060 Republic of Korea Telephone: +82-70-7011-6161 Fax: +82-31-386-8631 Homepage: http://www.vieworks.com European representative: DONGBANG ACUPRIME 1 Forrest Units, Hennock Road East, Marsh Barton, Exeter EX2 8RU, UK Telephone: +44(0)-1392-829500 Homepage: http://www.acuprime.com Rev.1.