Operating Instructions Radar sensor for continuous level measurement of liquids VEGAPULS 64 4 … 20 mA/HART - two-wire Document ID: 51141
Contents Contents About this document 1.1 Function............................................................................................................................ 4 1.2 Target group...................................................................................................................... 4 1.3 Symbols used................................................................................................................... 4 2 For your safety 2.1 Authorised personnel...................
Contents 8 Set up with other systems 8.1 DD adjustment programs................................................................................................ 57 8.2 Field Communicator 375, 475......................................................................................... 57 9 Diagnosis, asset management and service 9.1 Maintenance................................................................................................................... 58 9.2 Measured value and event memory..........
1 About this document 1 About this document 1.1 Function 1.2 Target group 1.3 Symbols used This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectification. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device. This operating instructions manual is directed to trained specialist personnel.
2 For your safety 2 2.1 For your safety Authorised personnel All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator. During work on and with the device the required personal protective equipment must always be worn. 2.2 Appropriate use VEGAPULS 64 is a sensor for continuous level measurement. You can find detailed information about the area of application in chapter "Product description".
2 For your safety 2.5 CE conformity The device fulfils the legal requirements of the applicable EC guidelines. By affixing the CE marking, we confirm successful testing of the product. You can find the CE Certificate of Conformity in the download section of our homepage. 2.6 NAMUR recommendations NAMUR is the automation technology user association in the process industry in Germany. The published NAMUR recommendations are accepted as the standard in field instrumentation.
2 For your safety • • • • • Flanges, process fittings and mounting accessories must ensure the microwave impermeability of the vessel and not let the radar signal escape to the outside If necessary, existing viewing windows in the vessel must be coated with a microwave impermeable material (e.g.
2 For your safety This device has been approved for both closed containers and openair environments with the following limitations: • • • • • • • • • • Closed Containers: For installations utilizing a tilt during installation: This device is limited to installation in a completely enclosed container made of metal, reinforced fiberglass or concrete to prevent RF emissions, which can otherwise interfere with aeronautical navigation, the maximum approved tilt angel is 10°.
2 For your safety • • • • • • • • • Cuves fermées : Pour les installations impliquant une inclinaison lors de l'installation : cet appareil ne doit être installé que dans une cuve totalement fermée en métal ou en béton, pour empêcher les émissions RF susceptibles d'interférer avec la navigation aéronautique. L'angle d'inclinaison maximum autorisé est de 10°.
3 Product description 3 Type label Product description 3.1 Configuration The type label contains the most important data for identification and use of the instrument: 1 2 3 12 4 5 6 7 11 10 9 8 Fig.
3 Product description • • Scope of delivery Hardware version from 1.0.0 Software version from 1.1.
3 Product description The emitted signal is reflected by the medium and received by the antenna as an echo. The frequency of the received signal always deviates from the actual emitting frequency. The frequency difference is proportional to the distance and thus to the filling height. This difference is calculated via special algorithms in the sensor electronics. The determined filling height is then converted into a corresponding output signal and outputted as the measured value. Packaging 3.
3 Product description parameter adjustment of these instruments, the adjustment software PACTware with VEGA-DTM is required. You can find further information in the operating instructions "Interface adapter VEGACONNECT" (Document-ID 32628). VEGADIS 81 The VEGADIS 81 is an external display and adjustment unit for VEGA plics® sensors. For sensors with double chamber housing the interface adapter "DISADAPT" is also required for VEGADIS 81.
3 Product description You can find additional information in chapter "Mounting" of this operating instruction. Electronics module Electronics module "VEGAPULS series 60" is a replacement part for radar sensors of VEGAPULS series 60. A different version is available for each type of signal output. You can find further information in the operating instructions "Electronics module VEGAPULS series 60" (Document-ID 36801).
4 Mounting 4 Protection against moisture Mounting 4.1 General instructions Protect your instrument against moisture ingress through the following measures: • • • Use the recommended cable (see chapter "Connecting to power supply") Tighten the cable gland Loop the connection cable downward in front of the cable gland This applies particularly to: • • • Outdoor mounting Installations in areas where high humidity is expected (e.g.
4 Mounting is a simple and effective way to align the sensor to the surface of the bulk solid material. The following versions are available: • • Mounting strap - Ceiling mounting Length 300 mm Length 170 mm The instrument is normally mounted vertically with a bracket on the ceiling. This allows swivelling the sensor up to 180° for optimal orientation and rotating for optimal connection. Fig. 3: Ceiling mounting via the mounting strap with length 300 mm Fig.
4 Mounting > 200 mm (7.87") Fig. 5: Wall mounting horizontally via the mounting strap with length 170 mm Fig.
4 Mounting 1 Fig. 7: Combi compression flange 1 Combi compression flange The adapter flange is available in different flange sizes. It is permanently connected with the radar sensor and sealed. 1 2 3 Fig. 8: Adapter flange 1 2 3 Connection screw Adapter flange Process seal You can find drawings of these mounting options in chapter "Dimensions". 4.
4 Mounting 1 2 Fig. 9: Mounting strap for screwing to the sensor 1 2 For angle of inclination in steps For angle of inclination, infinitely variable Depending on the selected variant, the sensor can be rotated in the strap: • • Polarisation Single chamber housing –– Angle of inclination in three steps 0°, 90° and 180° –– Angle of inclination 180°, infinitely variable Double chamber housing –– Angle of inclination 90°, infinitely variable –– Angle of inclination in two steps 0° and 90° 4.
4 Mounting these can be suppressed by an appropriate adjustment (see chapter "Setup"). If you cannot maintain this distance, you should carry out a false signal storage during setup. This applies particularly if buildup on the vessel wall is expected. In such cases, we recommend repeating the false signal storage at a later date with existing buildup. > 200 mm (7.87") Fig.
4 Mounting Fig. 13: Mounting of the radar sensor with inflowing medium The socket piece should be dimensioned in such a way that the antenna end protrudes at least 5 mm (0.2 in) out of the socket. ca. 5 mm Socket with threaded fitting Fig. 14: Recommended socket mounting h If the reflective properties of the medium are good, you can mount VEGAPULS 64 on sockets which are higher than the length of the antenna. You will find recommended values for socket heights in the following illustration.
4 Mounting Socket with flange connection Socket diameter d Socket length h 50 mm ≤ 250 mm 80 mm ≤ 300 mm 100 mm ≤ 400 mm 150 mm ≤ 500 mm Socket diameter d Socket length h 1½" ≤ 7.9 in 2" ≤ 9.9 in 3" ≤ 11.8 in 4" ≤ 15.8 in 6" ≤ 19.7 in The mounting socket should be as short as possible and its end rounded. This reduces false echoes from the vessel mounting socket. Fig.
4 Mounting Sensor orientation Socket diameter d Socket length h 150 mm ≤ 500 mm Socket diameter d Socket length h 3" ≤ 11.8 in 4" ≤ 15.8 in 6" ≤ 19.7 in In liquids, direct the sensor as perpendicular as possible to the product surface to achieve optimum measurement results. Fig. 18: Alignment in liquids Vessel installations The mounting location of the radar sensor should be a place where no other equipment or fixtures cross the path of the radar signals. Vessel installations, such as e.g.
4 Mounting Fig. 20: Agitators Foam generation Through the action of filling, stirring and other processes in the vessel, compact foam can form on the product surface, damping the emitted signals considerably. If foam is causing measurement errors, the largest possible radar antenna should be used. As an alternative, sensors with guided microwave can be used. These are unaffected by foam generation and are best suited for such applications. Flow measurement with rectangular overfall 4.
4 Mounting • • • • • • Install the sensor on the headwater side Installation in the centre of the flume and vertical to the liquid surface Distance to the overfall orifice Distance of orifice opening above ground Min. distance of the orifice opening to tailwater Min. distance of the sensor to max. storage level Flow measurement with Khafagi Venturi flume 3 ... 4 x hmax 90° hmax 2 1 B Fig. 22: Flow measurement with Khafagi-Venturi flume: hmax. = max.
5 Connecting to power supply 5 Safety instructions Connecting to power supply 5.1 Preparing the connection Always keep in mind the following safety instructions: Warning: Connect only in the complete absence of line voltage. • • Voltage supply The electrical connection must only be carried out by trained personnel authorised by the plant operator. If overvoltage surges are expected, overvoltage arresters should be installed. Power supply and current signal are carried on the same two-wire cable.
5 Connecting to power supply With plastic housing, the NPT cable gland or the Conduit steel tube must be screwed without grease into the threaded insert. Max. torque for all housings, see chapter "Technical data". Cable screening and grounding If screened cable is required, the cable screen must be connected on both ends to ground potential. In the sensor, the screen is connected directly to the internal ground terminal.
5 Connecting to power supply Fig. 23: Connection steps 5 and 6 - Single chamber housing Fig. 24: Connection steps 5 and 6 - Double chamber housing 6. Insert the wire ends into the terminals according to the wiring plan Information: Solid cores as well as flexible cores with wire end sleeves are inserted directly into the terminal openings. In case of flexible cores without end sleeves, press the terminal from above with a small screwdriver, the terminal opening is then free.
5 Connecting to power supply 9. Tighten the compression nut of the cable entry gland. The seal ring must completely encircle the cable 10. Reinsert the display and adjustment module, if one was installed 11. Screw the housing lid back on The electrical connection is finished. 5.3 Wiring plan, single chamber housing The following illustration applies to the non-Ex as well as to the Ex-ia version. Electronics and terminal compartment 2 3 4...20mA (+)1 2(-) 5 6 7 8 4 1 Fig.
5 Connecting to power supply Terminal compartment 2 4...20mA (+)1 3 Display 2(-) 5 6 7 8 4 1 Fig. 27: Terminal compartment, double chamber housing 1 2 3 4 Voltage supply, signal output For display and adjustment module or interface adapter For external display and adjustment unit Ground terminal for connection of the cable screen Information: Parallel use of an external display and adjustment unit and a display and adjustment module in the terminal compartment is not supported.
5 Connecting to power supply Terminal compartment 2 3 4...20mA (+)1 2(-) 5 6 7 8 4 1 Fig. 29: Terminal compartment, double chamber housing 1 2 3 4 Voltage supply, signal output For display and adjustment module or interface adapter For external display and adjustment unit Ground terminal for connection of the cable screen Information: Parallel use of an external display and adjustment unit and a display and adjustment module in the terminal compartment is not supported.
5 Connecting to power supply Assignment of the plug connector 4 3 1 2 Fig. 31: View to the plug connector M12 x 1 1 2 3 4 Wire assignment, connection cable Pin 1 Pin 2 Pin 3 Pin 4 Contact pin Colour connection cable in the sensor Terminal, electronics module Pin 1 Brown 5 Pin 2 White 6 Pin 3 Blue 7 Pin 4 Black 8 5.7 Wiring plan - version IP 66/IP 68, 1 bar 1 2 Fig.
6 Set up with the display and adjustment module 6 6.1 Set up with the display and adjustment module Insert display and adjustment module The display and adjustment module can be inserted into the sensor and removed again at any time. You can choose any one of four different positions - each displaced by 90°. It is not necessary to interrupt the power supply. Proceed as follows: 1. Unscrew the housing lid 2.
6 Set up with the display and adjustment module 1 2 Fig. 34: Installing the display and adjustment module in the double chamber housing 1 2 In the electronics compartment In the terminal compartment Note: If you intend to retrofit the instrument with a display and adjustment module for continuous measured value indication, a higher lid with an inspection glass is required. 6.
6 Set up with the display and adjustment module • • • –– –– –– –– Move to the menu overview Confirm selected menu Edit parameter Save value [->] key: –– Change measured value presentation –– Select list entry –– Select menu items in the quick setup menu –– Select editing position [+] key: –– Change value of the parameter [ESC] key: –– Interrupt input –– Jump to next higher menu Adjustment system The instrument is operated via the four keys of the display and adjustment module.
6 Set up with the display and adjustment module Selection of national language This menu item is used to select the national language for further parameter adjustment. You can change the selection via the menu item "Setup - Display, Menu language". With the "OK" key you move to the main menu. 6.4 Parameter adjustment - Quick setup To quickly and easily adapt the sensor to the application, select the menu item "Quick setup" in the start graphic on the display and adjustment module.
6 Set up with the display and adjustment module parameters to ensure optimum adjustment of the measurement. The procedure is described in the following. Setup - Measurement loop name Here you can assign a suitable measurement loop name. Push the "OK" key to start the editing. With the "+" key you change the sign and with the "->" key you jump to the next position. You can enter names with max. 19 characters.
6 Set up with the display and adjustment module • • Process/measurement conditions: –– Condensation –– Smooth product surface –– High requirements on measurement accuracy Properties, sensor: –– Low sensitivity to sporadic false echoes –– Stable and reliable measured values through averaging –– High accuracy –– Short reaction time of the sensor not required - Storage tank with product circulation: • Setup: large-volumed, upright cylindrical, spherical • Product speed: slow filling and emptying • Installat
6 Set up with the display and adjustment module • Properties, sensor: –– Higher measurement speed through less averaging –– Sporadic false echoes are suppressed - Dosing vessel: • Setup: all vessel sizes possible • Product speed: –– Fast filling and emptying –– Vessel is filled and emptied very often • Vessel: tight installation situation • Process/measurement conditions: –– Condensation, buildup on the antenna –– Foam generation • Properties, sensor: –– Measurement speed optimized by virtually no averag
6 Set up with the display and adjustment module - Open water (gauge measurement): • Rate of level change: slow level change • Process/measurement conditions: –– Large distance from sensor to water surface –– Extreme damping of output signal due to wave generation –– Ice and condensation on the antenna possible –– Spiders and insects build nests in the antennas –– Floating material and animals sporadically appear on water surface • Properties, sensor: –– Stable and reliable measured values through frequent
6 Set up with the display and adjustment module conditions, this menu item offers different options for vessel bottom and ceiling for certain applications. Enter the requested parameters via the appropriate keys, save your settings with [OK] and jump to the next menu item with the [ESC] and the [->] key. Vessel height/Measuring range Through this selection the operating range of the sensor is adapted to the vessel height, which considerably increases measurement certainty under different basic conditions.
6 Set up with the display and adjustment module 100% 2 5m (196.9") 0,5 m (19.68") 3 0% 1 Fig. 36: Parameter adjustment example min./max. adjustment 1 2 3 Min. level = max. measuring distance Max. level = min. measuring distance Reference plane If these values are not known, an adjustment with the distances of e.g. 10 % and 90 % is possible. Starting point for these distance specifications is always the sealing surface of the thread or flange.
6 Set up with the display and adjustment module 4. For the full vessel, enter the distance value in m matching the percentage value. 5. Save settings with [OK] and move with [ESC] and [->] to Min. adjustment. Setup - Min. adjustment Proceed as follows: 1. Select with [->] the menu item "Min. adjustment" and confirm with [OK]. 2. Edit the percentage value with [OK] and set the cursor to the requested position with [->]. 3. Set the requested percentage value with [+] and save with [OK].
6 Set up with the display and adjustment module Setup - Current output Min./Max. In the menu item "Current output Min./Max.", you determine the reaction of the current output during operation. The default setting is min. current 3.8 mA and max. current 20.5 mA. Lock/unlock setup - Adjustment In the menu item "Lock/unlock adjustment", you can protect the sensor parameters against unauthorized or inadvertent modification. The PIN is activated/deactivated permanently.
6 Set up with the display and adjustment module Display - Displayed value 1 and 2 In this menu item you can define the way measured values are indicated on the display. The default setting for the displayed value is "Distance". Display - Display format In this menu item you define the number of decimal positions with which the measured value is displayed. The default setting for the display format is "Automatically".
6 Set up with the display and adjustment module The selected curve is continuously updated. A submenu with zoom functions is opened with the [OK] key: • • • Diagnosis - Simulation "X-Zoom": Zoom function for the meas. distance "Y-Zoom": 1, 2, 5 and 10x signal magnification in "dB" "Unzoom": Reset the presentation to the nominal measuring range without magnification In this menu item you can simulate measured values via the current output. This allows the signal path to be tested, e.g.
6 Set up with the display and adjustment module With the adjustment software PACTware and the PC, the stored echo curves can be displayed with high resolution and used to recognize signal changes over time. In addition, the echo curve saved during setup can also be displayed in the echo curve window and compared with the current echo curve. Additional adjustments Date/Time In this menu item, the internal clock of the sensor is set to the requested time and time format.
6 Set up with the display and adjustment module well as measured value and echo curve memory is deleted. The event and parameter modification memories remain unaffected. Basic settings: Resets the parameter settings, incl. special parameters, to the default values of the respective instrument. Any stored false signal suppression or user programmable linearisation curve, as well as the measured value memory, is deleted. Order-related settings are not taken over into the current parameters after this reset.
6 Set up with the display and adjustment module Additional adjustments - The instrument settings are copied with this function.
6 Set up with the display and adjustment module Additional adjustments False signal suppression The following circumstances cause interfering reflections and can influence the measurement: • • • • High sockets Vessel internals such as struts Agitators Buildup or welded joints on vessel walls Note: A false signal suppression detects, marks and saves these false signals so that they are no longer taken into account in the level measurement.
6 Set up with the display and adjustment module Delete: An already created false signal suppression will be completely deleted. This is useful if the saved false signal suppression no longer matches the metrological conditions in the vessel. Extend: is used to extend an already created false signal suppression. This is useful if a false signal suppression was carried out with too high a level and not all false signals could be detected.
6 Set up with the display and adjustment module Change the settings of the special parameters only after having contacted our service staff. Info - Instrument name In this menu item, you can read out the instrument name and the instrument serial number: Info - Instrument version In this menu item, the hardware and software version of the sensor is displayed.
6 Set up with the display and adjustment module • • • All data of the menu "Setup" and "Display" In the menu "Additional settings" the items "Sensor-specific units, temperature unit and linearization" The values of the user programmable linearization curve 51141-EN-160616 The function can also be used to transfer settings from one instrument to another instrument of the same type.
7 Setup with PACTware 7 Via the interface adapter directly on the sensor Setup with PACTware 7.1 Connect the PC 2 1 3 Fig. 37: Connection of the PC directly to the sensor via the interface adapter 1 2 3 USB cable to the PC Interface adapter VEGACONNECT Sensor Via the interface adapter and HART 2 4 USB N OPE 3 TWIST LO CK 1 5 Fig.
7 Setup with PACTware also usually equipped with a sufficient current limiting resistance. In such cases, the interface converter can be connected parallel to the 4 … 20 mA cable (dashed line in the previous illustration). Prerequisites 7.2 Parameter adjustment For parameter adjustment of the instrument via a Windows PC, the configuration software PACTware and a suitable instrument driver (DTM) according to FDT standard are required.
7 Setup with PACTware In the full version there is also an extended print function for complete project documentation as well as a save function for measured value and echo curves. In addition, there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves. The standard version is available as a download under www.vega.com/downloads and "Software". The full version is available on CD from the agency serving you. 7.
8 Set up with other systems 8 8.1 Set up with other systems DD adjustment programs Device descriptions as Enhanced Device Description (EDD) are available for DD adjustment programs such as, for example, AMS™ and PDM. The files can be downloaded at www.vega.com/downloads under "Software". 8.2 Field Communicator 375, 475 Device descriptions for the instrument are available as EDD for parameter adjustment with the Field Communicator 375 or 475.
9 Diagnosis, asset management and service 9 Diagnosis, asset management and service 9.1 Maintenance If the instrument is used correctly, no maintenance is required in normal operation. In some applications, buildup on the antenna system can influence the measuring result. Depending on the sensor and application, take measures to avoid heavy soiling of the antenna system. If necessary, clean the antenna system in certain intervals. 9.
9 Diagnosis, asset management and service • • • PC with PACTware/DTM Control system with EDD Display and adjustment module • • PC with PACTware/DTM Control system with EDD Further echo curves: Up to 10 echo curves can be stored in a ring buffer in this memory section. Further echo curves are stored via: 9.3 Asset Management function The instrument features self-monitoring and diagnostics according to NE 107 and VDI/VDE 2650.
9 Diagnosis, asset management and service still valid. Plan in maintenance for the instrument because a failure is expected in the near future (e.g. due to buildup). This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD. Failure The following table shows the error codes in the status message "Failure" and gives information on the reason and rectification. Keep in mind that some information is only valid with four-wire instruments.
9 Diagnosis, asset management and service Code Cause Rectification DevSpec State in CMD 48 F125 –– Temperature of the electronics in the nonspecified range –– Check ambient temperature –– Isolate electronics –– Use instrument with higher temperature range Bit 7 of Byte 0…5 F260 –– Error in the calibration –– Exchanging the eleccarried out in the factory tronics –– Error in the EEPROM –– Send instrument for repair Text message Impermissible electronics temperature Error in the calibration F261 Bi
9 Diagnosis, asset management and service Code Cause Rectification S601 –– Danger of vessel overfilling –– Make sure that there is Bit 9 of Byte 14…24 no further filling –– Check level in the vessel S603 –– Operating voltage below –– Check electrical conspecified range nection –– If necessary, increase operating voltage Text message Overfilling Impermissible operating voltage Maintenance DevSpec State in CMD 48 Bit 11 of Byte 14…24 The following table shows the error codes and text messages in
9 Diagnosis, asset management and service Code Text message Cause Rectification M507 –– Error during setup –– Carry out reset and repeat setup Error in the instrument set- –– Error when carrying out a reset tings –– False signal suppression faulty 9.4 DevSpec State in CMD 48 Bit 7 of Byte 14…24 Rectify faults Reaction when malfunction occurs The operator of the system is responsible for taking suitable measures to rectify faults.
Level 9 Diagnosis, asset management and service 1 2 0 1 2 time Real level Level displayed by the sensor Notes: • Wherever the sensor displays a constant value, the reason could also be the fault setting of the current output to "Hold value" • If the level indication is too low, the reason could be a line resistance that is too high Measurement error with constant level 1. Measured value shows a too low or too high level Error pattern Level Fault description 2.
9 Diagnosis, asset management and service Measurement error during filling 4. Measured value remains unchanged during filling Error pattern 0 time Level 5. Measured value remains in the bottom section during filling 0 7. Measured value jumps towards 0 % during filling Level time 0 Rectification –– Echo from the tank bottom larger than the level echo, for example, with products with εr < 2.
9 Diagnosis, asset management and service Measurement error during emptying 12. Measured value jumps towards 0 % during emptying Error pattern Level 11. Measured value remains unchanged in the close range during emptying 0 0 time Level 13. Measured value jumps sporadically towards 100 % during emptying Cause Rectification –– False signal larger than the level echo –– Level echo too small –– Eliminate false signal in the close range.
9 Diagnosis, asset management and service In both cases, the serial number of the sensor is needed. The serial numbers are stated on the type label of the instrument, on the inside of the housing as well as on the delivery note. When loading on site, first of all the order data must be downloaded from the Internet (see operating instructions manual "Electronics module"). Caution: All user-specific settings must be entered again. Hence, you have to carry out a new setup after the electronics exchange.
10 Dismount 10 Dismount 10.1 Dismounting steps Warning: Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel or pipeline, high temperatures, corrosive or toxic products etc. Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order. 10.2 Disposal The instrument consists of materials which can be recycled by specialised recycling companies.
11 Supplement 11 Supplement 11.1 Technical data Materials and weights 316L corresponds to 1.4404 or 1.
11 Supplement Weights ƲƲ Instrument (depending on housing, process fitting and antenna) Torques Max. torques, threaded version ƲƲ G¾ ƲƲ G1½ approx. 2 … 17.2 kg (4.409 … 37.92 lbs) 30 Nm (22.13 lbf ft) 200 Nm (147.5 lbf ft) Max. torques, version plastic horn antenna ƲƲ Mounting screws, mounting strap on sensor housing ƲƲ Flange screws, compression flange DN 80 4 Nm (2.950 lbf ft) 5 Nm (3.689 lbf ft) ƲƲ Flange screws, adapter flange DN 100 7 Nm (5.
11 Supplement 1 3 2 Fig. 55: Data of the input variable 1 2 3 Reference plane Measured variable, max. measuring range Utilisable measuring range (depending on the antenna system) Max. measuring range 30 m (98.43 ft) Recommended measuring range (depending on the antenna system) ƲƲ Plastic horn antenna up to 30 m (98.43 ft) ƲƲ Thread with integrated horn antenna 1½" up to 20 m (65.62 ft) ƲƲ Flange from DN 80, 3" with encapsulated antenna system up to 30 m (98.
11 Supplement HART output values according to HART 7.01) ƲƲ PV (Primary Value) Lin. percent ƲƲ TV (Third Value) Meas. certainty ƲƲ SV (Secondary Value) Distance ƲƲ QV (Fourth Value) Electronics temperature Fulfilled HART specification Further information on Manufacturer ID, Device ID, Device Revision 7.
11 Supplement 10 mm (0.3937 in) 2 mm (0.079 in) 0 - 2 mm (- 0.079 in) 0,25 m (0.8202 ft) - 10 mm (- 0.3937 in) 1 2 3 Fig. 57: Deviation under reference conditions - plastic horn antenna 1 2 3 Reference plane Antenna edge Recommended measuring range 10 mm (0.3937 in) 2 mm (0.079 in) 0 - 2 mm (- 0.079 in) 0,25 m (0.8202 ft) 3 - 10 mm (- 0.3937 in) 1 2 Fig.
11 Supplement Characteristics and performance data Measuring frequency W-band (79 GHz technology) Measuring cycle time approx.
11 Supplement Antenna lens Seal Process temperature (measured on the process fitting) PTFE PTFE -40 … +130 °C (-40 … +266 °F) -40 … +200 °C (-40 … +392 °F) Ambient temperature - Process temperature 1 80 °C (176 °F) 2 0 °C (32 °F) -40 °C (-104 °F) 80 °C (176 °F) -40 °C (-104 °F) Fig.
11 Supplement A 80°C / 176°F 64°C / 147°F 1 56°C / 133°F 53°C / 127°F 48°C / 118°F 3 2 4 0°C / 32°F -40°C -40°F 50°C 122°F 80°C 176°F 100°C 212°F 150°C 302°F 200°C 392°F B -40°C / -40°F Fig.
11 Supplement A 80°C / 176°F 67°C / 159°F 1 60°C / 140°F 57°C / 135°F 52°C / 126°F 3 2 4 0°C / 32°F -40°C -40°F 50°C 122°F 80°C 176°F 100°C 212°F 150°C 302°F 200°C 392°F B -40°C / -40°F Fig.
11 Supplement Data on rinsing air connection Max. permissible pressure 6 bar (87.02 psig) Air volume, depending on pressure (recommended range) Plastic horn antenna Air volume Pressure Without reflux valve 0.2 bar (2.9 psig) 3.3 m3/h 0.4 bar (5.8 psig) 5 m3/h 0.6 bar (8.7 psig) 6 m3/h 1 m3/h 0.8 bar (11.6 psig) - 2.1 m3/h 1 bar (14.5 psig) - 3 m3/h 1.2 bar (17.4 psig) - 3.5 m3/h 1.4 bar (20.3 psig) - 4.2 m3/h 1.6 bar (23.2 psig) - 4.4 m3/h 1.8 bar (20.3 psig) - 4.
11 Supplement Wire cross-section (spring-loaded terminals) ƲƲ Massive wire, stranded wire 0.2 … 2.5 mm² (AWG 24 … 14) ƲƲ Stranded wire with end sleeve 0.2 … 1.5 mm² (AWG 24 … 16) Electromechanical data - version IP 66/IP 68 (1 bar) Options of the cable entry ƲƲ Cable gland with integrated connection cable M20 x 1.5 (cable: ø 5 … 9 mm) ƲƲ Blind plug M20 x 1.5; ½ NPT ƲƲ Cable entry ½ NPT Connection cable ƲƲ Wire cross-section 0.
11 Supplement Integrated clock Date format Day.Month.Year Time zone Ex factory CET Time format 12 h/24 h Rate deviation max. 10.5 min/year Additional output parameter - Electronics temperature Output of the temperature values ƲƲ Analogue Via the current output ƲƲ Digital Via the digital output signal - depending on the electronics version Range -40 … +85 °C (-40 … +185 °F) Resolution < 0.
11 Supplement Housing material Version IP-protection class NEMA protection Aluminium Single chamber IP 66/IP 68 (0.2 bar) Type 6P Double chamber IP 66/IP 67 Type 4X IP 68 (1 bar) Type 6P IP 68 (1 bar) IP 66/IP 68 (0.2 bar) Type 6P Type 6P Stainless steel, electropolished Single chamber IP 66/IP 68 (0.2 bar) Type 6P Stainless steel, precision casting Single chamber IP 66/IP 68 (0.
11 Supplement Aluminium housing ~ 116 mm (4.57") ~ 87 mm (3.43") ø 86 mm (3.39") ø 86 mm (3.39") 120 mm (4.72") 116 mm (4.57") M16x1,5 M20x1,5 M20x1,5/ ½ NPT M20x1,5/ ½ NPT 1 2 Fig. 65: Housing versions with protection rating IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher 1 2 Single chamber version Double chamber version Aluminium housing with protection rating IP 66/IP 68 (1 bar) ~ 105 mm (4.13") ~ 150 mm (5.91") ø 86 mm (3.
11 Supplement Stainless steel housing ~ 87 mm (3.43") ~ 69 mm (2.72") ø 79 mm (3.11") ~ 59 mm (2.32") ø 80 mm (3.15") ø 86 mm (3.39") M20x1,5/ ½ NPT M20x1,5/ ½ NPT 1 120 mm (4.72") 112 mm (4.41") 117 mm (4.61") M16x1,5 M20x1,5/ ½ NPT 2 3 Fig. 67: Housing versions with protection rating IP 66/IP 68 (0.2 bar) - with integrated display and adjustment module the housing is 9 mm/0.
11 Supplement 19 mm (0.75") 10,5 mm (0.41") 1 ø 107 mm (4.21") 126 mm (4.96") VEGAPULS 64, plastic horn antenna with compression flange ø 21 mm (0.83") ø 75 mm (2.95") ø 115 mm (4.53") ø 156 mm (6.14") ø 200 mm (7.87") Fig.
11 Supplement 126 mm (4.96") ø 107 mm (4.21") ø 21 mm (0.83") 1 ø 75 mm (2.95") ø 156 mm (6.14") 14,7 mm (0.58") 2 26,5 mm (1.04") 19 mm (0.75") 10,5 mm (0.41") VEGAPULS 64, plastic horn antenna with compression flange and rinsing connection 3 ø 200 mm (7.87") Fig.
11 Supplement 144 mm (5.67") VEGAPULS 64, plastic horn antenna with adapter flange 20 mm (0.79") 36,5 mm (1.44") 17,5 mm (0.69") 1 ø 75 mm (2.95") ø 98 mm (3.86") 2 Fig.
11 Supplement 1 2 34,7 mm (1.37") 20 mm (0.79") 62,5 mm (2.46") 170 mm (6.69") VEGAPULS 64, plastic horn antenna mit adapter flange und rinsing connection ø 75 mm (2.95") ø 98 mm (3.86") 3 Fig.
11 Supplement VEGAPULS 64, plastic horn antenna with mounting strap 125 mm (4.92") ø 107 mm (4.21") 9 mm (0.35") 98 mm (3.86") ø 115 mm (4.53") 9 mm (0.35") 85 mm (3.35") 12 mm (0.47") ø 75 mm (2.95") 6 mm (0.24") 8,5 mm (0.34") 19 mm (0.75") 170 mm (6.69") 300 mm (11.81") 2,5 mm (0.10") 12 mm (0.47") 88 VEGAPULS 64 • 4 … 20 mA/HART - two-wire 51141-EN-160616 Fig.
11 Supplement VEGAPULS 64, thread with integrated horn antenna SW 36 mm (1.42") 61 mm (2.40") G¾ ø 24 mm (0.94") 25 mm (0.98") 25 mm (0.98") 61 mm (2.40") SW 36 mm (1.42") TA ¾ NPT ø 24 mm (0.94") TB ø 42,5 mm (1.67") TC 101 mm (3.98") G1 ½ SW 55 mm (2.17") 33 mm (1.31") 33 mm (1.31") 101 mm (3.98") SW 55 mm (2.17") 1 ½ NPT ø 42,5 mm (1.67") TD 51141-EN-160616 Fig. 74: VEGAPULS 64, thread with integrated horn antenna TA TB TC TD G¾ (DIN 3852-E) ¾ NPT (ASME B1.20.
11 Supplement VEGAPULS 64, flange with encapsulated antenna system DN 50 2 19 mm (0.75") ø45 mm (1.77") 141 mm (5.55") 89 mm (3.50") 1 19 mm (0.75") ø102 mm (4.02") Fig. 75: VEGAPULS 64, encapsulated antenna system DN 50 1 2 Version up to 130 °C (266 °F) Version up to 200 °C (392 °F) VEGAPULS 64, flange with encapsulated antenna system DN 80 ø138 mm (5.43") 32 mm (1.25") 32 mm (1.25") ø75 mm (2.95") 181 mm (7.13") 2 129 mm (5.08") 1 1 2 90 51141-EN-160616 Fig.
11 Supplement 11.3 Industrial property rights VEGA product lines are global protected by industrial property rights. Further information see www.vega.com. VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte. Nähere Informationen unter www.vega.com. Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellectuelle. Pour plus d'informations, on pourra se référer au site www.vega.com.
INDEX INDEX A Adjustment 42, 43 Agitator 23 C Change the language 44 Check output signal 63 Connection 27 ––Cable 26 Copy sensor settings 49 Current output ––Adjust 49 ––Min./Max.
51141-EN-160616 Notes VEGAPULS 64 • 4 … 20 mA/HART - two-wire 93
Notes 51141-EN-160616 94 VEGAPULS 64 • 4 … 20 mA/HART - two-wire
51141-EN-160616 Notes VEGAPULS 64 • 4 … 20 mA/HART - two-wire 95
All statements concerning scope of delivery, application, practical use and operating conditions of the sensors and processing systems correspond to the information available at the time of printing. Subject to change without prior notice © VEGA Grieshaber KG, Schiltach/Germany 2016 VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com www.vega.