Operating Instructions Radar sensor for continuous level measurement of bulk solids VEGAPULS 69 4 … 20 mA/HART - two-wire Document ID: 47247
Contents Contents About this document 1.1 Function............................................................................................................................ 4 1.2 Target group...................................................................................................................... 4 1.3 Symbols used................................................................................................................... 4 2 For your safety 2.1 Authorised personnel...................
Contents 8 Set up with other systems 8.1 DD adjustment programs................................................................................................ 63 8.2 Field Communicator 375, 475......................................................................................... 63 9 Diagnosis, asset management and service 9.1 Maintenance................................................................................................................... 64 9.2 Measured value and event memory..........
1 About this document 1 About this document 1.1 Function 1.2 Target group 1.3 Symbols used This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectification. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device. This operating instructions manual is directed to trained specialist personnel.
2 For your safety 2 2.1 For your safety Authorised personnel All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator. During work on and with the device the required personal protective equipment must always be worn. 2.2 Appropriate use VEGAPULS 69 is a sensor for continuous level measurement. You can find detailed information about the area of application in chapter "Product description".
2 For your safety 2.5 CE conformity The device fulfils the legal requirements of the applicable EU directives. By affixing the CE marking, we confirm the conformity of the instrument with these directives. You can find the EU conformity declaration on our website under www.vega.com/downloads. 2.6 NAMUR recommendations NAMUR is the automation technology user association in the process industry in Germany. The published NAMUR recommendations are accepted as the standard in field instrumentation.
2 For your safety • • • • The instrument must be stationary mounted and the antenna directed vertically downward The instrument may only be used outside closed vessels in the version with G1½ or 1½ NPT thread with integrated horn antenna.
2 For your safety • • This device may not cause interference, and This device must accept any interference, including interference that may cause undesired operation of the device This device has been approved for both closed containers and openair environments with the following limitations: • • • • • • • • • • Closed Containers: For installations utilizing a tilt during installation: This device is limited to installation in a completely enclosed container made of metal or concrete to prevent RF em
2 For your safety • L’utilisateur de l’appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d’en compromettre le fonctionnement.
2 For your safety with the goal of continuously improving company environmental protection. The environment management system is certified according to DIN EN ISO 14001.
3 Product description 3 Type label Product description 3.1 Configuration The type label contains the most important data for identification and use of the instrument: 1 2 3 4 5 6 7 8 9 10 15 14 13 12 11 Fig.
3 Product description • • Scan the Data Matrix code on the type label of the instrument or Enter the serial number manually in the app Scope of this operating instructions manual This operating instructions manual applies to the following instrument versions: Scope of delivery The scope of delivery encompasses: • • • • • • Hardware version from 1.0.0 Software version from 1.1.
3 Product description via special algorithms in the sensor electronics. The determined filling height is then converted into a corresponding output signal and outputted as the measured value. Packaging 3.3 Packaging, transport and storage Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test based on ISO 4180. The packaging of standard instruments consists of environmentfriendly, recyclable cardboard.
3 Product description parameter adjustment of these instruments, the adjustment software PACTware with VEGA-DTM is required. You can find further information in the operating instructions "Interface adapter VEGACONNECT" (Document-ID 32628). VEGADIS 81 The VEGADIS 81 is an external display and adjustment unit for VEGA plics® sensors. For sensors with double chamber housing the interface adapter "VEGADIS adapter" is also required for VEGADIS 81.
3 Product description Flanges with plastic horn antenna For mounting the instrument with plastic horn antenna to a socket, two flange versions are available: the combi compression flange and the adapter flange You can find additional information in chapter "Mounting" of this operating instruction. Mounting strap with plas- A mounting strap is available for mounting the instrument with plastic tic horn antenna horn antenna to the wall or ceiling.
4 Mounting 4 Protection against moisture Mounting 4.1 General instructions Protect your instrument against moisture ingress through the following measures: • • • Use the recommended cable (see chapter "Connecting to power supply") Tighten the cable gland Loop the connection cable downward in front of the cable gland This applies particularly to: • • • Outdoor mounting Installations in areas where high humidity is expected (e.g.
4 Mounting Mounting strap 4.2 Mounting versions, plastic horn antenna The optional mounting strap allows simple mounting of the instrument on a wall, ceiling or boom. Especially in the case of open vessels, this is a simple and effective way to align the sensor to the surface of the bulk solid material. The following versions are available: • • Mounting strap - Ceiling mounting Length 300 mm Length 170 mm The instrument is normally mounted vertically with a bracket on the ceiling.
4 Mounting Fig. 4: Wall mounting with inclined wall via the mounting strap with length 300 mm Flange Two versions are available for mounting the instrument on a socket: • • Combi compression flange Adapter flange Combi compression flange The combi compression flange is suitable for different vessel flanges DN 80, ASME 3" and JIS 80. It comes not sealed against the radar sensor and can thus only be used unpressurized.
4 Mounting 1 2 3 Fig. 6: Adapter flange 1 2 3 Connection screw Adapter flange Process seal 4.3 Mounting preparations, mounting strap The mounting strap is supplied unassembled (optionally) and must be screwed to the sensor before setup with three hexagon socket screws M5 x 10 and spring washers. Max. torque, see chapter "Technical data". Required tools: Allen wrench size 4. There are two different variants of screwing the strap to the sensor, see following illustration: 1 2 Fig.
4 Mounting Polarisation 4.4 Mounting instructions Radar sensors for level measurement emit electromagnetic waves. The polarization is the direction of the electrical component of these waves. The polarization direction is marked by a nose on the housing, see following drawing: 1 Fig. 8: Position of the polarisation 1 Nose for marking the direction of polarisation Note: When the housing is rotated, the direction of polarization changes and hence the influence of the false echo on the measured value.
4 Mounting Fig. 10: Mounting of the radar sensor with inflowing medium 47247-EN-170315 Silo with lateral filling In bulk solids silos with lateral pneumatic filling the instrument should not be mounted above the filling stream, as the radar signal will be disrupted. The optimal mounting position is to the side of the filling aperture. To avoid heavy soiling, the distance to any filter or dust exhauster should be as large as possible.
4 Mounting Fig. 11: Mounting of the radar sensor with inflowing medium Type of socket Information: The mounting socket should be as short as possible and its end rounded. This reduces false echoes from the vessel mounting socket. Fig. 12: Recommended socket mounting If the medium has good reflective properties, you can also mount the VEGAPULS 69 on longer sockets. Approximate socket heights are shown in the following illustration.
h 4 Mounting d The below charts specify the max. pipe socket length h depending on the diameter d. Socket length h 80 mm ≤ 200 mm 100 mm ≤ 300 mm 150 mm ≤ 500 mm Socket diameter d Socket length h 3" ≤ 7.87 in 4" ≤ 11.8 in 6" ≤ 19.7 in In order to measure as much of the vessel volume as possible, the sensor should be aligned so that the radar signal reaches the lowest level in the vessel.
4 Mounting r 1/3 r...1/2 r Fig. 14: Mounting position and orientation With the help of an optional swivelling holder, the sensor can be easily aligned to the vessel centre. The necessary angle of inclination depends on the vessel dimensions. It can be easily checked with a suitable bubble tube or mechanic's level on the sensor.
4 Mounting d α a Fig. 15: Proposal for installation after orientation VEGAPULS 69 47247-EN-170315 The following table shows the necessary angle of inclination. It depends on the measuring distance and the distance "a" between vessel centre and installation position. Distance d 2° (m) 4° 6° 8° 10° 2 0.1 0.1 0.2 0.3 0.4 4 0.1 0.3 0.4 0.6 0.7 6 0.2 0.4 0.6 0.8 1.1 8 0.3 0.6 0.8 1.1 1.4 10 0.3 0.7 1.1 1.4 1.8 15 0.5 1 1.6 2.1 2.6 20 0.7 1.4 2.1 2.8 3.5 25 0.
4 Mounting Distance d 2° (m) 4° 6° 8° 10° 80 2.8 5.6 8.4 11.1 13.9 90 3.1 6.3 9.4 12.5 15.6 100 3.5 7 10.5 13.9 17.4 110 3.8 7.7 11.5 15.3 19.1 120 4.2 8.4 12.5 16.7 20.8 Example: In a vessel 20 m high, the installation position of the sensor is 1.4 m from the vessel centre. The necessary angle of inclination of 4° can be read out from this table. Proceed as follows to adjust the angle of inclination with the swivelling holder: 1.
4 Mounting Small, inclined sheet metal baffles above the installations scatter the radar signals and prevent direct interfering reflections. Fig. 17: Cover flat, large-area profiles with deflectors Material heaps Large material heaps are best measured with several instruments, which can be mounted on e.g. traverse cranes. For this type of application it is advantageous to orient the sensor perpendicular to the bulk solid surface. The sensors do not influence each other.
4 Mounting 2 3 max. 50 mm (1.97") 1 Fig. 19: Mounting the instrument on insulated vessels. 1 2 3 Mounting in multiple chamber silo Electronics housing Spacer Vessel insulation The walls of multiple-chamber silos are often made of profile material, e.g. profile sheeting, to ensure the required stability. If the radar sensor is mounted very close to a heavily structured vessel wall, substantial interfering reflections can be generated.
4 Mounting Fig. 21: Installation and orientation in multiple chamber silos Dust deposits - Rinsing air connection To avoid heavy buildup and dust on the antenna, the sensor should not be mounted close to the dust exhauster inside the vessel. To protect the sensor against buildup, particularly in case of strong condensation, air rinsing is recommended.
4 Mounting Fig. 23: Rinsing air connection with compression flange Fig. 24: Rinsing air connection with adapter flange You can find details on the rinsing air connection in chapter "Technical data".
5 Connecting to power supply 5 Safety instructions Connecting to power supply 5.1 Preparing the connection Always keep in mind the following safety instructions: Warning: Connect only in the complete absence of line voltage. • • Voltage supply The electrical connection must only be carried out by trained personnel authorised by the plant operator. If overvoltage surges are expected, overvoltage arresters should be installed. Power supply and current signal are carried on the same two-wire cable.
5 Connecting to power supply On plastic housings, the NPT cable gland or the Conduit steel tube must be screwed into the threaded insert without grease. Max. torque for all housings, see chapter "Technical data". Cable screening and grounding If screened cable is required, the cable screen must be connected on both ends to ground potential. In the sensor, the screen is connected directly to the internal ground terminal.
5 Connecting to power supply Fig. 25: Connection steps 5 and 6 - Single chamber housing Fig. 26: Connection steps 5 and 6 - Double chamber housing 6. Insert the wire ends into the terminals according to the wiring plan 47247-EN-170315 Information: Solid cores as well as flexible cores with wire end sleeves are inserted directly into the terminal openings. In case of flexible cores without end sleeves, press the terminal from above with a small screwdriver, the terminal opening is then free.
5 Connecting to power supply 9. Tighten the compression nut of the cable entry gland. The seal ring must completely encircle the cable 10. Reinsert the display and adjustment module, if one was installed 11. Screw the housing lid back on The electrical connection is finished. 5.3 Wiring plan, single chamber housing The following illustration applies to the non-Ex as well as to the Ex-ia version. Electronics and terminal compartment 2 3 4...20mA (+)1 2(-) 5 6 7 8 4 1 Fig.
5 Connecting to power supply Terminal compartment 2 4...20mA (+)1 3 Display 2(-) 5 6 7 8 4 1 Fig.
5 Connecting to power supply Terminal compartment - Radio module PLICSMOBILE Status SIM-Card Test USB (+)1 2(-) 1 Fig. 31: Terminal compartment - Radio module PLICSMOBILE 1 Voltage supply You can find detailed information on connection in the supplementary instructions "PLICSMOBILE GSM/GPRS radio module". Electronics compartment 5.5 Double chamber housing Ex d 2 4...20mA (+)1 2(-) 1 5 6 7 8 1 Fig.
5 Connecting to power supply Terminal compartment 2 3 4...20mA (+)1 2(-) 5 6 7 8 4 1 Fig. 33: Terminal compartment - double chamber housing 1 2 3 4 Electronics compartment Voltage supply, signal output For display and adjustment module or interface adapter For external display and adjustment unit Ground terminal for connection of the cable screen 5.6 Wiring plan, double chamber housing Ex d ia 2 4...20mA (+)1 2(-) 1 5 6 7 8 3 Fig.
5 Connecting to power supply Terminal compartment 4...20mA (+)1 2 2(-) 1 Fig. 35: Connection compartment - Ex-d-ia double chamber housing 1 2 Voltage supply, signal output Ground terminal for connection of the cable screen Plug M12 x 1 for external display and adjustment unit 4 3 1 2 Fig.
5 Connecting to power supply Electronics compartment 5.7 Double chamber housing with DISADAPT 1 2 3 Fig. 37: View to the electronics compartment with VEGADIS adapter for connection of the external display and adjustment unit 1 2 3 VEGADIS adapter Internal plug connection Plug connector M12 x 1 Assignment of the plug connector 4 3 1 2 Fig.
5 Connecting to power supply Wire assignment, connection cable 5.8 Wiring plan - version IP 66/IP 68, 1 bar 1 2 Fig. 39: Wire assignment in permanently connected connection cable 1 2 Brown (+) and blue (-) to power supply or to the processing system Shielding 5.9 Switch-on phase After connecting the instrument to power supply or after a voltage recurrence, the instrument carries out a self-check for approx.
6 Set up with the display and adjustment module 6 6.1 Set up with the display and adjustment module Insert display and adjustment module The display and adjustment module can be inserted into the sensor and removed again at any time. You can choose any one of four different positions - each displaced by 90°. It is not necessary to interrupt the power supply. Proceed as follows: 1. Unscrew the housing lid 2.
6 Set up with the display and adjustment module 1 2 Fig. 41: Installing the display and adjustment module in the double chamber housing 1 2 In the electronics compartment In the terminal compartment Note: If you intend to retrofit the instrument with a display and adjustment module for continuous measured value indication, a higher lid with an inspection glass is required. 6.
6 Set up with the display and adjustment module • • • –– –– –– –– Move to the menu overview Confirm selected menu Edit parameter Save value [->] key: –– Change measured value presentation –– Select list entry –– Select menu items in the quick setup menu –– Select editing position [+] key: –– Change value of the parameter [ESC] key: –– Interrupt input –– Jump to next higher menu Adjustment system The instrument is operated via the four keys of the display and adjustment module.
6 Set up with the display and adjustment module Selection of national language This menu item is used to select the national language for further parameter adjustment. You can change the selection via the menu item "Setup - Display, Menu language". With the "OK" key you move to the main menu. 6.4 Parameter adjustment - Quick setup To quickly and easily adapt the sensor to the application, select the menu item "Quick setup" in the start graphic on the display and adjustment module.
6 Set up with the display and adjustment module parameters to ensure optimum adjustment of the measurement. The procedure is described in the following. Setup - Measurement loop name Here you can assign a suitable measurement loop name. Push the "OK" key to start the editing. With the "+" key you change the sign and with the "->" key you jump to the next position. You can enter names with max. 19 characters.
6 Set up with the display and adjustment module Vessel height/Measuring range: The VEGAPULS 69 is a bulk solids radar sensor for high, slender vessels. It covers a measuring range up to 120 m. This menu item lets you limit the active measuring range in which the instrument searches for level echoes. Information: The min. adjustment must be carried out independently of this.
6 Set up with the display and adjustment module • –– Large distance to the medium –– Large angles of repose Properties, sensor: –– Lower averaging –– Very high measured value jumps are accepted Heap: • Sensor mounting on movable conveyor belts • Detection of the heap profile • Height detection during filling • Process/measurement conditions: –– Measured value jumps, e.g.
3 100% 2 35 m (1378") 1m (39.37") 6 Set up with the display and adjustment module 0% 1 Fig. 43: Parameter adjustment example min./max. adjustment 1 2 3 Min. level = max. measuring distance Max. level = min. measuring distance Reference plane If these values are not known, an adjustment with the distances of e.g. 10 % and 90 % is possible. Starting point for these distance specifications is always the sealing surface of the thread or flange.
6 Set up with the display and adjustment module 4. For the full vessel, enter the distance value in m matching the percentage value. 5. Save settings with [OK] and move with [ESC] and [->] to Min. adjustment. Setup - Min. adjustment Proceed as follows: 1. Select with [->] the menu item "Min. adjustment" and confirm with [OK]. 2. Edit the percentage value with [OK] and set the cursor to the requested position with [->]. 3. Set the requested percentage value with [+] and save with [OK].
6 Set up with the display and adjustment module Setup - Current output Min./Max. In the menu item "Current output Min./Max.", you determine the reaction of the current output during operation. The default setting is min. current 3.8 mA and max. current 20.5 mA. Lock/unlock setup - Adjustment In the menu item "Lock/unlock adjustment" you safeguard the sensor parameters against unauthorized or unintentional modifications.
6 Set up with the display and adjustment module Display - Displayed value 1 and 2 In this menu item you can define the way measured values are indicated on the display. The default setting for the display value is "Percent". Display - Backlight The display and adjustment module has a backlight for the display. In this menu item you can switch the lighting on or off. You can find the required operating voltage in chapter "Technical data". In delivery status, the lighting is switched on.
6 Set up with the display and adjustment module Diagnosis - Simulation In this menu item you can simulate measured values via the current output. This allows the signal path to be tested, e.g. through downstream indicating instruments or the input card of the control system. Select the requested simulation variable and set the requested value. Caution: During simulation, the simulated value is outputted as 4 … 20 mA current value and as digital HART signal.
6 Set up with the display and adjustment module Additional settings - Date/ In this menu item, the internal clock of the sensor is set to the requestTime ed time and time format. At the time of shipment from factory, the instrument is set to CET (Central European Time). Additional adjustments - Reset During a reset, the parameter settings carried out by the user are reset to the default values (see below table). Proceed as follows: 1.
6 Set up with the display and adjustment module Menu Menu item Default value Setup Measurement loop name Sensor Units Distance in m Application Medium: Crushed stones/gravel Temperature in °C Application: Silo Vessel bottom: Flat Vessel height/Measuring range: 120 m Display Min. adjustment 120 m Max. adjustment 0,000 m(d) Damping 0.0 s Current output mode Output characteristics: 4 … 20 mA Current output Min./Max. Min. current: 3.8 mA Failure mode: < 3.6 mA ▼ Max. current: 20.
6 Set up with the display and adjustment module • • The menu items "Reset, Date/Time" in the menu "Additional settings" The user-programmable linearization curve The copied data are permanently saved in an EEPROM memory in the display and adjustment module and remain there even in case of power failure. From there, they can be written into one or more sensors or kept as backup for a possible electronics exchange.
6 Set up with the display and adjustment module • • Agitators Buildup or welded joints on vessel walls Note: A false signal suppression detects, marks and saves these false signals so that they are no longer taken into account in the level measurement. This should be done with the lowest possible level so that all potential interfering reflections can be detected. Proceed as follows: 1. Select with [->] the menu item "False signal suppression" and confirm with [OK]. 2. Confirm again with [OK]. 3.
6 Set up with the display and adjustment module Delete: An already created false signal suppression will be completely deleted. This is useful if the saved false signal suppression no longer matches the metrological conditions in the vessel. Extend: is used to extend an already created false signal suppression. This is useful if a false signal suppression was carried out with too high a level and not all false signals could be detected.
6 Set up with the display and adjustment module Info - Instrument name In this menu item, you can read out the instrument name and the instrument serial number: Info - Instrument version In this menu item, the hardware and software version of the sensor is displayed. Info - Factory calibration date In this menu item, the date of factory calibration of the sensor as well as the date of the last change of sensor parameters are displayed via the display and adjustment module or via the PC.
6 Set up with the display and adjustment module • • The items "Sensor-specific units, temperature unit and linearisation" in the menu "Additional settings". The values of the user-programmable linearisation curve 47247-EN-170315 The function can also be used to transfer settings from one instrument to another instrument of the same type.
7 Setup with PACTware 7 Via the interface adapter directly on the sensor Setup with PACTware 7.1 Connect the PC 2 1 3 Fig. 44: Connection of the PC directly to the sensor via the interface adapter 1 2 3 USB cable to the PC Interface adapter VEGACONNECT Sensor Via the interface adapter and HART 4 1 2 3 Fig.
7 Setup with PACTware also usually equipped with a sufficient current limiting resistance. In such cases, the interface converter can be connected parallel to the 4 … 20 mA cable (dashed line in the previous illustration). Prerequisites 7.2 Parameter adjustment For parameter adjustment of the instrument via a Windows PC, the configuration software PACTware and a suitable instrument driver (DTM) according to FDT standard are required.
7 Setup with PACTware In the full version there is also an extended print function for complete project documentation as well as a save function for measured value and echo curves. In addition, there is a tank calculation program as well as a multiviewer for display and analysis of the saved measured value and echo curves. The standard version is available as a download under www.vega.com/downloads and "Software". The full version is available on CD from the agency serving you. 7.
8 Set up with other systems 8 8.1 Set up with other systems DD adjustment programs Device descriptions as Enhanced Device Description (EDD) are available for DD adjustment programs such as, for example, AMS™ and PDM. The files can be downloaded at www.vega.com/downloads under "Software". 8.2 Field Communicator 375, 475 Device descriptions for the instrument are available as EDD for parameterisation with Field Communicator 375 or 475.
9 Diagnosis, asset management and service 9 Diagnosis, asset management and service 9.1 Maintenance If the instrument is used correctly, no maintenance is required in normal operation. In some applications, buildup on the antenna system can influence the measuring result. Depending on the sensor and application, take measures to avoid heavy soiling of the antenna system. If necessary, clean the antenna system in certain intervals. 9.
9 Diagnosis, asset management and service • • • PC with PACTware/DTM Control system with EDD Display and adjustment module • • PC with PACTware/DTM Control system with EDD Further echo curves: Up to 10 echo curves can be stored in a ring buffer in this memory section. Additional echo curves are stored via: 9.3 Asset Management function The instrument features self-monitoring and diagnostics according to NE 107 and VDI/VDE 2650.
9 Diagnosis, asset management and service still valid. Plan in maintenance for the instrument because a failure is expected in the near future (e.g. due to buildup). This status message is inactive by default. It can be activated by the user via PACTware/DTM or EDD. The following table shows the error codes in the status message "Failure" and gives information on the reason and rectification. Keep in mind that some information is only valid for four-wire instruments.
9 Diagnosis, asset management and service Code Text message F261 Error in the instrument settings F264 Installation/Setup error F265 Measurement function disturbed Reason Removal • Error during setup • False signal suppression faulty • Error when carrying out a reset • Adjustment not within the vessel height/measuring range • Max.
9 Diagnosis, asset management and service Code Reason Removal DevSpec M501 • Hardware error EEPROM • Exchanging the electronics • Send instrument for repair Bit 1 of Byte 14 … 24 M502 • Hardware error EEPROM • Exchanging the electronics • Send instrument for repair Text message Error in the nonactive linearisation table Error in the event memory M503 Measurement reliability too low M504 Error at a device interface M505 No echo available M506 Installation/Setup error M507 Error in the instr
9 Diagnosis, asset management and service Error Reason 4 … 20 mA signal not stable 4 … 20 mA signal missing • • Removal Fluctuations of the measured variable Electrical connection faulty • Set damping appropriate to the instrument via the display and adjustment module or PACTware/DTM • Check connection according to chapter "Connection steps" and if necessary, correct according to chapter "Wiring plan" • Voltage supply missing • Check cables for breaks; repair if necessary • Operating voltage too •
9 Diagnosis, asset management and service 2. Measured value jumps towards 100 % Error pattern Level Fault description 0 time Cause • • Rectification Due to the process, the amplitude of the product echo decreases A false signal suppression was not carried out • Carry out a false signal suppression • Amplitude or position of a false • Determine the reason for the signal has changed (e.g.
9 Diagnosis, asset management and service Measurement error during emptying 6. Measured value remains unchanged in the close range during emptying Error pattern Level Fault description 0 time Cause • • False signal larger than the level echo Level echo too small Rectification • Eliminate false signals in the • • • 8.
9 Diagnosis, asset management and service • • In the factory Or on site by the user In both cases, the serial number of the sensor is needed. The serial numbers are stated on the type label of the instrument, on the inside of the housing as well as on the delivery note. When loading on site, the order data must first be downloaded from the Internet (see operating instructions manual "Electronics module"). Caution: All application-specific settings must be entered again.
10 Dismount 10 Dismount 10.1 Dismounting steps Warning: Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel or pipeline, high temperatures, corrosive or toxic products etc. Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order. 10.2 Disposal The instrument consists of materials which can be recycled by specialised recycling companies.
11 Supplement 11 Supplement 11.1 Technical data Note for approved instruments The technical data in the respective safety instructions are valid for approved instruments (e.g. with Ex approval). These data can differ from the data listed herein, for example regarding the process conditions or the voltage supply. Materials and weights 316L corresponds to 1.4404 or 1.4435 Materials, wetted parts with metal-jacketed lens antenna ƲƲ Process fitting 316L, Alloy C22 (2.4602), Alloy 400 (2.
11 Supplement Torques Max. torques, threaded version ƲƲ G¾ ƲƲ G1½ 30 Nm (22.13 lbf ft) 200 Nm (147.5 lbf ft) Max. torques, version plastic horn antenna ƲƲ Mounting screws, mounting strap on sensor housing ƲƲ Flange screws, compression flange DN 80 4 Nm (2.950 lbf ft) 5 Nm (3.689 lbf ft) ƲƲ Flange screws, adapter flange DN 100 7 Nm (5.163 lbf ft) Max. torques, version metal-jacketed lens antenna ƲƲ Terminal screws for swivelling holder 8 Nm (5.9 lbf ft) ƲƲ Plastic housing 10 Nm (7.
11 Supplement Starting current ≤ 3.6 mA; ≤ 10 mA for 5 ms after switching on Damping (63 % of the input variable), adjustable 0 … 999 s Load see load diagram under Power supply HART output values according to HART 7.02) ƲƲ PV (Primary Value) Lin. percent ƲƲ TV (Third Value) Meas. certainty ƲƲ SV (Secondary Value) ƲƲ QV (Fourth Value) Fulfilled HART specification Further information on Manufacturer ID, Device ID, Device Revision Distance Electronics temperature 7.
11 Supplement 30 mm (1.181 in) 5 mm (0.197 in) 0 - 5 mm (- 0.197 in) 0,15 m (0.49 ft) 1,0 m (3.281 ft) 3 - 30 mm (- 1.181 in) 1 2 Fig. 58: Deviation under reference conditions 1 2 3 Reference plane Recommended min. distance Measuring range end Repeatability Deviation with bulk solids ≤ ±1 mm The values depend to a great extent on the application. Binding specifications are thus not possible.
11 Supplement Emitted HF power (depending on the parameter adjustment)5) ƲƲ Average spectral transmission power density -3 dBm/MHz EIRP ƲƲ Max. power density at a distance of 1m < 3 µW/cm² ƲƲ Max. spectral transmission power density +34 dBm/50 MHz EIRP Ambient conditions Ambient, storage and transport tempera- -40 … +80 °C (-40 … +176 °F) ture Process conditions For the process conditions, please also note the specifications on the type label. The lowest value always applies.
11 Supplement A 80°C / 176°F 1 2/3 60°C / 140°F 51°C / 124°F 4 35°C / 95°F 0°C / 32°F 50°C 122°F -40°C -40°F 80°C 176°F 100°C 212°F 130°C 266°F B -40°C / -40°F Fig.
11 Supplement Vessel pressure ƲƲ Metal-jacketed lens antenna -1 … 3 bar (-100 … 300 kPa/-14.5 … 43.5 psig) ƲƲ Plastic horn antenna - Version with adapter flange from DN 100 PP or PP-GF 30 -1 … 1 bar (-100 … 100 kPa/-14.5 … 14.5 psig) ƲƲ Plastic horn antenna -1 … 2 bar (-100 … 200 kPa/-14.5 … 29.
11 Supplement Plastic horn antenna Air volume Pressure Without reflux valve With reflux valve 0.6 bar (8.7 psig) 6 m3/h 1 m3/h 0.8 bar (11.6 psig) - 2.1 m3/h 1 bar (14.5 psig) - 3 m3/h 1.2 bar (17.4 psig) - 3.5 m3/h 1.4 bar (20.3 psig) - 4.2 m3/h 1.6 bar (23.2 psig) - 4.4 m3/h 1.8 bar (20.3 psig) - 4.8 m3/h 2 bar (23.2 psig) - 5.
11 Supplement Electromechanical data - version IP 66/IP 68 (1 bar) Options of the cable entry ƲƲ Cable gland with integrated connection cable M20 x 1.5 (cable: ø 5 … 9 mm) ƲƲ Blind plug M20 x 1.5; ½ NPT ƲƲ Cable entry Connection cable ½ NPT ƲƲ Wire cross-section 0.5 mm² (AWG 20) ƲƲ Tensile strength < 1200 N (270 lbf) ƲƲ Wire resistance < 0.036 Ω/m ƲƲ Standard length 5 m (16.4 ft) ƲƲ Max. length 180 m (590.6 ft) ƲƲ Min. bending radius 25 mm (0.
11 Supplement Integrated clock Date format Day.Month.Year Time zone, factory setting CET Time format Rate deviation max. 12 h/24 h 10.5 min/year Additional output parameter - Electronics temperature Output of the temperature values ƲƲ Analogue Via the current output ƲƲ Digital Via the digital output signal - depending on the electronics version Range -40 … +85 °C (-40 … +185 °F) Resolution < 0.
11 Supplement Housing material Version IP-protection class NEMA protection Aluminium Single chamber IP 66/IP 68 (0.2 bar) Type 6P Double chamber IP 66/IP 68 (0.2 bar) Type 6P Stainless steel (electropolished) Single chamber IP 66/IP 68 (0.2 bar) Type 6P Stainless steel (precision casting) Single chamber IP 66/IP 68 (0.2 bar) Type 6P Double chamber IP 66/IP 68 (0.
11 Supplement Plastic housing ~ 69 mm (2.72") ø 79 mm (3.11") ~ 84 mm (3.31") ø 79 mm (3.11") M20x1,5/ ½ NPT 112 mm (4.41") 112 mm (4.41") M16x1,5 M20x1,5/ ½ NPT 1 2 Fig. 62: Housing versions with protection rating IP 66/IP 67 - with integrated display and adjustment module the housing is 9 mm/0.35 in higher 1 2 Single chamber version Double chamber version Aluminium housing ~ 116 mm (4.57") ~ 87 mm (3.43") ø 86 mm (3.39") ø 86 mm (3.39") 120 mm (4.72") 116 mm (4.
11 Supplement Aluminium housing with protection rating IP 66/IP 68 (1 bar) ~ 105 mm (4.13") ~ 150 mm (5.91") ø 86 mm (3.39") ø 86 mm (3.39") 120 mm (4.72") 116 mm (4.57") M16x1,5 M20x1,5 M20x1,5 M20x1,5/ ½ NPT 1 2 Fig. 64: Housing version with protection rating IP 66/IP 68 (1 bar) - with integrated display and adjustment module the housing is 9 mm/0.35 in higher 1 2 Single chamber version Double chamber version Stainless steel housing ~ 87 mm (3.43") ~ 69 mm (2.72") ø 79 mm (3.
11 Supplement Stainless steel housing with protection rating IP 66/IP 68, 1 bar ~ 93 mm (3.66") ~ 103 mm (4.06") ø 80 mm (3.15") ~ 105 mm (4.13") ø 79 mm (3.11") ø 86 mm (3.39") 120 mm (4.72") 112 mm (4.41") 117 mm (4.61") M16x1,5 M20x1,5 M20x1,5/ ½ NPT 1 M20x1,5/ ½ NPT 2 3 Fig. 66: Housing version with protection rating IP 66/IP 68 (1 bar) - with integrated display and adjustment module the housing is 9 mm/0.
11 Supplement 126 mm (4.96") ø 107 mm (4.21") ø 21 mm (0.83") 1 ø 75 mm (2.95") ø 156 mm (6.14") 14,7 mm (0.58") 2 26,5 mm (1.04") 19 mm (0.75") 10,5 mm (0.41") VEGAPULS 69, plastic horn antenna with compression flange and rinsing connection 3 ø 200 mm (7.87") Fig.
11 Supplement 36,5 mm (1.44") 17,5 mm (0.69") 1 20 mm (0.79") 144 mm (5.67") VEGAPULS 69, plastic horn antenna with adapter flange ø 75 mm (2.95") ø 98 mm (3.86") 2 ø 180 mm (7.09") ø 220 mm (8.66") Fig.
11 Supplement 170 mm (6.69") VEGAPULS 69, plastic horn antenna mit adapter flange und rinsing connection 2 34,7 mm (1.37") 20 mm (0.79") 62,5 mm (2.46") 1 ø 75 mm (2.95") 3 ø 98 mm (3.86") ø 180 mm (7.09") ø 220 mm (8.66") Fig.
11 Supplement VEGAPULS 69, plastic horn antenna with mounting strap 125 mm (4.92") ø 107 mm (4.21") 9 mm (0.35") 98 mm (3.86") ø 115 mm (4.53") 47247-EN-170315 9 mm (0.35") 85 mm (3.35") 12 mm (0.47") ø 75 mm (2.95") 6 mm (0.24") 8,5 mm (0.34") 19 mm (0.75") 170 mm (6.69") 300 mm (11.81") 2,5 mm (0.10") 12 mm (0.47") Fig.
11 Supplement VEGAPULS 69, metal-jacketed lens antenna 2 63 mm (2.48") G 1/8" x 8mm 144 mm (5.67") 11,5 mm (0.45") 1 ø 98 mm (3.86") Fig. 72: VEGAPULS 69, metal-jacketed lens antenna. Flange thickness acc. to drawing, flange dimensions acc. to DIN/ASME/JIS. 1 2 Version up to 130 °C (266 °F) Version up to 200 °C (392 °F) VEGAPULS 69, metal-jacketed lens antenna with rinsing connection 43 mm (1.69") 2 1 5 35 mm (1.38") 3 4 5 27,5 mm (1.08") 4 43 mm (1.69") 144 mm (5.67") 3 27,5 mm (1.
11 Supplement VEGAPULS 69, metal-jacketed lens antenna with swivelling holder max. 10° 2 144 mm (5.67") 63 mm (2.48") 1 Fig.
11 Supplement VEGAPULS 69, metal-jacketed lens antenna with swivelling holder and rinsing connection max. 10° 2 63 mm (2.48") 43 mm (1.69") 35 mm (1.38") 3 4 43 mm (1.69") 35 mm (1.38") 5 3 4 144 mm (5.67") 1 5 Fig.
11 Supplement 11.4 Industrial property rights VEGA product lines are global protected by industrial property rights. Further information see www.vega.com. VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte. Nähere Informationen unter www.vega.com. Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellectuelle. Pour plus d'informations, on pourra se référer au site www.vega.com.
INDEX INDEX A Adjustment 43, 48, 49 C Change the language 50 Check output signal 68 Connection ––Cable 31 Copy sensor settings 54 Current output ––Adjust 55 ––Min./Max.
47247-EN-170315 Notes VEGAPULS 69 • 4 … 20 mA/HART - two-wire 97
Notes 47247-EN-170315 98 VEGAPULS 69 • 4 … 20 mA/HART - two-wire
47247-EN-170315 Notes VEGAPULS 69 • 4 … 20 mA/HART - two-wire 99
All statements concerning scope of delivery, application, practical use and operating conditions of the sensors and processing systems correspond to the information available at the time of printing. Subject to change without prior notice © VEGA Grieshaber KG, Schiltach/Germany 2017 VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com www.vega.