Philips Healthcare 9896 002 16623 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 DoseAware Base Station Package INSTRUCTIONS FOR USE Document version 1.
Published by Philips Medical Systems Nederland B.V. Philips Medical Systems Nederland B.V. reserves the right to make changes to both this document and to the product it describes. Product specifications are subject to change without notice. Nothing contained within this document is intended as any offer, warranty, promise or contractual condition, and must not be taken as such. 0900a44780045a3a © Koninklijke Philips Electronics N.V. 2009 All Rights are reserved.
Contents 1 2 Philips Healthcare 9896 002 16623 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 Introduction.............................................................................................. 1-1 1.1 About the DoseAware System.................................................... 1-1 1.2 About these Instructions for Use................................................ 1-2 1.
3.1 Introduction.............................................................................. 3-1 3.2 Getting started........................................................................... 3.2.1 Starting the wall mounted Base Station........................ 3.2.2 Starting the MCS mounted Base Station...................... Base Station features.................................................................. 3.3.1 Online view – view current dose rates ................................................
Contents 6 7 8 9 9896 002 16623 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 Philips Healthcare 10 Using the Cradle...................................................................................... 6-1 6.1 Introduction.............................................................................. 6-1 6.2 Getting started...............................................................
10-4 10-4 10-4 10-5 10-5 10-6 9896 002 16623 0900a44780045a3a 10.7 Technical Specifications............................................................. 10.7.1 Radio communication.................................................. 10.7.2 Base Station................................................................ 10.7.3 PDM........................................................................... 10.7.4 Cradle.......................................................................... 10.
1 1.1 Introduction About the DoseAware System Philips Healthcare 9896 002 16623 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 Figure 1.1 DoseAware System overview with Base Station Package components highlighted Legend 1 Base Station 2 PDM rack 3 Ethernet 4 Radio 5 Dose Manager 6 Dose View 7 PDM 8 USB 9 Cradle DoseAware Base Station Package Document version 1.
About these Instructions for Use About these Instructions for Use 1.2 These Instructions for Use are intended to assist users in the safe and effective operation of the product described. Before attempting to operate the product, you must read these Instructions for Use, noting and strictly observing all WARNINGS and CAUTION notices. Pay special attention to all the information given and procedures described in the Safety section.
Intended use of the DoseAware System 1.3 These Instructions for Use describe the most extensive configuration of the product, with the maximum number of options and accessories. Not every function described may be available on your product. These Instructions for Use describe the Base Station Package. In order to get an immediate hands-on experience of the Base Station Package, we recommended that you interact with the Base Station, DoseView, and Cradle in parallel to reading these Instructions for Use.
Compatibility The DoseAware System may not be used as a legal staff dose recording solution. The DoseAware System is not intended for patient use. • • Do not expose the PDMs in direct X-ray beam. They are designed to be exposed to scattered radiation only. The Dose Manager, DoseView, Cradle and PDM Rack are not intended to be used inside examination rooms. The DoseAware product is not a replacement for a TLD (ThermoLuminescent Dosimeter) or similar product.
Compliance 1.5 Changes and/or additions to the product should only be carried out by Philips Medical Systems or by third parties expressly authorized by Philips Medical Systems to do so. Such changes and/or additions must comply with all applicable laws and regulations that have the force of law within the jurisdiction(s) concerned, and with best engineering practice. WARNING 1.
System requirements 1.7 System requirements DoseView has the following system requirements: • Operating systems: Windows Vista or Windows XP • .NET 3.
2 2.1 Safety Important safety directions If the DoseAware system is not functioning correct or damage is visible, inform a Philips service engineer, which will take appropriate actions in order not to harm personnel or patients. Handle the hardware and software with care. Make sure that the hardware and software is used and stored in a secured environment to prevent unauthorized access.
Electrical safety Adequate training • Do not use the product for any application until you have received adequate and proper training in its safe and effective operation. If you are unsure of your ability to operate this product safely and effectively, DO NOT USE IT. Operation of this product without proper and adequate training could lead to fatal or other serious personal injury. It could also lead to clinical misdiagnosis/clinical mistreatment.
Mechanical safety 2.3 Covers or cables should only be removed by qualified and authorized service personnel. Use this product in rooms or areas that comply with all applicable laws (or regulations having the force of law) concerning electrical safety for this type of product. Electrically isolate this product from the mains electrical supply before cleaning, disinfecting or sterilizing it. 2.3 Mechanical safety WARNING Do not remove covers from this product.
Electrostatic discharge (ESD) Fire regulations for the type of medical area being used should be fully applied, observed and enforced. Fire extinguishers should be available for both electrical and non-electrical fires. WARNING Only use extinguishers on electrical or chemical fires, which are specifically labeled for those purposes. Using water or other liquids on an electrical fire can lead to fatal or other serious personal injury.
Electromagnetic Compatibility (EMC) 2.7 ESD damage is cumulative and may not be apparent at first, as indicated by a hard failure, but can cause degraded performance. Therefore, always use proper ESD handling procedures. ESD can result from low humidity conditions, use of electrical equipment on carpeting, linens, and clothing. 2.
Electromagnetic Compatibility (EMC) This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions: (1) This device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence.
Modality specific safety 2.8 The customer or the user of the DoseAware system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the DoseAware system as recommended below, according to the maximum output power of the communications equipment: • A minimum distance of 20 cm between a PDM and a mobile phone or regular electronic device (e.g. a computer).
Network safety, security and privacy 2.9 Network safety, security and privacy Customer Role in the Product Security Partnership As with any computer-based system, protection must be provided such that firewalls and/or other security devices are in place between the medical system and any externally accessible systems. The USA Veterans Administration has developed a widely used Medical Device Isolation Architecture for this purpose.
Toxic or hazardous substances and elements 2.10 http://www.healthcare.philips.com/main/support/productsecurity Toxic or hazardous substances and elements 2.10 The following table details the toxic or hazardous substances and elements which are present in the DoseAware systems.
Equipment label overview http://www.healthcare.philips.com/main/about/Sustainability 2.11 Equipment label overview This section describes the DoseAware product labels and their locations. NOTE Some of the information (such as frequency, type, date of manufacture, and other markings) on the labels for your DoseAware product may vary from the examples shown below.
Equipment label overview Figure 2.2 2.11 Base Station label for products sold in the EU Cradle product label The Cradle product label is located on the bottom of the Cradle. Philips Healthcare 9896 002 16623 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 Figure 2.3 DoseAware Base Station Package Document version 1.
Equipment label overview PDM product label Figure 2.4 PDM label PDM product label for products sold in the EU 9896 002 16623 The following PDM product label is located on the rear side of the PDM for products sold in European Union (EU) countries. 0900a44780045a3a The following PDM product label is located on the rear side of the PDM. Figure 2.5 Safety Philips Healthcare 4 5 6 2.
3 3.1 Using the Base Station Introduction The Base Station is the display unit in the DoseAware System. It communicates wireless with PDMs within radio range in order to collect, present and store PDM dose data.
Getting started 3.2 3.2.1 Getting started Starting the wall mounted Base Station 1 Connect the output connector of the power adaptor to the back of the Base Station, if needed. 2 Connect the input connector to an AC electrical outlet socket. The Base Station screen is a touch screen. Operate the Base Station by tapping buttons and graphs on the screen. NOTE WARNINGS PDMs are shown on a first come first serve principle (see “reserved slots menu” in section “Admin Settings menu” on page 3-12).
Base Station features 3.2.2 3.3 Starting the MCS mounted Base Station 1 Switch on the X-ray system and the Base Station will automatically switch on. First a Philips start up screen will be displayed for a few seconds and after 30 seconds the system is ready to use. The Online View will display a maximum of 8 PDMs at a time. Only PDMs that are in range of the Base Station will be displayed. The Base Station screen is a touch screen. Operate the Base Station by tapping buttons and graphs on the screen.
Base Station features • A Walk-Up View (see section “Walk-Up view – view detailed dose data ” on page 3-7 ) where you can view historical dose data for a PDM chosen from the OnLine View. • A Base Station Settings View (see section “Base Station Settings view” on page 3-11 ) for administrator settings. NOTE Online view – view current dose rates 9896 002 16623 3.3.1 When the Base Station memory is full, the oldest data will be overwritten with newer data.
Base Station features 3.3 The Online View displays the current personal dose rates and peak dose rate indicators, for up to 8 PDMs within range of the Base Station. NOTE PDMs are shown on a first come first serve principle (see “reserved slots menu” in section “Admin Settings menu” on page 3-12). PDMs that are not shown in the Online View still measure and store dose data.
Base Station features The| Displayed Name| button identifies the PDMs. It may be truncated on the button due to lack of space. When a PDM moves out of range of the Base Station, its button will remain visible in the Online View for two minutes. The button will then remain grayed out for another eight minutes before it disappears. This feature allows people to temporarily leave the room without loosing their position in the list. When a button is grayed out, it is not possible to access the Walk-Up View.
Base Station features 3.3.2 3.3 Walk-Up view – view detailed dose data Figure 3.3 Personal Dose Meter Information panel Enter the Walk-Up View of a PDM by tapping the |Displayed Name| button of that PDM in the Online View. On the left hand side of the WalkUp View a Personal Dose Meter Info panel is displayed (see figure above).
Base Station features The Walk-Up View also consists of the Personal Dose Overview (see section “Personal Dose Overview” on page 3-8), which is the default WalkUp view. It has two sub views: • the Annual Personal Dose view (see section “Annual Personal Dose” on page 3-9) • the Personal Dose History view (see section “Personal Dose History” on page 3-10) 9896 002 16623 0900a44780045a3a Personal Dose Overview Figure 3.
Base Station features 3.3 last Dose Session reset, measured in Sv. Use the reset buttons (2) to reset a Dose Session to zero. Information about dose session reset will appear as an event in Dose Manager. • Total Personal Dose since the last Dose History reset. • The current day’s Personal Dose History graph. You can also access the Personal Dose History sub view (3), see section “Personal Dose History” on page 3-10.
Base Station features • The annual dose limit distributed over the full year (dashed line). As long as the accumulated dose stays below this value, the annual dose limit will not be exceeded for the full year. • Accumulated dose in relation to the annual dose limit for the current year (blue/orange line). When the accumulated dose exceeds the distributed annual dose limit, the color will change from blue to orange in order to alert the user to take actions.
Base Station features 3.3 In the Day/Week time spans you can view accumulated dose value bars. Each bar represents the accumulated dose during one/four hours, respectively. Within the Day and Week time spans, tapping a dose bar in the diagram will zoom into the larger underlying time scale (that is Week -> Day and Day -> Hour). Within each time span (Hour/Day/Week) you can step forward and backward in time with the arrow buttons on the bottom of the screen.
Base Station features 0900a44780045a3a User Settings menu The User Settings menu The User Settings menu consists of one screen. Adjust the backlight level of the screen by moving the bar horizontally. This change will take effect immediately. 9896 002 16623 Figure 3.7 Admin Settings menu The Admin Settings menu consists of four screens. Access these screens by tapping the arrow buttons in the lower right corner of each screen.
Base Station features Figure 3.8 3.3 The Base Station Name and Base Station Location menu Reserved Slots menu Philips Healthcare 9896 002 16623 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 Figure 3.9 DoseAware Base Station Package Document version 1.
Base Station features This is an optional setting, which you can use to reserve a number of slots in the Online View for the clinical roles Doctor, Nurse and Technician (see figure above). If you, for example, have reserved three slots for Doctors but the Base Station detects only one, there will be two empty slots before the first Nurse appears.
Base Station features 3.3 If you change this information, the Base Station requires a restart. NOTE Network setup is only applicable if you are using Dose Manager. Time and Date menu 9896 002 16623 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 Figure 3.11 The Time and Date menu Set time, date, time zone and manage daylight saving time for the Base Station (see figure above).
Shutting down the Base Station 3.4 3.4.1 Shutting down the Base Station Shutting down a wall mounted Base Station • Unplug the power adaptor. 3.4.2 Shutting down an MCS mounted Base Station 9896 002 16623 0900a44780045a3a • Switch off the X-ray system and the Base Station will be switched off. Philips Healthcare 4 5 6 3.
4 4.1 Using DoseView Introduction to DoseView The DoseView application lets you, for one PDM at a time: • View the PDM’s dose history. • Change the PDM options, for example full name and displayed name. Philips Healthcare 9896 002 16623 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 Figure 4.1 DoseAware Base Station Package Document version 1.
Getting started with DoseView The following items are available when you start DoseView: DoseView information 4.2 Description 1 Application menu Access the DoseView options dialog, see section “Specify password and language” on page 4-15. 2 Home toolbar Access tools to navigate in dose graphs and dose tables, see section “Home Toolbar overview” on page 4-3. 3 PDM panel Show and manage options for a PDM that is placed in a Cradle.
Home Toolbar overview 4.3 4.3 Home Toolbar overview Figure 4.2 Home toolbar The home toolbar is where you find tools to navigate in dose graphs and dose tables. The following items are available in the home toolbar: Table 4.
Home Toolbar overview Function Description Month button View dose history for the current month. Year button View dose history for the current year. Table 4.3 Advanced Navigation Group Function Description Start time button The viewed time range start day. End time button The viewed time range end day. Table 4.4 Data Filter Group — select which information is visible in the Function Description Accumulated dose check box Show/hide the accumulated dose in the graph.
PDM panel 4.4 4.4 PDM panel Philips Healthcare 9896 002 16623 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 Figure 4.3 4.4.1 PDM panel PDM information The following information is visible when a PDM is placed in the Cradle: Table 4.5 DoseAware Base Station Package PDM information Item Description Full name Full name of the person using the PDM.
Description Clinical role One of Doctor, Nurse, Technician or Other. PDM A unique PDM serial number. PDM position One of Head, Torso, Hand, Belly, Leg or Other. Battery status The PDM's battery status: • Green: normal use. • Yellow: normal use. • Red: the PDM need to be replaced in 4-6 months at normal use. • Crossed battery: there is no battery left. The PDM does not measure radiation and will not communicate with Base Stations. Dose information Table 4.6 NOTE 4.4.
4.4.4 PDM panel 4.4 Using DoseView 4-7 Change PDM options 9896 002 16623 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 PDM options dialog box 1 Philips Healthcare Figure 4.4 DoseAware Base Station Package Document version 1.
0900a44780045a3a PDM panel PDM options dialog box 2 1 Follow the instructions below to change PDM options: Make sure that a Cradle is connected to your computer’s USB port. 2 Insert a PDM in the Cradle. The computer will detect the PDM automatically and the PDM information will appear in DoseView. 3 Access the PDM options dialog by clicking the| PDM options| (see figure above). The PDM options dialog consists of two tabs as shown below.
PDM panel Philips Healthcare 4.4 PDM options 9896 002 16623 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 Figure 4.
0900a44780045a3a PDM panel PDM Options dialog box — Settings In the Settings tab you can: • Set Base Station Mode to Show/Hide, which makes the PDM to be shown/ not shown on a Base Station. • Turn Power Mode On/Off, which puts the PDM in operating or power saving mode. In operating mode, communication with Base Stations will take place and registration of dose data will occur. In power saving mode no communication with Base Stations will take place and no registration of dose data will occur.
PDM panel WARNING NOTES 4.4 Make sure that the computer’s clock is correct, otherwise the dose history data will be shifted in time and therefore not accurate anymore. • • • If you need to change hour-portion of time, the dose history will have to be reset. This is done automatically; you just need to confirm the action. When you synchronize PDM’s time the internal clock will be synchronized to the same time as the host PC, including the Windows time zone settings.
View dose graph View dose graph 9896 002 16623 0900a44780045a3a 4.5 Figure 4.8 Using DoseView Philips Healthcare 4 5 6 4.5 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 4-12 55 56 57 Dose graph with selection (vertical red dotted lines) DoseAware Base Station Package Document version 1.
View dose graph 4.5 Use the data filter panel in the home menu toolbar to select which information to be visible in the graph: • Accumulated Dose graph – blue graph • Dose Rate graph – orange graph, displays dose rate samples for every second, where such data is available • Show details • Checked: The graphs will display dose rate samples for every second, where such data is available.
View dose table View dose table 9896 002 16623 0900a44780045a3a 4.6 Figure 4.9 Dose table with selection In the Dose Table tab (see figure above) you can view a table of: • Accumulated Dose values. • Dose Rate values. The tables are covering a time span that you choose either from the: • Navigation panel. • Or the Advanced Navigation panel. Using DoseView DoseAware Base Station Package Document version 1.2 Philips Healthcare 4 5 6 4.
Specify password and language 4.7 You can also select one or several rows to be summarized in the Selection Summary field of the PDM panel. The Selection Summary field will provide information about start time, end time, time span, accumulated dose and mean dose rate. With the Show details check box checked, the table will list second data, where such data is available. 4.
Specify password and language In the Application Menu -> Program Options dialog box (see figure above) you can: • Specify a password to protect the access to the PDM options dialog. The password is only used when you are making PDM options changes. Others can still view the PDM data. • Change application language. • View the DoseView program version. • • Contact your local administrator for password guidelines.
5 5.1 Using the PDM Introduction Figure 5.1 Philips Healthcare The PDM The PDM is an active dose meter designed for maintenance-free usage throughout its lifetime. You can personalize the PDM’s appearance by attaching one of the 8 inlays of different color, which are delivered together with the PDM (see the PDM Quick Guide).
Getting started The PDM now records dose values and transmits them to Base Stations within range. You can also read out the recorded dose values by using DoseView via the Cradle (see section “Getting started with DoseView” on page 4-2). 9896 002 16623 0900a44780045a3a See the “Technical Data” chapter to learn more about PDM memory and data transfer between PDM and Base Stations. Philips Healthcare 4 5 6 5.
6 6.1 Using the Cradle Introduction Figure 6.1 The Cradle The Cradle is a dock station that lets you connect a PDM to a computer for data read out as well as PDM Options writing into the PDM. Use the Cradle in combination with the DoseView and/or Dose Manager applications.
Getting started 6.2 Getting started NOTE Do not connect the Cradle to the computer unless DoseView and the Cradle driver are installed on the computer. 1 Start the DoseView application. 2 Connect a Cradle to your computer’s USB port. 3 Put a PDM in the Cradle. 9896 002 16623 DoseView will start loading the PDM dose history. This may take up to a few minutes. You can follow the progress on the progress bar at the bottom of the DoseView window.
7 7.1 Philips Healthcare Maintenance Calibrating the Base Station screen If the touch screen does not respond correctly to user interaction, a touch screen re-calibration may be needed. 1 2 7.2 Follow the instructions below to perform a re-calibration: Tap and hold on the Philips start up screen that appears during Base Station startup until the Setup view appears. Follow the instruction on the bottom of the screen and the subsequent instructions in the next views.
Cleaning and disinfection Chrome parts should only be cleaned by rubbing down with a dry woolen cloth. Do not use abrasive polishes. To preserve the finish, use non-abrasive wax. Disinfection 7.2.2 CAUTION Do not use flammable or potentially explosive disinfecting sprays. Such sprays create vapors, which can ignite, causing fatal or other serious personal injury.
8 8.1 Troubleshooting Base Station Table 8.1 8.2 Base Station troubleshooting Problem Solution The PDM does not appear in the Online View. Use DoseView or Dose Manager to check that the PDM’s: • Battery Status is OK (see section “Change PDM options” on page 4-7). • Power Mode is set to ‘On’ (see section “Change PDM options” on page 4-7). • Base Station Mode is set to ‘Show’ (see section “Change PDM options” on page 4-7). DoseView and Cradle Table 8.
9896 002 16623 0900a44780045a3a DoseView and Cradle Philips Healthcare 4 5 6 8.2 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 8-2 55 56 57 Troubleshooting DoseAware Base Station Package Document version 1.
9 9.1 Product disposal Introduction Philips Medical Systems is concerned to help protect the natural environment, and to help ensure continued safe and effective use of this product, through proper support, maintenance and training. Therefore Philips products are designed and manufactured to comply with relevant guidelines for environmental protection. As long as the product is properly operated and maintained, it presents no environmental risks.
Final disposal of the system Existing users are strongly advised to seek advice from their local Philips Medical Systems representative before committing themselves to passing on any product. Alternatively, contact the manufacturer. 9.3 Final disposal of the system Final disposal is when the user disposes of the product in such a way that it can no longer be used for its intended purpose(s).
Fitting, removing, and disposing of batteries 9.4 For advice and information, contact your Philips Service Organization first, or otherwise the manufacturer. 9.4 Fitting, removing, and disposing of batteries NOTE Batteries harm the environment; dispose of the old batteries in an environmentally sound way. For information about disposal of the product, batteries, and hazardous materials, see the Philips Medical Systems product sustainability website at: http://www.healthcare.philips.
9896 002 16623 0900a44780045a3a Fitting, removing, and disposing of batteries Philips Healthcare 4 5 6 9.4 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 9-4 55 56 57 Product disposal DoseAware Base Station Package Document version 1.
10 10.1 Technical data PDM radiology Table 10.
PDM memory 10.2 PDM memory The PDM has two dose data memories: • The accumulated dose memory, where the PDM stores accumulated dose values every hour for the entire lifetime of the PDM. • The dose rate memory, where the PDM stores dose rate samples. When the radiation exceeds 40 µSv/h, the PDM stores one sample per second. This memory is limited to 3600 second-samples. Base Station memory The Base Station stores dose information and PDM Info for PDMs that have been connected to it.
PDM and Base Station/DoseView memories 10.5 10.5 PDM and Base Station/DoseView memories Dose rate samples that have been overwritten in the PDM’s dose rate memory may still be available in the Base Station. If there are no dose rate samples neither in the Base Station, nor in the PDM dose rate memory, the Base Station and DoseView will instead display mean dose rate values based on accumulated dose values (see figure below).
4 5 6 10.6 Time management 7 8 9 10 10.6 11 12 The PDM logs dose history in local time with no daylight saving time 13 adjustments. Daylight saving time adjustment is done in the Base Station, 14 DoseView or Dose Manager when the dose history is presented. 15 16 The following happens when the daylight saving time is changed: 17 • When going to summertime, one extra hour with no dose data will be 18 19 added to the dose log.
Technical Specifications 10.7.3 Philips Healthcare Characteristic Measure Backlight life time Approximately 50000 hours Maximum heat dissipation 25 W Network Ethernet 10/100 USB 1.1 host for USB Mass Storage Device Table 10.3 10.7.4 10.
Security and privacy requirements 10.8 Characteristic Measure Power via USB Communication with computer USB 2.0 Security and privacy requirements Antivirus An AVS solution is not installed on this Windows CE device because of the security measures taken to reduce the attack surface. TCP/UDP Port number Protocol Additional note TCP 8070 gSOAP httpd 2.