8000 SERIES OPERATION MANUAL IMPORTANT NOTICE THIS PRODUCT WILL REQUIRE AN UNLOCK CODE AFTER THE EVALUATION PERIOD HAS EXPIRED. (60 Days After Invoice Date) AFTER THE EVALUATION PERIOD HAS EXPIRED THE OPERATION OF THE PRODUCT IS LOCKED AND THE DISPLAY SHOWS A NUMBER WHICH MUST BE QUOTED TO TRANSMILLE TO RECEIVE THE UNLOCK CODE THE UNLOCK CODE IS AVAILALBLE FROM TRANSMILLE ONLY AFTER PAYMENT HAS BEEN RECEIVED. (This code is only entered once in the life of the instrument.
8000 SERIES OPERATION MANUAL DECLARATION OF CONFORMITY Manufacturer’s Name: Manufacturer’s Address: Transmille Ltd.
8000 SERIES OPERATION MANUAL TABLE OF CONTENTS 8000 SERIES MULTIMETER INTRODUCTION ....................................................................................6 MAIN FEATURES ...................................................................................................................................6 ACCURACY AND FUNCTIONALITY ...........................................................................................................7 MULTI PRODUCT MULTIMETER...................................
8000 SERIES OPERATION MANUAL USING THE 8081 – BEST MEASUREMENT PRACTICE. .................................................................36 Nulling or Zero .............................................................................................................................36 Zeroing the Voltage ranges. ........................................................................................................36 Understanding thermal EMF voltages. .......................................................
8000 SERIES OPERATION MANUAL 8000 Series Multimeter Introduction The 8000 series range of multimeters offer maximised capabilities from a highly accurate 4ppm / 9ppm advanced design. Utilising the precision electronics to their fullest extent, the 8000 Series provides high performance core functionality combined with advanced operation modes in a single instrument.
000 SERIES OPERATION MANUAL Accuracy And Functionality The 8000 Series multimeters are available in 2 accuracy grades - the 8081 / 8080 4ppm model and the 8071 of 9ppm model. The appearance of these units is the same, however the model is indicated on the front & rear panel and shown on the display on power up. Multi Product Multimeter Designed to provide an accurate cost effective portable instrument for the calibration of a wide range of signal sources including pressure, temperature and more.
8000 SERIES OPERATION MANUAL Multi Interface Support All functions and ranges of the series 8000 multimeter are fully programmable over the multiple interfaces. The support of RS232, USB and Ethernet saves the cost of fitting GPIB cards to the PC, and also allows easy connection to portable PC’s, reducing the set up time for on site calibration. GPIB is also available where use of this type of interface is already implemented. TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL Preparing the Multimeter For Use Initial Inspection After shipment the multimeter should be inspected for any signs of external damage. Should external damage be found contact the carrier immediately. Do not connect a damaged instrument to the line power as this may result in internal damage. Please keep the original box as this can be used when returning the multimeter for service and recalibration.
8000 SERIES OPERATION MANUAL Positioning the Multimeter. The multimeter can be used free standing on a bench or mounted in a standard 19” rack enclosure. The multimeter can be operated at any angle, the two front feet have tilt legs for bench operation. A 2” (5cm) space behind the instrument is also required for line and interface connections. Minimum 2” (5cm) Clearance TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL Rear Panel Connections and controls Connections on the rear panel are for Line Power via a 3 Pin IEC connector incorporating the Line fuse and on-off switch -note the mains inlet is filtered. The multiple interface connections are available including RS232, USB, GPIB and Ethernet (LAN) These interfaces are optically isolated from the multimeter intput. On 8081 units a set of rear panel terminals are fitted as standard, including voltage, current (to 1A) and a guard terminal.
8000 SERIES OPERATION MANUAL Setting and checking the Line Voltage. WARNING THE LINE POWER CORD MUST HAVE AN EARTH CONDUCTOR TO AVOID RISK OF SHOCK. THIS INSTRUMENT MUST BE CORRECTLY EARTHED. The multimeter has been designed to work from either 100-120 Volt line supply or 200 - 240 Volt line supply. Check Supply voltage as marked on the rear panel before connecting to power line. Connecting the multimeter to the wrong supply will cause internal damage to the instrument.
8000 SERIES OPERATION MANUAL Connecting to a computer RS232 Interface Connection Configuration BAUD RATE PARITY DATA BITS STOP BITS Cable Type Male to Female Serial Cable Straight Though Pin Connection (Not Null MODEM) N/A Software Driver : : : : 9600 NONE 8 1 USB Interface Connection Configuration N/A Cable Type Software Driver USB ‘A’ Type connector to USB ‘B’ Connector FTDI USB Driver (Supplied) TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL Ethernet (LAN) Interface Connection Configuration 8000 Series Menu Setup : Press MENU Press Ä until ETHERNET is displayed Press Æ *For automatic IP Address (DHCP Host required*) Select ENABLE DHCP *To set a manual IP Address Select SET IP and enter IP Address in xxx.xxx.xxx.xxx format. Cable Type Software Driver * For further information consult your network administrator.
8000 SERIES OPERATION MANUAL Powering up the multimeter After connecting line power, the multimeter can be switched on with the line power switch above the mains inlet socket on the rear panel. The front panel displays will illuminate and indicate if the instrument is in evaluation mode for a period of 5 seconds, giving the user a chance to unlock the instrument.
8000 SERIES OPERATION MANUAL Output Connections WARNING THE LINE POWER CORD MUST HAVE AN EARTH CONDUCTOR TO AVOID RISK OF SHOCK. THIS INSTRUMENT MUST BE CORRECTLY EARTHED. Input sockets are all 4mm safety type, the voltage pairs contacts are low thermal gold plated for minimum thermal EMF. Voltage / 2-Wire Resistance Current (to 1A) / 4-Wire Resistance High Current (to 30A) Guard TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL It is recommended that the voltage and low current leads be high quality screened cable with gold plated 4mm plugs fitted. The cable must be able to withstand 1025 volts AC and have an insulation resistance greater than 1 TOhm to avoid introducing any shunting effect on the high resistance ranges.
8000 SERIES OPERATION MANUAL Operation Safety Warnings WARNING : THE INFORMATION IN THIS SECTION IS INTENDED ONLY FOR QUALIFIED PERSONNEL. THE USER MUST AT ALL TIMES BE ADEQUATELY PROTECTED FROM ELECTRIC SHOCK. QUALIFIED PERSONNEL MUST ENSURE THAT OPERATORS OF THE EQUIPMENT ARE ADEQUATELY INSULATED FROM CONNECTION POINTS. A SOFT CARRY-CASE AND A HARD TRANSIT CASE ARE AVAILABLE FOR REGULAR TRANSPORTATION OF THE MULTIMETER.
8000 SERIES OPERATION MANUAL Controls & Functions To familiarise yourself with the 8000 Series multimeters, it is advised to learn a selection of the basic operations of the front panel controls before use.
8000 SERIES OPERATION MANUAL Preparing the Multimeter For Use The multimeter front panel displays and function buttons will illuminate, with the function selected remaining lit once the start up sequence is complete.
8000 SERIES OPERATION MANUAL Connections for Voltage Measurement to 1kV Connections for Low Current Measurement to 1A Connections for Low Current Measurement to 30A TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL Connections for 2 Wire Resistance Measurement Connections for 4 Wire Resistance Measurement TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL Electrometer Input : High Resistance Measurement Electrometer Input : Low Current Measurement TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL Front Panel Controls and Indicators Front Panel Sections Dual Display Input Terminals & Indicator LEDs Function Keys Range & Readback Controls Digital Control Dial Input & Menu Controls Electrometer I/O Input Terminals & Indicator LEDs Voltage (1025V) 2 Wire Ohms 4 Wire Ohms Sense Low Current (1A) 4 Wire Ohms Current High Current (30A) 4 Wire Ohms Current Guard Input Voltage / 2 Wire Resistance Terminal LED Indicator Low Current / 4 Wire Resistance Terminal LED Indicat
8000 SERIES OPERATION MANUAL Function Keys DCV ACV ACDCV Ω 2WIRE FREQ DCI ACI ACDCI Ω 4WIRE USER DC Voltage AC Voltage ACDC Voltage (Coupled) 2 Wire Resistance Frequency DC Current AC Current ACDC Current (Coupled) 4 Wire Resistance User Defined Function Range & Readback Controls RANGE UP RANGE DOWN AUTO RANGE CONFIG TRIG DIGITS NULL FILTER Select Next Range Up Select Next Range Down Set Auto Range Mode Enter Configuration Menu Select Trigger Setting Select Number of Digits Set Null Point Select Filter
8000 SERIES OPERATION MANUAL Menu Dial Multi Purpose Use – Context Specific : Range Up / Down Menu Navigation Menu Item Selection Numeric Increment / Decrement Electrometer I/O Terminals Vout IMeas Voltage output for Electrometer high resistance measurement function Current measurement input for Electrometer function Front panel Keyboard The front panel of the 8000 Series Multimeter utilises a high quality custom rubber keyboard with tactile feel buttons and integral dual display window.
8000 SERIES OPERATION MANUAL Dual Display Main Display Area [Measurement Data] Secondary Data Display [Additional Information / Settings Data] The 8000 Series multimeters incorporate clear, bright dual vacuum fluorescent displays to present the user with a large amount of information in an easy to read configuration. The main display shows the reading, function selected and various indicators depending on the mode of operation (for example, an ‘s’ to indicate a sample being taken).
8000 SERIES OPERATION MANUAL Operating the multimeter Selecting a Range Range selection can be set to automatic AUTORANGE mode or a manual range can be set using MANUAL RANGING. If the input signal is greater than the range full scale the multimeter will indicate overrange For frequency measurements, ranging is automatic. TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL Settings Menu Map TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL Settings Menu Descriptions Menu Item Description of Use Set reading and key beep modes on or off View / Set current date and time Set decimal separator type (.
8000 SERIES OPERATION MANUAL Configuration Menu Map TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL Configuration Menu Descriptions Menu Item Description of Use Displays function, range and terminal settings.
8000 SERIES OPERATION MANUAL User Menu Map (Part 1) TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL User Menu Map(Part 2) TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL User Menu Descriptions Menu Item Description of Use Configure probe inputs for PRT measurement Rear input configuration available with Rear Terminal Option only Configure pressure modules for pressure measurement Configure thermocouple type and cold junction compensation for thermocouple measurement Configure dual input voltage and current input mode for shunt measurement Configure dual input voltage and current input mode for DC power measurement Select electrometer low curr
8000 SERIES OPERATION MANUAL Using the 8081 – Best measurement practice. The 8000 series DMM offers several easy to use features to help the user make high accuracy measurements. The following information gives advice on using these features and how to avoid common errors. Nulling or Zero It is always good measurement practice on DC Voltage, DC Current and Resistance ranges to ‘zero’ the range first before making a reading.
8000 SERIES OPERATION MANUAL Understanding thermal EMF voltages. Thermal EMF voltages are generated wherever there is a metal to metal junction, an electrical connection between two different metals. The voltage generated is dependant on the metals and the temperature of the junction. Voltages up to many 10’s of micro volts can easily be generated by a simple banana plug connection, each connection being effectively a thermocouple.
8000 SERIES OPERATION MANUAL Zeroing Current Ranges Current ranges are easily zeroed by simply open circuiting the terminals and nulling the range CURRENT TERMINALS LEFT OPEN CIRCUIT FOR CURRENT RANGE NULL Diagram : Current Null using open circuit terminals 2-Wire null method (all ranges) [Except Electrometer] The recommended method is to short together the ‘measurement’ end of the test leads and zeroing the DMM. Using this technique the resistance of the test leads will also be zeroed out.
8000 SERIES OPERATION MANUAL 8000 Series Functions DC Voltage Press DCV to enter the DCV measurement mode – the status display will indicate the settings and range information. Status Display Section MANUAL Description Display / Selection of DC Voltage range – Use menu dial or RANGE UP RANGE DOWN buttons to change range.
8000 SERIES OPERATION MANUAL AC Voltage Press ACV to enter the ACV measurement mode – the status display will indicate the settings and range information. Press FREQ to enter the FREQUENCY measurement mode (1Hz resolution) : ACV and FREQ will both illuminate. Press FREQ to switch back to ACV display Status Display Section MANUAL Description Display / Selection of AC Voltage range – Use menu dial or RANGE UP RANGE DOWN buttons to change range.
8000 SERIES OPERATION MANUAL DC Current Press DCI to enter the DCI measurement mode – the status display will indicate the settings and range information. Status Display Section MANUAL Description Display / Selection of DC Current range – Use menu dial or RANGE UP RANGE DOWN buttons to change range. Selected range indicated by a * Ranging mode : AUTO or MANUAL FRONT FRONT or REAR terminals in use (Rear to 1A Max.
8000 SERIES OPERATION MANUAL DC Current (Electrometer Mode) (Option) Press DCI to enter the DCI measurement mode – the status display will indicate the settings and range information. Selecting ranges 10nA , 100nA, 1uA, 10uA will activate the electrometer mode. Inputs are measured from the IMeas BNC terminal on the right hand side of the multimeter. An LED indicates the BNC terminal set is active.
8000 SERIES OPERATION MANUAL AC Current Press ACI to enter the ACI measurement mode – the status display will indicate the settings and range information. Press FREQ to enter the FREQUENCY measurement mode (1Hz resolution) : ACV and FREQ will both illuminate. Press FREQ to switch back to ACV display Status Display Section MANUAL Description Display / Selection of AC Current range – Use menu dial or RANGE UP RANGE DOWN buttons to change range.
8000 SERIES OPERATION MANUAL 2-Wire Resistance Press Ω 2WIRE to enter the Ω 2WIRE measurement mode – the status display will indicate the settings and range information. Status Display Section X RANGE eg. *10m Ω DC Description Display / Selection of 2-Wire Resistance range – Use menu dial or RANGE UP RANGE DOWN buttons to change range.
8000 SERIES OPERATION MANUAL 2-Wire Resistance (Electrometer Mode) (Option) Press Ω 2WIRE to enter the Ω 2WIRE measurement mode – the status display will indicate the settings and range information. Selecting ranges above 10 MOhms will activate electrometer mode - this is displayed in terms of current (100uA, 10uA, 1uA, 100nA, 10nA). Inputs are measured from the IMeas BNC terminal on the right hand side of the multimeter. An LED indicates the BNC terminal set is active.
8000 SERIES OPERATION MANUAL 4-Wire Resistance Press Ω 4WIRE to enter the Ω 4WIRE measurement mode – the status display will indicate the settings and range information. Status Display Section X RANGE eg. *1Ω DC Description Display / Selection of 4-Wire Resistance range – Use menu dial or RANGE UP RANGE DOWN buttons to change range.
8000 SERIES OPERATION MANUAL Frequency Press FREQ to enter the FREQUENCY measurement mode (.1Hz resolution) : ACV and FREQ will both illuminate. Status Display Section X RANGE eg. *1V Description Display / Selection of AC Voltage range – Use menu dial or RANGE UP RANGE DOWN buttons to change range.
8000 SERIES OPERATION MANUAL Thermocouple Measurement : Connection Diagrams (Option) Configuration A Process control calibrator connection. Configuration B (Requires Rear Terminal Option) Process Control Calibrator / Thermocouple Connection with Cold Junction Compensation using external PRT probe in bath. TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL Thermocouple Measurement : Multimeter Setup (Option) Press SHIFT then DCV to select thermocouple measurement.
8000 SERIES OPERATION MANUAL PRT Measurement : Connection Diagrams (Option) Configuration A Single probe measurement Configuration B Dual probe configuration with ratio comparison of known and unknown probes TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL PRT Measurement : Multimeter Setup (Option) Press SHIFT then Ω 4WIRE to select PRT measurement.
8000 SERIES OPERATION MANUAL Guide to setting up a PRT probe 1. Use digital control dial to choose Probe Configuration Ä PRT Coefficients Ä Primary Display Ä PRT Coefficients 2. Select Linearisation / Coefficient location ITS-90 Front ITS-90 Rear IEC 751 Front IEC 751 Rear 3.
8000 SERIES OPERATION MANUAL Pressure Measurement : Connection Diagram (Option) Pressure module connection TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL Pressure Measurement : Multimeter Setup (Option) Press SHIFT then USER to select PRESSURE measurement. Select SELECT TRANSDUCER TPMxx Select UNIT BAR PSI Pa mmHg cmH20 Atm kg/sq.cm oz/sq.
8000 SERIES OPERATION MANUAL Shunt Measurement : Connection Diagram (Option) Example shunt measurement configuration using 3000 Series multi product calibrator as current source TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL Shunt Measurement : Multimeter Setup (Option) The precision shunt measurement function allows accurate measurement of current shunts by monitoring the current supplied to the shunt and voltage measured across the shunt. By monitoring both the current and voltage the 8000 Series can provide an accurate measurement of the current shunt connected.
8000 SERIES OPERATION MANUAL DC Power / Dual Input V/30A (Option) The DC Power / Dual Input V/30A function allows the 8000 Series to automatically switch between the voltage and current inputs and display a calculated value in Watts.
8000 SERIES OPERATION MANUAL Using the Math Functions Status Display Section CLEAR NULL Description Clear stored null (zero reference point) NULL Store latest input value as zero reference point Offset measured input using entered offset value MX+C Offset is the addition value (C) in the MX+C formula OFFSET eg. If C=0.1, then a 10V input reads 10.1V eg.
8000 SERIES OPERATION MANUAL Remote Programming WARNING THE 8000 SERIES MULTIMETERS CAN MEASURE HIGH VOLTAGES UP TO 1025V AND CURRENT UP TO 30A AND MUST BE PROGRAMMED WITH DUE CAUTION. ANY PROGRAMS SHOULD BE EXTENSIVELY TESTED TO MAINTAIN SAFE OPERATION AND INCLUDE SAFEGUARD’S SUCH AS ERROR CATCHMENT AND HANDLING TO ENSURE THAT ANY COMMANDS SENT TO THE MULTIMETER PERFORM AS EXPECTED AND ANY THAT DO NOT ARE SAFELY HANDLED TO ENSURE USER SAFETY.
8000 SERIES OPERATION MANUAL Programming Commands Overview The 8000 series is controlled by a set of simple high level commands which can be used either individually or as part of a command sequence to setup the 8000 Series multimeter to the required state. The commands can be joined together using the / (forward slash) character.
8000 SERIES OPERATION MANUAL SCPI Command words are separated by a colon. A space is required between the command words and the parameters. Command words are not case sensitive.
8000 SERIES OPERATION MANUAL General Command Transmit Next Reading Transmit Currently Displayed Reading Use Rear Terminals Use Front Terminals Set Range Set Filter Range Dynamic Filter Format r f R F FD Set Digits Show Status Set Null Clear Null Display Date Version Information Internal Temperature Thermocouple D s n N a v TEMP? RANGE:THERMOCOUPLE: where TC_TYPE = {B,E,J,K,N,R,S,T} PRT (ITS 90 Linearisation) RANGE:PRT:ITS
8000 SERIES OPERATION MANUAL Setup Command Display Brightness Format B Enter Password k Set Password Leave Password-protected mode Set GPIB Address K k0 Set Date (Requires System Password or Unlocked Instrument) TRANSMILLE LTD.
8000 SERIES OPERATION MANUAL Calibration Command Format Parameters Enter Password k Set Password K Password = 1 to 10 digits Requires calibration mode to be active If a command includes a value which cannot be set due to, for example, the value being higher than the range maximum, the multimeter will reject the command and stay set as it is (the multimeter will return a *9 code).
8000 SERIES OPERATION MANUAL Calibration Calibration Overview To adjust the 8000 Series multimeter the multimeter can either be connected to a computer via the available interfaces. Calibration constants stored within the multimeter can then be adjusted using the built-in menus via the 8000 Series front panel. To prevent unauthorised access to calibration menus, a password is required before access is granted. Adjustment can be completed without disassembly of the multimeter.
8000 SERIES OPERATION MANUAL The calibration menus will change depending on the selected function : Function Menus Available Notes DC Voltage DC Current Calibration Zero Calibration +ve Full Scale Calibration -ve Full Scale Value Reset Set zero calibration point Set +FS calibration point Set -FS calibration point Calibrate FS to set value Reset range to default points AC Voltage AC Current Calibration 20% FS Calibration +ve Full Scale Value Reset Set 20% FS calibration
8000 SERIES OPERATION MANUAL Example 1 : DC Voltage 10V Range Calibration Point Actions DC Voltage ZERO Apply low thermal shorting link Select Calibration Zero Press ENTER DC Voltage +ve Full Scale Apply +10V signal Select +ve Full Scale Press ENTER DC Voltage -ve Full Scale Apply -10V signal Select -ve Full Scale Press ENTER Example 2 : AC Voltage 10V Range Calibration Point Actions AC Voltage 20% FS Apply 20% FS signal (2V) Select 20% Full Scale Press ENTER Frequency Points Frequency Respons
8000 SERIES OPERATION MANUAL General Maintenance WARNING The information in this section is intended only for qualified personnel. The user must at all times be adequately protected from electric shock. The 8000 series multimeters maintenance requirements are listed below. Please note that the multimeter does not require any regular internal servicing or adjustment.
8000 SERIES OPERATION MANUAL Guarantee and service Transmille Ltd. guarantees this instrument to be free from defects under normal use and service for a period of 1 year from purchase. This guarantee applies only to the original purchaser and does not cover fuses, or any instrument which, in Transmille’s opinion, has been modified, misused or subjected to abnormal handling or operating conditions.
8000 SERIES OPERATION MANUAL Transmille Ltd. Unit 4, Select Business Centre Lodge Road Staplehurst Kent. TN12 0QW United Kingdom. Tel Fax : +44 0 1580 890700 : +44 0 1580 890711 Email : sales@transmille.com Web : www.transmille.com 8000 Series Fax Back Form Your 8000 Series Precision Multimeter is fitted with a security system which requires a security code to be entered to allow continued operation of the unit beyond the 65 Day evaluation period.