Alpha COND 560 Controller mS / µS / ºC / ºF
R AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher.
Preface This manual serves to explain the use of the Alpha COND 560 Controller. The manual functions in two ways, firstly as a step by step guide to help the user operate the instrument. Secondly, it serves as a handy reference guide. This instruction manual is written to cover as many anticipated applications of the Alpha COND 560 Controller. If you have doubts in the use of the instrument, please do not hesitate to contact your nearest Alpha Authorised Distributor.
TABLE OF CONTENTS 1 1.1 1.2 1.3 1.4 INTRODUCTION 1 Before You Begin ................................................................................... 1 Intended Use ......................................................................................... 1 Safety Instructions ................................................................................. 2 Taking Out of Service / Correct Disposal of the Unit .............................. 2 2 2.1 2.2 2.3 2.4 2.
10.3 10.4 Appendix 3 General Instructions Concerning Controller Setting ........... 40 Appendix 4 - Abbreviations Used in LCD ...............................................
1 INTRODUCTION 1.1 Before You Begin Thank you for purchasing the Alpha COND 560 Controller. The construction of the Alpha COND 560 Controller employs leading edge technology and complies with safety regulations currently in force. Notwithstanding this, improper use could lead to hazards for the user or a third-party, and/or adverse effects on the plant or other equipment.
1.3 Safety Instructions • The Alpha COND 560 Controller should be installed and operated only by personnel familiar with the instrument and who are qualified for such work. • A defective Conductivity Controller must neither be installed nor put into service. • The Alpha COND 560 must only be operated under the specified operating conditions (see section 6). • The Alpha COND 560 must not be repaired by the customer. • No modifications to the Alpha COND 560 are allowed.
2 GETTING STARTED 2.1 Description of Instrument The Eutech Instruments’ Alpha CON 560 Controller is used for measuring conductivity and temperature values. The controller can be used for applications such as water treatment and controlling, galvanic-decontamination, chemical processing, food processing, clean or wastewater control and neutralization processes.
2.
2.3 2.3.1 Connecting Peripherals Connection Terminals Remove Back Cover: Remove the screws from the four corners at the back of the conductivity controller. Remove the back cover. The connectors are exposed on the back PCBA as shown in the Figure 1 below. Connectors: • J11 – 24V DC power connector • J8 – 9V DC power connector (1.2mm diameter) • J9 – Relay connections • J10 - Conductivity electrode & Temperature probe connections (wiring has to be done in the detachable connector J11 Connections 1.
2.3.2 Switching Between PT100 & PT1000 Temperature Sensors The controller supports both Pt100 & Pt1000 (2-wire or 3-wire) temperature sensors. The default factory setting is Pt100. If you need to use Pt1000 temperature sensor, you have to change the jumper setting (J7) as described below. Remove Back Cover: Remove screws from the four corners at the back of the conductivity controller. Remove the back cover. Remove Back PCBA: Remove the screw located center of the back PCBA (Figure 1).
2.3.3 Connecting Conductivity Electrode Any industrial conductivity probe (2-cell or 4-cell) can be connected to the conductivity controller. 2-Cell Conductivity Probe: 1. Short Pin 1 & 2 of J10 connector using a jumper wire. 2. Connect CON low input wire (white) to Pin 2 of J10 connector. 3. Connect CON high input wire (black) to Pin 3 of J10 connector. 4. Short Pin 3 & 4 of J10 connector using a jumper wire. 4-Cell Conductivity Probe: 5.
2.4 2.4.1 Installation Mechanical Dimensions Alpha COND 560 Alpha pH 600 ESC CAL ENT PH/ORP RF Transmitter Conductivity Monitor 2.4.
2.4.3 Panel Mount 1 Prepare panel cut-out of 92.0 mm X 92.
2.5 Display & Keypad 2.5.1 Display Overview The Liquid Crystal Display (LCD) of Alpha COND 560 conductivity controller has two alpha-numerical displays (Upper and a Lower). • Upper display: Measured conductivity value is displayed when the controller is in normal operation (measurement) mode. • Lower display: Measured temperature value is displayed when the controller is in normal operation (measurement) mode. In conductivity calibration mode, measured conductivity values are displayed here.
Mode Indicators MEAS Measurement mode (Refer Section 3.1 ) SETUP Setup mode (Refer Section 4.1) CAL Calibration mode (Refer Section 4) Status Annunciators READY Appears when the reading is stable HOLD Appears in Setup mode and Calibration mode to indicate that the relay function is frozen ATC Appears when Automatic Temperature Compensation (ATC) is enabled. Not visible when Manual Temperature Compensation (MTC) is enabled. Flashes if the temperature probe is faulty in its ATC mode.
2.5.2 Key Functions Alpha COND 560 lpha CON560 ESC CAL ENT Conductivity Monitor Key CAL Description Enter Calibration mode Enter Setup mode ENT Access sub screens (parameters) within a group of settings in Setup mode Confirm (save) setup parameters and numerical values Start/Confirm calibration in Calibration mode. Select a group of settings in Setup mode. Select parameters and increment/decrement numerical values in Setup and Calibration modes.
3 OPERATION 3.1 Measurement mode When the conductivity controller is powered on, the display shows all the LCD segments briefly, and then automatically enters into the measurement mode. MEAS µS READY 25.0 °C ATC The mode indicator ‘MEAS’ at the top of the display indicates that the conductivity controller is in measurement mode. The upper alpha-numerical display shows the measured conductivity value, while the lower display shows the temperature value.
3.
4 4.1 CALIBRATION MODE About Calibration Calibration should be carried out each time a new electrode is attached to the controller or when you suspect that the controller/electrode is out of calibration. The controller allows you to perform temperature calibration and conductivity calibration. About Temperature Calibration It is important to ensure that temperature calibration is carried out prior to conductivity calibration since temperature readings affect the accuracy of conductivity measurements.
4.2 Calibration The controller requires one-point calibration. Conductivity calibration is always carried out in the specific measuring range selected for the controller. Make sure, the controller is set to desired measuring range before you enter calibration mode (Refer Section 5.6 for selecting measuring ranges). Select a suitable standard conductivity solution so that its conductivity value is within 10% to 100% of the full scale of the selected measuring range of the controller.
uncalibrated (default) reading of the solution. The buffer annunciator appears in LCD. The ‘READY’ annunciator appears when the reading is stable. Once the reading stabilizes press the ▲ or ▼ key to adjust the upper display value to that of the standard solution. Press the ENT key to accept the value. NOTES: The acceptable calibration window is ±40% of the displayed (default) value. For example, if the display is 1000 µS, the values to which it can be adjusted is 600 to 1400 µS.
5 5.1 SETUP MODE Enter Setup mode The setup mode allows you to customize the settings of the conductivity controller to suite your requirements. While in measurement mode, press the ENT key to access setup mode. LCD shows ‘SETUP’ mode indicator and the first page of setup (rAng – Range settings). Press ▲ or ▼ key to access other pages of the setup mode.
5.2 Temperature Coefficient Settings Setting the temperature coefficient to match your process/calibration liquids allows controller to achieve conductivity readings with higher accuracy. 1 SETUP HOLD 4 3 2 ENT SETUP SETUP ENT SETUP % HOLD HOLD ENT % ENT HOLD 2 SETUP ENT HOLD 1 2 3 4 From measurement mode press ENT key to enter setup mode as described in section 5.1. The LCD shows the first screen of setup mode (rAng).
5.3 Temperature Settings Automatic Temperature Compensation (ATC): The conductivity value of a sample is affected by temperature. Use ATC feature of the controller to compensate for conductivity changes when the temperature of the sample or process liquid fluctuates. Connect a separate temperature probe (Pt100/pt1000) to the controller if an integrated conductivity/Temperature probe is not available.
Press ▲ or ▼ key to enable (ATC On) or disable (ATC OFF) automatic temperature compensation. Press ENT key to confirm your selection. If ATC enabled (ATC On): 4 Selecting temperature offset: The upper display shows the last configured temperature offset value (if any), otherwise the default is zero. The lower display shows currently measured temperature reading (including last configured offset value). LCD shows ‘ATC’ annunciator in lower-right corner.
5.4 Relay 1/Relay 2 Settings NOTE: If controller type is set to ‘OFF’ (Refer section 5.5 for controller settings), the ‘Relay 1/Relay 2 Setting’ is not available. The SP1 settings determine the operating parameters for Relay 1; while SP2 settings define the operating parameters for Relay 2. Since these two settings have identical setup parameters, only SP1 is described here. Note that when you change measuring range of the controller, the values you have set for SP1 & SP2 will not be affected.
NOTE: 4 This parameter lets you choose the relay function. Select ‘LO’ to activate the relay when the conductivity reading undershoots the set point (SP1) or select ‘HI’ to activate the relay when the conductivity reading overshoots the set point (SP1). SP1 and SP2 can be selected as ‘LO/LO’, ‘LO/HI’, ‘HI/LO’, or ‘HI/HI’. Setting hysteresis value: The lower display shows ‘HYS’. The upper display shows the last configured value, otherwise the default value 20 µS.
5.5 Controller Settings The controller settings allow you to set the relay type and relay status of the controller. 1 SETUP 2 ENT SETUP ENT HOLD HOLD 3 2 SETUP ENT HOLD SETUP ENT HOLD 3 SETUP ENT HOLD 1 From measurement mode press ENT key to enter setup mode as described in section 5.1. The LCD shows the first screen of setup mode (rAng). Press ▲ or ▼ key to select controller settings screen (Cntr). Press ENT key to access controller settings (Cntr).
If Controller type was set to Off (OFF): The controller reverts to Cntr screen. If Controller type was set to On (L.Ct): 3 Select the relay status under Non-Alarm condition: The lower display shows ‘rEL’. The upper display shows the last configured relay status. (‘dEEn’ or ‘En’). Press ▲ or ▼ key to select the desired relay status. dEEn: De-energized En: Energized Press ENT key to confirm your selection. The controller reverts to Cntr screen.
5.6 Measuring Range Settings The conductivity controller has 6 different measuring ranges. You need to select a range which is the most suitable for conductivity reading of your applications/samples. The factory default is 2000 µS (0 to 2000 µS/cm). The measuring range you select here will affect the way the controller behaves in the measurement mode and calibration mode.
5.7 Configuration Settings Configuration settings let you to enable/disable backlight of the LCD, select between 2-cell & 4-cell and reset the controller to factory defaults. 1 SETUP SETUP SETUP SETUP ENT R. ENT HOLD HOLD HOLD 4 3 2 ENT 2 This screen appears only when 2-cell is selected and measuring range is set to 0.0 to 200.
of the LCD. Press ENT key to confirm your selection. 3 Select cell type of the sensor: The lower display shows ‘CEL’. The upper display shows the last configured cell type (‘2’ or ‘4’). Press ▲ or ▼ key to select the cell type of your conductivity probe. Press ENT key to confirm your selection. NOTE: If you use a 2-cell conductivity sensor, make sure that the pin 1 and 2 are shorted and pin 3 and 4 are shorted on the J10 connector.
Press ▲ or ▼ key to access other setup screens or press ▲ and ▼ key simultaneously (escape) to return to measurement mode. To exit from any intermediate steps, press ▲ and ▼ keys simultaneously (escape). The controller returns to the first screen configuration settings (Cnfg).
5.8 Viewing Electrode Properties Each time you calibrate the conductivity sensor, the controller re-calculates the calibration factor of the sensor and shown in the LCD at the end of calibration. The setup mode allows you to view the last saved calibration factor and cell constant at any time. 2 1 SETUP HOLD 1 2 3 ENT SETUP ENT HOLD SETUP ENT HOLD From measurement mode press ENT key to enter setup mode as described in section 5.1. The LCD shows the first screen of setup mode (rAng).
6 TECHNICAL SPECIFICATIONS General Specification Conductivity Measuring Range: 0.000 - 2.000 µS/cm 0.00 - 20.00 µS/cm 0.0 - 200.0 µS/cm 0- 2000 µS/cm 0.00 - 20.00 mS/cm 0.0 - 200.
Electrical Data and Connections Conductivity Input Connection terminal Screw Terminal (3.5mm pitch) 8-pin, 2-pin & 5-pin terminal blocks, DC power connector (1.2mm diameter) EMC Specifications Emission Susceptibility EN 61 36 EN 61326 Power Input Universal +9 V DC Adaptor and 24V DC Supply 96mm x 96mm x 66 mm 210g 0 to 40 oC 80% up to 31 oC decreasing linearly to 50% at 40oC Other Dimensions (W x H x D) Weight (Estimated) Ambient Temp.
Hysteresis On-delay time Off-delay time Relay 2 (SP2) Set point value Relay function Hysteresis On-delay time Off-delay time 20 µS 0 seconds 0 seconds 1900 µS High 20 µS 0 seconds 0 seconds Display Backlight On 33
7 7.1 LIST OF ACCESSORIES Thermo Scientific Order Codes Conductivity Controller Replacement and Accessories Item Description Order Code Alpha CON 560 Controller Conductivity cell, Epoxy body, Graphite sensor, w/3-wire Pt100, k = 0.1 Conductivity cell, Epoxy body, Graphite sensor, w/3-wire Pt100, k = 1.0 Conductivity 2 Cell type probe, 0.1 - 200μS; Cell constant, K=0.1 with integrated Pt 100, Material SS316 and 25ft cable (open-ended) Conductivity 2 Cell type probe, up to 200 mS; Cell constant, K=1.
8 TROUBLESHOOTING Problem Cause Solution Power on, but no display a) b) Loose connections Incorrect output voltage of the power adaptor Cables not in correct polarity (+ and – position) a) Air bubbles in the probe Dirty probe Probe not deep enough in sample a) c) Unstable conductivity reading a) b) c) b) c) b) c) Electrical noise interference Tap probe to remove air bubbles Clean probe and recalibrate Ensure that the sample entirely covers the sensor of the probe Ensure shield wire is prope
9 9.1 GENERAL INFORMATION Warranty This Controller is supplied with a one-year warranty against significant deviations in material and workmanship from date of purchase and a six-month warranty for probe. Each instrument will have a warranty card with a specific serial number. The warranty card must be endorsed by the Authorized Distributor at the point of sale.
protection. Include a brief description of any faults suspected for the convenience of Customer Service Dept., if possible. 9.4 Maintenance and Cleaning Maintenance The Alpha COND 560 Controller contains no user repairable components. Please contact your authorized distributor if there is any problem with the unit. Cleaning To remove dust, dirt and spots, the external surfaces of the Conductivity Controller may be wiped with a damp, lint-free cloth. A mild household cleaner can also be used if necessary.
10 APPENDICES 10.1 Appendix 2 Conductivity of various aqueous solutions At Temperature 25 oC / 77 oF Solution Conductivity Resistivity 0.055 µS/cm 18.18 MΩ-cm 0.05 - 1 µS/cm 1 - 18 MΩ-cm 0.5 µS/cm 2 MΩ-cm 0.1 - 10 µS/cm 0.1 - 10 MΩ-cm 1 - 80 µS/cm 0.01 - 1 MΩ-cm Mountain Water 10 µS/cm 0.1 MΩ-cm Drinking Water 0.5 - 1 mS/cm 1 - 2 kΩ-cm Waste-water 0.9 - 9 mS/cm 0.1 - 1 kΩ-cm 1.5 mS/cm 0.7 kΩ-cm Brackish Water 1 - 80 mS/cm 0.
10.2 Appendix 2 Graphical Representation of the Function of Hysteresis SP1 = 100 µS & Set to LO SP2 = 1900 µS & Set to HI HYSTERESIS BAND Default = 20 µS HYSTERESIS BAND Default = 20 µS RELAY ON µS 100 120 RELAY OFF 1880 1900 SP2 SP1 Forward Direction Reverse Direction The relay of the conductivity controller activates when the set-point is reached (forward direction). In the reverse direction, it does not de-activate when the value reaches the set-point.
10.
10.4 Appendix 4 - Abbreviations Used in LCD Abbreviation Description AtC Automatic Temperature Compensation CAL Calibration mode C.ºC Calibration temperature CdAt Electrode properties CEL Cell CnFg Configuration Cntr Controller C.tC Temperature coefficient for calibration liquid dEF Default values dEEn De-energized En Energized FCt Factory (defaults) HI High limit HOLD Relay functions are hold HyS Hysteresis K Cell constant L.Ad Linear resistance L.
Set Setting SETUP Setup mode SP1 1st Set point SP2 2nd Set point tC Temperature Coefficient tyP Type Ur Reading is under range 42
Water Analysis Instruments North America 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031 Europe P.O. Box 254, 3860 AG Nijkerk Wallerstraat 125K, 3862 CN Nijkerk, Netherlands Tel: (31) 033-2463887 Fax: (31) 033-2460832 Asia Pacific Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949 Tel: 65-6778-6876 Fax: 65-6773-0836 www.thermo.com/process © 2009 Thermo Fisher Scientific Inc. All rights reserved.