testo Saveris Inbound Cold Chain data logger 0572 2800 Instruction manual
Content Content 1 About this document ............................................................................ 3 1.1 Use .........................................................................................................3 1.2 Icons ......................................................................................................3 2 Safety and waste disposal ................................................................... 4 2.1 Intended purpose ...........................................
1 About this document 1 About this document Please read this instruction manual through carefully and spelling yourself with the product before use. 1.1 Use The instruction manual is an integral part of the instrument. Pay particular attention to the safety instructions and warning notices in order to prevent injuries and damage to the product. Keep this documentation on hand so that you can refer to it when necessary. Always use the complete original instruction manual.
2 Safety and waste disposal 2 Safety and waste disposal 2.1 Intended purpose The testo Saveris Food Logistics Inbound data logger is used to store and read out series of measurements during the transportation of products that are subject to cold chain requirements. . Configuration is customer-specific and is carried out during production at Testo. No subsequent modification of the configuration is possible. The data logger is designed for single use only.
3 Description of the instrument 2.4 Waste disposal Dispose of faulty rechargeable batteries and spent batteries in accordance with the valid legal specifications. At the end of its useful life, dispose of the instrument via separate collection for electrical and electronic devices. Please observe local regulations regarding waste disposal, or alternatively return the product to Testo for disposal. WEEE Reg. No. DE 75334352 3 Description of the instrument 3.
4 Product use 3.3 Operating elements of the product Element 2 Start/Check button 3 LED (red) 5 Serial number/barcode (integrated) 4 LED (green) 6 Order number and configuration 7 Refer to instruction manual 4 Product use 4.1 Commissioning 1 6 Element 1 Serial number/barcode (removable) Switch the logger on.
4 Product use 2 The logger’s green and red LEDs flash briefly to acknowledge successful activation. 3 LED flashes green: system test successful Data logger starts recording reading LED flashes red: system test unsuccessful Data logger must not be used. Each time the Start/Check button is pressed again, this produces a status display.
4 Product use 5 Enclose data logger with the shipment. 8 6 Optional: attach location tape to the outside of the packaging. 7 Optional: attach location sticker to the packaging so that it is clearly visible from the outside.
4 Product use 4.2 Reading out data 1 Remove the data logger and immediately press the Start/Check button. While the button is pressed, the data logger indicates whether the limit values were adhered to/violated. - LED lights up green: limit values adhered to - LED lights up red: limit values violated CAUTION If too much time elapses between removing the data logger and pressing the button, this can lead to a limit value violation due to a deviating ambient temperature.
4 Product use If the connection for the data transfer is interrupted, the red LED flashes. The data logger attempts to establish the connection again (step 8). 4 Any additional press of the button indicates whether the limit values were adhered to or violated. 5 The used data loggers can be assembled in the box provided and returned to Testo. Please clean the loggers before assembling them in the box. 4.
4 Product use Status LED Cause After being switched on, the data logger flashes red at one-second intervals for a total of 10 seconds. The LED lights up green while the button is pressed. The LED lights up red while the button is pressed. The data logger flashes red for 10 seconds during/after the data transfer. An error occurred during the system test. Data logger must not be used! Green and red LEDs flash simultaneously at one-second intervals after removal and pressing the Start/Check button.
5 Technical data 5 Technical data Feature Value Measuring range Accuracy Resolution Sensor type -30 °C to +50 °C / -22 °F to +122 °F ± 0.5°C/°F 0.
6 Authorizations 6 Authorizations The use of the wireless module is subject to the regulations and stipulations of the respective country of use, and the module may only be used in each case in countries for which a country certification has been granted. The user and every owner undertake to adhere to these regulations and prerequisites for use, and acknowledge that the re-sale, export, import, etc. in particular in, to or from countries without wireless permits, is their responsibility. Product Mat.-No.
6 Authorizations Europa + EFTA The EU Declaration of Conformity can be found on the testo homepage www.testo.com under the product specific downloads.
6 Authorizations IC Warning RSS-Gen & RSS-247 statement: This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence.
6 Authorizations FCC warning statement This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.
Testo SE & Co. KGaA Testo-Strasse 1, D-79853 Lenzkirch, Germany Phone: +49 7653 681-0 Fax: +49 7653 681-100 E-Mail: info@testo.de Website: www.testo.