Tempris Lyophilisation User Manual Tempris- Temperature Measurement Systems T.LAB and T.PRO Tempris Lyophilisation Monitor (TLM) Version 10.1 Version 10, 2021-05-26 Copyright © 2016 - 2021 Tempris GmbH The copying, distribution and utilization of this document as well as the communication of its contents to others without expressed authorization is prohibited. All rights reserved in the event of patent, utility model or design patent registration.
Tempris Lyophilisation User Manual Important Information User Manual TLM Tempris-Temperature Measurement System T.PRO Tempris Lyophilization Monitor (TLM) Version 10.1 Version 10, 2021-05-26 Publisher: Tempris GmbH, Holzkirchen Software version Modell 10.1 TIRU3 The information contained in this documentation is published without regard to possible patent protection. Trademark names are used without guarantee of free usability.
Tempris Lyophilisation User Manual Important Information Change History Version Name Date Change V1 Franziska Söltl 2019-11-11 Initial creation V2 Franziska Söltl 2020-01-15 Preparation of safety instructions and regulations V3 Franziska Söltl 2020-02-13 Creation of the software and hardware description with pictures and screenshots V4 Franziska Söltl 2020-08-03 Change from software version 10.0 to 10.
Tempris Lyophilisation User Manual Important Information Table of Contents 1 Important Information ............................................................................................................... 7 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 1.10 2 Introduction ............................................................................................................................... 19 2.1 2.2 3 Tempris for wireless temperature measurement without batteries .............................
Tempris Lyophilisation User Manual Important Information 4.5 4.6 4.7 4.8 4.9 5 4.4.3 Disconnect .................................................................................................................................................................................. 35 4.4.4 Login ............................................................................................................................................................................................. 36 4.4.5 Logout .........
Tempris Lyophilisation User Manual Important Information 1 Important Information This documentation is part of the Tempris-System. It contains necessary instructions for the use of the Tempris system and the Tempris Lyophilization Monitor Software and must be thoroughly studied before using Tempris. The operator may only put the Tempris system into operation after the persons responsible for operation have been instructed in its handling on the basis of the Tempris documentation.
Tempris Lyophilisation User Manual Important Information 1.1.3 Pictures The images used in the Tempris documentation (photos, screen images, etc.) may differ from the original provided that the correct description of the functions illustrated by the image is not affected. 1.1.4 Duplications All rights reserved, including photomechanical reproduction and storage in electronic media. Commercial use of the models and works shown in this documentation is not permitted. 1.1.
Tempris Lyophilisation User Manual Important Information 1.2 Symbols and Conventions 1.2.1 Product Names The following Tempris product names are used in this documentation (Table 1-1). Table 1-1: Product names Product name Description TIRU TIRU2 TIRU3 Type designation for Tempris Interrogation Unit - interrogation unit of the Tempris system. TIRU2 is the second-generation interrogation unit. TIRU3 is the thirdgeneration interrogator. T.LAB T.LAB-A1 T.LAB-A2 T.LAB-A3 T.LAB-A4 T.
Tempris Lyophilisation User Manual Important Information 1.2.
Tempris Lyophilisation User Manual Important Information 1.3 Intended Use Tempris is a system for wireless and battery-free product temperature measurement during freeze drying processes in the pharmaceutical sector. Measurement records obtained with Tempris can be used for the development, optimisation and monitoring of lyophilisation cycles, for hot and cold spot determination as well as for scale-up and transfer of lyophilisation cycles. Tempris was developed for process development and optimization.
Tempris Lyophilisation User Manual Important Information 1.6 User Responsibility When Tempris is used in a productive lyophilisation system, the user is solely responsible for the safety and quality of the products produced.
Tempris Lyophilisation User Manual Important Information 1.6.3 Radio transmission in the freeze-drying chamber During radio transmission in the freeze-drying chamber / liquid phase, the following must be observed: • Due to its physical properties, liquid media* (such as pharmaceutical products) absorb electromagnetic waves in the 2.4 GHz ISM band used.
Tempris Lyophilisation User Manual Important Information 1.7 Safety Information 1.7.1 General Safety Information • • • • • • • • • • • • • • • • A prerequisite for safe working is compliance with all specified safety instructions and instructions for action. Therefore read the instructions thoroughly and follow them. Observe all warning notices. Read the operating instructions carefully before starting any work.
Tempris Lyophilisation User Manual Important Information 1.7.2 Safety Information for Installation and Operation of the System • • • • Install in a place without vibration Only install in environments with humidity <80% Rh (non-condensing) Storage Temperature: 0-60°C; ideal 25+-10°C Operating temperature: 0°C-45°C 1.7.3 Power Source War The correct power source for the Tempris components can be found on the component labels.
Tempris Lyophilisation User Manual Important Information 1.7.6 CE-Conformity Declaration of conformity with regard to the EU Directive 2014/53/EU (RED Directive) This declaration applies exclusively to configurations (combinations of software, firmware and hardware) supported or provided by Tempris GmbH for use in the EU. The use of software or firmware not supported by Tempris GmbH may result in the device no longer fulfilling the legal requirements.
Tempris Lyophilisation User Manual Important Information 1.7.7 Federal Communications Commission - United States CFR 47 Part 15.105 (a) This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.
Tempris Lyophilisation User Manual Important Information 1.8 Liability Limitation The use of Tempris is the responsibility and risk of the operator or user.
Tempris Lyophilisation User Manual Introduction 2 Introduction 2.1 Tempris for wireless temperature measurement without batteries The essential quality feature in the production of pharmaceuticals, especially in freeze-drying, is the monitoring of the product temperature. Since this could not previously be measured directly in the manufacturing process, the proof was provided by several validation runs. This is associated with a high expenditure of time and money.
Tempris Lyophilisation User Manual Introduction 2.2 Tempris Operating Principle The Tempris system consists of a control unit, the Tempris Interrogation Unit (TIRU), a TIRU antenna and a sensor with associated sensor antenna (Figure 2-1). Depending on the configuration, the control unit is equipped with up to four integrated antenna interfaces for connection and operation with up to four TIRU antennas.
Tempris Lyophilisation User Manual Tempris Device Setup 3 Tempris Device Setup 3.1 Tempris System Components A typical T.PRO or T.
Tempris Lyophilisation User Manual Tempris Device Setup 3.2 TIRU3 Connections and Interfaces The TIRU3 interrogation unit offers a PoE+ (Power-over-Ethernet Plus) interface, which is used both as a power supply and as a network connection. The unit also has up to four antenna interfaces in the form of SMA sockets for connecting and operating Tempris interrogation antenna systems (Figure 3-1).
Tempris Lyophilisation User Manual Tempris Device Setup On the side opposite the operating connections there is a service connection which is intended exclusively for use by TIRU3 service technicians. Figure 3-3 shows the service connection of the TIRU3 interrogator unit.
Tempris Lyophilisation User Manual Tempris Device Setup One side contains a status LED with the following meaning (Figure 3-4 – LED 1): • • • Green: Standby Orange: Measurement running (second interval), Fast flashing (boot process) Red: Error (rapid flashing) For Backup there is an additional LED on the same side in the middle, at the bottom.
Tempris Lyophilisation User Manual Tempris Device Setup 3.3 Tempris Lyophilisation Monitor Installation In case the TIRU3 is delivered with a laptop, the Tempris Lyophilization Monitor Software will already be installed on it. You should therefore only install Tempris Lyophilization Monitor software when updating software or on externally used computers. 3.3.1 Tempris Lyophilization Monitor System Requirements The Tempris Lyophilization Monitor Software can be used with both T.LAB and T.PRO.
Tempris Lyophilisation User Manual Tempris Device Setup 3.3.2 New-Installation The Tempris systems usually includes a USB memory stick with the following TLM setup directories: Tempris-Lyophilization-Monitor (TLM-Setup for 32-bit Version) Tempris reserves the right to provide these setup directories alternatively via ZIP archives or download links from the Tempris websites.
Tempris Lyophilisation User Manual Tempris Device Setup 3.3.3 Update-Installation You can update any Tempris Lyophilization Monitor version to the latest Tempris Lyophilization Monitor version by following the instructions in chapter 3.3.2 . You can use the ABOUT… function from the HELP menu to find out the version number of the Tempris Lyophilization Monitor software currently installed on your Windows system.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4 Tempris Lyophilization Monitor User Interface and Program Functions This section describes the Tempris Lyophilization Monitor (TLM) software. The information in this section applies to PC-based Tempris systems and does not necessarily apply to PLC-based Tempris systems. Detailed instructions for controlling Tempris via an automation interface can be found in our additional manual "Automation".
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.1.1 Tempris Lyophilization Monitor Software Licenses Table 4-1 lists the software licenses available with the Tempris Lyophilization Monitor software. Table 4-1: Tempris Lyophilization Monitor Software Licenses.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.2 Program Start and Overview The Tempris Lyophilization Monitor is started by double-clicking the Tempris Lyophilization Monitor program icon Figure 4-1 shows an overview of the Tempris Lyophilization Monitor software. The software consists of the menu items: File, Modules, Settings, Options, Help and Service.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.3 Status Display and Error Reporting 4.3.1 Status Display Bar The following figure shows an example of the displays in the status bar of the Tempris Lyophilization Monitor software (Figure 4-2). Figure 4-2: TLM Status Display Bar Table 4-2 contains an overview of the status display elements of the TLM software.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.3.2 Display of Status and Error Messages Status messages, warnings and error messages are usually displayed in the field for displaying messages in the status display bar (see chapter 4.3.1) or in the main display area in the overlay display mode. If a message must be acknowledged by user confirmation, a popup window is activated to display the message with the corresponding user prompt.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.4 Menu – File 4.4.1 Configuring Connections The communication between your PC and your TIRU3 interrogation unit takes place via the integrated TIRU3 Ethernet interface. Select the CONNECTION… icon in the FILE menu to establish a new connection with the TIRU. You can define one or more TIRU3 connections in the dialog box "Manage Network Connections" (Figure 4-3).
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.4.2 Establish a Connection Make sure that the TIRU3 connection you want to use has been previously set up with the correct IP address and port number (see chapter 4.4.1) via FILE - CONNECTIONS... To establish a connection, go to FILE and select CONNECT. From the drop-down menu you can now select one of your previously configured connections (Figure 4-4).
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.4.4 Login If the TLM-USERADMIN option is activated on your TIRU3 system, you must log on to the TIRU3 with your user name and password (Figure 4-5) using the FILE - LOGIN function after successfully connecting to the TIRU3 in order to be able to use further functions such as MESSUNG STARTEN.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.4.7 Close Application To close the application, you can either choose the CLOSE icon in the upper right corner of the application or QUIT APPLICATION in the FILE menu. To exit the program, confirm the entry with QUIT (Figure 4-6).
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.5 Menu Modules 4.5.1 Start Measurement Before you start a measurement, you should configure your sensor settings in the menu item SETTINGS under SYSTEM - SENSOR SETTINGS… and upload the calibration data of your sensors to the TLM software. The pertaining calibration date has been supplied on a USB stick, to the TLM software. Detailed instructions can be found in chapter 4.6.1.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.5.3 Start Logging To start a recording, go to MODULES and START LOGGING. The “Audit Trail Comment” dialog box now opens (Figure 4-8). This serves to better reflect your measurement within the AuditTrail. By selecting the Include comment in log file name suggestion button you can accept your input for saving the log file.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.5.4 Stop Logging If you have recorded your data successfully and want to end your logging, go to the menu item MODULES to STOP LOGGING. Your logging will now be terminated and saved in the Windows-dialog under the name you selected when you started logging. Confirm the termination of logging by clicking YES in the dialog box (Figure 4-10).
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.5.5 Report In the menu item MODULES under REPORT, you can create a CSV file, a log file report, select the entire measurement data statistics of the logging or a screen section of same. In this section, the functions under Report are explained in more detail. CSV-Export By clicking MODULES - REPORT and then CSV-EXPORT…, you can export the measurement data of the log file as CSV-file.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions Log File Report The Log File Report function allows you to create a PDF from your log file. Um eine Log-Datei laden zu können müssen Sie abgemeldet und von der TIRU3 getrennt sein. The file name, date and time of the output date and system information (such as sensor settings, serial numbers, software versions, users) are automatically inserted in each log file.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions Under “Header Settings” you can upload your company logo (as png, jpeg, bmp or gif) and choose a text for the header. By ticking the box Filename with full path you add the currently used storage location to your document. In the “Footer Settings” you can add a footer text to the document.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions Total Measurement Data Statistics To output the measurement data statistics, select the menu item MODULES under REPORT and MEASUREMENT TOTAL DATA STATISTICS… (Figure 4-17). Now the window "Measured Data Statistics" opens. Here you can view the number of measurement intervals. In addition, you will receive an evaluation of the obtained data points for each sensor active during the measurement.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions Measurement Display Range Data Statistics To output the measurement data statistics of a selected screen section, select the menu item MODULES under REPORT and MEASUREMENT DISPLAY RANGE DATA STATISTICS…. The "Measurement Data Statistics"-window now opens (Figure 4-18). Here you can view the number of measurement intervals of the image section selected under Dock-Panel GRAPHIC.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.5.6 Audit Trail The Tempris Lyophilization Monitor option TLM-AUDITTRAIL ("Audit Trail") records essential system events as well as all operator inputs and actions that create, modify, delete, or process system or record entries on the TIRU3 interrogation unit. These audit trails are stored in an unalterable, secure log file on the internal file system of the TIRU3 interrogation unit.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.5.7 Backup The TIRU3 interrogation unit has an integrated backup system which can be activated with the TLMBACKUP software license. If the TLM-BACKUP license is activated on the TIRU3 system, a backup log file with the metadata of the measurement run and the recorded measurement data is automatically stored on the backup system of the TIRU3 for each measurement run.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.6 Menu – Settings 4.6.1 System Sensor Settings The sensors are configured before the measurement via the menu item SETTINGS in the subtopic SYSTEM under the item SENSOR SETTINGS…. First you should upload the calibration files supplied by Tempris on a data carrier under "Calibration file options".
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions You can now select the number and position of the antenna ports according to your system (corresponding to the numbering in Chapter 3.2 Figure 3-1). Under "Sensor settings" you can add a sensor by clicking on the icon. We do recommend to only add sensors used by the system to save system resources.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions License Management To operate various options in the software, you will need licenses for each TLM software option. You can view and manage the TLM software licenses of the TIRU3 interrogation unit currently connected to your Tempris Lyophilization Monitor software using the LICENSE MANAGEMENT… function from the SETTINGS - SYSTEM menu. If the TLM-USERADMIN license (user administration, see Chapter 4.6.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.6.2 User In the menu item SETTINGS under USER there are two sub-items: On the one hand the USER ADMINISTRATION and the option CHANGE PASSWORD.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions If you are logged in as an administrator you can open the "Tempris User Administration" dialog for access control (Figure 4-26). Here you will find the login name, the description, the role, the creation date, the expiration date as well as status information on whether a user has been deleted or blocked. Figure 4-26: Tempris User Management You can add, edit, or delete users in this panel.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions Login Settings You can also set global login and password settings for all users. In the login settings you can set the duration after which a user is automatically logged out. This can be used to prevent unwanted access. You can also set a maximum number of retry count after which a user will be blocked. You can specify whether the user should be locked permanently or until the next restart.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions Function Access Matrix Table 4-4 shows the access matrix for the functions of the Tempris Lyophilization Monitor Software. Depending on the context, some menu items may not be selectable (grayed out). Under Notes you can read which functions are necessary to activate menu items.
Listener Logged off OPTIONS LANGUAGE GERMAN ENGLISH CHINESE TEMPLATES FIRMWARE UPDATE HELP MANUAL GERMAN ENGLISH AUTOMATION MANUAL ENGLISH ABOUT TIRU3 ABOUT • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • • Config. PRODUCT SETTINGS… Admin.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions Change Password To change your password, go to the menu item SETTINGS under USER and select the option CHANGE PASSWORD.... The "New Password" dialog opens (Figure 4-28). Here you can assign a new password. Confirm your password by clicking the SAVE button.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.6.3 Define chamber dimensions To enter the chamber dimensions of your freeze dryer, click on the button SET CHAMBER DIMENSIONS... in the menu item SETTINGS or enter SETTINGS on the CHAMBER tile via the dock panel. The "Chamber Layout" dialog box now opens. (Figure 4-29).
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.6.4 Product Management By pressing the "Product" button or in the SETTINGS menu item of the PRODUCT MANAGEMENT... button, the "Product settings" dialog opens. Here you enter your product name, the vial size and the vial filling quantity. The field "Supported sensor size" now shows you which sensors you can use for the measurement of this product.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.7 Menu – Options 4.7.1 Languages In the menu item OPTIONS under LANGUAGE you can currently choose between the languages English and German. The user interface with menu, dock-panel and status display is then changed to the desired language. 4.7.2 Templates In the menu item OPTIONS under TEMPLATES…, you can create a template that you can use when you start logging.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions By clicking the SAVE icon you can save the created template under the desired name, here tempris example template (Figure 4-32). Figure 4-32: saving a template under a desired name If you now start logging, your created and loaded template appears in the "Data Entry" dialog box (Figure 4-33). You can now fill in the fields Product, Processor and Department here and confirm with OK.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.8 Help 4.8.1 User Manual The manual of the Tempris system for lyophilization can be found in digital form in the TLM software under HELP and MANUAL. The manual is currently available in German and English. 4.8.2 About TIRU3 Once a connection is established with a TIRU3 system, you can call up a window with information on the connected TIRU3 system in the HELP menu under ABOUT TIRU3….
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.9 Dock-Panel The individual windows of the Dock-Panel (TEXT, GRAPHICS, SETTINGS) can be moved to another position or dragged out of the window into a new one by simply dragging the respective panel. The Dock-Panel consists of three units: TEXT Display, GRAPHIC Display, SETTINGS. 4.9.1 Text Display During the measurement the Dock-Panel shows an overview of the currently active sensors.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.9.2 Graphic Display After connecting to the TIRU3 and starting a measurement, the Dock-Panel GRAPHIC displays the temperatures measured by the sensors in °C over time (Figure 4-37).
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions Sensor Display In this setting you select whether you want all, only those with data collected or no sensor to be displayed. You can also switch between LINE and MARKER display (Figure 4-39). If you are in the marker display, you can also select the SIZE of the markers (between 1 and 6). Figure 4-39: Dock-Panel Graphic View - Sidebar for setting the graphic display.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions Color settings of the sensor channels You can change the colors for each sensor by clicking on the color box of the respective sensor. Here you can choose from the standard colors (Figure 4-41) or in the Advanced tab (Figure 4-42) from any color value.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions Time Display It is also possible to change the time display. Here you can choose between RELATIVE (time since start of measurement) and ABSOLUTE time (date and time), as well as between LOCAL and UTC time system. (Figure 4-43). Figure 4-43: Dock-Panel Graphic View - Sidebar for setting the graphic display. Display of the measurement data in absolute (left) or relative (right).
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions Tile Size Text Display As a further setting you have the possibility to adjust the TILE SIZE. This helps you to keep an overview of all sensors with a large number of sensors (see Figure 4-45). Figure 4-45: Dock Panel Graphic View - Links: Text Size Small; Middle: Text Size Medium; Right: Text Size Large.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions Display Settings in the Graphic You can change the following settings directly in the graphic using the mouse: - Left Mouse Key: display of sensor data as text at picked time (Figure 4-47) Right Mouse Key: Moving the measurement points (by right-clicking on the time or temperature axis, the range is only moved along the time or temperature).
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions 4.9.3 Settings You cannot make any settings in the Dock-Panel during a measurement. To make settings, you must first stop the measurement.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions Automation The Tempris option TLM-AUTOMATION (Tempris Lyophilization Monitor Automation) implements a freely parameterizable Modbus-over-TCP/IP automation interface for communication with SCADA, PLC and data recorder systems. A detailed description can be found in chapter 756. License The LICENSE function takes you to the license management of the TLM software.
Tempris Lyophilisation User Manual Tempris Lyophilization Monitor User Interface and Program Functions System Information Under SYSTEM INFO, the system gives you an overview of the TIRU3 (Figure 4-51), the antenna ports, the software licenses, the user accounts and the sensors (with measuring channel, sensor channel, serial number, ID, minimum and maximum temperature and the next upcoming calibration date). You can export the complete system information of the TIRU3 as a PDF file.
Tempris Lyophilisation User Manual Additional Modules 5 Additional Modules 5.1 Automation For configuration with your automation interface, please request our "Automation" manual at info@tempris.com. 5.2 INET To activate the function INET you need the license key TLM-INET. As soon as you have logged in with the corresponding license, the INET button appears under SETTINGS. (Figure 5-1).
Tempris Lyophilisation User Manual Additional Modules To set up the TIRU INET, click on the INET button. Now the dialog box "TIRU INET Settings" opens. (Figure 5-2). Here you can operate several TIRU3 together to increase the number of antennas that can be used. To do this, you must first define a TIRU3 as slave (example IP: 192.168.1.101). To do this, log in with a TIRU3 and select Slave under Own function as shown inFigure 5-2.
Tempris Lyophilisation User Manual Additional Modules The adjustment of the antennas and sensors is done via the button SENSOR in the Dock-Panel "Settings". The general settings, calibration file, options and the sensor settings remain as described in chapter 4.6.1 under SENSOR SETTINGS…. Only the selection of antenna ports is now increased from 4 to 8 (Figure 5-4). On the left side select the connected antennas for your master TIRU3 (1-4) on the right side the slave TIRU3 (5-8).
Tempris Lyophilisation User Manual Figure Index 6 Figure Index Figure 2-1: Structure of the Tempris system consisting of the control unit (TIRU) with TLM software, TIRU antenna and quartz-based sensor with associated sensor antenna. ................................................... 20 Figure 3-1: TIRU3 - Connections................................................................................................................................... 22 Figure 3-2: Torque Wrench (delivery option) ...........
Tempris Lyophilisation User Manual Figure Index Figure 4-32: saving a template under a desired name ........................................................................................ 60 Figure 4-33: data input when starting logging ....................................................................................................... 60 Figure 4-34: system information in menu item About TIRU3 ............................................................................
Tempris Lyophilisation User Manual Table Index 7 Table Index Table 1-1: Product names ................................................................................................................................................. 9 Table 3-1: Tempris Lyophilization Monitor System Requirements .................................................................. 25 Table 4-1: Tempris Lyophilization Monitor Software Licenses. .........................................................................