RTR-5W RTR5W User's Manual Thank you for purchasing our product. Carefully read this instruction manualbefore using this unit. #OPYRIGHT 4 $ #OR PORATION !LL RIGHTS RESER VED 2006.
ɡ Notices about this User's Manual In order to properly use this product, please carefully read this manual before using. T&D Corporation accepts no responsibility for any malfunction of and/or trouble with this product or with your computer that is caused by the improper handling of this product and will deem such trouble or malfunction as falling outside the conditions for free repair outlined in the attached warranty. -All rights of this Userȧs Manual belong to T&D Corporation.
Software User's Agreement ɡ Escape Clauses -T and D Corporation does not guarantee the operation of ThermoRecorder for Windows -T and D Corporation shall not accept any responsibility for any damage whether direct or indirect that results from the usage of ThermoRecorder for Windows. -Specifications of ThermoRecorder for Windows may be subject to change and service may be terminated without advance notice to the user.
Safety Precautions and Instructions To ensure safety be sure to obey all of the following warnings. The following items should be strictly obeyed for the safe usage of this unit, and for protecting yourself and other people from bodily harm and/or damage to property.Before using make sure to carefully read, understand and follow the safety rules and precautions for our products as outlined below.
! Warnig Do not take apart, repair or modify the main unit. This may cause fire or electrocution. When installing and using this product, make sure to follow all warnings and directions from your computer manufacturer. Be careful of igniting fire. If water or a foreign body enters into this unit, immediately unplug the AC adaptor and stop using. Continued use may cause fire or electrocution. Do not use this unit in wet or humid places, such as a bathroom.
! Caution We are not responsible for any damage, malfunction or trouble, whether direct or indirect, caused by the use of our product.Please be fully aware of this before using our product. This product has been designed for private or industrial use only. It is not for use in situations where strict safety precautions are necessary such as in connection with medical equipment, whether directly or indirectly. This unit is not waterproof. If the unit gets dirty, wipe it with a clean cloth dipped in alcohol.
Do not use or store the unit in places such as listed below: Doing so may cause electrocution, fire and/or other adverse effects to the device and/or your computer. - Areas exposed to direct sunlight This will cause the inside of the device to become overheated and may cause fire, deformation, and/or other damage including malfunction. - Areas prone to strong magnetic fields This may cause damage including malfunction.
Contents ɡ Notices about this User's Manual --------------------------i ɡ Using Help ---------------------------------------------------- 25 Software User's Agreement ---------------------ii ɡ Network Settings Utility Functions ---------------------- 26 ɡ Escape Clauses ----------------------------------------------- ii Network Initialization Settings -----------------27 ɡ Copyright-------------------------------------------------------- ii Detailed Network Settings ----------------------29 Safety P
ɡ Change Graph Colors ------------------------------------- 60 Warning Monitoring ------------------------------90 ɡ Copy Display to Clipboard -------------------------------- 61 How to use the RTR-5W Web Viewer -------94 Operating the Graph Display ------------------62 ɡ Opening the RTR-5W Web Viewer --------------------- 94 ɡ Returning to Original Size -------------------------------- 62 ɡ RTR-5W Web Viewer Functions ------------------------ 95 ɡ Zooming In and Out --------------------------------
What is Web Wing WL RTR-5W ? The RTR-5W not only serves as a base station for the wireless downloading of recorded data from our compact waterproof Wireless Communication RTR-5 Series Data Loggers, but, moreover, the [Web Wing WL RTR-5W] serves as a network base station that allows you to monitor and view current readings and gather recorded data via LAN or Internet connection. œŰŶŵŦų ŊůŵŦųůŦŵ ŐŧŧŪŤŦ ōłŏ 㧨 㧪 HILD ഩ࡙ ·ςͺ 2 ̜ .
ɡ Basic Functions Gathering and Managing Remote Unit Data via a Network The RTR-5W is able to gather via wireless communication the measured and recorded data from any RTR-5x Series Data Logger and using a Local Area Network (LAN) or Internet connection makes it possible to view current readings and / or download the gathered recorded data.
ɡ Basic Functions of RTR-5W for Windows The software is made up of the following 5 applications: [Network Settings Utility], [RTR-5W for Windows], [Temperature /Humidity Graph], [Multi-Scale Graph] and [Event Viewer]. Also, by using your web browser to access RTR-5W, it is possible to view data via the Internet. *Only compatible with Internet Explorer Ver.6 or higher using Windows 2000 / XP ȺNetwork Settings UtilityȻ Make the necessary network settings for RTR-5W.
ȺTemperature / Humidity GraphȻ/ȺMulti-Scale GraphȻ Temperature and Humidity are displayed in the Temperature and Humidity Graph In the Multi-Scale Graph, voltage, pulse, temperature and humidity data can be viewed in Graph form. In Multi-Scale Graph, it is possible to make settings for scale and unit of measurement to match the type of date being downloaded. View Graphs and Print It is possible to view the downloaded data as a list and print.
ȺEvent ViewerȻ View and print Event data recorded with the RVR-52A Data Logger. View List Up to 64 channels of downloaded data can be viewed in one list. Print Preview and Print Printing can be carried out after checking the preview window. Creating Text File It is possible to convert the data for a specified range (time period) to common text file format (CSV type format), so that it can be exported to spreadsheet software such as Excel or Lotus.
Package Contents The following items are included in the package: #!54)/. Software CD-ROM x 1 LAN Cable LN-20W x 1 AC Adaptor AD-0605 x 1 RISK OF ELECTRIC SHOCK DRY LOCATION USE ONLY. SHOCK HAZARD NONSERVICEABLE PARTS INSIDE. Web Wing WL RTR-5W x 1 unit ൵ව΄ͼΡ #!54)/. RISK OF ELECTRIC SHOCK D R Y L O C AT I O N U S E O N LY. SHOCK HAZARD NONS E R V I C E A B L E PA RT S I N S I D E .
RTR-5W Part Names and Functions ɡ Part Names ȬFRONTȭ Optical Communication Area POWER LED While the power is ON, the lamp will be on. Alarm LED When a warning occurs, the lamp will blink. Monitor LED While wireless communication is occurring with a Remote Unit, the lamp will blink. Optical ommunication Area When you wish to communicate directly, not by wireless, with a data logger (Remote Unit), it is necessary to place the logger (Remote Unit) face down on this area for communication to occur.
ȬLEFT SIDEȭ ȬRIGHT SIDEȭ CF interface slot Ethernet Jack Reset Button AC Adaptor Jack Ethernet Jack AC Adaptor Jack CF Interface Slot RESET Button 10Base-T/100Base-TX Ethernet ConnectorConnect with LAN Cable Connect using the supplied AC adaptor Insert a Wireless LAN Card ɦ 1 NOTE: Do not insert or try to connect anything other than a CF communication card Press to return to the factory default settings *1ȇFor information and updates concerning which wireless LAN cards can be used please see our Hom
Outline of Use ɡ Basic Operation ȺGetting ReadyȻ 1.Prepare the Data Loggers you wish to use as Remote Units. Get the units ready for measuring by connecting the proper sensors and installing the proper batteries. 2.Getting the RTR-5W Ready to Use Get the unit ready for communication by connecting the network cable and AC adaptor. 3.
ȺInitialization SettingsȻ 1.Network Initialization Settings Ȥfrom the Network Settings Utility To connect to a network, it is necessary to enter an IP address and Subnet Mask that are appropriate for your network. Under Network Initialization Settings make the appropriate settings for your ȨIP addressȩ and ȨSubnet Maskȩ. More detailed settings can be made in [Detailed Network Settings] 2.
ȺGraph DisplayȻ Temp / Humidity Graph Temperature and Humidity data can be displayed in the same Temperature and Humidity Graph (Up to 8 channels of data) You can view the data displayed in the graph window as a list and also choose to print. It is also possible to convert the data to common text file format (CSV type format). Multi-scale Graph In the Multi-Scale Graph, voltage, pulse, temperature and humidity data can be viewed in the same Graph.
ȺOther FunctionsȻ Warning Monitoring SettingsȤ from RTR-5W for Windows Monitoring at each Location for warnings is carried out and if any of the gathered data exceeds the set limit, a notification can be sent via e-mail to your computer or cellular phone.Also, because there is a built-in external output terminal, it is possible to connect to an external device, such an alarm or light, so that a notification of a warning can be seen or heard onsite.
ɡ Software Operations Table ȺRTR-5W for WindowsȻ ȶFileȷ menu RTR-5W for Windows ȶViewȷ menu ȶRegistration / Administrationȷ menu ȶSettingȷ menu ȶDataȷ menu IJĴ Temp/Humidity Graph Multi-scale Graph Event Viewer Text Data Output Settings Read Data Collector Registration Info File Output Registration Info File for Data Collector Expand to Whole Tree Programmed Auto Download View Log Store in Task Tray Display Format Settings Create Locations / Properties Create Groups / Properties Remote Unit Registr
ȺNetwork Settings UtilityȻ Network Initialization Settings Network Settings Utility Network Detailed Settings ȶMenuȷ Menu IP Address Subnet Mask Login ID Password DNS SNTP SMTP Mail Server IP Block Wireless LAN Port Number Time Difference Rebooting the Communication Functions (Restarting the System) Login History – ON / OFF Clear Login History Communication Time Settings ȺRTR-5W Web ViewerȻ Web Viewer Monitor Current Readings Graph IJĵ
Getting the RTR-5W Ready to Use ɡ Connect the supplied AC adaptor *To ensure a proper connection make sure that the plug is completely inserted NOTE: - Do not use an AC adaptor other than the one that is supplied with the product. Doing so may cause fire or other trouble. - Insert the AC adaptor plug into an AC 100V socket. Inserting the plug into a socket with different voltage may cause fire or other trouble.
Communicating with the Data Logger It is possible to communicate between an RTR-5W base station and Data Loggers RTR-51/52/53ȆRVR-52A by two methods: [Wireless Communication] and [Optical Communication]. ɡ Communicating with a Data Logger via Optical Communication Optical communication can be carried out by placing an RTR-51/52/53 or RVR-52A Data Logger face down on the RTR-5W so that the optical communication areas meet.
External Output ɡ Connecting the Signal Wire 1. Prepare a ϫ 0.5 ȡ 0.9 single type wire and remove the covering to about 10mm from the tip. 2. Using a screwdriver or other such tool, while pressing down on the terminal button at the back of the unit, insert the wire into the hole . =A? =B? ȴEnlarged view of Wire Connection Terminal ȵ ɟ External Output : With Warning ON Voltage when OFF: AC/DC less than 50V Current when ON: less than 0.
Checking your Operating Environment To properly use the logger, the following operational environment is necessary. ɡ PC Operating Environment OS OS Microsoft Windows 2000 Microsoft Windows *For installation, it is necessary to have Administrator (Computer Administrator) rights.
Connecting to a Network The following are some ways to connect an RTR-5W to a network. Please select the connection method that suits your network environment. ɡ Connecting to a HUB for in-company Communication *To ensure a proper connection make sure that the plug is completely inserted Connection Example ,!. CABLE 㧔,. 7 242 7 㧔 㧕 (5" 0# 㧔 㧕 ɡ Connecting Directly to PC for Communication Do not use the supplied cable; please purchase a cross LAN cable separately.
ɡ Connecting to a Wireless LAN By inserting a Wireless LAN card into the RTR-5W, it is possible to carry out communication in places where LAN wiring is difficult or troublesome. *To ensure a proper connection make sure that the plug is completely inserted NOTE: - Before you insert or take out the LAN card, make sure that the AC adaptor is unplugged. - For information and updates concerning which wireless LAN cards can be used please see our Homepage. Connection Example 242 7 7IRELESS ,!.
Connecting to the Internet In order to connect the RTR-5W to the Internet you must first set up an Internet connection environment by making arrangements with a provider for a line and get a global IP address and domain. Also, if necessary, make all domain name server and / or router settings. For more details about various settings, it is best, if present, to contact the network administrator. To get more details about a global IP address and domain, please contact your provider.