2 electrode free chlorine sensor Type CAA320X Co/Gold version Type CAA330X Co/Platinum version Installations and commissioning instructions Reference: DOC0514 Rev :1.
General information Page 2/24 General information: SYCLOPE Electronique 2020® Manual of 06/04/2020 Rev: 1.0 Free chlorine sensor Installation and commissioning instructions (Ref: DOC514) Editor: SYCLOPE Electronique S.A.S. Z.I. Aéropole Pyrénées Rue du Bruscos 64 230 SAUVAGNON - France – Tel: +33 559 337 036 Fax: +33 559 337 037 Email: syclope@syclope.fr Internet: http://www.syclope.fr © 2020 by SYCLOPE Electronique S.A.S.
Content Page 3/24 Content I. Generality .................................................................................................................. 4 1) 2) 3) 4) 5) 6) II. Use of the document ......................................................................................................... 4 Symbols and signs ............................................................................................................. 4 Storage and transport ................................................
Generality Page 4/24 I. Generality 1) Use of the document Please read carefully the entire document before starting the installation and the commissioning of the controller device, in order to ensure the safety of swimmers, users and equipment’s. The information provided in this document must be strictly observed. SYCLOPE Electronique S.A.S. declines all responsibility in cases where failure to comply with the instructions of this documents.
Generality Page 5/24 3) Storage and transport It is important to store and to transport the electrode free chlorine sensor in its original packaging in order to minimize risk of damage. Furthermore, the package must be stored in an environment that is protected against humidity and exposure to chemical products.
Generality Page 6/24 6) FCC conformity The device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference (2) this device must accept any interference received, including interference that may cause undesired operation FCC Regulations state that unauthorized changes or modifications to this equipment may void the user’s authority to operate it.
Safety and environmental instructions Page 7/24 II. Safety and environmental instructions Please: Read this manual carefully before the unpacking, the installing or the commissioning of this equipment Take into account all the hazards and of recommended precautionary measures The failure to respect these procedures can result in serious injury to users or damaging the device.
Safety and environmental instructions Page 8/24 Make sure that the probe used with the device matches the measured and used quantities. Refer to the individual technical instructions for each probe. In case of doubt, immediately contact our technical service or your authorized installer. Probes are sensitive and have consumable parts. They must be monitored, maintained and calibrated regularly using specific reference products not supplied with the equipment.
Safety and environmental instructions Page 9/24 5) Probe rating plate identification Cu/Pt CAA330X 7 10 1 11 2 12 3 8 4 5 9 6 1 Manufacturer’s label 9 Identification of the manufacturer 2 Model of the product 10 CE approved 3 Measuring range 11 Conformity with the FCC part 15 Class B 4 Range of current 12 FCC ID 5 Reference of the product 6 Serial number 7 Particular risks.
Safety and environmental instructions Page 10/24 6) FCC conformity The Sensor complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference (2) this device must accept any interference received, including interference that may cause undesired operation FCC Regulations state that unauthorized changes or modifications to this equipment may void the user’s authority to operate it.
Safety and environmental instructions Page 11/24 7) Disposal and conformity The recyclable packaging of the probe equipment must be disposed of according to current regulations. Elements such as paper, cardboard, plastic or any other recyclable elements must be taken to a suitable sorting center. DEEE: According to European directive 2012/19/EC, this symbol means that as of 4 July 2012 electrical appliances cannot be thrown out together with household or industrial waste.
Technical specifications Page 12/24 III.
Mechanical and hydraulic installation Page 13/24 IV.
Mechanical and hydraulic installation Page 14/24 2) Installation in the probe housing The CAA320X et CAA330X probes ares installed in the SYCLOPE measuring chambers (modular type BMT or BMT block) whose dimensions allow the installation of standard 1” probes. Hydraulic connections are made using PE flexible tubes or compatible connections.
Mechanical and hydraulic installation Position the probe ring from the top of the probe. Position the probe ring from the top of the probe. Step 2 Page 15/24 Insert the probe into the measuring chamber. Position the clamping nut from the top of the probe Position the clamping nut from the top of the probe Step 5 Step 4 Step 3 Insert the probe into the measuring chamber. Lock the tightening nut until the O-ring seals. Lock the tightening nut until the O-ring seals.
Electrical connections Page 16/24 VI. Electrical connections 1) Electrical connections Electrical installations must be carried out according to the standards and by authorized personnel! Before making the connections, turn off the power supplies! 2) Analog power and output operational connections The probe must always remain on. It must never be stored off in water. In case of extended use shutdown, remove the probe from the passage chamber and store it in its original dry box.
Calibration Page 17/24 VII. Calibration CAA320X et CAA330X probes require a calibration of the slope of the sensor and, if necessary (low chlorine measurement), a calibration of the zero point. Depending on the physio-chemical parameters of the water, the gain multiplier switch can be used. 1) Calibration of the sensor slope A calibration of the slope of the sensor is required after the first commissioning (about 3h) and on a regular basis for normal chlorination operation.
Maintenance Page 18/24 VIII. Maintenance 1) Disassembly the probe of the measuring chamber Before dismantling the probe from its passage chamber, close the shut-off valves upstream and downstream of the sampling circuit. Remove pressure from the system by opening the drain located under the passage chamber. For a probe already in place, unscrew the 1’1/4 nut and remove the probe from its housing after disconnecting the electrical wires using the screwdriver provided in the storage box..
Maintenance Page 19/24 Position the bead-filled cap under the probe and fill it back on the probe by turning clockwise and locking the packing cap on the O-ring above the copper electrode. 3) Change of the copper counter-electrode The change of the copper counter-electrode must be done every year. This period may vary depending on water quality and flow rate on electrodes.
Maintenance Page 20/24 IX.
Maintenance Page 21/24 X. Maintenance Probe must always be powered on. It must never be stored off in water. In case of extended use shutdown, remove the probe from the measuring chambre, dry it and store it in its original dry box. Repairs can only be carried out by qualified technicians and must be carried out exclusively in our factory. For every other problem not mentioned in this manual or for any other information, please contact your reseller or authorized installer or our after-sales services.
Notes Page 22/24 NOTES Installation and commissioning instructions
SYCLOPE Electronique S.A.S. Z.I. Aéropole pyrénées Rue du Bruscos 64 230 SAUVAGNON - France – Tel: +33 559 337 036 Fax: +33 559 337 037 Email: syclope@syclope.fr Internet: http://www.syclope.fr © 2020 by SYCLOPE Electronique S.A.S.