English Instruction and operation manual S430 Pitot tube flow sensor
Dear Customer, Thank you for choosing our product. The operating instructions must be read in full and carefully observed before starting up the device. The manufacturer cannot be held liable for any damage which occurs as a result of non-observance or noncompliance with this manual. Should the device be tampered with in any manner other than a procedure which is described and specified in the manual, the warranty is cancelled and the manufacturer is exempt from liability.
Table of contents 1. 2. 3. 4. Safety instructions.......................................................................5 Application.................................................................................7 Features.....................................................................................8 Technical Data............................................................................9 4.1 General.................................................................................9 4.
9. Calibration................................................................................32 10. Maintenance............................................................................32 11. Disposal or waste.....................................................................32 12. Warranty................................................................................32 Appendix......................................................................................
1. Safety instructions 1. Safety instructions Please check if this instruction manual accords to the product type. Please observe all notes and instructions indicated in this manual. It contains essential information which have to be observed before and during installation, operation and maintenance. Therefore this instruction manual has to be read carefully by the technician as well as by the responsible user / qualified personnel.
1. Safety instructions WARNING! Permitted operating parameters! Observe the permitted operating parameters, any operation exceeding this parameters can lead to malfunctions and may lead to damage on the instrument or the system. • Do not exceed the permitted operating parameters. • Make sure the product is operated in its permitted limitations. • Do not exceed or undercut the permitted storage and operation temperature and pressure.
1. Safety instructions Storage and transportation • Make sure that the transportation temperature of the sensor without display is between -30°C ... 70°C and with display between -10°C ... 60°C. • For transportation it is recommended to use the packaging which comes with the sensor. • Please make sure that the storage temperature of the sensor is between -10°C ... 50°C. • Avoid direct UV and solar radiation during storage. • For the storage the humidity has to be <90%, no condensation. 2.
3. Features 3. Features 8 • Flow and consumption measurement in wet air or high mass flow / velocity applications. • Measurement at compressor outlet possible. • Insertion type flow sensor for easy installations under pressure through a ball valve. • High temperature application up to 200°C. • Tube diameters from 1” up to 10”, larger diameters on request. • No mechanical wear parts. • All parts which are in contact with the medium are made of stainless steel.
4. Technical Data 4. Technical Data 4.
4. Technical Data 4.2 Electrical Data Power supply 24 VDC, 150 mA 4.3 Output-Signals Analog output Signal: 4 ... 20 mA Scaling: 0 to max. volume flow / velocity Accuracy: 0.06 mA Pulse output 1 pulse per m3, isolated switch, max. 30 VDC, 200 mA (active signal) (pulse length: 30 … 120 ms, depends on flow rate) Modbus output (default setting) See chapter 8.4 M-Bus output (default setting) See chapter 8.5 4.4 Accuracy Accuracy Flow: ±(1.5% of reading+0.3% full scale) Pressure: 0.
4. Technical Data Cut-off flow of S430 is pressure depending. Following chart shows the relationship at tube of 100 mm, 50°C and 90% humidity. For other tube and conditions please download “Flow range” software from the homepage of the manufacturer.
5. Dimensional drawing 5.
6. Installation 6. Installation Before you install the sensor, make sure that all components listed below are included in your package. Qty Description Item No. 1 S430 Sensor S695 4300/S695 4302 1 Sealing ring No P/N 1 Alignment key No P/N 2 M12 plug C219 0059 1 Instruction manual No P/N 1 Calibration certificate No P/N 6.1 Installation Requirements To install the sensor a ball valve or a nozzle is needed. • The inner thread has to be G 1/2”.
6. Installation ATTENTION! Wrong measurement is possible, if the sensor is not installed correctly. • Careful attention must be paid to the design of the inlet and outlet section. Obstructions can cause counter-flow turbulence as well as turbulence in the direction of the flow. • The sensor is for indoor use only! At an outdoor installation, the sensor must be protected from solar radiation and rain. 6.2.
6. Installation 6.2.2 Insertion angles Please install the sensor only in the recommended area like you can see in the picture. A horizontal installation is not permitted. Please consider that the distance to the horizontal level must be >10°. Installation in a vertical pipe is possible, required the flow is from bottom up and the sensor has at least a 10 degree angle (recommended is 20 degree). ATTENTION: Do not install in vertical pipes if the flow is from top to down.
6.3 Calculating the insertion depth Generally, the sensor tip must be placed in the centre of the pipe. To facilitate this, the sensor is designed with a scale engraved on its shaft. To determine the right position, calculate the insertion depth as described below. Note: If the pipe diameter is greater than the shaft length, it is feasible to install the sensor out of centre of the pipe. For more information, please contact the manufacturer. 6.3.
6.3.2 Non-centre installation In bigger pipe diameter (> DN150) situations, the non-centre installation is applicable. In the non-centre installation, the sensor is inserted to exact 100 mm-depth of pipes.
6.4 Installation of the sensor First please observe the flow direction indicated on the housing or on the shaft. It must match the flow direction of the compressed air or gas. 1. Ball valve has to be closed. 2. The sensor head has to be completely covered by the connection thread (see photo on the left). 3. Underlay the “O-ring” at the thread of the ball valve before screwing the flow sensor. 4. Screw the connection thread tightly to the ball valve and align flow sensor to flow direction. 5.
Maximum angle deviation of a proper installation: 6.5 Performing the zero flow calibration The zero flow calibration is a compulsory operation during the installation of the sensor. Use one of the following methods to perform the calibration: • The service App S4C-FS, available on the Google Play store or our Website for free download • The optional local display 6.5.
Sensors that are within the reach of your Bluetooth signal while not connected with other S4C-FS Apps are displayed on the home screen. Note: A red Wireless Connection icon is displayed behind a connected sensor, which indicates that the sensor is successfullyconnected but accessible in the read-only mode. In this mode, you can only view the online parameter values and existing settings. 4. Click the Scan QR Code button, and then scan the QR code provided on the calibration certification.
6.5.3 Using the local display See steps 4 and 5 described in the following Figure 3. Figure 2: Steps to expose the sensor to the system pressure Figure 3: Steps to perform the zero flow calibration 6.
4. Close the ball valve. 5. Unscrew the connection head and pull the flow sensor out of the ball valve. 6.7 Making electrical connection The flow sensor is equipped with two connector plugs “A” and “B”. The cables are connected to the sensor through the M12 connector.
Legend to pin assignment GND: Ground for Modbus SW: Isolated pulse switch input/output SDI: Digital signal (internal use) D+: Modbus data + -VB: D-: Modbus data - +VB: Positive supply voltage M: M-Bus data +l: Positive 4 ... 20 mA signal N/A: Not applicable -I: Negative 4 ... 20 mA signal Negative supply voltage ATTENTION! Do not screw the M12 plug using force. Otherwise, it may damage the connecting pins. 7.
7. Configuration • • Installation FAD: Free Air Delivery calculated based on the total measured volumetric flow at programmed intake conditions (for example, 25°C, 980 hPa, 30 %rH) Working Condition Flow: Working volumetric Flow at fluid pressure and temperature. Select an installation method • Center Installation • 100mm insertion depth 7.1.2 Unit settings Configure units for flow, pressure, temperature, and consumption. 7.1.
7. Configuration 7.2 Configuration using the service App S4C-FS This is the most convenient way especially when the sensor is not equipped with the optional local display. The App is available for Android systems. You can download the service App from the Google Play store or our Website. To change settings using the App, you must scan the QR code on the calibration certificate. This ensures that only valid users can change the sensor settings.
7. Configuration 7.3 Configuration using the optional sensor display With the Sensor display it is possible to show the value of the velocity, the flow and the consumption. Moreover it shows error messages and it is possible to change the configuration setting of the sensor. "Enter key” = “Up key” = "Down key” = 7.3.1 Starting process After power up, the display starts automatically with a initialisation procedure.
7. Configuration Icons are shown in the status bar indicating status or warnings for the sensor in service. The following table provides descriptions for these icons. Icon Description Calibration expired Temperature over operating range Flows over measuring range Pressure over operating range Pressure sensor damaged Temperature sensor damaged Flow direction 7.3.3 Configuration instructions 1. Press the Enter key (>3s). 2. Enter the unlock code:12 using the Up and Down keys, and then enter Enter key. 3.
7. Configuration 8. Sensor signal outputs 8.1 Analog output The sensor has an analogue output range of 4...20 mA. This output can be scaled to match a desired measuring range. Standard is the range from 0 to max. flow depending on the pipe diameter. The corresponding flow in different pipe sizes can be found in the Appendix section. For other ranges, please contact the manufacturer. 8.2 Pulse output The sensor will send out one pulse per consumption unit.
8. Sensor signal outputs In case the flow rate is higher than 50 m3/s, l/s of ft3/s, the S430 can not output the pulses with default settings (one pulse per consumption unit). For this the pulse can be set by our service software or a connected display to 1 pulse per 10 consumption units or 1 pulse per 100 consumption units. For example, if set to 1 pulse per 10 m3, the sensor will send one pulse each 10 m3. Example: S430 Volume flow [m3/h] Pulse length [ms] Max.
8. Sensor signal outputs 8.3 Pulse Connection Diagram Using the isolated pulse switch (Connector B Pin 4 and 5) Variant 1: Variant 2: 8.4 Modbus output Mode : RTU Baud rate : 19200 Device address : 1 Framing / parity / stop bit : 8, E, 1 Response time : 1 second Response delay : 0 ms Inter-frame spacing Remarks : 7 char • 30 Modbus communication settings can be changed by the service software.
8. Sensor signal outputs Inde Channel description x Unit Resolutio n Format Length Modbus address 0 Gas temperature °C 0.1 FLOAT 4 Byte 0 1 Pressure bar 0.001 FLOAT 4 Byte 2 2 Velocity m/s 0.1 FLOAT 4 Byte 4 3 Flow m3/h 0.1 FLOAT 4 Byte 6 4 Consumption m3 1.0 UNIT32 4 Byte 8 5 Reverse velocity m/s 0.1 FLOAT 4 Byte 10 6 Reverse flow m3/h 0.1 FLOAT 4 Byte 12 7 Reverse consumption m3 1.0 UNIT32 4 Byte 14 1.
9. Calibration 9. Calibration The sensor is calibrated ex work. The exact calibration date is printed on the certificate which is supplied together with the sensor. The accuracy of the sensor is regulated by the on site conditions, parameters like oil, high humidity or other impurities can affect the calibration and furthermore the accuracy. However we recommend to calibrate the instrument at least once per year. The calibration is excluded from the instruments warranty.
12. Warranty The warranty is cancelled: • If the user opens the measurement instrument without a direct request written in this instruction manual. • If repairs or modifications are undertaken by third parties or unauthorised persons. • If the serial number has been changed, damaged or removed. Other claims, especially those for damage occurring outside the instrument are not included unless responsibility is legally binding. Warranty repairs do not extend the period of warranty.
Appendix Appendix Scaling table analogue output (standard range): Medium: Air at ISO 1217; 20°C; 1000mbar inch 1/2” 3/4” 1” 1 1/4” 1 1/2” 2” 2 1/2” 3” 4” 5” 6” 8” 10” 12” 34 Tube PLF mm 0.671 16.10 0.722 21.70 0.746 25.00 0.750 26.00 0.756 27.30 0.761 28.50 0.767 30.00 0.776 32.80 0.784 36.00 0.785 36.30 0.792 39.30 0.793 40.00 0.797 41.80 0.800 43.10 0.804 45.80 0.810 50.00 0.811 51.20 0.812 53.10 0.813 54.50 0.818 57.50 0.820 60.00 0.823 64.20 0.824 65.00 0.825 68.90 0.827 70.30 0.827 71.10 0.
S430 35
FCC Statement: This device complies with part 15 of the FCC rules Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. NOTE: The manufacturer is not responsible for any radio or TV interference caused by unauthorized modifications to this equipment. Such modifications could void the user’s authority to operate the equipment.
SUTO iTEC GmbH Werkstr. 2 79426 Buggingen Germany Tel: +49 (0) 7631 936889-0 Fax: +49 (0) 7631 936889-19 Email: sales@suto-itec.com Website: http://www.suto-itec.com All rights reserved © SUTO iTEC (ASIA) Co., Ltd. Room 10, 6/F, Block B, Cambridge Plaza 188 San Wan Road, Sheung Shui, N.T. Hong Kong Tel: +852 2328 9782 Fax: +852 2671 3863 Email: sales@suto-itec.asia Website: http://www.suto-itec.com Modifications and errors reserved.