Power-LOAD™ Cot Fastener System Model 6390 Operations/Maintenance Manual For parts or technical assistance: USA: 1-800-327-0770 2011/10 B.4 6390-009-001 REV B www.stryker.
Table of Contents Symbols and Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Warning/Caution/Note Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Introduction . .
Table of Contents Loading a Power-PRO Cot into a Vehicle (Model 6500/6506 & 6510/6516 with the Power-LOAD Option) . . . . 55 Unloading a Power-PRO Cot from a Vehicle (Model 6500/6506 & 6510/6516 with the Power-LOAD Option) . . 57 Unloading a Cot From a Vehicle Manually (Power-LOAD Power Loss or System Error) . . . . . . . . . . . . . . . . . 59 Loading a Cot into a Vehicle Manually (Power-LOAD Power Loss or System Error) . . . . . . . . . . . . . . . . . . . .
Symbols and Definitions SYMBOLS Warning, consult accompanying documentation Safe Working Load Maximum Weight Dangerous Voltage Pinch Point Class II Equipment: equipment in which protection against electric shock does not rely on basic insulation only, but in which additional safety precautions such as double insulation or reinforced insulation are provided, there being no provision for protective earthing or reliance upon installation conditions.
Symbols and Definitions WARNING/CAUTION/NOTE DEFINITION The words WARNING, CAUTION and Note carry special meanings and should be carefully reviewed. WARNING Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may also describe potential serious adverse reactions and safety hazards.
Introduction This manual is designed to provide instructions for the operation and maintenance of the Stryker 6390 Power-LOAD™. Read it thoroughly before installing or using the equipment or beginning any maintenance on it. PRODUCT DESCRIPTION The Stryker 6390 Power-LOAD is designed to assist the user in loading the cot into a vehicle, unloading the cot from the vehicle, and to restrict the movement of the cot as it is being transported in the vehicle patient compartment under normal conditions.
Introduction IMPORTANT CONTACT INFORMATION For information about Federal Ambulance Specification KKK−A−1822, contact: Chief, Automotive & Commodity Management Branch (QMDAA) Office of Motor Vehicle Management General Services Administration 2200 Crystal Drive, Suite 1006 Arlington, VA 22202 Telephone: 703-605-2277 For more information about AMD standards, contact: Ambulance Manufacturers Division (National Truck Equipment Association) 37400 Hills Tech Drive Farmington Hills, MI 48331−3414 STRYKER CONTACT I
Introduction SPECIFICATIONS Safe Working Load Note: Safe Working Load represents the sum of 870 lb 395 kg the cot total weight and patient Maximum Lift Capacity (patient and accessories) 700 lb 318 kg Overall Length/Minimum Length/Width 95 in / 85 in / 24.5 in 241 cm / 216 cm / 62 cm Weight Floor Plate Assembly 16.5 lb 7.5 kg Anchor Assembly 23 lb 10.5 kg Transfer Assembly 67 lb 30.
Introduction SPECIFICATIONS (CONTINUED) Environmental Conditions Temperature Operation Storage 130 0F (54 0C) -30 0F (-34 0C) 158 0F (70 0C) -30 0F (-34 0C) 93% 93% Relative Humidity 0% 0% Return To Table of Contents 10 6390-009-001 REV B www.stryker.
Summary of Safety Precautions Carefully read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel. WARNING • • • • • • • • • • • • • • • • • • • • • • • • • • • • • Do not operate Power-LOAD using a voltage inconsistent with the rating on the unit. To avoid the risk of equipment damage or smoke hazard, do not use Power-LOAD above its duty cycle. To ensure performance and prevent power hazards, connect Power-LOAD to a 12.8V-15.
Summary of Safety Precautions WARNING (CONTINUED) • • • • • • • • • • • • • • • • • Keep hands and extremities clear of the Power-LOAD trolley lifting arms and the cot base during powered loading and unloading. Return damaged batteries to a service center for recycling. Do not dispose of the battery. To reduce the risk of electric shock, do not remove the battery when Power-LOAD in operation.
Summary of Safety Precautions CAUTION (CONTINUED) • • • • • • • To avoid the risk of equipment damage, do not slam the cot into the trolley when loading. To avoid the risk of equipment damage, do not push the cot into the vehicle patient compartment until the cot base is fully retracted. Do not let go of the manual release until the cot locks into position at the head end. If you let go too early, then the cot base may prevent the cot from properly locking into Power-LOAD.
Vehicle Cot Compatibility Information The Stryker Model 6390 Power-LOAD system is designed to be fully compatible with cots with the Power-LOAD compatible option only.
Installation Setup Procedures FLOOR PLATE ASSEMBLY INSTALLATION COMPONENT CHECKLIST Follow this component checklist to ensure that you have all of the required components to properly install the PowerLOAD system. For installation instructions, see “Installing the Floor Plate (with a Wood Floor)” on page 21. Figure 4: Floor Plate Installation Kit Components (6390-001-055) Return To Table of Contents www.stryker.
Installation Setup Procedures POWER-LOAD ASSEMBLY INSTALLATION COMPONENT CHECKLIST Follow this component checklist to ensure that you have all of the required components to properly install the PowerLOAD system. For installation instructions, see “Installing the Power-LOAD System” on page 27. Figure 5: Power-LOAD Assembly Installation Kit Components (6390-001-054) Return To Table of Contents 16 6390-009-001 REV B www.stryker.
Installation Setup Procedures OPTIONAL WHEEL GUIDE ASSEMBLY INSTALLATION COMPONENT CHECKLIST Follow this component checklist to ensure that you have all of the required components to properly install the PowerLOAD system. For installation instructions, see “Installing The Optional Wheel Guide” on page 34. Figure 6: Optional Wheel Guide Assembly Installation Kit Components (6390-027-000) Return To Table of Contents www.stryker.
Quality System Regulation for Installation QUALITY SYSTEM REGULATION The U.S. Food and Drug Administration (FDA) Code of Federal Regulations (CFR) Title 21 provides guidance regarding the installation of devices, such as Power-LOAD. To comply with these federal regulations, each device must be verified to have been properly installed by trained* individuals by following the inspection criteria in the “Power-LOAD Installation Checklist” on page 38.
Floor Plate Installation Guidelines Select your configuration (see Figure 7, Figure 8, or Figure 9 for reference). Note: All measurements are shown in inches. 1 2 3 4 Choose one hole to drill Ø 1/2 in (1.3 cm) through for electrical input. Router depth to be 9/16 in (1.4 cm) for the floor plate pocket. Floor plate may be cut shorter to either indicated length. Note: The foot end of the floor plate should be 2-3/4 in (7 cm) from the location where the door sill first drops off.