Floor Plate Installation Guidelines 2 1 4 3 3 Sill Edge Sill Figure 8: Power-LOAD Floor Plate with Standard Wall Mount Cot Fastener Wall Plate (Optional) 2 1 4 Sill Edge Sill 3 3 Figure 9: Power-LOAD Floor Plate with Dual Standard Floor Mount Cot Fastener Floor Plates (Optional) Return To Table of Contents 20 6390-009-001 REV B www.stryker.
Installation Guide INSTALLING THE FLOOR PLATE (WITH A WOOD FLOOR) Floor Plate Assembly Kit Components (6390-001-055) Ǩɣ (6) Flat Head Cap Screw (p/n 0001-194-000) Ǩɣ (4) Sub Anchor Attachment Bracket Retainer (p/n 6390-001-110) Ǩɣ (1) Rubber Grommet (p/n 0037-247-000) Ǩɣ (1) Anchor to Vehicle Cable (p/n 6390-001-135) Ǩɣ (1) Wire Protection Loom (p/n 6390-001-153) Ǩɣ (1) Drain Tube (p/n 6390-001-170) Ǩɣ (6) P-Style Clamp (p/n 6390-001-202) Tools Required: Ǩɣ Pencil Ǩɣ Router Ǩɣ Drill Ǩɣ Tape Measure Ǩɣ 1/2”
Installation Guide INSTALLING THE FLOOR PLATE (WITH A WOOD FLOOR) (CONTINUED) WARNING During installation, always wear safety glasses and a face mask while operating a router. 9. Using a router and 1/2” to 1” router bit, router out the floor by following the lines marked onto the floor in step 7. Note: Set the router bit depth by using the floor plate as a depth gauge to 1/2 in (1.3 cm). Use a firm stance when using the router to avoid damaging the floor. 10.
Installation Guide INSTALLING THE FLOOR PLATE (WITH A WOOD FLOOR) (CONTINUED) 19. Using a drill and a 9/16” drill bit, drill the drain hole through the floor plate and the floor (as selected in step 18). 20. Press the drain tube (6390-001-170) onto the drain from the underside of the vehicle. 21. Route the drain tube under the vehicle away from the exhaust system ensuring that the open end of the tube is lower than the drain for proper drainage.
Installation Guide INSTALLING THE FLOOR PLATE (WITH A WOOD FLOOR) (CONTINUED) Install the Anchor Assembly 34. Insert a floor plate attachment bracket (6390-001-108) at the foot end of the vehicle patient compartment to install the safety hook (6390-001-148) (Figure 10). Line the notches with the tabs to insert, then slide the bracket into the floor plate assembly. WARNING Failure to install the safety hook can cause injury to the patient or operator.
Installation Guide INSTALLING THE FLOOR PLATE (WITH A WOOD FLOOR) (CONTINUED) 38. Locate the first floor plate attachment bracket (Figure 13) location (near the foot end). This location provides the most wheel clearance. Notes: Ǩɣ To shift the cot closer to the head end of the vehicle patient compartment, select another location. Ǩɣ Make sure that the location is not further from the first location than the value calculated in step 37 on page 24.
Installation Guide INSTALLING THE FLOOR PLATE (WITH A WOOD FLOOR) (CONTINUED) 42. Carefully place the anchor assembly (A) on top of the floor plate assembly (B) as shown in Figure 15, laying the excess wire into the floor plate channel. Align the four holes with the floor plate attachment brackets. CAUTION When installing the anchor assembly, ensure that the wires rest inside of the floor plate assembly channel and are not pinched by the anchor assembly.
Installation Guide INSTALLING THE POWER-LOAD SYSTEM Power-LOAD Assembly Kit Components (6390-001-054) Ǩɣ (1) Socket Head Cap Screw (p/n 0004-658-000) Ǩɣ Ǩɣ (6) Button Head Cap Screw (0004-665-000) Ǩɣ Ǩɣ (4) Anchor Mounting Post (p/n 6390-001-150) Ǩɣ Ǩɣ (1) Transfer Trim, HE, Right (p/n 6390-001-211) Ǩɣ Ǩɣ (2) Dead Stop Bumper (p/n 6390-001-243) Ǩɣ Ǩɣ (2) Dead Stop Block, Threaded (p/n 6390-001-246) Tools Required: Ǩɣ 3/8” Allen Wrench Ǩɣ T25 Torx Driver (4) Socket Head Cap Screw (p/n 0004-662-000) (1) Trol
Installation Guide INSTALLING THE POWER-LOAD SYSTEM (CONTINUED) Install the Transfer Assembly WARNING To avoid the risk of injury, two installers are required when lifting and positioning the transfer assembly. 3. Position the transfer assembly (A), so the open end is facing the anchor assembly (B) as shown in Figure 18. C B A Figure 18: Install Transfer Assembly 4. Installer 1 (Foot End) - Lift the transfer assembly.
Installation Guide INSTALLING THE POWER-LOAD SYSTEM (CONTINUED) Install the Trolley Assembly 6. 7. Using a 5/32” Allen wrench, install the supplied socket head cap screw (0004-658-000) (A) to secure the supplied trolley magnet activator (6390-0001-106) (B) as shown in Figure 19. Using a T25 Torx driver, install four supplied button head cap screws (0004-665-000) (C) to attach the supplied foot end transfer wear pad (6390-001-225) (D) to the transfer assembly as shown in Figure 19.
Installation Guide INSTALLING THE POWER-LOAD SYSTEM (CONTINUED) 9. Position the trolley assembly (A) between the transfer assembly (B) and head end of the anchor assembly (C) on the floor as shown in Figure 21. 10. Installer 1 (Foot End) - Slide the rollers of the trolley assembly into the side channel of the transfer assembly. 11. Installer 2 (Head End) - Lift the trolley assembly to align the second set of rollers and slide the rollers until the trolley is near the middle of the transfer. 12.
Installation Guide INSTALLING THE POWER-LOAD SYSTEM (CONTINUED) 14. Assemble the bumper block by aligning the supplied dead stop block (6390-001-244) (with through holes) (A) with the supplied dead stop bumper (6390-001-243) (B) and the supplied inner dead stop block (6390-001246) (with threaded holes) (C) as shown in Figure 23. A B C D B A D Figure 23: Assemble Bumper Block 15. Slide the assembled dead stop bumper into the end of the transfer assembly.
Installation Guide INSTALLING THE POWER-LOAD SYSTEM (CONTINUED) 18. Using a T25 Torx driver, install the two supplied button head cap screws (0004-665-000) (A) to attach the supplied patient left transfer head end trim (6390-001-210) and supplied patient right transfer head end trim (6390-001-211) (B) to the transfer assembly as shown in Figure 25. A B Figure 25: Attach Head End Trim Return To Table of Contents 32 6390-009-001 REV B www.stryker.
Installation Guide INSTALLING THE POWER-LOAD SYSTEM (CONTINUED) 19. Press the main power button (A), as shown in Figure 26, on the patient left side of the trolley assembly to turn power on to the unit. Notes: Ǩɣ If you wired Power-LOAD into a circuit with a switch, make sure that the switch is turned on. Ǩɣ When pressing the main power button, a green battery power LED (B) also turns on to indicate that the Power-LOAD system is on.
Installation Guide INSTALLING THE OPTIONAL WHEEL GUIDE Optional Wheel Guide Kit Components (6390-027-000) Ǩɣ (2) Flat Head Cap Screw (p/n 0001-195-000) Ǩɣ (1) Wheel Guide Assembly (p/n 6390-001-017) Ǩɣ (5) Wheel Turner Support (p/n 6390-001-174) Ǩɣ (1) Wheel Kicker Rail (p/n 6390-001-176) Ǩɣ (1) Wheel Turner Nut (p/n 6390-001-191) Tools Required: Ǩɣ Tape Measure Ǩɣ Drill Ǩɣ 3/8” Drive Ratchet Ǩɣ 5/32” Allen Wrench Ǩɣ Ǩɣ Ǩɣ Ǩɣ Ǩɣ (2) Pan Head Machine Screw (p/n 0023-296-000) (5) Spacer (p/n 6390-001-173) (
Installation Guide INSTALLING THE OPTIONAL WHEEL GUIDE (CONTINUED) 2. Using the measurement from step 1, determine if you will need to install the spacers (6390-001-173) (D) for the wheel guide assembly. Measurement from the Center Line (Floor Plate) Greater than 17-1/2 in (44.45 cm) from the wall 13-5/16 in (33.81 cm) to 17-1/2 in (44.45 cm) Exactly 17-1/2 in (44.45 cm) Exactly 13-5/16 in (33.81 cm) Action Required The optional wheel guide is not required. Spacers (6390-001-173) (D) are required.
Installation Guide INSTALLING THE OPTIONAL WHEEL GUIDE (CONTINUED) 3. 4. 5. 6. 7. With the vehicle door closed, align the wheel kicker rail (6390-001-176) (G) against the wall (cabinet). Using a tape measure, measure 1-1/2 in (3.81 cm) from the foot end of the wheel kicker rail and the door panel. Using a pencil, make a small mark on the wall (cabinet) noting the location of the back of the wheel guide rail.
Installation Guide INSTALLING COT WHEEL COVERS Tools Required: Ǩɣ 5/8” Socket Ǩɣ 3/8” Drive Ratchet Ǩɣ 11/16” Combination Wrench Procedure: 1. Tilt the cot onto its head end. Note: Make sure that the head section is extended and locked. 2. 3. 4. 5.
Installation Guide POWER-LOAD INSTALLATION CHECKLIST Note: Allow the battery to charge for 20 minutes prior to starting the Power-LOAD functional check. The battery power LED should flash green while charging.
Installation Guide POWER-LOAD INSTALLATION CHECKLIST (CONTINUED) ______ 20. Load the Power-LOAD compatible cot without using the LOAD functions to simulate manual loading of cots into the vehicle patient compartment. Ensure that the cot locks into place. See “Loading a Cot into a Vehicle Manually (Power-LOAD Power Loss or System Error)” on page 61. ______ 21. Press and hold the release lever at the foot end of the Power-LOAD system and pull to remove the cot from the vehicle patient compartment.