® Wireless Sentry™ Multiple Bed and Chair Monitoring System Operator’s Manual for Receiver WR-3 Transmitter WT-2 Charger WC-1 Pendant WP-2 Floor Mat WFM-1 Repeater WSR-1 UNIVERSAL MEDICAL PRODUCTS, INC. 1620 N. 20th Circle Lincoln, NE 68503-1103 Tel.: 888 716-0077 (Toll Free) or 718 499-0020 Fax: 732 583-6923 or 718 499-6494 E-Mail: information@u-m-p.com WebSite: WWW.U-M-P.
Table of Contents Introduction · · · · · · · · · · · · · · · · · · · · · · · · · 1 Location of Controls Receiver and Transmitter · · · · · · · · · · · · · · 2 Pendant and Repeater· · · · · · · · · · · · · · · · 3 Unpacking· · · · · · · · · · · · · · · · · · · · · · · · · · 4 Setting Up · · · · · · · · · · · · Receiver · · · · · · · · · Transmitter and Charger · Pendant · · · · · · · · · Wireless Floor Mat· · · · Repeater · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
Introduction The Wireless Sentry is a comprehensive multi-patient monitoring system designed to alert care givers at a central location. Using 900 MHz spread-spectrum digital technology, this completely wireless system provides extended range performance for up to sixteen locations. The system utilizes five main components – the 16-Station Receiver unit, the Wireless Transmitter with Charging Cradle (Charger), the Wireless Pendant, the Wireless Repeater and the Wireless Floor Mat.
Location of Controls 2
Wireless Pendant (WP-2) LED Battery Compartment (on back of unit) Alarm Button Repeater (WSR-1) 3
Unpacking In the carton marked Wireless Sentry™ Receiver Model WR-3, you should find the following items: 1 Receiver unit (WR-3) 1 Pedestal 2 wall-mounting screws 1 AC Adapter with cable 1 Index Card 1 Set of Instructions In the carton marked Wireless Sentry™ Transmitter Model WT-2, you should find the following items: 1 Transmitter unit (WT-2) with rechargeable battery installed 1 Charging Cradle (WC-1) 1 AC Adapter with cable 1 Set of Instructions The Wireless Pendant™ Model WP-2 carton should contain the
Setting Up Receiver Setup 1. Plug the adapter cable into the back of the Receiver, then plug the adapter into the wall outlet. 2. Turn on the unit by pressing the POWER button. The red LED power indicator will go on. 3.. Program the Receiveras described below under Programming. Station Index Card: Use your fingernail or a coin to remove the clear plastic cover from the front of the Receiver.
Setting up the Charging Cradle: Plug the adapter cable into the Charging Cradle (Charger) and then into the wall outlet. Setting up the Transmitter before programming: 1. Before using a particular Transmitter unit for the first time, charge it for at least 14 hours by placing it into the Charging Cradle. The red LED on the Charging Cradle will light when the Transmitter is properly inserted. This procedure will help lengthen the life of the battery. 2.
Wireless Floor Mat Setup Replacing the Batteries: The Floor Mat’s transmitter comes with new batteries installed. When it becomes necessary to replace them, use the following procedure: 1. Use a coin to remove the battery cover on the transmitter by placing it in the slot and turning it slightly in a clockwise direction. 2. Pry out the old batteries with a pen or small screwdriver. 3. Place three new batteries into position, being sure to align the polarities correctly.
The Repeater can also be set up on a desk or counter by attaching the back of the unit to the plastic Pedestal that is included in the package. This sets the unit at a convenient angle on a flat surface. 5. Once the Repeater is set up, no further attention is required since all alarm signals will be passed along automatically to the Receiver. Programming Before programming, turn on the Receiver by pressing the POWER button. The red LED power indicator will light.
To Program a Pendant Location: 1. Press the PROGRAM button on the Receiver for 2 seconds. The green LED will light and the LED indicating the first empty station location will flash. 2. Press the PROGRAM button to select the desired empty location. The LED of each empty station will light up in turn each time the button is pushed. 3. Press and hold the ALARM button on the Pendant until you hear a long beep from the Receiver and the station LED stops flashing. 4.
To Program a Repeater: NOTE: Repeaters can only be programmed into the first four station locations of the Receiver. 1. Press the PROGRAM and CONFIRM buttons on the Receiver for two seconds. The green LED will light. The first available station LED (of Locations 1 through 4) will flash. 2. Press the CONFIRM button on the Repeater until you hear a long beep from the Receiver and the station LED stops flashing. 3. Press CONFIRM on the Receiver to return to normal monitor mode.
Checking for Programmed Locations To see which stations are active and thus being monitored at any time, press and hold the CONFIRM button on the Receiver. The LEDs of the stations being actively monitored will light up. Erasing a Programmed Location 1. Press PROGRAM and ALARM RESET together on the Receiver and hold for 2 seconds. The first programmed location LED will flash. 2. Press PROGRAM to select the next location. 3. Press ALARM RESET to erase a location.
3. If more than one alarm is received, pressing the ALARM RESET button again will select the next alarm location. 4. Press the CONFIRM button to select the station to visit. The beep at the Receiver will stop but the LED will remain on. 5. When the problem causing the Alarm situation is corrected, the Transmitter will send a signal to the Receiver to turn off the LED. NOTE: If the alarm was from a Pendant, the LED will be off at this point.
Out-of-Range Monitoring (for Wireless Transmitter only) After being programmed into the Receiver, the Transmitter will send silent signals to the Receiver approximately every 90 seconds to indicate that it is within a safe range. If the Transmitter is carried beyond the safe range (Out-of-Range), the Receiver will emit light and sound indications to alert the care giver. These alerting signals can be reset in the same way as a normal alarm.
2. Set the Transmitter into the Charger, being sure that the charging contacts on the bottom of the Transmitter touch the corresponding contacts in the Charger. 3. The red LED on the Transmitter will flash when the battery is fully charged. Monitoring While Charging: NOTE: SPECIAL FEATURE! The Transmitter can monitor a patient even while it is recharging in the Charging Cradle. To do this: 1. Disconnect the monitoring pad and re-connect it to the jack on the Charger. 2. Set the Transmitter into the Charger.
Safety Instructions: Transmitter WT-2 Battery Pack CAUTION TO REDUCE THE RISK OF FIRE OR INJURY, READ AND FOLLOW THESE INSTRUCTIONS: 1. Use only a battery pack supplied by Universal Medical Products, Inc., in Transmitter WT-1 of the UMP Wireless Sentry™ monitoring system. 2. Do not dispose of the battery pack in a fire. The cells may explode. Check with local codes for possible special disposal instructions. 3. Do not attempt to open or mutilate the battery pack.
FCC Warning Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. Radio Frequency Interference This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
Limited Warranty UNIVERSAL MEDICAL PRODUCTS, INC., guarantees the quality and workmanship of the Wireless Sentry™ Receiver and each Transmitter, Charger, Pendant, Floor Mat and Repeater for a period of one (1) year from date of purchase. This warranty is in force provided there is no evidence of misuse or improper installation. Under the conditions of this warranty, UNIVERSAL MEDICAL PRODUCTS, INC.