Table of contents Dentsply Sirona Instructions for Use Primeprint Table of contents 1 General data ........................................................................................................... 1.1 1.2 1.3 1.4 Dear Customer,............................................................................................ Contact data................................................................................................. General information about this operating manual ....................
Dentsply Sirona Instructions for Use Primeprint 3.4 Technical description (component and interfaces) ...................................... 3.4.1 Major components ......................................................................... 3.4.1.1 Front view ..................................................................... 3.4.1.2 Connections.................................................................. 3.4.1.3 Build chamber............................................................... 3.4.1.
Table of contents Dentsply Sirona Instructions for Use Primeprint 5.3.6 6 7 Installing and configuring the unit .................................................. 5.3.6.1 Installing the unit........................................................... 5.3.6.2 Configuring the device.................................................. 43 43 45 Operation ................................................................................................................ 47 6.1 6.2 Referencing run ............
Dentsply Sirona Instructions for Use Primeprint 1 General data 1.1 Dear Customer, 1 General data 1.1 Dear Customer, ® Thank you for your purchase of this Primeprint unit from Dentsply Sirona. This unit enables you to use computer aided manufacturing of dental applications. Improper use and handling can create hazards and cause damage. Please therefore read and follow these operating instructions carefully. Always keep them within easy reach.
1 General data 1.3 General information about this operating manual Dentsply Sirona Instructions for Use Primeprint 1.3 General information about this operating manual Observe the Operating Instructions Please familiarize yourself with the unit by reading through these Operating Instructions before putting it into operation. It is essential that you comply with the specified warning and safety information.
Dentsply Sirona Instructions for Use Primeprint 1 General data 1.4 General conventions and structure of the document 1.4 General conventions and structure of the document 1.4.1 Structure of the document 1.4.1.1 Identification of the danger levels To prevent personal injury and material damage, please observe the warning and safety information provided in these instructions for use. Such information is highlighted as follows: DANGER An imminent danger that could result in serious bodily injury or death.
1 General data 1.5 Scope of these Operating Instructions Dentsply Sirona Instructions for Use Primeprint 1.4.1.2 Formats and symbols used The formats and symbols used in this document have the following meaning: Prerequisite Requests you to do something. 1. First action step 2. Second action step or ➢ Alternative action Result ➢ Individual action step See "Formats and symbols used [→ 8]" Identifies a reference to another text passage and specifies its page number. ● List Designates a list.
Dentsply Sirona Instructions for Use Primeprint 1 General data 1.6 Warranty and liability 1.6 Warranty and liability Maintenance In the interest of the safety and health of patients, users or third parties, it is necessary that maintenance work is carried out at fixed time intervals to ensure the operational safety and reliability of your product. The operator must ensure the implementation of the maintenance work.
1 General data 1.7 Legend Dentsply Sirona Instructions for Use Primeprint 1.7 Legend Year of manufacture 20XX Product disposal symbol (see "Disposal [→ 66]"). Accompanying documents This symbol can be found on the rating plate on the unit. Meaning: Observe the Operating Instructions when operating the unit. This symbol can be found on the rating plate on the unit. Meaning: The accompanying documents are available on the Dentsply Sirona homepage.
Dentsply Sirona Instructions for Use Primeprint 1 General data 1.7 Legend Symbols on the packaging Take note of the following symbols on the packaging: Top Protect from moisture Fragile; handle with care Do not stack Temperature during storage and transport Relative humidity during storage and transport Air pressure during storage and transport 67 45 173 D3691 D3691.201.01.01.02 04.
2 Safety instructions 2.1 Basic safety information Dentsply Sirona Instructions for Use Primeprint 2 Safety instructions 2.1 Basic safety information 2.1.1 Prerequisites NOTE Important information on building installation In order to prevent the risk of an electric shock, this unit must only be connected to a supply mains with a ground wire. The building installation must be performed by a qualified expert in compliance with the national regulations.
Dentsply Sirona Instructions for Use Primeprint 2 Safety instructions 2.2 Electromagnetic compatibility 2.1.4 Accessories In order to ensure product safety, this device may be operated only with original Dentsply Sirona accessories or third-party accessories expressly approved by Dentsply Sirona. In particular, only the power cable also supplied or the corresponding original spare part may be used with the unit. The user is responsible for any damage resulting from the use of non-approved accessories. 2.
2 Safety instructions 2.2 Electromagnetic compatibility Dentsply Sirona Instructions for Use Primeprint 2.2.2 Immunity to interference The UNIT is intended for operation in the electromagnetic environment specified below. The customer or user of the UNIT should make sure that it is used in such an environment.
Dentsply Sirona Instructions for Use Primeprint 2 Safety instructions 2.2 Electromagnetic compatibility Interference immunity tests DIN EN 61326-1 Test level Compliance level Electromagnetic environment – guidelines Conducted RF interference IEC 61000-4-6 3 Veff 150 kHz to 80 MHz 3 Veff d= [1.2] √P Radiated RF interference IEC 61000-4-3 3 V/m 80 MHz to 800 MHz 3 V/m d= [1.2] √P at 80 MHz to 800 MHz 3 V/m 800 MHz to 2.5 GHz 3 V/m d= [2.3] √P at 800 MHz to 2.
2 Safety instructions 2.2 Electromagnetic compatibility Dentsply Sirona Instructions for Use Primeprint 2.2.3 Working clearances Recommended working clearances between portable and mobile RF communication devices and the UNIT Rated maximum output power of transmitter [W] The UNIT is intended for operation in an electromagnetic environment, where radiated RF interference is checked.
Dentsply Sirona Instructions for Use Primeprint 2 Safety instructions 2.3 Electrostatic charge 2.3 Electrostatic charge 2.3.1 ESD protective measures ESD ESD stands for ElectroStatic Discharge. ESD protective measures ESD protective measures include: ● Procedures for preventing electrostatic charge build-up (e.g.
2 Safety instructions 2.4 Connecting the unit Dentsply Sirona Instructions for Use Primeprint Background Integrated circuits (logical circuits and microprocessors) are used to implement a wide variety of functions in dental/X-ray/CAD/CAM systems. The circuits must be miniaturized to a very high degree in order to include as many functions as possible on these chips. This leads to structure thicknesses as low as a few ten thousandths of a millimeter.
Dentsply Sirona Instructions for Use Primeprint 2 Safety instructions 2.7 Product-specific safety instructions 2.7 Product-specific safety instructions CAUTION Status LED illuminated in red In the event of any faults, the status bar lights up red. ➢ Turn off the machine at the main switch before you reach into the processing area. You can restart the machine once the error is eliminated.
3 General information on the device 3.1 Standards/ approvals/ certifications Dentsply Sirona Instructions for Use Primeprint 3 General information on the device 3.1 Standards/ approvals/ certifications CE mark This product bears the CE mark in accordance with the provisions of Council Directives 2014/53/EU (RED), 2006/42/EC (Machinery Directive), 2014/30/EC (EMC Directive) and 2011/65/EU (RoHS) including 2015/863/EC as Annex II. As such, the following standards apply: ETSI EN 301 489-3 V2.1.
Dentsply Sirona Instructions for Use Primeprint 3 General information on the device 3.1 Standards/ approvals/ certifications Compliance statement Changes or modifications not expressly approved by the manufacturer could void the user’s authority to operate the equipment. This device complies with Part 15 of the FCC rules.
3 General information on the device 3.2 Intended use Dentsply Sirona Instructions for Use Primeprint 3.2 Intended use The Dentsply Sirona 3D printer is designed to produce threedimensional dental applications from printable viscous materials. It hereby represents a part of an overall system of Computer Aided Imaging (data acquisition), Computer Aided Design (modelling of the application) and Computer Aided Manufacturing (production). The product only covers the initial manufacturing process.
Dentsply Sirona Instructions for Use Primeprint 3 General information on the device 3.3 Scope of supply 3.3 Scope of supply The detailed scope of supply is specified in the document "Checklist". 3.4 Technical description (component and interfaces) 3.4.1 Major components 3.4.1.1 Front view B C A 67 45 173 D3691 D3691.201.01.01.02 04.
3 General information on the device 3.4 Technical description (component and interfaces) Dentsply Sirona Instructions for Use Primeprint 3.4.1.2 Connections 24 A USB 1 type A D Fuse cover B USB 1 Type B E Main switch 1 = ON, 0 = OFF C LAN F Power connection Connected loads: 100 V AC to 240 V AC 50/60 Hz 2.0 A – 0.85 A Fusing: T5A, H250V 67 45 173 D3691 D3691.201.01.01.02 04.
Dentsply Sirona Instructions for Use Primeprint 3 General information on the device 3.4 Technical description (component and interfaces) 3.4.1.3 Build chamber CAUTION Risk of burning The Primeprint 3D printer uses a heating system integrated into the build chamber. Make sure that you do not touch this when loading and or unloading with the material unit or when reaching into the build chamber. The areas marked in red in the adjacent figure become hot. A B C 67 45 173 D3691 D3691.201.01.01.02 04.
3 General information on the device 3.4 Technical description (component and interfaces) Dentsply Sirona Instructions for Use Primeprint 3.4.1.4 Primeprint box NOTE Replace damaged transport containers Damaged (broken, torn etc.) transport containers must not be used any further and must be replaced by a corresponding spare part (REF 6744887). A B 26 A+B Primeprint box (REF 6744895) A Build platform B Transport container (REF 6744887) 67 45 173 D3691 D3691.201.01.01.02 04.
Dentsply Sirona Instructions for Use Primeprint 3 General information on the device 3.4 Technical description (component and interfaces) 3.4.1.5 Primeprint material unit A B C 67 45 173 D3691 D3691.201.01.01.02 04.
3 General information on the device 3.4 Technical description (component and interfaces) Dentsply Sirona Instructions for Use Primeprint 3.4.
Dentsply Sirona Instructions for Use Primeprint 3 General information on the device 3.5 Technical data 3.5 Technical data Type designation Primeprint Rated line voltage 100 V AC to 240 V AC Rated power frequency 50/60 Hz Nominal current 2.0 A - 0.
4 Transportation and installation 4.1 unpacking Dentsply Sirona Instructions for Use Primeprint 4 Transportation and installation 4.1 unpacking All products from Dentsply Sirona are carefully checked prior to shipment. Please perform an incoming inspection immediately after delivery. 1. Check the delivery note to ensure that the consignment is complete. 2. Check whether the product shows any visible signs of damage.
Dentsply Sirona Instructions for Use Primeprint 4 Transportation and installation 4.3 Installation requirements 4.3 Installation requirements 4.3.1 Requirements on the place of installation Make sure that the unit is installed on an even, horizontal surface of approx. 530x515 mm (W x D). The load carrying capacity should be at least 50 kg. The height of the machine is 670 mm. Install the machine so that access to the power plug is guaranteed at all times.
4 Transportation and installation 4.4 Connecting the unit to the power supply and existing network Dentsply Sirona Instructions for Use Primeprint 4.4 Connecting the unit to the power supply and existing network Connecting the unit to the power supply Only use the power cord supplied in the package for connecting to the mains. The unit must be connected directly to a fixed socket and must not be bridged via an extension cable. ➢ Connect the unit to the power supply.
Dentsply Sirona Instructions for Use Primeprint 4 Transportation and installation 4.6 Repacking 4.6 Repacking ü The Primeprint material unit is removed. ü The Primeprint box is removed. ü The main switch on the back side of the unit is set to the 0 (OFF) position. 1. Disconnect the power cable and the connecting cable from the back side of the unit and stow them away. 2. Check the unit for completeness according to the scope of supply! 3. Pack the unit securely and use the transport locks provided.
5 Commissioning 5.1 Putting the inLab 4 PC into operation Dentsply Sirona Instructions for Use Primeprint 5 Commissioning 5.1 Putting the inLab 4 PC into operation Observe the following steps in order to put the unit into operation using the inLab 4-PC (optional). 5.1.1 Startup prerequisites 5.1.1.1 Required accessories Supplied parts: ● inLab PC with power cable ● Keyboard ● Mouse ● Power cable ● Ethernet cable for connection of the machine with the PC.
Dentsply Sirona Instructions for Use Primeprint 5 Commissioning 5.1 Putting the inLab 4 PC into operation 5.1.1.2 Making connections NOTE Do not connect and turn on the device yet! First, connect the PC to the monitor, keyboard, and mouse. Follow the instructions in this document exactly in order to successfully perform start-up of your device. You must install the user software prior to connecting the device to the PC.
5 Commissioning 5.1 Putting the inLab 4 PC into operation Dentsply Sirona Instructions for Use Primeprint 5.1.2 Safety The inLab 4-PC is equipped with Windows 10. The Windows Firewall is activated. The software Microsoft Security Essentials is also preinstalled by default. Please activate the automatic update function in the settings of this software for optimal protection.
Dentsply Sirona Instructions for Use Primeprint 5 Commissioning 5.1 Putting the inLab 4 PC into operation 5.1.3 Installing the software NOTE Initial installation without device Perform the initial installation of the software without the connected device. The software requires the 2.00 firmware version of the license stick. Update the firmware version if necessary. For additional information, refer to the "License manager" section in the user manual inLab CAM.
5 Commissioning 5.2 Connection to the PC of the acquisition unit Dentsply Sirona Instructions for Use Primeprint 5.2 Connection to the PC of the acquisition unit 5.2.1 Connecting to the PC via WLAN (option) Making the connection Connect access point ➢ Connect the LAN port A of the production unit and the access point, using the network cable (10 m, REF 6151521). Ä The access point is pre-configured at the factory for this application. 38 67 45 173 D3691 D3691.201.01.01.02 04.
Dentsply Sirona Instructions for Use Primeprint 5 Commissioning 5.2 Connection to the PC of the acquisition unit Positioning the access point 1. As a test, place the access point near the production unit at head level or higher. 2. Perform a communication test as described in the separate instructions (see "Operating MC XL via WLAN in infrastructure mode", section "Final work, analyzing connection quality"). If applicable, follow the instructions on changing channels. 3.
5 Commissioning 5.3 Putting the unit into operation Dentsply Sirona Instructions for Use Primeprint 5.3 Putting the unit into operation 5.3.1 Functional elements 5.3.1.1 Touch display A Unit status B Material unit management C Configuration These operating instructions describe how to operate the unit by executing and confirming commands via the touch display. All inputs that are required on the unit can be made here. 5.3.1.
Dentsply Sirona Instructions for Use Primeprint 5 Commissioning 5.3 Putting the unit into operation 5.3.2 Removing the transport safety device and putting the build chamber into operation The Primeprint unit is equipped with a transport safety device that must be removed during commissioning. To do this, proceed as follows: 1. Open the door. 2. Take out the transport safety device. 5.3.
5 Commissioning 5.3 Putting the unit into operation Dentsply Sirona Instructions for Use Primeprint 5.3.5 Switching the unit ON and OFF NOTE Do not put the unit into operation at low temperatures! If you move the unit to the operating site from a cold environment, condensation may form and result in a short circuit. Within the machine, grease depots are included for lubricating components that can cause error messages at low temperatures. ü Install the unit at room temperature.
Dentsply Sirona Instructions for Use Primeprint 5 Commissioning 5.3 Putting the unit into operation 5.3.6 Installing and configuring the unit 5.3.6.1 Installing the unit You must connect the unit to the PC before putting it into operation. This is described in the section "Connecting the PC/interfaces". IMPORTANT The Primeprint unit can be operated with both a static as well as an automatically assigned IP address. The Primeprint unit is always configured for operation ex works with a static IP address.
5 Commissioning 5.3 Putting the unit into operation Dentsply Sirona Instructions for Use Primeprint 5.3.6.1.3 Updating devices With the "Refresh Devices" button you can ● display the status; e.g. check whether a machine has in the meantime finished producing, or ● Check the current availability of a device. 5.3.6.1.4 Remove the unit If you no longer require a unit (e.g. a unit is replaced), you can remove it. ü The unit is not in operation. 1.
Dentsply Sirona Instructions for Use Primeprint 5 Commissioning 5.3 Putting the unit into operation 5.3.6.2 Configuring the device In the "Machine and Instrument tray Management" software "inLab CAM" area you can make subsequent amendments to the various settings for your unit. 1. Click the "Configuration" button in the system menu. 2. Click on the "Machine and Instrument tray Management" button. 3. Click on the unit that you wish to configure. 5.3.6.2.
5 Commissioning 5.3 Putting the unit into operation Dentsply Sirona Instructions for Use Primeprint 4. Confirm the new network settings with "Ok" or press "Cancel" in order not to save the changed settings. 5. If you have changed the setting to "Automatic" you must now disconnect the Ethernet connection to the PC and connect the machine to the network socket in your local network. 6. If you wish to change the settings back again, then change over to "Manual IP settings".
Dentsply Sirona Instructions for Use Primeprint 6 Operation 6.1 Referencing run 6 Operation 6.1 Referencing run NOTE Observe the Operator’s Manual Please also observe the information in the Operator's Manual for inLab CAM SW software. The referencing run is used for the function check of the sensors and the position check of movable parts in the build chamber. The referencing run takes place automatically via the firmware.
6 Operation 6.2 Production process Dentsply Sirona Instructions for Use Primeprint 6.2 Production process 3D printing is done on the build platform, which is placed in the 3D printer with the Primeprint material unit. 6.2.1 Equipping the Primeprint material unit The Primeprint material unit consists of the lower material unit vat and the upper material unit cover. The desired cartridge is inserted into the Primeprint material unit.
Dentsply Sirona Instructions for Use Primeprint 6 Operation 6.2 Production process 2. Open the door of the unit. 3. Optional: Unlock the material unit lock by turning it counterclockwise. 4. Optional: Remove the currently used material unit. 5. Insert a correct material unit. Make sure that the material unit is correctly guided in the guide rails provided. 6. The 3D printer recognizes and checks the Primeprint material unit used. 7.
6 Operation 6.2 Production process Dentsply Sirona Instructions for Use Primeprint 6.2.3 Loading with Primeprint box CAUTION Risk of injury when reaching into the build chamber There is the risk of cut and crush injuries caused by sharp edges and movable parts. ➢ Be sure not to touch the sharp edges and movable parts. During each change, ensure clean contact surfaces of all parts; otherwise, the fixation and the correct positioning cannot be guaranteed.
Dentsply Sirona Instructions for Use Primeprint 6 Operation 6.2 Production process 5. Rotate the Primeprint box back into the starting position. 6. The 3D printer recognizes and checks the inserted unit and, if necessary, the touch display prompts you to insert a correct build platform into the unit. 7. To start the 3D printer, close the door and follow the instructions on the touch display. 6.2.4 Process start Follow the instructions that appear on the touch display. 67 45 173 D3691 D3691.201.01.01.