Table of contents Dentsply Sirona Operating Instructions CEREC Primemill Table of contents 1 2 3 Dear Customer,....................................................................................................... 5 1.1 Contact data................................................................................................. 5 General data ........................................................................................................... 6 2.1 2.2 2.
Dentsply Sirona Operating Instructions CEREC Primemill Table of contents 5.4.3 5.4.4 Touch display description .............................................................. Illumination of the processing chamber, LED light strip, and On/ Off button ....................................................................................... 5.4.5 Using the processing chamber screen........................................... 5.4.6 Installation....................................................................
Table of contents 7.4 Instruments .................................................................................................. 7.4.1 Overview of materials/instruments................................................. 7.4.2 Changing instruments .................................................................... Care, cleaning agents, and disinfectants ..................................................... Cleaning surfaces ...............................................................................
Dentsply Sirona Operating Instructions CEREC Primemill 1 Dear Customer, 1.1 Contact data 1 Dear Customer, ® Thank you for your purchase of this CEREC Primemill unit from Dentsply Sirona. This device enables you to produce dental restorations, e.g. from ceramic material with a natural appearance (CEramic REConstruction). Improper use and handling can create hazards and cause damage. Please therefore read and follow these operating instructions carefully. Always keep them within easy reach.
2 General data 2.1 Identification of the danger levels Dentsply Sirona Operating Instructions CEREC Primemill 2 General data Please read this document completely and follow the instructions exactly. You should always keep it within reach. Original language of the present document: German. 2.1 Identification of the danger levels To prevent personal injury and material damage, please observe the warning and safety information provided in these operating instructions.
Dentsply Sirona Operating Instructions CEREC Primemill 2 General data 2.2 Formats and symbols used 2.2 Formats and symbols used The formats and symbols used in this document have the following meaning: Prerequisite Requests you to do something. 1. First action step 2. Second action step or ➢ Alternative action Result ➢ Individual action step See "Formats and symbols used [→ 7]" Identifies a reference to another text passage and specifies its page number. ● List Designates a list.
3 General description 3.1 Certification Dentsply Sirona Operating Instructions CEREC Primemill 3 General description 3.1 Certification CE mark Sirona Dental Systems GmbH hereby declares that the CEREC Primemill radio system type complies with Directive 2014/53/EU. The full text of the EU Declaration of Conformity is available at Dentsply Sirona Internet homepage. This product bears the CE mark in accordance with the provisions of Council Directive 2006/42/EC (machinery directive).
Dentsply Sirona Operating Instructions CEREC Primemill 3 General description 3.2 Intended use 3.2 Intended use The Sirona Dental CAD/CAM System is intended for use in partially or fully edentulous mandibles and maxillae in support of single or multipleunit cement retained restorations. For the BH 3.0 S, SSO 3.5 L and SBL 3.3 L titanium bases, the indication is restricted to the replacement of single lateral incisors in the maxilla and lateral and central incisors in the mandible.
3 General description 3.2 Intended use Manufacturer Biomet 3i Dentsply Sirona Operating Instructions CEREC Primemill Name of Implant System Implant Size Platform Diameter Length Osseotite 3.4 3.25 8.5mm – 18mm 4.1 3.75 8.5mm – 20mm 4.1 3/4 5.0 5.0 7mm – 18mm 4/5 Certain 3.4 3.25 8.5mm – 18mm 4/3 3/4/3 4.1 4.0 8.5mm – 20mm 4/5/4 5/4 5.0 5.0 8.5mm – 15mm 4/5 Zimmer Thommen Medical Osstem/Hiossen Tapered Screw-Vent Thommen Medical Implants Osstem TS Implant System 3.5 3.7/4.
Dentsply Sirona Operating Instructions CEREC Primemill Manufacturer BioHorizons (Internal Connection) 3 General description 3.2 Intended use Name of Implant System Implant Size Platform Diameter Length Tapered 3.0, Tapered plus 3.0 3.0/3.4/3.8 9mm – 18mm 3.0 10.5mm – 15mm 4.6 7.5mm – 15mm Tapered internal, Tapered internal tissue level 3.0/3.8 9mm – 18mm Internal dental implant 3.5 9mm – 15mm Single stage dental implants 3.5/4.0 9mm – 15mm 5.8 7.
3 General description 3.3 Further use of Sirona Dental CAD/CAM system Dentsply Sirona Operating Instructions CEREC Primemill WARNING Small diameter implants and angled abutments are not recommended for the posterior region. CAUTION Federal Law (USA) restricts the sale of this device to or on the order of a physician, dentist, or licensed practitioner. 3.
Dentsply Sirona Operating Instructions CEREC Primemill 3 General description 3.4 Legend 3.4 Legend This symbol can be found on the rating plate on the unit. Meaning: See warning notice in section "Replacing the main fuses [→ 60]". This symbol can be found on the door of the unit. Meaning: See warning in section "Opening the processing chamber door during the machining process [→ 16]". This symbol can be found on the rating plate on the unit.
3 General description 3.4 Legend Dentsply Sirona Operating Instructions CEREC Primemill Symbols on the packaging Take note of the following symbols on the packaging: Top Protect from moisture Fragile; handle with care Stack limit Temperature during storage and transport Relative humidity during storage and transport Air pressure during storage and transport 14 67 19 756 D3439 D3439.201.01.01.02 09.
Dentsply Sirona Operating Instructions CEREC Primemill 4 Safety 4.1 Basic safety information 4 Safety 4.1 Basic safety information 4.1.1 Prerequisites NOTE Important information on building installation In order to prevent the risk of an electric shock, this unit must only be connected to a supply mains with a ground wire. The building installation must be performed by a qualified expert in compliance with the national regulations.
4 Safety 4.2 Opening the processing chamber door during the machining process Dentsply Sirona Operating Instructions CEREC Primemill 4.1.4 Accessories and consumables In order to ensure reliable, high-quality results, product safety, and durability, our range of CEREC Primemill production units must only be operated with original accessories and consumables from Dentsply Sirona or approved accessories and consumables from third-party suppliers.
Dentsply Sirona Operating Instructions CEREC Primemill 4 Safety 4.3 Electrostatic charge 4.3.2 ESD protective measures ESD ESD stands for ElectroStatic Discharge. ESD protective measures ESD protective measures include: ● Procedures for preventing electrostatic charge build-up (e.g.
4 Safety 4.4 Wireless phone interference with equipment Background Dentsply Sirona Operating Instructions CEREC Primemill Integrated circuits (logical circuits and microprocessors) are used to implement a wide variety of functions in dental/X-ray/CAD/CAM systems. The circuits must be miniaturized to a very high degree in order to include as many functions as possible on these chips. This leads to structure thicknesses as low as a few ten thousandths of a millimeter.
Dentsply Sirona Operating Instructions CEREC Primemill 4 Safety 4.5 Disturbance of data transmission 4.5 Disturbance of data transmission Note on wireless communication Data communication between the acquisition unit and the CEREC Primemill production unit should preferably be established via the wireless interface CEREC Radio Device or WLAN. As for all wireless connections (e.g. cell phones), heavy utilization of the available radio channels or shielding caused by building installations (e.g.
5 Installation and startup 5.1 Transport and unpacking Dentsply Sirona Operating Instructions CEREC Primemill 5 Installation and startup 5.1 Transport and unpacking All products from Dentsply Sirona are carefully checked prior to shipment. Please perform an incoming inspection immediately after delivery. 1. Check the delivery note to ensure that the consignment is complete. 2. Check whether the product shows any visible signs of damage.
Dentsply Sirona Operating Instructions CEREC Primemill 5 Installation and startup 5.3 Installation site 5.3 Installation site 150 0 59”mm 1500mm 59” 0mm 15059” 1500mm 59” 1500mm 59” 1500mm 59” 1500mm 59” 1500mm 59” 1500mm 59” 1500mm 59” 1500mm 59” CAUTION Install out of the reach of patients! Do not install or operate the production unit in the vicinity of the patient (place it at least 1.5 m away from the patient). The production unit requires a level base of approx. 729 mm x 465 mm (W x D).
5 Installation and startup 5.3 Installation site Dentsply Sirona Operating Instructions CEREC Primemill NOTE Installation in a cabinet If the unit is installed in a cabinet, you must provide for adequate heat exchange. The ambient temperature surrounding the unit must be between 5 °C (41 °F) and 40 °C (104 °F). CAUTION Risk of injury and damage to the unit The unit can be tilted when the drawer is extended. ➢ Install the unit so that the front does not project beyond the base.
Dentsply Sirona Operating Instructions CEREC Primemill 5 Installation and startup 5.4 Commissioning 5.4 Commissioning NOTE Important information on initial startup Observe the software installation instructions! 5.4.1 Functional elements Unit overview F A B E D C H G Production unit overview 67 19 756 D3439 D3439.201.01.01.02 09.
5 Installation and startup 5.4 Commissioning Dentsply Sirona Operating Instructions CEREC Primemill Ports on the back side B A C D E F G H Connections A Main switch I = ON, 0 = OFF E LAN B Fuse cover F USB B C Power connection G USB 1 USB 2 D 24 Communications interface for suction H Connection for suction 67 19 756 D3439 D3439.201.01.01.02 09.
Dentsply Sirona Operating Instructions CEREC Primemill 5 Installation and startup 5.4 Commissioning Processing chamber B A B A E F C G D Processing chamber 67 19 756 D3439 D3439.201.01.01.02 09.
5 Installation and startup 5.4 Commissioning Dentsply Sirona Operating Instructions CEREC Primemill 5.4.2 Supplied tools 5.4.2.1 Instruments The following instruments are available for milling and grinding purposes. When replacing instruments, ensure the permitted instrument combinations are used (see "Permitted instrument combinations [→ 44]"). 5.4.2.2 Calibration pins The calibration pins are used when calibrating the instrument sets (see "Calibrating the unit [→ 41]"). 5.4.2.
Dentsply Sirona Operating Instructions CEREC Primemill 5 Installation and startup 5.4 Commissioning 5.4.3 Touch display description In these Operating Instructions, operation is described in such a manner that you can execute and confirm commands via your PC or the touch display, such as "Start", "Stop", "Cancel" or "OK". Other possible commands also have buttons on the touch display. 5.4.
5 Installation and startup 5.4 Commissioning Dentsply Sirona Operating Instructions CEREC Primemill Machining operation Sub-process Processing chamber LED light strip ON/OFF button Maintenance Action required – Close door, change tool White, constant Yellow, constant Green, constant Longer actions; e.g. calibration White, constant Progress bars Green, constant displayed in blue and white, fully in green Shorter actions; e.g.
Dentsply Sirona Operating Instructions CEREC Primemill 5 Installation and startup 5.4 Commissioning 5.4.6 Installation 5.4.6.1 Connecting to the PC via LAN An Ethernet port is located on the rear of the unit, which can be used to connect the PC to the production unit. Use a network cable to do this (LAN connection). Using a network cable Connect the PC to the LAN connection of the unit. 5.4.6.
5 Installation and startup 5.4 Commissioning Dentsply Sirona Operating Instructions CEREC Primemill 5.4.6.3.2 Process steps that have to be performed on the acquisition unit or the PC Automatic unit search ü The unit is connected to the PC via a LAN cable or via WLAN. 1. Click the "Configuration" button in the system menu. 2. Click on the "Devices" button. 3. Click on the "Scan for New Devices" button. Ä All units connected to the PC are recognized.
Dentsply Sirona Operating Instructions CEREC Primemill 5 Installation and startup 5.4 Commissioning 5.4.6.4 Connecting the suction device (optional) 5.4.6.4.1 Connecting the suction tube A B CAUTION Trip/fall hazard If the suction tube is routed poorly, there may be a risk of tripping. ➢ To prevent injuries caused by tripping, route the suction tube so that there is no risk of tripping. 1. Connect one end of the suction tube to the available connection point on the suction device (A). 2.
5 Installation and startup 5.4 Commissioning Dentsply Sirona Operating Instructions CEREC Primemill 5.4.6.4.2 Connecting the power cord NOTE Grounded power outlet The unit must be connected to a grounded power outlet. 1. Insert the power cable into the relevant socket on the suction system. 2. Plug the other end into an appropriate power socket with a protective ground terminal.
Dentsply Sirona Operating Instructions CEREC Primemill 5 Installation and startup 5.4 Commissioning 5.4.6.4.4 Automatic mode ü The interface cable is connected. ➢ Set the on/off switch to the position Auto. Note on the automatic mode: The production unit monitors the correct connection (interface cable and suction tube) and the operation of the suction system during the running processes. 5.4.6.4.5 Setting the suction power Use the control dial (A) to set the suction power.
5 Installation and startup 5.4 Commissioning Dentsply Sirona Operating Instructions CEREC Primemill 5.4.6.5 Connecting to the PC via WLAN (option) Making the connection A Accesspoint Connect access point ➢ Connect the LAN port A of the production unit and the access point, using the network cable (10 m, Order No.: 61 51 521). Ä The access point is pre-configured at the factory for this application. Positioning the access point 1.
Dentsply Sirona Operating Instructions CEREC Primemill 5 Installation and startup 5.4 Commissioning 5.4.6.6 Operating several production units at one access point For the operation of several production units on one access point, you will also require the following components: ● 1x LAN switch (e. g. Netgear ProSave 5 Port Gigabit Switch, Model GS 105) ● 1x LAN network cable (10 m, order No.: 61 51 521). A 10m B Accesspoint 10m LAN Switch 10m 1.
5 Installation and startup 5.4 Commissioning Dentsply Sirona Operating Instructions CEREC Primemill 5.4.6.7.1 CEREC Radio Device Creating a network The supplied network devices have not as yet been part of a network. To enable several networks from various operators to be created next to one another or in the same area, new network devices must first form a network. This is described in the following section. To create a new network with several new network devices, perform the following steps: 1. 2. 3.
Dentsply Sirona Operating Instructions CEREC Primemill 5 Installation and startup 5.4 Commissioning 5.4.7 Filling the water tank NOTE Using the tank cap opener If you find the tank cap, tank drain or filter insert hard to open by hand, use the tank cap opener (see "Using the tank cap opener"). NOTE Coolant Use distilled or demineralized water.
5 Installation and startup 5.4 Commissioning Dentsply Sirona Operating Instructions CEREC Primemill NOTE Damage to surfaces! In the undiluted state, the coolant additive DENTATEC disintegrates plastic surfaces and can cause discoloration. ➢ Do not place DENTATEC on the unit. ➢ Do not spill DENTATEC. 4. Add approx. 75 ml of DENTATEC to the tank. 5. Fill the tank with water until the filter insert is completely immersed (up to the bottom edge of the cover thread, approx. 3.5 liters). 6.
Dentsply Sirona Operating Instructions CEREC Primemill 5 Installation and startup 5.5 Repacking Ä If no hub is found in the network, you can manually enter the IP settings. 4. Confirm the date and time. 5. Activate the suction settings if necessary. 6. Press the "Back" button to change a setting or press "Save" to conclude setup. Switching the unit off ü The unit has finished the machining operation. ➢ Press the On/Off button on the front. Ä The unit then switches off.
6 Operation 6.1 Configure Dentsply Sirona Operating Instructions CEREC Primemill 6 Operation CAUTION Risk of injury on calibration pins/instruments If you reach into the processing chamber (e.g.: when inserting/ removing a ceramic block, changing instruments, or inserting/ removing a calibration phantom), you may injure your hand on the calibration pins/instruments. Be careful not to brush against the calibration pins/instruments with your hand.
Dentsply Sirona Operating Instructions CEREC Primemill 6 Operation 6.2 Calibrating the unit 6.2 Calibrating the unit NOTE Use only the supplied calibration tools Use only the supplied calibration pins and the corresponding calibration phantom when calibrating the unit. Unit calibrated ex works The unit is calibrated at the factory. No additional calibration is required during initial startup. Proceed as described below when performing a subsequent calibration.
6 Operation 6.2 Calibrating the unit Dentsply Sirona Operating Instructions CEREC Primemill Calibration procedure 1. Select the symbol shown on the left in the touch display. 2. Select "Calibration" and then "Start". 3. Register the calibration phantom on the machine (scanning the code) and insert it in the machine. 4. Close the door. 5. Register the calibration pins (set 1) on the machine (RFID) and insert them in the machine. 6. Close the door. Ä The calibration for set 1 is performed.
Dentsply Sirona Operating Instructions CEREC Primemill 6 Operation 6.3 Machining process 6.3 Machining process Before every process, make sure that all necessary connections are correctly coupled. IMPORTANT Note the processing instructions of the respective material manufacturer for all manufacturing processes. NOTE Check the processed restorations after completion of the process and finishing (e.g. sintering, stain & glaze, polishing) for any defects.
6 Operation 6.3 Machining process Dentsply Sirona Operating Instructions CEREC Primemill 6.3.1.2 Milling For milling purposes, use the following instruments as well as the appropriate torque wrench: When doing so, pay attention to the connection geometry of the torque wrench. Instrument REF Usage Color Connection geometry of the force transmission Bur 2.5 ZrO2 CS 6713940 Milling of zirconium oxide (wet and dry) Yellow Square Bur 1.
Dentsply Sirona Operating Instructions CEREC Primemill 6 Operation 6.3 Machining process 6.3.3 Starting the machining processes ü The touch display of the production unit shows the HOME screen and the device door is closed. 1. The production unit positions the tools as soon as the door is closed. 2. If a data matrix code is present: The Webcam is activated and you can scan in the data matrix code of the block (see "").
6 Operation 6.3 Machining process Dentsply Sirona Operating Instructions CEREC Primemill 6.3.4 Ending the machining processes 1. When the machining process has been completed, open the processing chamber door. 2. Remove the restoration. WARNING Risk of injury on the remainder of the ceramic block The remaining portion of the ceramic block may have sharp edges (e.g. A) that could injure you if it is not removed carefully. Always grasp the remainder of the ceramic block by its metal holder. 3.
Dentsply Sirona Operating Instructions CEREC Primemill 6 Operation 6.4 Manual block clamp 6.4 Manual block clamp Keep the clamp tool in the associated holder. NOTE Wear of the ball pressure screw The high clamping forces cause wear of the ball pressure screw. ➢ Replace the ball pressure screw every 500 clamping operations. Blocks with 6 mm diameter block holders NOTE Fasten the block tightly If the block is not sufficiently tightened, it can lead to unsuitable results and ceramic breakages.
6 Operation 6.4 Manual block clamp Dentsply Sirona Operating Instructions CEREC Primemill Ä The block is pressed laterally against the contact surface of the block fixing and simultaneously pulled in axially. The plate of the block holder thus rests on the block fixing. Removing the adapter sleeve 1. Loosen the ball pressure screw. 2. Place the adapter sleeve removal tool in the inner groove (D) and pull out the adapter sleeve.
Dentsply Sirona Operating Instructions CEREC Primemill 7 Service 6.4 Manual block clamp 7 Service NOTE Observe country-specific Regulations! Some countries have legal regulations which require regular safety inspections of electrical devices or systems by the operator. NOTE Annual maintenance Have maintenance performed on your unit annually by trained technical personnel / a service engineer. The touch display shows a reminder notification when it is time for maintenance.
7 Service 7.1 Using the cleaning hose and the wet cleaning process Dentsply Sirona Operating Instructions CEREC Primemill 7.1 Using the cleaning hose and the wet cleaning process The processing chambers of devices used for dry milling of zirconium oxide should be cleaned regularly in order to prevent deposits of zirconium oxide dust in the chamber.
Dentsply Sirona Operating Instructions CEREC Primemill 7 Service 7.2 Changing filter bags and HEPA filters 6. After vacuuming the chamber you should perform the wet cleaning of the machine. This function can be started via the "Routine Actions" are (symbol on left) by selecting the "Cleaning Program" item. The process runs for 15 minutes, but can be canceled at any time if the desired result is reached. 7.
7 Service 7.2 Changing filter bags and HEPA filters Dentsply Sirona Operating Instructions CEREC Primemill Changing HEPA filters After every fourth filter bag change, the touch display automatically displays a notification to replace the HEPA filter. This number may differ depending on the amount of zirconium oxide materials milled and suctioned. If the HEPA filter is full before the notification, the touch display might display a warning at low pressure, meaning that the HEPA filter needs to be replaced.
Dentsply Sirona Operating Instructions CEREC Primemill 7 Service 7.2 Changing filter bags and HEPA filters 5. Take the perforated sheet out. 6. Remove the dusty HEPA filter and insert a new HEPA filter. 7. Put the perforated sheet back on and screw it down with the two Phillips screws. 8. Put the filter bag back on again. 9. Put the maintenance cover on and screw it down with the two knurled nuts. NOTE Do not jam the filter bag Make sure that the cap is sealed properly and the filter bag is not jammed.
7 Service 7.3 Changing the water Dentsply Sirona Operating Instructions CEREC Primemill 7.3 Changing the water 7.3.1 General information NOTE Coolant Use distilled or demineralized water. If a water change is due, a notification appears on your touch display. Preventing odors All coolant additives contain a biodegradable preservative. Despite this, however, odors may still develop under unfavorable conditions. Observe the following: ● Change the water at least once a week.
Dentsply Sirona Operating Instructions CEREC Primemill 7 Service 7.3 Changing the water 7.3.2 Changing the water NOTE Change the filter insert regularly! Replace the filter insert with a new one on every fourth water change at the latest. Water tank 67 19 756 D3439 D3439.201.01.01.02 09.
7 Service 7.3 Changing the water Dentsply Sirona Operating Instructions CEREC Primemill 7.3.2.1 Procedure NOTE Disposal Dispose of the contents of the container in accordance with local, national, and international regulations. To change the water, proceed as follows: ü The unit is switched on. ü No machining process is running. ü Open the drawer. 1. 2. 3. 4. 5. 6. 7. 8. 9. Pull the water tank out. Open the drain opening (D). Empty two thirds of the water from the tank.
Dentsply Sirona Operating Instructions CEREC Primemill 7 Service 7.4 Instruments 7.4 Instruments 7.4.1 Overview of materials/instruments Additional instrument combinations are displayed in the touch display. 7.4.2 Changing instruments NOTE Instrument replacement Change the instruments as soon as the system prompts you to do so. CAUTION Risk of injury on calibration pins/instruments If you reach into the processing chamber (e.g.
7 Service 7.4 Instruments Dentsply Sirona Operating Instructions CEREC Primemill 7. If an application is selected requiring the replacement of more than one instrument, repeat the aforementioned process for the remaining instruments. NOTE Cleaning cooling water nozzles The cooling water nozzles in the processing chamber must be free from limescale, grinding dust/milling dust deposits. The corresponding cooling water jet must always strike the instrument accurately! ü The cooling water nozzles are dirty.
Dentsply Sirona Operating Instructions CEREC Primemill 7 Service 7.5 Care, cleaning agents, and disinfectants 7.
7 Service 7.7 Replacing the main fuses Dentsply Sirona Operating Instructions CEREC Primemill 7.7 Replacing the main fuses WARNING Electric shock Disconnect the power plug at the unit end before replacing the fuses. NOTE Fuse type Use only fuses of the same type in the fuse holder! Fuse holder A Cover C Fuse holder B Main switch D Fuse Fuses: ü T3.15H250V Order no. 64 45 378 The power plug must be disconnected. 1.
Dentsply Sirona Operating Instructions CEREC Primemill 7 Service 7.8 Changing the filter 7.8 Changing the filter NOTE Change the filter insert regularly! Replace the filter insert with a new one once per week or on every fourth water change. You must also replace with a new filter cartridge when the "water pressure is too low" message appears. NOTE Filter Only use filter inserts approved by Dentsply Sirona! Water tank 67 19 756 D3439 D3439.201.01.01.02 09.
7 Service 7.9 Removing water from the unit Dentsply Sirona Operating Instructions CEREC Primemill 7.8.1 Procedure for all materials ü The tank is drained, see "Removing water from the unit [→ 62]". 1. Open the drawer of the unit. 2. Pull the water tank out of the unit's housing with the handle on the bottom front side of the tank. 3. Turn the tank cap counter-clockwise and take it off. If you find the tank cap hard to open by hand, use the tank cap opener. 4. Take the filter insert out of the tank. 5.
Dentsply Sirona Operating Instructions CEREC Primemill 7 Service 7.10 Using the tank cap opener 7.10 Using the tank cap opener NOTE Risk of damage to the tank Use the tank cap opener only for opening the tank cap. Do not use the tank cap opener for closing the tank cap. It is sufficient to tighten the tank cap clockwise by hand. Opening the tank cap ü The water tank has been pulled out and drained.
8 Technical description 8.1 System requirements Dentsply Sirona Operating Instructions CEREC Primemill 8 Technical description 8.1 System requirements ● CEREC SW 5.1 and higher versions ● inLab CAM SW 21.0 and higher versions 8.2 Production unit 8.2.1 General technical description ● Digital feed control with force monitoring for extremely sensitive processing ● Process-controlled tool drives ● Processing repeatability: +/- 25 μm Grinding instruments ● Diamond 1.4 CS (white) ● Diamond 1.
Dentsply Sirona Operating Instructions CEREC Primemill 8 Technical description 8.2 Production unit 8.2.2 Technical data Type designation CEREC Primemill or CEREC Primemill US Nominal mains voltage 100 V–240 V ~ Nominal mains frequency 50/60 Hz Nominal current 2.1 – 4.
9 Disposal Dentsply Sirona Operating Instructions CEREC Primemill 9 Disposal In accordance with Directive 2012/19/EU and national disposal regulations regarding old electrical and electronic devices, please be advised that such items must be disposed of in a special way within the European Union (EU). These regulations require the environmentally friendly recycling/disposal of old electrical and electronic devices. Such items must not be disposed of as domestic refuse.
Dentsply Sirona Operating Instructions CEREC Primemill 10 Consumable 10 Consumable Instrument REF Usage Color Connection geometry of the force transmission Diamond 1.4 CS 6714088 Grinding White Triangular Diamond 1.2 CS 6714070 Grinding White Diamond 1.0 CS 6714062 Grinding White Diamond 0.6 CS 6714054 Grinding White Instrument REF Usage Color Connection geometry of the force transmission Bur 2.
10 Consumable Dentsply Sirona Operating Instructions CEREC Primemill REF Designation 6718410 Block clamp tool, spare part 6479856 HT torque wrench, spare 6479849 Torque wrench, spare 6623792 Adapter sleeve 6704790 Adapter sleeve removal tool 6718444 Calibration set, spare part 6732528 Calibration pin (2 x), spare part 6258987 Ball pressure screw set (5 x), spare 6718451 Cleaning set, spare part 68 Illustration 67 19 756 D3439 D3439.201.01.01.02 09.
Dentsply Sirona Operating Instructions CEREC Primemill REF Designation 6718469 Screen insert, spare part 6429950 Filter MC/MCX (pack of 6 units) 6151562 10 Base-T crossover cable 10m 67 19 756 D3439 D3439.201.01.01.02 09.
Index Dentsply Sirona Operating Instructions CEREC Primemill Index B I Base, 21 Installation site, 22 Building installation, 15 Installing the unit automatic, 30 C manual, 30 Calibration tools Calibration phantom, 41 removal, 30 Instruments, 58, 64 Calibration pins, 41 Changing a defective instrument, 58 Care, cleaning, and disinfecting agents, 59 CE mark, 8 M Connection Main switch, 24 Ethernet, 29 Maintenance, 15 LAN, 29 WLAN (Wi-Fi), 34 Connection for suction, 24 Regulations, 49 Manu
Dentsply Sirona Operating Instructions CEREC Primemill Index T Temperature range, 65 Transport, 20 Type designation, 65 U Unpacking, 20 V Ventilation slots, 19 W Water, 65 Water tank Changing the water, 56 Filling, 37 Odors, 54 Overview, 37, 55, 61 Removing water from the unit, 62 water change, 54 Weight, 65 67 19 756 D3439 D3439.201.01.01.02 09.