Service Manual SDS1000D Digital Storage Oscilloscope June 2011 SIGLENT Technologies Co., Ltd.
Declaration Copyright © SIGLENT Technologies Co.,Ltd. All rights reserved. Contents in this Manual are not allowed to copy, extract and translate before being allowed by SIGLENT Company.
General Safety Summary Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified. Only qualified personnel should perform service procedures. 1. To Avoid Fire or Personal Injury ● Use Proper Power Cord. Use only the power cord specified for this product and certified for the country of use. ● Connect and Disconnect Properly.
● Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present. ● Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel. ● Do Not Operate in Wet/Damp Conditions. ● Do Not Operate in an Explosive Atmosphere. ● Keep Product Surfaces Clean and Dry. Power Line Conducted Emission Limits (Class B) Measuring standard: EN61326:1998+A1, 2002+A2, 2003 2.
Document Summarize Chapter 1 Specifications List the specifications and general specifications of SDS1000D series digital oscilloscope. Chapter 2 Prepare for Use Introduce the preparatory work should be done before using the oscilloscope. Chapter 3 Performance Test Introduce how to execute the performance test to understand current performance status of the oscilloscope. Chapter 4 Calibration Introduce how to calibrate the oscilloscope.
Catalogue Document Summarize.......................................................................................................................6 Chapter 1 Specifications ...................................................................................................................8 Specifications............................................................................................................................8 Generic Specifications ..........................................................
Chapter 1 Specifications All specifications apply to 10X probe and SDS1000D Digital Storage Oscilloscopes. To verify that the oscilloscope meets specifications, the oscilloscope must first meet the following conditions: ● The oscilloscope must have been operating continuously for thirty minutes within the specified operating temperature. ● You must perform the Do Self Cal operation, accessible through the Utility menu, if the operating temperature changes by more than 5° C.
100MHz(SDS1102D) 60MHz(SDS1062D) 40MHz(SDS1042D) 25MHz(SDS1022D) Single-shot Bandwidth 200MHz(SDS1202D) 150MHz(SDS1152D) 100MHz(SDS1102D) 60MHz(SDS1062D) 40MHz(SDS1042D) 25MHz(SDS1022D) Lower frequency limit (AC -3dB) ≤10Hz(at input BNC) DC Gain Accuracy <±3.0%: 5mv/div to 10V/div in Fixed Gain Ranges <±4.0%:typical for 2mv/div and Variable Gain Ranges DC Measurement Accuracy: All Gain settings≤ 100mv/div ±[3%X(reading+offset)1% of offset +0.
Modes Time Accuracy base ±50ppm measured over 1ms interval Horizontal Range Scan 2.5nS/DIV - 50S/DIV Roll: 100mS/DIV ~50S/DIV (1-2.
Hard Ware Frequency Counter Reading resolution 6 Bytes Range DC Couple, 10Hz to MAX Bandwidth Signal Types Satisfying all Trigger signals(Except Pulse width trigger and Video Trigger) Control Panel Function Auto Set Auto adjusting the Vertical, Horizontal system and Trigger Position Save/Recall Support 2 Group referenced Waveforms,20 Group setups,20 Group captured Waveforms internal Storage/Recall function and USB flash driver storage function.
Generic Specifications Display System Display Mode Color TFT 5.7in.
Chapter 2 Prepare for Use This chapter includes the following contents: ● ● ● ● ● General Inspecting Power-On Inspecting Connect the Probe Probe Compensation Auto Setting Service Manual
General Inspecting After receiving a new SDS1000D series oscilloscope, please inspect the instrument as follows: 1. .Inspect the shipping container for damage. Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically. 2. .Inspect the whole instrument.
Power-On Inspection Normal operating voltage for DS1000CA series digital oscilloscope is the range of 100-240VRMS with the frequency from 45Hz to 440Hz. Connect one terminal of the power cord to the socket in left side of the oscilloscope and the other to the power source. Picture 2-1 Connect power line Press the button on top of oscilloscope, some keys on the front panel will be lighted for about 6 seconds until the normal display appears. And then you can operate the oscilloscope.
Probe Compensation You are suggested to compensate probe before first using so as to match it to the input channel. Measurement error may generate without the compensation or because of deviation compensation. To compensate the probe, follow the steps below. 1.Set the probe option attenuation in the channel menu to 10X. Set the switch to 10X on the probe.
Auto Setup SDS1000D Series Digital Storage Oscilloscopes have auto setup function, which can identify the type of waveform and adjust controls to produce a usable display of the input signal. AUTO is the auto setup button. Table2-1 Auto setup function menu: Option (Multi-cycle sine) Introduction Auto set the screen and display several cycle signal. Set the screen and auto display single cycle signal. (Single-cycle sine) Auto set and show the rising time. (Rising edge) Auto set and show the falling time.
Bandwidth Limit Off(full) V/div Adjusted VOLTS/DIV adjustability Coarse Signal inverted Off Horizontal position Center S/div Adjusted Trigger type Edge Trigger source Auto detect the channel which has the input signal Trigger slope Rising Trigger mode Auto Service Manual
Chapter 3 Performance Test This chapter includes the following contents: ■ Interface Test USB Host USB Device RS-232 Pass/Fail ■ Specification Test Service Manual
Interface Test USB Host Test Purpose: Test if the USB Host interface works normally through U disc. Tools: ● An Unit SDS1000D digital oscilloscope ● An U disc Steps: 1.Insert a U disc into the interface of USB Host on the front panel of the oscilloscope. Picture 3-1 Connect the U disc 2.Wait until a prompt message “USB Flash Drive Plug In!” appeared on the screen. 图 3-2 USB host connect Successfully 3.Press Save/Recall Save to file, to name renewably or delete the information in U disc.
USB Device Test Purpose: Test if the USB Device interface works normally through EasyScope for SDS1000D series oscilloscope. Tools: ● An Unit SDS1000D digital oscilloscope ● A computer with USB interface ● A standard USB cable(Type AB) ● EasyScope software for SDS1000D series Steps: 1.Install the EasyScope to computer. 2.Connect the oscilloscope with PC using an USB cable and install the driver program step by step following the instruction.
RS-232 Test Purpose: Test if the RS-232 interface works normally through EasyScope for SDS1000D series oscilloscope. Tools: ● An Unit SDS1000D digital oscilloscope ● A computer with RS-232 interface ● A standard RS-232 cable ● EasyScope software for SDS1000D series. Steps: 1. Install the EasyScope to computer. 2. Connect the oscilloscope with PC using an USB cable Picture 3-5 Connect the RS232 interface 3. Run EasyScope,click “Setting” “Connect Setting” , set the IO as below.
Picture 3-6 EasyScope connect setting 4. Press “UTILITY” page3/4 IO Setting, select the Baud as 38400 5. Click “Connect”, if successful, the two point lamps at the top right corner of the software will be changed from red to green.; if failed, a prompt message will be appeared as below.
Pass/Fail Test Purpose: Test if the Pass/Fail interface works normally through the signal from the back BNC of the oscilloscope. Tools: ● An Unit SDS1000D digital oscilloscope , another digital oscilloscope ● A BNC cable ● A Probe Steps: 1. Power-On two oscilloscopes. 2. Connect the Pass/Fail interface of SDS1000D with the channel BNC of another oscilloscope using the BNC cable. 3. Connect the probe to channel 1 on SDS1000D, and press AUTO button. 4.
Specification Test Purpose: Test if some specifications are in the specified range. Table 3-2 require instruments Description Minimum requirements Examples DC Voltage Source -6 mV to 30 V, ±0.
Volt/div setting DC voltage output level Accuracy limits for Vdiff 10v/div +30V, -30V 58.2 V~61.8 V 5v/div +15V, -15V 29.1 V~30.9 V 2v/div +6V, -6V 11.64 V~12.36 V 1v/div +3V, -3V 5.82 V~6.18 V 500mv/div +1.5V, -1.5V 2.91 V~3.09 V 200mv/div +600mV, -600mV 1.164 V~1.236 V 100mv/div +300mV, -300mV 582 mV~618 mV 50mv/div +150mV, -150mV 291 mV~309 mV 20mv/div +60mV, -60mV 116.4 mV~123.6 mV 10mv/div +30mV, -30mV 58.2 mV~61.8 mV 5mv/div +15mV, -15mV 29.1 mV~30.
Test Trigger Sensitivity Steps: : 1、Press the button “DEFAULT/SETUP”, and set the VOLT/DIV to 500mv/div, set SEC/DIV to 25ns/div. 2、Connect the oscilloscope channel 1 to the standard sinewave generator 3、Set the standard sinewave generator frequency to 10 MHz 3、Set the generator output level so the Vpp measurement is about 500 mV. 4、Press the button “set to 50%”. Adjust LEVEL as necessary and then check that triggering is stable.
Chapter 4 Calibration Self Calibration Self Calibration procedure can optimize the signal path at the most measurement precision. You can run this procedure at any time. If the operating temperature changes by more than 5℃ or the units run less than thirty minutes, you should do the self calibration. Steps: 1. Cut off all the probes and lead.
Picture 4-2 Calibrating 4. After the calibration finished, Press “SINGLE” key to exit. Picture 4-3 Calibration succeed interface Generally speaking, the self calibration will take about 3-5 minutes, if the calibration does not Pass after this time, or the progress bar stop at one of the calibration item, there may be a trouble in the instrument.
Chapter 5 Disassembly and Assembly This chapter includes the following topics: ● ● ● ● ● ● ● ● ● ● ● Notice Disassemble and Assemble the on-off and knobs Disassemble and Assemble back rind Disassemble and Assemble back cover board Disassemble and Assemble top cover board Disassemble and Assemble fan Disassemble and Assemble power board Disassemble and Assemble metal shelf Disassemble and Assemble main board Disassemble and Assemble LCD Disassemble and Assemble keyboard Service Manual
Notice Notices: : ● ● ● ● ● ● Don’t disassemble the products except the work need. Disassemble only can be done by qualified person. Cut the power before disassembling. Take static glove when disassembling. Use proper tools and follow the disassemble sequence. Prevent metal parts from transfiguration and avoid being scratched when disassembling.
Disassemble and Assemble On-off and Knobs Picture 5-1 Disassemble and Assemble the on-off and knobs Disassemble steps: 1. Nip on-off ○ 1 button by forceps, and then pull up. 2. Cock knobs③、④ gently by forceps Assemble steps: Please operate as reverse orders. The same is as below.
Disassemble and Assemble Back Rind Picture 5-2 Disassemble and Assemble back rind 1 图 5-3 Disassemble and Assemble back rind 2 Disassemble steps: 1. Back out the two bolts⑤ and two bolts⑥. 2. Back out the two bolts○ 7 . 3. Remove the back rind.
Disassemble and Assemble Back Cover Board Picture 5-4 Disassemble and Assemble back cover board Disassemble steps: 1. Remove the back cover board by the forceps.
Disassemble and Assemble Top Cover Board Picture 5-5 Disassemble and Assemble top cover board Disassemble steps: 1. Back out the two bolts○ 9 . 2. Pull on the top cover board.
Disassemble and Assemble Fan Picture 5-6 Disassemble and Assemble fan Disassemble steps: 1. Back out the four bolts ○ 11 .
Disassemble and Assemble power board Picture 5-7 Disassemble and Assemble power board Disassemble steps: 1. Back out the four bolts○ 13 . 2. Pull out the power line from the main board and remove the power board.
Disassemble and Assemble Metal Shelf Picture 5-8 Disassemble and Assemble metal shelf Disassemble steps: 14 , and remove the metal shelf ○ 15 . 1. Back out the four bolts ○ 2. Take off the silica gel key○ 16 .
Disassemble and Assemble Main board Picture 5-9 Disassemble and Assemble main board Disassemble steps: 1. Back out the five bolts○ 17 . 2. Pull out the screen line and keyboard line from main board. 3. Back out the three big nuts from the BNC. 4. Pull out the main board○ 18 slowly.
Disassemble and Assemble LCD Picture 5-10 Disassemble and Assemble LCD Disassemble steps: 1. Break off the LCD○ 20 with fingers and thumb.
Disassemble and Assemble Keyboard Picture 5-11 Disassemble and Assemble keyboard Disassemble steps: 1. Back out the four bolts ○ 21 and remove the keyboard..
Chapter 6 Prompting Messages and Troubleshooting Prompting messages: ■ Trig level at limit! : Mention you that the trigger Level is at a limit when you turn the Trig level knob. ■ Horizon position at limit! : Mention you that the horizontal position is at a limit when you turn the horizon position knob. ■ Volts/Div at limit! : Mention you that the vertical voltage have already touched the Min 2MV/div or the Max 5V/div.
Drive. ■ Store Data Success! : Save setup data, waveform data or picture data to the internal of the oscilloscope or USB flash successful. ■ Ready Data Success! : Read setup data or waveform data from the internal of the oscilloscope or USB flash successful. ■ Please set back USB to printer! : Press the “S/div” knob will appear this information on the screen when the “Print Key” option is set to “Print Picture” and the “Back USB” option is set to “Computer”.
Troubleshooting 1. After the oscilloscope is powered on, ,the screen remains dark ,please do as following steps: : ⑴ Check the power cable’s connection. ⑵ Ensure the power switch is turned on. ⑶ After the inspections above,restart the oscilloscope. ⑷ If the oscilloscope is still not used after the checking, please connect with SIGLENT company 2.
Chapter 7 Updating Parts SIGLENT Provide some replaceable parts in order to maintain or update for users. Please see as the following table. Note that the numbers in the table. Table 7-1 Updating Parts Item No. Name Specification 1 On-Off button 2 Back Rind 3 Big knobs 4 Small knobs 5 Bolts PM2.