JPD-FR400 0482 Infrared Thermometer Instruction Manual Shenzhen Jumper Medical Equipment Co., Ltd Manual Ver.: 1.
Introduction Thank you for choosing the Dual-mode Digital Infrared Thermometer of Jumper Medical, Infrared thermometer for short. The infrared thermometer JPD-FR400 can be used to read the body temperature by measure the ear and forehead temperature, which is suitable for adult and children(ear test mode only for the child above 3 months). Please read the instructions carefully before using the product, and put it in a safe and secure place for reference.
Contents 1. The advantages of infrared thermometer JPD-FR400 ............ 1 2. Necessary safety instructions ................................... 2 3.Instructions for product designs .................................. 3 4.How to measure the forehead temperature ........................ 4 5.How to measure the ear temperature ............................. 4 6.Instructions for display and operation ............................. 6 7.Cleaning and Disinfection ....................................... 9 8.
1. The advantages of infrared thermometer JPD-FR400 Multiple use (measurement of forehead temperature and ear temperature) Dual-mode design, you are free to take body temperature by detecting the infrared heat given off by the forehead or eardrum,The Measuring range of the thermometer in forehead /ear mode is between 32.0°F 212.0°F (0 °C- 100 °C) . The product consists of ABS plastics, temperature sensor, Infrared temperature measuring element, microcomputer controlled circuit, LCD , backlight and buzzer.
Operating principle The infrared temperature sensor detects infrared energy emitted by the eardrum. A built-in lens focuses the collected energy, which is then converted into a temperature reading by the thermopiles and measurement circuits. 2. Necessary safety instructions ●The device is not used for newborn baby. ●The device is not continuous monitoring device. ●JPD-FR400 is not waterproof. Please do not douse it into water or other liquids.
●Battery Installation: Put the two AAA batteries into battery compartment in correct polarities. Push the battery cover horizontally along the arrow. Notes: ☆ If you are the first time using the thermometer, please pull out the Plastic sheet . ☆ Battery polarities should be correctly installed. Otherwise, damage may be caused to the device. ☆ Please put in or remove batteries in right order, or may cause damage to the device bracket.
Display description 1. Ambient temperature 2. Ready for measurement 3. Ear mode 4. Head mode 5. ˚C / ˚F 6. Low power indicator 7. Memory mode 8. Temperature value 4.How to measure the forehead temperature ▀ Spot measurement 1. With the sensor/probe cover attached, position the thermometer at the center of the forehead, just above the eyebrow. Ensure that the thermometer is in contact with the forehead. 2. Press and release the HEAD button.The temperature will be displayed on the screen instantly.
2. Insert the ear probe into the ear canal. 3. Press and release the EAR button. The temperature will be displayed on the screen instantly. Note: Children under 1 year: Pull the ear straight back. Children aged 1 year to adult: Pull the ear up and back. In order to avoid the inaccuracy: (1)Please make sure that there is no dirt in on the temperature sensor (2)Please make sure that the device only be used in room, and there is no strongly conversation of wind.
Connecting the thermometer to an APP application of a mobile phone The thermometer has a built-in Bluetooth module and can be connected to a smartphone by using the Bluetooth 4.0 wireless technology. View the measurement data in real time by using mobile phone software "MyThermo". The software automatically saves the measurement data and draws a temperature curve graph, which helps view the temperature change status in real time.
Forehead measurement mode Eardrum measurement mode The measured value exceeds 212 °F/100°C. The measured value falls below 32.0°F/0°C. 3 short beeps with a red backlit for 3 seconds. 3 short beeps with a red backlit for 3 seconds. Inquiry for memory data, Storing 20 temperature readings LCD Operational method and instruction for display displays Sound and backlit Press and hold button EAR for 3-6 seconds and the LCD will display “- - -” with the M None signal blinking.
The LCD will display “- - -” with the M signal blinking if there is no test data. None °C/°F conversion LCD Operational steps Sound and display backlit When the thermometer is off, press and hold the EAR button for 6-9 seconds. The screen will display “- - - ” with “M” at the bottom right. Continue to hold the button until “- - -°C/°F” appears on the screen. Release the EAR None button and the “- - -°C/°F” temperature unit will start to blink.
local regulation requirements when the products and batteries are not available. ●Please take out the battery if the device unused for long time. 7.Cleaning and Disinfection Cleaning 1) Clean the temperature probe with a soft cloth. Clean the lens of the temperature probe with a cotton swab. 2) Wipe the thermometer body with a slightly damp soft cloth, and gently dry the body with a piece of tissue paper. Keep water off the lens during the cleaning process. Otherwise, the lens may be damaged.
3) 4) the storage and transportation environments meet the requirements Check whether safety risks exist on a regular basis. Remove the batteries if the thermometer will not be used for more than two months. 9. Replace the battery. Slide the battery cover off along the marked direction. Put two AAA batteries correctly into the compartment. Remove the batteries if the thermometer will not be used for more than two months. 10.
Symptom Possible Cause Solution (50.0°F) or higher than temperature between 40°C (104°F). 10°C (50.0°F) and 40°C (104°F). An error occurs when Contact Jumper. data is being read from or "ErC" is displayed. written into the memory, or the temperature correction is not complete. The lens of the Clean the lens using a temperature probe is cotton swab. dirty. The temperature The distance between the Move the thermometer reading is lower temperature probe and closer to the target.
3.Measurement range Forehead temperature mode:0.0℃-100.0℃ /32.0F – 212.0 F Ear temperature mode:0.0℃-100.0℃ /32.0F – 212.0 F 4.Accuracy (Laboratory) ±0.2℃ /±0.4F(36℃-39℃/96.8℉-102.2℉) ±0.3/±0.5℉(less than 36 °C (96.8°F) or greater than 39 °C (102.2 °F) 5.Display resolution 0.1℃/0.1℉ 6.Measure time About 1 second 7.Latency Time About 1 second 8.Measuring distance Forehead temperature:1~5 mm(0.04~0.2 inches) 9.
13. Security type The signal indicates that the thermometer is type BF applied part. 14. Authorized European Representative: Wellkang Ltd Suite B, 29Harley Street, London, W1G9QR,UK. 15. Symbols Symbol Description Type BF applied part. Attention must be paid. Information about a manufacturer, such as name and address. Please read the instructions carefully. Waste electrical materials should be sent to a dedicated collection point for recycling.
. Appendix A: EMC Information-Guidance and Manufacture’s Declaration CAUTION: ● JPD-FR400 Infrared Thermometer needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided for in the ACCOMPANYING DOCUMENTS. ● Portable and mobile RF communications equipment can affect JPD-FR400 Infrared Thermometer ● The JPD-FR400 Infrared Thermometer should not be used adjacent to or stacked with other equipment.
discharge (ESD) IEC 61000-4-2 Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 contact ±8 kV air ±8 kV air 3 A/m 3 A/m concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %. Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m).b Field strengths from fixed RF transmitters, as determined by an Electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
The JPD-FR400 Infrared Thermometer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the JPD-FR400 Infrared Thermometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the JPD-FR400 Infrared Thermometer as recommended below, according to the maximum output power of the communications equipment.
FCC IC Caution: This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
Shenzhen Jumper Medical Equipment Co., Ltd. Address: D Building, No. 71, Xintian Road, Fuyong Street, Baoan,Shenzhen,Guangdong,China Tel: +86-755-26692192, 26696279 Fax: +86-755-26852025 Website:www.jumper-medical.