JPD-FR409-BT 0482 Infrared Thermometer Instruction Manual Shenzhen Jumper Medical Equipment Co., Ltd Manual Version: 1.0 Date of Issue: 2019.
Product Information Product Name: Infrared Thermometer Model: JPD-FR409-BT Manufacturer: Shenzhen Jumper Medical Equipment Co., Ltd Address: D Building, No. 71, Xintian Road, Fuyong Street, Baoan, Shenzhen, Guangdong, China. Copyright Copyright © Jumper Medical. All rights reserved. Statement Shenzhen Jumper Medical Equipment Co., Ltd. owns and reserves all of the rights comprised in the copyright of this document.
Contents FCC Statement................................................................................................ 1 Unpacking Check............................................................................................ 2 Package Contents............................................................................................ 2 Safety Precautions...........................................................................................2 Warning....................................................
FCC Statement This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference. (2) This device must accept any interference received, including interference that may cause undesired operation. Notes: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
Unpacking Check Please open the package carefully before use, check whether all accessories are available or not and whether any component is damaged during transportation, and perform installation and operation following this user manual. In case of any damage or operation problem, please contact the dealer or contact Jumper directly. You will need the following information when making your claim: device model, serial number, purchase date, and your contact information and address. Package Contents No.
Remove the batteries if the thermometer will not be used for more than two months. Do not immerse the thermometer in water or expose it to direct sunlight. Do not subject the thermometer to vibration or impact. The normal body temperature varies from person to person. Tracking a person's body temperature will help determine whether they have fever. Do not take body temperature readings within 20 minutes after you do physical exercises or get excited.
earwax, if any. The ambient temperature must not be extremely high or low. To make sure accurate readings, keep the thermometer under room temperature for more than 30 minutes before use. Do not use the thermometer under an ambient temperature higher than 40°C (104°F) or lower than 10°C (50°F), which is beyond the operating temperature range of the thermometer. Risk of pollution! The user is recommended to send the overdue thermometer to local garbage disposal site or send it back to us.
It is dangerous to make a self-diagnosis or self-treatment based on the obtained measurement results. For such purposes, please consult a doctor. Do not charge an alkaline dry-cell battery or throw it in fire. Otherwise, the battery may explode. Do not disassemble the thermometer or attempt to repair it. Otherwise, the thermometer may be damaged permanently. Do not take temperature measurements on body parts other than forehead and ears. Otherwise, the temperature readings may be inaccurate.
Symbol Description Information about the manufacturer. Date of manufacture. Consult the instructions for use. This product complies with the MDD93/42/EEC requirements. Waste electrical materials should be sent to a dedicated collection point for recycling. A personal injury or damage to the thermometer may Warning occur if the thermometer is not used correctly. Inaccurate reading or damage to the thermometer may Attention occur if the thermometer is not used correctly.
To determine if an individual is experiencing an elevated body temperature and/or having a fever, it is critical to know the individual’s normal body temperature when he/she is well. Take multiple readings to obtain the normal body temperature range and note the specific body site measured, for example: forehead or eardrum temperature. Body Site Normal Temperature Range Forehead 34.7°C-37.3°C Eardrum 35.8°C-38.0°C (96.4°F-100.4°F) Mouth 35.5°C-37.5°C (95.9°F- 99.5°F) Armpit 34.7°C-37.3°C (94.
Product Description 1) Overview Infrared Thermometer JPD-FR409-BT measures the body temperature based on the infrared energy emitted from the eardrum or the forehead. Users can quickly get measurement results after positioning properly the temperature probe in the ear canal or on the forehead. 2) Structure The thermometer consists of a shell, an LCD, a measure button, a beeper, an infrared temperature sensor, and a Microprocessor.
Do not use the thermometer if the ear is infected with otitis or suppuration. Features 01 Good safety ● Passive infrared receiving technology.
Product structure (1) (2) (3) (4) (5) (6) (7) (8) LCD display screen Indicator Power button / Measure button Adult mode / Child mode Mode button (Forehead / Ear / Object) Probe (take off the cover when measuring the ear temperature) Unit switch button(℃/℉) Probe cover (put the cover on when measuring the forehead temperature) (9) Memory button / Sound switch (10) Battery cover Display description 1. Object temperature mode 2. Forehead temperature mode 3. Child mode 4. Ear temperature mode 5.
Sounds and indicator color instructions Range Indicator Sounds Color Forehead temperature (Adult / Child) 35.0℃-37.5℃/95.0℉-99.5℉ A long beep Green 37.6℃-42.2℃/99.6℉-108.0℉ 3 short double beeps Red 35.0℃-37.5℃/95.0℉-99.5℉ A long beep Green 37.6℃-42.0℃/99.6℉-107.6℉ 3 short double beeps Red A long beep Green Ear temperature (Adult / Child) Object temperature 0℃-100℃/32.0℉-212℉ Note: When the forehead and the ear temperature is between 35.0°C/95.0℉ and 37.5°C/99.
Screen Display Operating Instructions Displayed State Sound and Indicator Color Take off the probe cover, press and release the Power button for 1 second to power on the thermometer. Press the Mode button, the thermometer enters the Ear mode. The symbol “Ear” is See the table in the displayed on the screen. “Sounds and Switch to the adult or child indicator color measurement mode by switching instructions” the side adult/child button section. Ear temperature according to your measurement for adult needs.
Screen Display Operating Instructions Displayed State Put the cover on the probe, press and release the Power button for 1 second to power on the thermometer. Press the Mode button, the thermometer enters the Forehead mode. The “Head” symbol is displayed on the screen. Switch to the adult or child measurement mode by switching the side adult/child button Forehead according to your measurement temperature for needs. adult Point the thermometer to the center of the forehead.
Operating Instructions Screen Display Displayed State Sound and Indicator Color Press and release the Power button for 1 second to power on the thermometer. Then press the Mode button. The thermometer enters the See the table in the Object mode. The “House” symbol “Sounds and indicator color is displayed on the screen. instructions” Point the thermometer to the center section. of the object. Press and release the Measure button. The temperature will be displayed on the screen.
Operating Instructions Screen Display Displayed State Sound and Indicator Color In a power-on state, press the Memory button enter the memory mode. When the Memory button is released, 01 will be shown, followed by the recorded reading. Press the Memory button again for the next recorded data. 02 will be shown, followed by the recorded Silent,the indicator reading. light is green. A maximum of 20 temperature readings can be recalled.
Screen Display Operating Instructions Displayed State The display is as shown , when there is no more data checked while recalling memories. Remove 2 dry batteries and after 10 seconds re-install the power to clear all memory data. Sound and Indicator Color With a long beep, the green indicator light shows up and turn red, when device is powered on. Switching between mute and un-mute In the power-on state, press and hold the sound switch button for about 2 seconds to switch the sound on or off.
Screen Display Operating Instructions Displayed State Press the Mode button to switch between object temperature and body temperature. Body temperature contains the Forehead temperature and Ear temperature. Sound and Indicator Color The silent green indicator light Switching between ºF/ºC In the power off state, remove the battery cover, and toggle the °C/°F unit in the battery compartment to switch the toggle device. Toggle switch for ℃ upwards and downwards toggle switch as the ℉ unit.
Screen Display Operating Instructions Displayed State Sound and Indicator Color Error information & low battery The ambient temperature is higher A long beep with than 40.0ºC (104.0ºF) or lower than the red indicator light. 10.0ºC (50.0ºF). An error occurs when data is being A long beep with read from or written to the memory, the red indicator or the temperature correction is not light. complete. When the battery voltage is lower than 2.4V ± 0.1V, the low battery symbol will appear on the display.
2. Take the probe cover off from the thermometer before measuring the ear temperature. 3. Press the Power button to power on the thermometer. 4. Press the Mode button, the thermometer enters the Ear mode. The “ ” symbol is displayed on the screen. 5. When measuring an adult, switch the button to the "adult" mode. When measuring children, switch the button to the "child" mode and the symbol “ ”is displayed on the screen. 6. Insert the temperature probe into the ear canal. 7.
Do not force the thermometer into the ear canal. Otherwise, the ear canal may get injured. When taking the temperature on an adult, gently pull the ear up and back to make sure the ear canal is straight, so that the temperature probe can receive an infrared ray from the eardrum. Be careful when taking temperature on a child, whose ear canal is small. Measuring Forehead Temperature 1. Put the cover on the probe of thermometer. 2. Press the Power button to power on the thermometer. 3.
6. Press and release the Measure button for 1 second. The temperature reading will be displayed on the screen instantly. 7. If no measurement is conducted, the thermometer will be powered off automatically in 10 seconds. Measuring Object Temperature 1. Put the cover on the probe of thermometer. 2. Press the Power button to power on the thermometer. 3. Press the Mode button, the thermometer enters the Object mode. 4.
and put the thermometer in a dry and well-ventilated place. It is dangerous to make a self-diagnosis or self-treatment based on the obtained measurement results. For such purposes, please consult a doctor. Notes: (1) The thermometer is suitable for an indoor environment without strong air convection (for example, winds from a fan, an air-conditioner, or a heater) between the thermometer and the person. (2) Make sure that the ear canal is clean and dry before starting a measurement.
measurement data in real time by using APP "JUMPER HEALTH". The software automatically saves the measurement data and draws a temperature curve graph, which helps view the temperature change status in real time.For a method for connecting the thermometer to the smart phone,see Operation Instruction. Replacing Batteries 1、Slide the battery cover off along the marked direction and take it off. 2、Insert the two AAA batteries into the compartment according to the stated polarities.
Cleaning and Disinfection Cleaning Recommended detergents: * Medical detergents; * Home use mild detergents; Cleaning steps: (1) Take the batteries out before cleaning. (2) Clean the temperature probe with a soft cloth. Clean the lens of the temperature probe with a cotton swab. (3) Wipe the thermometer body with a slightly damp soft cloth. Keep water out off the lens during the cleaning process. Otherwise, the lens may be damaged.
Disinfecting steps: 1) Wet the clean soft cloth with a small quantity of disinfectant, wipe the thermometer and quickly dry it. 2) Disinfect the thermometer body and the area around the temperature probe with a cloth slightly moistened with 75% medical alcohol. Do not use hot steam or ultraviolet radiation for disinfection. Otherwise, the thermometer may be damaged or quickly aged.
than two months. Troubleshooting Problem The thermometer fails to power on. "Er1" is displayed. The temperature reading is lower than the typical body temperature range. The temperature reading is higher than the typical body temperature range. Possible Cause Solution Low battery Change the batteries. Polarities of the batteries are reversed. Make sure that the batteries are installed correctly. The thermometer is damaged. Contact the manufacturer.
Specifications Product Name Infrared Thermometer Product Model JPD-FR409-BT Power Supply Mode Internal power supply Operating Voltage DC 3V Battery Model AAA x 2 Battery Life Alkaline dry battery for around 20,000 measurements Operating Mode Continuous operating Display Segment LCD Measure time About 1 second Latency Time About 3 seconds Measuring Range Forehead mode:35.0°C–42.2°C (95.0°F–108.0°F) Ear mode:35.0°C–42.0°C (95.0°F–107.6°F) Object mode:0.0°C–100.0°C (32.0°F–212.
The infrared thermometer has been tested and conforms to the standard ASTM E1965-98. ASTM laboratory accuracy requirements in the display range of 96.8°F to102.2°F (36°C-39°C) for ear canal IR thermometers is ±0.4°F (±0.2°C). Note that for mercury-in-glass and electronic thermometers, the requirement per ASTM Standards E667-86 and E1112-86 is ±0.2°F (±0.1°C). Security Class Type of protection against electric shock: internally powered equipment.
EMC Information-Guidance and Manufacture’s Declaration CAUTION: ● The Infrared Thermometer JPD-FR409-BT needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided for in the ACCOMPANYING DOCUMENTS. ● Portable and mobile RF communications equipment can affect Infrared Thermometer JPD-FR409-BT. ● The Infrared Thermometer JPD-FR409-BT should not be used adjacent to or stacked with other equipment.
equipment. The Infrared Thermometer JPD-FR409-BT is suitable for use in all establishments, RF emissions CISPR 11 including domestic establishments and Class B those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
frequency fields should be at levels (50/60 Hz) characteristic of a typical magnetic location in a typical field commercial or hospital IEC 61000-4-8 environment. Guidance and manufacturer´s declaration – Electromagnetic immunity –for equipment and systems that are not life-supporting Guidance and manufacturer´s declaration – Electromagnetic immunity The Infrared Thermometer JPD-FR409-BT is intended for use in the electromagnetic environment specified below.
Recommended separation distance where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an Electromagnetic a site survey,a should be less than the compliance level in each frequency range.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection from structures, objects and people. a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
the maximum output power of the communications equipment. Rated Separation distance according to frequency of transmitter maximum output m 80 MHz to 800 MHz 800 MHz to 2.5 GHz power of transmitter W 0.01 0.12 0.23 0.1 0.38 0.73 1 1.2 2.3 10 3.8 7.
The batteries, the packaging, and any damage caused by improper use are not covered by the warranty. Excluding the following user-caused failures: 1.Failure resulting from unauthorized disassembly and modification. 2.Failure resulting from an unexpected dropping during application or transportation. 3.
Shenzhen Jumper Medical Equipment Co., Ltd. Address: D Building, No. 71, Xintian Road, Fuyong Street, Baoan, Shenzhen, Guangdong, China Tel:+86-755-26696279 Fax:+86-755-26852025 Website: http://www.jumper-medical.