All-in-One Health Monitor User Instruction Manual
User Manual for All-in-One Health Monitor This manual is written for the PC-303 All-in-One Health Monitor. The manual describes, in accordance with the All-in-One Health Monitor’s features and requirements, the main structure, functions, specifications, correct methods for transportation, installation, usage, operation, repair, maintenance and storage, etc. as well as the safety procedures to protect both the user and equipment. Refer to the respective chapters for further details.
User Manual for All-in-One Health Monitor Instructions for use Dear Customers, Thank you for purchasing the PC-303 All-in-One Health Monitor. Please read the following information before using the device. These instructions describe the operating procedures which are to be strictly followed, read these instructions carefully before using the All-in-One Health Monitor. Failure to follow these instructions can cause monitoring abnormalities, damage to the monitor and personal injury.
User Manual for All-in-One Health Monitor If the monitor is connected with other devices, the total leakage current may exceed the limitation and as a result this can cause potential danger to the user. Although biocompatibility tests have been performed on all the applied parts, under exceptional circumstances, allergic patients may have anaphylaxis. Do NOT use the monitor on patients with anaphylaxis.
User Manual for All-in-One Health Monitor radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation.
User Manual for All-in-One Health Monitor Table of contents CHAPTER 1 OVERVIEW ............................................................................................................... I 1.1 Features .......................................................................................................................... I 1.2 Product Name and Model ................................................................................................ II 1.3 Intended Use ....................................
User Manual for All-in-One Health Monitor 2.2.2 Starting the Monitor............................................................................................VII 2.3 Taking Measurements ................................................................................................. VIII 2.3.1 Blood Pressure Measurement ........................................................................... VIII 2.3.2 SpO2 Measurement ...................................................................................
User Manual for All-in-One Health Monitor 4.1 Blood Pressure Measurement...................................................................................... XLI 4.2 SpO2 Measurement ................................................................................................... XLII 4.3 Pulse Rate Measurement ........................................................................................... XLII 4.4 Temperature Measurement.......................................................................
User Manual for All-in-One Health Monitor 7.3 Cleaning and Disinfection of Accessories ...................................................................... LIV 7.4 Storage ..................................................................................................................... LIV 7.5 Transportation ..............................................................................................................
1.
User Manual for All-in-One Health Monitor 1.2 Product Name and Model Name: All-in-One Health Monitor Model: PC-303 1.3 Intended Use The All-in-One Health Monitor is a device designed for measuring the patient’s physiological parameters, such as Non-Invasive Blood Pressure (NIBP), Oxygen saturation (SpO2), Pulse Rate (PR) and Body Temperature (TEMP); Additionally, the device is available to communicate with the compatible Blood Glucose Monitoring System and ECG monitor to make the measurement.
User Manual for All-in-One Health Monitor CHAPTER 2 OPERATION INSTRUCTIONS 2.1 Appearance 2.1.1 The Front Panel Figure 2.
User Manual for All-in-One Health Monitor Description: 1/2. / up/down key: on the setup display screen, a short press will change the parameter value step by step, press and hold to change the parameter values quickly; on the review display screen, short press to review the history data records one by one, 3. press and hold to recall the history data records quickly.
User Manual for All-in-One Health Monitor 2.1.2 The Right and Upper Sides of the Device Figure 2.
User Manual for All-in-One Health Monitor The power switch and external DC power input socket are on the right side of the monitor as shown in figure 2.2 left. The signal input/output ports are on the upper side of the monitor as shown in figure 2.2 right. Description: : External DC power input socket. 1. 2. : Power switch (press and hold to turn the monitor on/off). 3. Battery cover. 4. : USB data interface. 5. ↑↓GLU: Connector to link to the Blood Glucose Monitoring System. 6. NIBP: Cuff connector.
2.2 Installation User Manual for All-in-One Health Monitor 2.2.1 Power Supply 1. Internal power supply by the built-in battery When the battery indicator " " displays full grids, the built-in battery is fully charged. When it blinks, the battery voltage is low, and the user should charge the battery by connecting the device to the AC power adapter or a USB power source via USB cable. When the grids of the battery indicator are rolling circularly, the battery is being charged. 2.
User Manual for All-in-One Health Monitor 2.3 Taking Measurements 2.3.1 Blood Pressure Measurement 1. Applying the cuff; unfold the cuff and wrap it around the upper arm evenly to the appropriate tightness. The correct cuff position is shown in figure 2.3. Figure 2.3 Cuff position 2. Connect the hose from the cuff to the connector on the upper-side of the device where marked “NIBP”. 3. Press the start/cancel button" " to begin the blood pressure measurement.
User Manual for All-in-One Health Monitor Blood pressure measurement is prohibited to those who have severe hemorrhagic tendencies or with sickle cell disease, as partial bleeding may be caused. An appropriate cuff should be selected according to the age and arm circumference of the patient. The cuff width should be 2/3 of the length of the upper arm. The inflatable part should be long enough to permit wrapping approximately 80% of the limb.
User Manual for All-in-One Health Monitor Continuous measurements may result in purpura, neuralgia and lack of blood. Do NOT wrap the cuff on limbs with transfusion tubes, intubations or skin lesions on the area, as damage may be caused to the limbs. The equipment can be used on pregnant or pre-eclamptic patients, but should not be used on neonatal patients.
User Manual for All-in-One Health Monitor The environmental or operational factors which can affect the performance of the device and/or its blood pressure reading (e.g. common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, age, pregnancy, pre-eclampsia, renal diseases, patient motion, trembling, shivering). The performance of the equipment can be affected by extreme temperature, humidity and altitude.
User Manual for All-in-One Health Monitor 2.3.2 SpO2 Measurement Operation procedures: 1. Connect the smart SpO2 probe to the connector on the upper-side of the device marked "SpO2". When disconnecting the connector, be sure to hold the head of the connector firmly and pull. 2. The red blinking light inside the clip of the SpO2 probe indicates a successful connection. 3.
User Manual for All-in-One Health Monitor Safety instructions for SpO2 measurements Continuous use of the SpO2 probe may result in discomfort or pain, especially for those with microcirculatory problems. It is recommended that the probe should NOT be applied to the same place for over two hours, change the measurement site periodically and when necessary.
User Manual for All-in-One Health Monitor The SpO2 sensor cannot be immersed into water, liquid or cleanser. The SpO2 sensor can be repeatedly used. Please clean and disinfect before reuse. ☞ Connectors with the label "SpO2" can only be connected with the smart SpO2 probe. Note: The ECG and SpO2 functions cannot be used simultaneously.
User Manual for All-in-One Health Monitor 2.3.3 Temperature Measurement The infrared temperature probe is a delicate transducer. To operate please follow these steps and procedures. Failure to accurately operate may cause damage to the probes. 1. The infrared temperature probe Please place the infrared temperature probe to a stable ambient temperature for 30min before measuring. If the patient sweats, please wipe the sweat. Please begin to take measurement when the temperature is stable.
User Manual for All-in-One Health Monitor Operation procedure: Step 1: Connect the infrared temperature probe to the connector on the upper side of device with mark of "TEMP". When Monitor LCD screen displays " ", it means that the probe is connected successfully. Step 2: Power on the probe by pressing the measuring key on the probe firstly. The user can take measurement when the probe screen shows " ". Step 3: Insert the measuring tip of temperature probe into the earhole and press the power on key.
User Manual for All-in-One Health Monitor The infrared temperature probe will stand by automatically if no operation for 1 min. If you need to make re-measurement, please press the measuring key and repeat step 2 and step 3. The measuring result will be displayed, auto stored, and can be reviewed on the All-in-One Health Monitor. Normal body temperature varies in a range.
User Manual for All-in-One Health Monitor b. Operation outside of the manufacturer specified operating temperature and humidity ranges. c. Storage outside of the manufacturer specified ambient temperature and humidity ranges. d. Mechanical shock. e. Manufacturer defined soiled or damaged infrared optical components. Do NOT take a measurement when the patient is moving. Patients with tympanitis and otitis problems should NOT use this device.
User Manual for All-in-One Health Monitor 1. Test strip (The Meter will turn on automatically after inserting test strip) 2. Data interface (Link with the All-in-One Health Monitor) 4. Ejector (Remove the used strip by push this key) 3. LCD display screen (Display measuring result and other information) 5. " 6. " " Key " Key (Used on setting mode) (Used to Reviewing record or as an navigator key during setting) Figure 2.
User Manual for All-in-One Health Monitor Appearance and key functions of the Blood Glucose Monitoring System: 1. Test strip slot: when the strip is inserted into the slot, the device will automatically turn on. 2. LCD display. 3. Data interface: can be used to connect the All-in-One Health Monitor for data transmitting. 4. Ejector: remove the used strip. 5. 6. key: power on/off, also for memory recalling mode. key: Setting mode.
User Manual for All-in-One Health Monitor Step 3: Take a new test strip, insert it into the test slop of the device, then the device will automatically turn on , and display the current temperature first then show the blood icon. Step 4: Refer to the following operations for lancing device and blood lancet. When the blood drop icon appears, apply the blood drop to the front edge of the test strip, put the blood sample on the injecting hole.
User Manual for All-in-One Health Monitor Figure 2.7 A Measuring result Operations for the Lancing Device and Blood Lancet 1. Unscrew the lancing device by turning the end cap counter clockwise. 2. Insert a new lancet firmly into the lancet holder. 3. Twist off the protective tip of the lancet. Close the end cap of the lancing device. 4. Wipe the finger with alcohol cotton and wait it dry, then collect blood. Fingertip is preferred to be the blood collecting spot.
User Manual for All-in-One Health Monitor Proper blood volume Figure 2.7B Operation for Lancing Device and Blood Lancet Refer to the provided user guide for the "Blood Glucose Monitoring System" for further detailed instructions. Safety Instructions for Blood Glucose Measurement The provided test strips should be used with the Blood Glucose Monitoring System. Do NOT clean or disinfect the finger with iodine.
User Manual for All-in-One Health Monitor The test strip will draw blood at one end automatically. Do NOT press or scrape the bleeding finger. The test strip should be used as soon as possible after unpacking, and the unused strips should be kept in an airproof bottle. Only take one measurement per minute. ☞ If the monitor is connected with both the temperature probe and the blood glucose monitoring system, the monitor screen will show " ".
User Manual for All-in-One Health Monitor ECG Monitor has begun to take the ECG measurement. 5. 30 Seconds later, the result will display on the screen of the host device, and the measurement will terminate. Figure 2.8A ECG Monitor Start / Stop: Start/Stop ECG measurement.
User Manual for All-in-One Health Monitor Figure 2.8B Palm measurement Figure 2.8C Chest measurement Figure 2.8D Leg measurement To obtain a clear and high quality ECG signal, the lead wire measurement can be used. Connect the lead wire firmly to the lead wire socket of the ECG Monitor. Place the electrodes and connect the lead wires as shown in Figure 2.8E to obtain the Lead II ECG signal.
User Manual for All-in-One Health Monitor 1. Check the ECG Monitor to make sure that there is no visible damage that may affect the user’s safety and the measurement performance. If there is obvious damage, stop using the monitor. 2. Do NOT make a diagnosis by oneself according to the measurement results, always consult a doctor if abnormal information is presented frequently. 3. Do NOT use the device in a bathroom or humid environment.
User Manual for All-in-One Health Monitor Between the left edge of the breast bone and Rib 5 L(Yellow)/ F (Green)/ F (Green)/ LA(Black) LL(Red) LL(Red) 2.4 Pressure Accuracy Verification Operation procedure: 1. Unscrew the M3x6 screw from the battery compartment on the back of the All-in-One Health Monitor, as shown in figure 2.9. 2. Take a Cuff connector plug from the battery cover, as shown in figure 2.10. (Note: there are two plugs but you will only need one.) 3.
User Manual for All-in-One Health Monitor 5. Turn on the All-in-One Health Monitor. Press the menu button to go to the settings. Press and hold the large NIBP measurement button to enter the Pressure Accuracy Verification Mode. 6. Start pumping, and check if the pressure reading on the All-in-One Health Monitor matches the pressure meter reading. Cuff Connector Plug M3x6 Screw Figure 2.9 Figure 2.
User Manual for All-in-One Health Monitor Air tube Cuff Connector Figure 2.11 Figure 2.12 Pressure Accuracy Verification is a function to inspect the accuracy of pressure measurement by the NIBP module inside the device. Technician or equipment manager should do pressure accuracy verification every half year or year in order to check if the pressure measurement still conforms to the requirement of product performance.
User Manual for All-in-One Health Monitor Pressure accuracy verification must be operated by technician or equipment manager. Doctor or nurse is not allowed to do the verification, it is very dangerous especially when the pressure cuff is still on patients.
User Manual for All-in-One Health Monitor 2.
User Manual for All-in-One Health Monitor Type BF applied parts SN Charger or USB data interface Connector to link with blood glucose Serial Number monitoring system Refer to manual Follow WEE regulations for disposal TEMP Connector for temperature probe SpO2 Connector to smart SpO2 probe NIBP Connector for cuff ECG Connector for ECG accessory Page XXXIII/69
User Manual for All-in-One Health Monitor CHAPTER 3 MONITORING SCREEN DISPLAY 3.1 Measuring Screen Figure 3.1A Measuring screen Figure 3.
User Manual for All-in-One Health Monitor Figure 3.1C Measuring screen Figure 3.
User Manual for All-in-One Health Monitor Screen Description: 1. : wireless transmission icon; " ": means that the wireless transmission function is on; when the icon is blinking, the wireless connection set up is unsuccessful; when this icon is steady, the wireless connection set up is successful; " ": the wireless transmission function is off. 2. : Beep sound indicator; : pulse beep is on; : pulse beep is off. 3. : Battery voltage indicator.
User Manual for All-in-One Health Monitor 13. SpO2: the value of SpO2 with unit of %. : Pulse bar-graph. 14. 15. PR: Pulse rate with unit of bpm. 16. : the heart beat symbol, which flashes with heart beat. 17. TEMP/GLU: the current displayed temperature with an option of ℃ for Celsius, or °F for Fahrenheit. When the optional GLU is chosen, the blood glucose value will be displayed with the default unit of mmol/L. 3.
User Manual for All-in-One Health Monitor Operation Instructions: 1. Press and hold the" "key, and release after hearing one beep, to enter into the setup screen. When the patient ID blinks, the setup function is available. 2. A short press of the " / " key enables or disables the wireless transmission function. 3. A short press of the " " key confirms the setting. The beeping mark " 4. A short press of the " / 5. A short press of the " " key confirms the setting.
User Manual for All-in-One Health Monitor 3.3 History Data Review Screen On the measurement display screen, press and hold the " " key to recall the stored data records, as shown in figure 3.3 Operation instructions: 1. Press and hold the" " key, release the key after hearing one beep. The memory mark" " will appear (i.e. entering to review display screen). The patient's ID number will blink at the same time. 2. A short press of the " / " key will browse the patient's ID numbers. 3.
User Manual for All-in-One Health Monitor Note: when selecting the patient ID, the screen only displays patients with history data records. 3.4 Data Uploading 1. When the wireless transmission function is on, the monitor can communicate with a host for viewing and management. a. Open the host ’s wireless function and procedure and start to scan the PC-303 monitor. b. The host will pair with the PC-303 monitor at a moment. c.
User Manual for All-in-One Health Monitor CHAPTER 4 TECHNICAL SPECIFICATIONS 4.1 Blood Pressure Measurement 1. Technique: Oscillometric 2. Pressure measuring range (Adult/ Pediatric): 0mmHg-300mmHg 3. Accuracy of pressure measurement: ±3mmHg 4. Overpressure protection limit (Adult/ Pediatric): ≤300mmHg 5. Blood pressure measurement range (Adult/ Pediatric): SYS: 60mmHg-255mmHg DIA: 30mmHg-195mmHg 6.
User Manual for All-in-One Health Monitor 4.2 SpO2 Measurement 1. Technique: optical with dual-wavelength Wavelength: Red light: 663nm, Infrared light: 890nm Maximal optical output power: less than 2mW maximum average 2. SpO2 displayed range: 35%-100% 3. SpO2 measuring accuracy: Arms is not greater than 3% for SpO2 range from 70% to 100% Note: Arms is defined as root-mean-square value of deviation according to ISO 80601-2-61. 4. SpO2 display update: every second 5.
User Manual for All-in-One Health Monitor 4.4 Temperature Measurement 1. Measuring range: 32.0℃-43.0℃ 2. Measuring accuracy: ±0.2℃ is for TEMP range from 36.0℃ to 39.0℃, and ±0.3℃ is for the rest; ±0.4℉ is for TEMP range from 96.8℉to 102.2℉, and ±0.5℉ is for the rest. 3. Response time: ≤5s 4.5 Blood Glucose Measurement (Optional) 1. Technique: Amperometric, glucose oxidase 2. Measuring range: 1.1mmol/L-33.3mmol/L (20-600mg/dL) 3. Measuring time: ≤10 seconds 4.6 ECG Measurement (Optional) 1.
User Manual for All-in-One Health Monitor 3. Display scale: 5.0mm/mV±10% 4. Common-mode rejection ratio (CMRR): ≥60dB 4.7 Others 4.7.1 Operating Environment 1. Operating temperature: 5℃-40℃; Atmospheric pressure: 70.0kPa-106.0kPa; Relative humidity: 30%-80%, non-condensing; Power supply: AC100V-AC240V, 50/60Hz, 15VA; Internal power supply: DC3.7V (rechargeable Lithium battery); 2. The device should be situated in a place protected against direct sunlight, to prevent overheating inside of the equipment.
User Manual for All-in-One Health Monitor 7. Only monitor a single person at a time. Warning: Do not use any other adapters than those provided by Creative. 4.7.2 Classification 1. Protection against electric shock: Class II equipment and internally powered equipment 2. The degree of protection against electric shock: Type BF applied part 3. Define apply part: cuff, SpO2 probe, temperature probe, ECG lead wires (optional). 4.
User Manual for All-in-One Health Monitor CHAPTER 5 Trouble Cannot turn on the device No blood pressure result No SpO2 result No TEMP result TROUBLESHOOTING Possible reason Solution The built-in battery is drained Battery is not installed Recharge by connecting the power supply adapter Install the Lithium battery Some parts provided by other manufacturers are inserted to the connector The cuff is wrapped around the arm incorrectly The windpipe is not correctly inserted to the NIBP jack The SpO2 prob
User Manual for All-in-One Health Monitor Taking measurements before Do not take a measurement until "READY" "READY" appears on the appears on the temperature probe screen temperature probe screen CHAPTER 6 Item PACKING LIST Description Quantity Check 1 All-in-One Health Monitor One piece OK 2 Handbag One piece OK 3 User Manual One piece OK 4 Cuff One piece OK 5 USB cable One piece OK 6 Charger (with USB socket) One piece 7 Temperature probe One piece 8 Smart SpO2 probe One
User Manual for All-in-One Health Monitor and link cable ) Optional 10 Blood glucose test strips One pack 11 Disposable adhesive ECG electrodes Six pieces 12 ECG accessory One piece 13 ECG lead wire (snap) One piece Page XLVIII/69
User Manual for All-in-One Health Monitor CHAPTER 7 MAINTENANCE AND SERVICE The All-in-One Health Monitor should be properly maintained to ensure its maximum performance and long service life. In addition to the warranty period, the company also offers long-term service for each customer. It is important that the user reads and follows the operation instructions, important information and maintenance measures. 7.1 Technical Maintenance 7.1.
User Manual for All-in-One Health Monitor 7.1.2 Routine Maintenance The All-in-One Health Monitor is designed to have a life expectancy of at least 5 years. In order to ensure precise readings, it is recommended to test the pressure accuracy every year. Please see section 2.4 to perform the Blood Pressure Accuracy Check Method or contact your supplier for more information.
User Manual for All-in-One Health Monitor 7.1.3 Battery Maintenance Please pay attention to the polarity of the battery, do NOT insert into the battery compartment with reversed polarities. In order to avoid damaging the battery, do NOT use other power supply devices to charge the battery. After use, dispose of the battery according to local regulations, do NOT throw into fire. Do NOT hit or strike the battery with force. Do NOT use this battery in other devices.
User Manual for All-in-One Health Monitor battery is automatic and seamless. If the battery is damaged, please replace it with a battery with "CCC" or "CE" mark. The model and specifications of the battery should be the same as the original battery. The user must ensure that the battery meets all applicable safety codes. The user can also contact the distributor for service. 7.2 Cleaning and Disinfection of the Main Unit The device should be cleaned on a regular basis.
User Manual for All-in-One Health Monitor It is recommended to regularly clean the outer shell and screen of the monitor. Only use a non-corrosive cleanser such as clear water. Wipe the surface of the monitor and transducers with an alcohol impregnated wipe, and dry with a clean cloth or just air-dry. Dilute the cleaner. Do NOT use scrubbing materials. The monitor can be disinfected. To avoid damage do not let liquid cleaner flow into the connector jack of the monitor.
User Manual for All-in-One Health Monitor 7.3 Cleaning and Disinfection of Accessories It is recommended to clean and disinfect the accessories with a piece of gauze soaked in 75% Alcohol or 70% Isopropanol. Do not use damaged accessories. Accessories cannot be entirely immersed into water, liquid or cleanser. Do NOT use radiation, steam or epoxyethane to disinfect accessories. Wipe off any remaining residue of alcohol or isopropanol after disinfection.
User Manual for All-in-One Health Monitor Relative humidity: 10%~95%, non-condensing Atmospheric pressure: 53.0kPa~106.0kPa 7.5 Transportation The monitor should be transported by land (vehicle or railway) or air in accordance with the contractual terms. Do NOT hit or drop with force. Warranty Clause 1. This monitor has a warranty of 12 months (including rechargeable battery) and 6-month for all accessories, from the date of purchase. 2.
User Manual for All-in-One Health Monitor If the serial number is deliberately damaged, torn off or unreadable 5. If the monitor is non-functional outside of the warranty period, the manufacturer or distributor will offer an estimate for repair.
User Manual for All-in-One Health Monitor Caution: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.
User Manual for All-in-One Health Monitor interference that may cause undesired operation. Any Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
User Manual for All-in-One Health Monitor Device Information: Name Model Serial Number: Date Shop User Information: Name Postcode Tel: Add: Repair Record Date Repairing Item Personnel Page LIX/69
User Manual for All-in-One Health Monitor Patent State Intellectual Property Office of the P.R.C. patented and certificated Creative All-in-One Health Monitor on March 26th, 2014. Patent Number: ZL 2013 2 0615696.X Shenzhen Creative Industry Co., Ltd. 2/F, Block 3, Nanyou Tian'an Industry Town, 518054 Shenzhen, GD, P.R. China Tel: +86-755-2643 3514 Fax: +86-755-2643 0930 E-mail: info@creative-sz.com Website: www.creative-sz.